K203006 · R2 Technologies, Inc. · QPZ · Aug 5, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203006
Device Name
R2 Dermabrasion Tip (for Dermal Cooling System)
Applicant
R2 Technologies, Inc.
Product Code
QPZ · General, Plastic Surgery
Decision Date
Aug 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4340
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal. The Dermal Cooling System is intended to be used by trained healthcare professionals.
Device Story
Dermal Cooling System is a cryosurgical device for skin cooling without cryogenic gases/liquids. Uses thermoelectric cooler (TEC) with integrated aluminum plate to lower skin temperature. Handpiece serves as handle for manual R2 Dermabrasion Tips. Used by trained healthcare professionals in clinical settings. Dermabrasion tips facilitate manual dermabrasion for scar/tattoo removal. Cooling effect provides pain/swelling relief post-procedure. Healthcare providers use device to perform aesthetic skin treatments and manage post-surgical symptoms. Benefits include non-invasive lesion removal and temporary pain/inflammation reduction.
Clinical Evidence
No clinical data. Bench testing only. Dermabrasion performance test (n=3) and thermal insulation test (n=3) conducted to validate accessory safety and effectiveness. Results showed no unexpected side effects and skin temperature maintained at or above +20 °C.
Technological Characteristics
Thermoelectric cooler (TEC) with integrated aluminum plate for surface contact cooling. Manual dermabrasion tips fabricated from metal with textured surface. Standalone device. No cryogenic gases or liquids used.
Indications for Use
Indicated for dermatologic procedures to remove benign skin lesions and for cooling to reduce pain, swelling, inflammation, and hematoma from minor surgical procedures. When used with R2 Dermabrasion Tips, indicated for general dermabrasion, scar revision, acne scar revision, and tattoo removal. For use by trained healthcare professionals.
Regulatory Classification
Identification
A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 5, 2021
R2 Technologies, Inc. Ragan Reppond Sr. Director, HR & Corporate Affairs 2603 Camino Ramon. Suite 200 San Ramon, California 94583
Re: K203006
Trade/Device Name: Dermal Cooling System Regulation Number: 21 CFR 878. 4340 Regulation Name: Contact Cooling System for Aesthetic Use Regulatory Class: Class II Product Code: QPZ, GED Dated: June 9, 2021 Received: June 11,2021
Dear Ragan Reppond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdth/cfdocs/cfpmn/pmn.cfm identifies coombination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Leam (https://www.fda.gov/training-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203006
Device Name Dermal Cooling System
#### Indications for Use (Describe)
The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.
When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal.
The Dermal Cooling System is intended to be used by trained healthcare professionals.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203006
## 510(k) Summary
- I. SUBMITTER R2 Technologies, Inc. 2603 Camino Ramon, Suite 200 Bishop Ranch 3 San Ramon, CA 94583
- Contact: Ms. Ragan Reppond Vice President of Corporate Affairs & Compliance R2 Technologies, Inc. Phone: 925-378-4400
rreppond@r2technologies.com
Date Prepared: July 29, 2021
#### II. SUBJECT DEVICE and ACCESSORY
| Device Name: | Dermal Cooling System |
|------------------------|---------------------------------------------|
| Device Common Name: | Contact cooling for skin lesion pain relief |
| Accessory Name: | R2 Dermabrasion Tips |
| Accessory Common Name: | Brush, Dermabrasion, Manual |
| Regulatory Class : | II |
| Product Code : | QPZ, GED |
| Regulation : | 21 CFR 878.4340, 21 CFR 878.4800 |
#### III. PREDICATE AND REFERENCE DEVICES
The predicate device is the Dermal Cooling System, K201260.
| Name of Device: | Dermal Cooling System |
|-------------------|-----------------------------------|
| Common Name: | Cryosurgical unit and accessories |
| Regulatory Class: | II |
| Product Code: | GEH, MLY |
| Regulation: | 21 CFR 878.4350 |
The reference device is the Derma Paddle (Livra) attachment (Class I, 510(k) exempt), manufactured by Procell Therapies.
#### IV. DEVICE DESCRIPTION
{4}------------------------------------------------
# K203006
The Dermal Cooling System is identical to the device cleared in K201260. The Dermal Cooling System is a cryosurgical device used to cool the skin, without the use of cryogenic gases or liquids, for the removal of benign skin lesions and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures. Surface contact cooling is achieved using a thermoelectric cooler (TEC), with an integrated aluminum plate, to lowerthe temperature of the skin. For dermabrasion, the Dermal Cooling System handpiece is intended to serve as a handle to facilitate manual movement of the R2 Dermabrasion Tips.
The R2 Dermabrasion Tips are optional accessories for the Dermal Cooling System that may be attached to the distal end of the handpiece to facilitate manual dermabrasion. The R2 Dermabrasion Tips may be used with a commercially available topical or water to facilitate movement of the handpiece across the treatment area.
### V. INDICATIONS FOR USE
The Dermal Cooling System is a cryosurgical instrument intended for use in dermatologic procedures for the removal of benign lesions of the skin and for use when cooling is intended for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures.
When the Dermal Cooling System is used with the R2 Dermabrasion Tips, the intended use includes general dermabrasion, scar revision, acne scar revision, and tattoo removal.
The Dermal Cooling System is intended to be used by trained healthcare professionals.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE
The predicate device, the Dermal Cooling System, has not been modified for use with the R2 Dermabrasion Tips and proposed expanded indication, i.e., general dermabrasion, scar revision, acne scar revision, and tattoo removal.
The reference device, the Derma Paddle (Livra) attachment, is included to substantiate the proposed expanded indications for use for the R2 Dermabrasion Tips as the intended use for the Derma Paddle (Livra) attachment (510(k) exempt) includes "general dermabrasion, scar revision, acne scar revision, and tattoo removal". The R2 Dermabrasion Tips and the Derma Paddle (Livra) attachment have these same fundamental scientific technological elements:
The R2 Dermabrasion Tips and the reference device, Derma Paddle (Livra) attachment have these similarities:
- used for aesthetic purposes;
- fabricated from metal; ●
{5}------------------------------------------------
- textured surface (roughness);
- intended for manual dermabrasion;
- classification (generic description)
The key differences between the R2 Dermabrasion Tips and the Derma Paddle (Livra) attachment are:
- geometry
- mechanism of attachment
The differences do not raise new questions of safety and effectiveness with respect to the R2 Dermabrasion Tips and proposed expanded indication.
### VII. PERFORMANCE DATA
Performance data is provided in support of the substantial equivalence determination, as summarized in the table below.
Dermabrasion performance test demonstrated the R2 Dermabrasion Tips can perform its intended use safely and effectively when used with the Dermal Cooling System, while the thermal insulation test supported the conclusion no additional risk was introduced when the R2 Dermabrasion Tips are used with the Dermal Cooling System.
| Test | Test method/Requirement | Acceptance criteria | Results |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|---------|
| Accessories<br>Validation | Dermabrasion Performance Test:<br>3 subjects tested with the R2<br>Dermabrasion Tips (20 Amps<br>EDM) and 3 subjects tested with<br>(80 Amps EDM) | No unexpected side effects<br>observed immediately post test,<br>and 2-days post test | Passed |
| | Thermal Insulation Test:<br>3 subjects tested with the R2<br>Dermabrasion Tips (20 Amps<br>EDM and 80 Amps EDM) | Skin temperature to remain at or<br>above +20 °C for the full duration | Passed |
No preclinical or clinical testing was performed.
### VIII. CONCLUSIONS
An optional accessory, manual dermabrasion tips, is added to the Dermal Cooling System cleared in K201260. Performance tests demonstrated the dermabrasion tips can perform as intended when used with the Dermal Cooling System without introducing additional risks for the intended use. No new question of safety or effectiveness is raised. The subject device is therefore considered as substantially equivalent to the predicate device K201260.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.