ACIST CVi1 Contrast Delivery System, ACIST CVi Contrast Delivery System
K203004 · Acist Medical Systems, Inc. · DXT · Dec 17, 2020 · Cardiovascular
Device Facts
Record ID
K203004
Device Name
ACIST CVi1 Contrast Delivery System, ACIST CVi Contrast Delivery System
Applicant
Acist Medical Systems, Inc.
Product Code
DXT · Cardiovascular
Decision Date
Dec 17, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
ACIST CVi®1 Contrast Delivery System The ACIST CVi®1 Contrast Delivery System is indicated for controlled administration of radiopaque contrast media and saline to human subjects while undergoing angiographic procedures. The CVi®1 Syringe Kits, Manifold Kit and AngioTouch® Hand Controller Kit must be discarded after each patient procedure. The CVi@1 Syringe Kits are also indicated for single patient use with ACIST CVi® Contrast Delivery Systems. The ACIST CVi@1 Contrast Delivery System is to be used only by and under quasi-continuous supervision of trained health care professionals in an appropriate licensed health care facility, in a room designated for radiological procedures that involve intravascular administration of a contrast agent. ACIST CVi® Contrast Delivery System The ACIST CVi® Contrast Delivery System is indicated for controlled administration of radiopaque contrast media and saline to human subjects while undergoing angiographic procedures. The ACIST CVi® Contrast Delivery System is specifically indicated for use in angiographic procedures for the delivery of ISOVUE (Iopamidol Injection) contrast media as supplied in an Imaging Bulk Package (IBP), for a maximum of ten (10) hours. The Syringe Kit must be discarded after six (6) patient procedures. The Manifold Kit and AngioTouch Kit must be discarded after each patient procedure. The ACIST CVi® Contrast Delivery System is to be used only by and under quasi-continuous supervision of trained health care professionals in an appropriate licensed health care facility, in a room designated for radiological procedures that involve intravascular administration of a contrast agent.
Device Story
The ACIST CVi/CVi1 Contrast Delivery System is a software-controlled, motor-driven pump used in interventional cardiology, radiology, and vascular surgery. It delivers radiopaque contrast media and saline at user-determined flow rates and volumes to patients via hospital-provided catheters and ACIST-supplied consumable kits (including the AngioTouch Hand Controller). The system synchronizes with X-ray imaging equipment to assist physicians in producing angiograms. The device is operated by trained healthcare professionals in catheterization and radiological laboratories. The current modification involves the AngioTouch Hand Controller, specifically replacing braided polyurethane high-pressure tubing with coextruded polyamide/polyurethane tubing and adding rotating luers. The system provides a hemodynamic pathway from the patient to the hospital's monitoring system. The device benefits patients by enabling precise, controlled contrast delivery during diagnostic and interventional procedures.
Clinical Evidence
Bench testing only. Testing included burst, functional, life, pressure, bond pull, flow, and durability. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity. Results demonstrate the modified device meets specifications and performs as intended.
Technological Characteristics
Motor-driven pump system; software-controlled. AngioTouch Hand Controller components: housings, bladders, twin tubing, and high-pressure tubing (coextruded polyamide/polyurethane). Connectivity: synchronizes with X-ray imaging systems. Sterilization: Ethylene Oxide. Single-use consumables.
Indications for Use
Indicated for controlled administration of radiopaque contrast media and saline to human subjects undergoing angiographic procedures. Used by trained healthcare professionals in licensed facilities for radiological procedures involving intravascular contrast administration.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
ACIST CVi®/CVi®1 Contrast Delivery System (K191060)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
December 17, 2020 ACIST Medical Systems, Inc. Jeff Koll Sr. Principal Regulatory Affairs Specialist 7905 Fuller Rd Eden Prairie, Minnesota 55344
### Re: K203004
Trade/Device Name: ACIST CVi1 Contrast Delivery System, ACIST CVi Contrast Delivery System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: November 23, 2020 Received: November 24, 2020
## Dear Jeff Koll:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203004
#### Device Name
ACIST CVi®1 Contrast Delivery System ACIST CVi® Contrast Delivery System
#### Indications for Use (Describe)
ACIST CVi®1 Contrast Delivery System
The ACIST CVi®1 Contrast Delivery System is indicated for controlled administration of radiopaque contrast media and saline to human subjects while undergoing angiographic procedures. The CVi®1 Syringe Kits, Manifold Kit and AngioTouch® Hand Controller Kit must be discarded after each patient procedure. The CVi@1 Syringe Kits are also indicated for single patient use with ACIST CVi® Contrast Delivery Systems. The ACIST CVi@1 Contrast Delivery System is to be used only by and under quasi-continuous supervision of trained health care professionals in an appropriate licensed health care facility, in a room designated for radiological procedures that involve intravascular administration of a contrast agent.
#### ACIST CVi® Contrast Delivery System
The ACIST CVi® Contrast Delivery System is indicated for controlled administration of radiopaque contrast media and saline to human subjects while undergoing angiographic procedures. The ACIST CVi® Contrast Delivery System is specifically indicated for use in angiographic procedures for the delivery of ISOVUE (Iopamidol Injection) contrast media as supplied in an Imaging Bulk Package (IBP), for a maximum of ten (10) hours. The Syringe Kit must be discarded after six (6) patient procedures. The Manifold Kit and AngioTouch Kit must be discarded after each patient procedure. The ACIST CVi® Contrast Delivery System is to be used only by and under quasi-continuous supervision of trained health care professionals in an appropriate licensed health care facility, in a room designated for radiological procedures that involve intravascular administration of a contrast agent.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 2.4
# 510(k) Summary [21 CFR 807.92]
| Submitter's<br>Name and<br>Address | ACIST Medical Systems, Inc.<br>7905 Fuller Road<br>Eden Prairie, MN 55344<br>Phone: 952.656.2409<br>Fax: 952.941.4648 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and Information | Jeffrey L. Koll<br>Sr. Principal Regulatory Affairs Specialist<br>Phone: 952.656.2409<br>Fax: 952-941-4648<br>E-mail: jeff.koll@acistmedical.com |
| Date Prepared | 29 September 2020 |
| Proprietary<br>Name(s) | ACIST CVi®1 Contrast Delivery System<br>ACIST CVi® Contrast Delivery System |
| Common Name | Contrast Delivery System |
| Product Code | DXT |
| Classification | Class II, 21 CFR Part 870.1650, Angiographic injector and syringe |
| Predicate Device | ACIST CVi®/CVi®1 Contrast Delivery System (K191060) |
| Device<br>Description | The CVi/CVi1 System is designed to aid the physician in the controlled<br>infusion of radiopaque contrast media. Radiographic imaging devices are<br>used in conjunction with the delivery of contrast media to produce<br>angiograms. Operating environments for the CVi/CVi1 System are<br>catheterization and radiological laboratories. The CVi/CVi1 System<br>contains a software-controlled motor-driven pump that delivers contrast<br>media at a user-determined flow rate and volume via the ACIST provided<br>consumable kits and a hospital provided angiographic patient catheter. The<br>CVi/CVi1 System is also equipped to synchronize with commercially<br>available X-ray imaging systems. The CVi/CVi1 System is used in<br>interventional cardiology, radiology, and vascular surgical procedures. The<br>CVi/CVi1 System device modification that is the subject of this 510k<br>premarket notification is a material component change to the AngioTouch |
{4}------------------------------------------------
| Indications<br>for Use | ACIST CVi®1 Contrast Delivery System<br>The ACIST CVi®1 Contrast Delivery System is indicated for controlled<br>administration of radiopaque contrast media and saline to human subjects<br>while undergoing angiographic procedures.<br>The CVi1 Syringe Kits, Manifold Kit and AngioTouch® Hand Controller<br>Kit must be discarded after each patient procedure. The CVi1 Syringe Kits<br>are also indicated for single patient use with ACIST CVi® Contrast<br>Delivery Systems.<br>The ACIST CVi®1 Contrast Delivery System is to be used only by and<br>under quasi-continuous supervision of trained health care professionals in<br>an appropriate licensed health care facility, in a room designated for<br>radiological procedures that involve intravascular administration of a<br>contrast agent |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ACIST CVi® Contrast Delivery System<br>The ACIST CVi® Contrast Delivery System is indicated for controlled<br>administration of radiopaque contrast media and saline to human<br>subjects while undergoing angiographic procedures.<br>The ACIST CVi® Contrast Delivery System is specifically indicated<br>for use in angiographic procedures for the delivery of ISOVUE<br>(Iopamidol Injection) contrast media as supplied in an Imaging Bulk<br>Package (IBP), for a maximum of ten (10) hours. The Syringe Kit must<br>be discarded after six (6) patient procedures. The Manifold Kit and<br>AngioTouch Hand Controller Kit must be discarded after each patient<br>procedure.<br>The ACIST CVi® Contrast Delivery System is to be used only by and<br>under quasi-continuous supervision of trained health care professionals in<br>an appropriate licensed health care facility, in a room designated for<br>radiological procedures that involve intravascular administration of a<br>contrast agent. |
| Substantial<br>Equivalence /<br>Comparison of<br>Technological<br>Characteristics | The proposed AngioTouch Hand Controller modifications to the CVi/CVi1<br>system includes:<br>Replacing current braided polyurethane high-pressure tubing with<br>coextruded polyamide/ polyurethane high-pressure tubing. Replace fixed luers with rotating luers on the high-pressure tubing Extend shelf life of the modified AngioTouch Hand Controller Kit These modifications do not introduce or raise different questions regarding<br>the safety or effectiveness of the device. The fundamental technological<br>characteristics and principle of operation of the modified device are<br>unchanged from the predicate device. The CVi/CVi1 system with modified<br>AngioTouch Hand Controller is substantially equivalent to the predicate<br>device in intended use, design, performance, and technological<br>characteristics. |
{5}------------------------------------------------
| Characteristic | Predicate Device<br>AngioTouch Hand Controller<br>K191060 | Proposed Device<br>AngioTouch Hand Controller (modified)<br>This 510(k) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Intended Use | For controlled administration of<br>radiopaque contrast media and saline to<br>human subjects while undergoing<br>angiographic procedures. | Same |
| Usability | Single Use | Same |
| AngioTouch<br>Hand Controller<br>Kit Components | Housings (2)<br>Bladders (2)<br>Twin Tubing<br>High Pressure Tubing | Same |
| High Pressure<br>Tubing Length | 54 inches<br>65 inches | 65 inches |
| High Pressure<br>Tubing Function | Deliver contrast to patient<br>Deliver saline to patient<br>Provides a hemodynamic pathway from<br>the patient to the hospital's hemodynamic<br>monitoring system | Same |
| High Pressure<br>Tubing Material | Polyurethane | Polyurethane/Polyamide |
| Sterilization<br>Method | Ethylene Oxide | Same |
### Consumable Kit Comparison- AT-P Hand Controller
Performance The modified device was subjected to bench and biocompatibility testing. Bench testing Data included burst, functional, life, pressure, bond pull, flow, and durability. Test results demonstrate that the modified device meets specification and performs as intended. No new safety or performance issues were raised during the testing. The modified device is substantially equivalent to the predicate device.
The following biocompatibility tests were completed:
- ·Cytotoxicity
- ·Sensitization
- •Irritation
- Acute Systemic Toxicity
- •Material Mediated Pyrogenicity
#### Conclusion
The CVi/CVi1 Contrast Delivery System with modified AngioTouch Hand Controller Kit (AT X) is substantially equivalent in design, performance, and technological characteristics to the predicate device for its intended purpose.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.