Bausch + Lomb ABT12 multi-purpose solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
Device Story
ABT12 multi-purpose solution is a sterile isotonic aqueous solution for soft and silicone hydrogel contact lens care. It functions as a chemical disinfectant, cleaner, and storage medium. The solution contains a triple disinfection system (polyaminopropyl biguanide, polyquaternium, and alexidine) and dual surfactants (poloxamer 181 and poloxamine 1107). Users apply the solution to lenses for cleaning and rinsing, followed by a minimum 4-hour disinfection soak. The device is intended for over-the-counter use by contact lens wearers. It benefits patients by maintaining lens hygiene, removing protein deposits, and ensuring safe lens storage, thereby supporting comfortable lens wear and reducing the risk of ocular complications associated with contaminated lenses.
Clinical Evidence
A 3-month, multicenter, randomized, masked, parallel, bilateral clinical study (n=252) compared ABT12 to the predicate. Primary endpoints included slit lamp findings (safety) and comfort, dryness, and lens deposits (effectiveness). Results showed ABT12 was non-inferior to the predicate. No serious adverse events or study discontinuations related to the solution occurred. Visual acuities and lens performance parameters were similar between groups.
Technological Characteristics
Sterile isotonic aqueous solution containing polyaminopropyl biguanide (0.0005%), polyquaternium (0.00015%), and alexidine (0.00025%). Surfactants: Poloxamer 181 and Poloxamine 1107. Features a triple disinfection system. Minimum disinfection time: 4 hours. Packaging: Clear plastic bottle with tamper-evident seal. Shelf life: 90 days after opening. Storage period: 30 days.
Indications for Use
Indicated for daily cleaning, conditioning, protein removal, rinsing, chemical disinfection, and storage of soft (hydrophilic) and silicone hydrogel contact lenses for patients requiring contact lens care.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
COMPLETE multi-purpose solution Easy Rub Formula (K014202)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 28, 2021
Bausch & Lomb Incorporated Melissa Thomas Director, Regulatory Affairs 1400 North Goodman Street Rochester, NY 14609
Re: K202932
Trade/Device Name: ABT12 multi-purpose solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated: April 27, 2021 Received: April 28, 2021
Dear Melissa Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
{1}------------------------------------------------
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202932
Device Name ABT12 multi-purpose solution
Indications for Use (Describe)
Bausch + Lomb ABT12 multi-purpose solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silicone hydrogel contact lenses, as recommended by your eye care practitioner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| General Information | |
|---------------------------------|---------------------------|
| Submitter: | Contact Person: |
| Bausch & Lomb Incorporated | Melissa Thomas |
| 1400 North Goodman Street | 1400 North Goodman Street |
| Rochester, NY 14609 | Rochester, NY 14609 |
| General Telephone: 585-338-5800 | 585-338-6045 (phone) |
| | 585-338-0702 (fax) |
| | Melissa.Thomas@bausch.com |
#### Premarket Notification Number: K202932
Preparation Date: May 27, 2021
| 1. Names | |
|------------------------|-----------------------------------------------|
| Device Name: | Bausch + Lomb ABT12 multi-purpose solution |
| Classification Name: | Accessories, Soft Lens Products |
| Common Name: | Soft (hydrophilic) Contact Lens Care Solution |
| Product Codes: | LPN |
| Device Classification: | Class II (21 CFR §886.5928) |
### 2. Predicate Devices
K014202 COMPLETE multi-purpose solution Easy Rub Formula cleared on February 20, 2002.
#### 3. Product Description
The subject of this 510(k) submission is for the new product Bausch + Lomb ABT12 multi-purpose solution which is substantially equivalent to the COMPLETE multi-purpose solution Easy Rub Formula which is currently commercialized.
Bausch + Lomb ABT12 multi-purpose solution is a sterile isotonic aqueous solution [containing polyaminopropyl biguanide (0.0005%), polyquaternium (0.00015%), and alexidine (0.00025%)] for disinfecting, cleaning, conditioning, rinsing, protein removal, and storing soft (hydrophilic) and silicone hydrogel contact lenses. The sterile solution is packaged in a plastic bottle with a tamper evident seal and labeled with a lot number and expiration date.
#### 4. Indications for Use
Bausch + Lomb ABT12 multi-purpose solution is indicated for use in the daily cleaning, conditioning, removing protein deposits, rinsing, chemical (not heat) disinfection, and storage of soft (hydrophilic) contact lenses, including silcone hydrogel contact lenses, as recommended by your eye care practitioner.
#### 5. Summary of Technological Characteristics
The technological characteristics of Bausch + Lomb ABT12 multi-purpose solution are substantially equivalent to those of the predicate devices.
{4}------------------------------------------------
| Features | Predicate Device<br>COMPLETE multi-purpose<br>solution Easy Rub Formula | Subject Device<br>ABT12 multi-purpose solution |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K014202 | K202932 |
| | Classification | |
| Classification | Class II | Class II |
| Product Code | LPN | LPN |
| | Indication | |
| Indicated for Soft<br>Contact Lenses | Yes | Yes |
| Indicated for<br>Silicone Hydrogel<br>Lenses | No | Yes |
| | Indicated for the care of soft<br>(hydrophilic) contact lenses. Use<br>this product, as recommended by<br>your eye care professional, to: •<br>Chemically (NOT HEAT) Disinfect<br>• Clean • Rinse • Store • Remove<br>Protein | Indicated for use in the daily cleaning,<br>conditioning, removing protein deposits,<br>rinsing, chemical (not heat) disinfection, and<br>storage of soft (hydrophilic) contact lenses,<br>including silicone hydrogel contact lenses,<br>as recommended by your eye care<br>practitioner. |
| Intended Use | | |
| | Formulation / Regimen | |
| Rub Regimen | 3 Or More Drops for 20 seconds | 3 Drops for 20 seconds |
| Optional No-Rub<br>Regimen | No | No |
| Rinse Required for<br>Cleaning | Rinse for 5 seconds on each side | Thoroughly rinse for 5 seconds on each side |
| Minimum<br>Disinfection Time | 6 hours | 4 hours |
| Triple Disinfection<br>System | No | Yes |
| Preservatives /<br>Disinfectants | Polyhexamethylene biguanide | Polyaminopropyl Biguanide<br>Polyquaternium<br>Alexidine |
| Dural Surfactants | No | Yes |
| Surfactants | Poloxamer 237 | Poloxamer 181<br>Poloxamine 1107 |
| Lens Storage<br>Period | 30 Days | 30 Days |
| Discard After<br>Opening | 90 Days | 90 Days |
| Primary Container | White Plastic Bottle, Multiple<br>Sizes | Clear Plastic Bottle, Multiple Sizes |
| Tamper Resistant | Yes | Yes, includes a tamper evident seal |
{5}------------------------------------------------
# 6. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that Bausch + Lomb ABT12 multi-purpose solution is substantially equivalent to the predicate devices.
# Performance Data
A series of preclinical and clinical testing was performed to demonstrate the safety and effectiveness of Bausch + Lomb ABT12 multi-purpose solution as described in Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997. A brief summary of the test results is provided below:
### Biocompatibility
Cytotoxicity, ocular irritation, oral toxicity, sensitization and in-vivo ocular biocompatibility studies were completed for Bausch + Lomb ABT12 multi-purpose solution in accordance with FDA Guidance Premarket Notification (510(K)) Guidance Document for Contact Lens Care Products. Testing was conducted under various conditions including on ABT12 multi-purpose solution itself, the ABT12 solution manufactured with the maximum preservative concentrations, and testing with extracts of eight different lenses (USAN materials) cycled 30 times in ABT12 multi-purpose solution. In relevant studies the predicate device K014202 was used for side by side comparison. ABT12 multi-purpose solution uses identical packaging materials previously cleared under K083757, testing previously conducted on the packaging components was leveraged to support biocompatibility. The test results demonstrated the biocompatibility of Bausch + Lomb ABT12 multi-purpose solution.
### Microbiology
The sponsor conducted a series of studies according to EN ISO 14729:2001 Ophthalmic optics – Contact lens care products – Microbiological requirements and test methods for products and regimens for hygienic management of contact lenses, including Amendment 1 (2010) and EN ISO 14730:2014 Ophthalmic optics – Contact lens care products – Antimicrobial preservative efficacy testing and guidance on determining discard date and demonstrate Bausch + Lomb ABT12 multi-purpose solution exceeds the criteria for disinfection and preservative efficacy.
#### Lens Compatibility
The results of lens compatibility studies demonstrate Bausch + Lomb ABT12 multi-purpose solution is compatible with soft contact lenses including silicone hydrogel contact lenses.
# Cleaning Efficacy
The cleaning efficacy of the solution was evaluated through the determination of the Critical Micelle Concentration (CMC). The surfactant concentrations are well above the CMC for the individual surfactants. In-vitro laboratory studies demonstrated the cleaning properties of Bausch + Lomb ABT12 multi-purpose solution with contact lenses artificially deposited with protein.
# Clinical Study
Bausch + Lomb conducted a controlled clinical study with soft (hydrophilic) contact lenses, including silicone hydrogel lenses, comparing the safety and effectiveness of Bausch + Lomb ABT12 multi-purpose solution to COMPLETE multi-purpose solution Easy Rub Formula. The results of the study support a substantial equivalence determination.
A total of 252 subjects were enrolled in a three-month multicenter, randomized, masked, parallel, bilateral study conducted in the USA. Subjects were randomized to receive Bausch + Lomb investigational
{6}------------------------------------------------
#### K202932
ABT12 multi-purpose solution (Test), or COMPLETE® Multi-Purpose Solution (Control). Eligible subjects were enrolled into one of five lens groups based on their habitual contact lenses. In total, 240 of the 252 subjects enrolled completed the study.
#### Subject Demographics:
Of the 127 subjects in the ABT12 group, 39 (30.7%) were male and 88 (69.3%) were female, with a mean (SD) age of 34.4 (9.72) years (range: 18 - 67 years). Of the 125 subjects in the COMPLETE group, 34 (27.2%) were male and 91 (72.8%) were female, with a mean (SD) age of 34.1 (9.79) years (range: 18 - 65 years).
With regard to race, in the ABT12 group 101 (80.2%) subjects were White, 13 (10.3%) were Black/African American, 11 (8.7%) were Asian, and 1 (0.8%) was Multiple. In the COMPLETE group 103 (82.4%) subjects were White, 11 (8.8 %) were Black/African American, 10 (8.0%) were Asian, and 1 (0.8 %) was Multiple.
### Safety results:
The primary endpoint of interest for slit lamp findings was achieved. The ABT12 group demonstrated noninferiority to the COMPLETE group Over All Follow-up Visits for slit lamp findings greater than Grade 2.
In the test device ABT12 multi-purpose solution there were no eyes with serious adverse events, adverse device effects, or significant non-serious adverse events, and there were no eyes discontinued from the study due to an AE. None of the AEs in either treatment group were considered related to the study solution.
There were no corneal infiltrates reported and no changes to pre-existing corneal scars. There were no notable differences between the ABT12 and COMPLETE groups for any of the lens wear or lens performance parameters at any of the study visits. Visual acuities were similar between eyes of both groups.
In conclusion, graded slit lamp findings; the comparison of AEs; and the comparison of visual acuities indicate that the ABT12 multi-purpose solution is safe for use with soft contact lenses, including silicone hydrogel lenses.
#### Effectiveness results:
For each of the three primary effectiveness endpoints (overall comfort averaged over all follow-up visits; dryness averaged over all follow-up visits; and optimal [{none or light} lens deposits at all follow-up visits), the treatment differences indicated that the ABT12 solution is non-inferior to the COMPLETE solution. There were no notable differences between the two treatment groups with regard to the additional effectiveness assessments of symptoms/complaints, worn lens characteristics, and dispensed lens characteristics. In conclusion, the three primary effectiveness endpoints indicate that the ABT12 multi-purpose solution is effective for use with soft contact lenses, including silicone hydrogel lenses.
#### Substantial Equivalence
The cumulative results of laboratory, in-vivo, and clinical testing sponsored by Bausch + Lomb demonstrate that the safety, effectiveness and performance of Bausch + Lomb ABT12 multi-purpose solution are substantially equivalent to COMPLETE multi-purpose solution Easy Rub Formula for soft contact lenses, including silicone hydrogel contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.