K202927 · Syncthink, Inc. · QEA · Oct 2, 2021 · Neurology
Device Facts
Record ID
K202927
Device Name
EYE-SYNC
Applicant
Syncthink, Inc.
Product Code
QEA · Neurology
Decision Date
Oct 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1455
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K202927 · Oct 2, 2021
EYE-SYNC
Syncthink, Inc.
Clinical validation study cohort (1,069 subjects); SCAT-5 clinical reference standard assessments
Retrospective analysis of clinical data from 1,069 athletes was used to validate the EYE-SYNC device's performance (sensitivity, specificity, and predictive values) as an aid in the diagnosis of concussion.
The EYE-SYNC® is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects. The EYE-SYNC® is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI), within three days of sport-related head injury in patients 17-24 years of age in conjunction with a standard neurological assessment, for use by medical professionals qualified to interpret the results of a concussion assessment examination. A negative EYE-SYNC® classification corresponds to eye movements that are consistent with a lack of concussion. A positive EYE-SYNC® classification corresponds to eye movements that may be present in patients with concussion.
Device Story
EYE-SYNC is a portable, head-mounted eye-tracking system used by medical professionals in clinical settings to aid concussion diagnosis. The device consists of a modified VR headset (with two high-speed infrared cameras and 12 LEDs at 850nm) and an Android tablet. It captures eye gaze data using pupil-corneal reflection. The system guides subjects through four assessment paradigms: smooth pursuit, saccades, vestibular-ocular reflex (VOR), and VOR cancellation (VORx). Data is processed to generate visual synchronization metrics and an automated classification (positive/negative for concussion). The tablet manages patient records, device control via Bluetooth, and cloud synchronization. By providing objective eye-tracking data, the device assists clinicians in identifying visual tracking impairments associated with mTBI, supporting clinical decision-making alongside standard neurological assessments.
Clinical Evidence
Validation study of 1,069 athletes (ages 17-24) compared EYE-SYNC output against SCAT-5 clinical reference standard. Results: 82% sensitivity (74-89% CI), 93% specificity (91-94% CI), 98% NPV, 56% PPV. Test-retest reliability study (n=150) showed high ICC values (0.78-0.86) across metrics (tangential error, radial error, phase error).
Indicated for patients 17-24 years of age with suspected sport-related concussion (mTBI) within three days of injury. Used as an aid to diagnosis in conjunction with standard neurological assessment by qualified medical professionals.
Regulatory Classification
Identification
A traumatic brain injury eye movement assessment aid is a prescription device that uses a patient's tracked eye movements to provide an interpretation of the functional condition of the patient's brain. This device is an assessment aid that is not intended for standalone detection or diagnostic purposes.
Special Controls
In combination with the general controls of the FD&C Act, the traumatic brain injury eye movement assessment aid is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance data under anticipated conditions of use must evaluate tracked eye movement in supporting the indications for use and include the following:
(i) Evaluation of sensitivity, specificity, positive predictive value, and negative predictive value using a reference method of diagnosis;
(ii) Evaluation of device test-retest reliability; and
(iii) A description of the development of the reference method of diagnosis, which may include a normative database, to include the following:
(A) A discussion of how the clinical work-up was completed to establish the reference method of diagnosis, including the establishment of inclusion and exclusion criteria; and
(B) If using a normative database, a description of how the “normal” population was established, and the statistical methods and model assumptions used.
(2) Software verification, validation, and hazard analysis must be performed. Software documentation must include a description of the algorithms used to generate device output.
(3) Performance testing must demonstrate the electrical safety and electromagnetic compatibility (EMC) of the device.
(4) The patient-contacting components of the device must be demonstrated to be biocompatible.
(5) A light hazard assessment must be performed for all eye-tracking and visual display light sources.
(6) Labeling must include:
(i) A summary of clinical performance testing conducted with the device, including sensitivity, specificity, positive predictive value, negative predictive value, and test-retest reliability;
(ii) A description of any normative database that includes the following:
(A) The clinical definition used to establish a “normal” population and the specific selection criteria;
(B) The format for reporting normal values;
(C) Examples of screen displays and reports generated to provide the user results and normative data;
(D) Statistical methods and model assumptions; and
(E) Any adjustments for age and gender.
(iii) A warning that the device should only be used by trained healthcare professionals;
(iv) A warning that the device does not identify the presence or absence of traumatic brain injury or other clinical diagnoses;
(v) A warning that the device is not a standalone diagnostic; and
(vi) Any instructions to convey to patients regarding the administration of the test and collection of test data.
GN Otometric 1068 A/S Type 1085 ICS Impulse (K151504)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2021
SyncThink, Inc. % Allison Kumar Regulatory Consultant Arina Consulting, LLC 27 Hilltop Dr San Carlos, California 94070
Re: K202927
Trade/Device Name: EYE-SYNC Regulation Number: 21 CFR 882.1455 Regulation Name: Traumatic brain injury eye movement assessment aid Regulatory Class: Class II Product Code: QEA Dated: August 18, 2020 Received: September 29, 2020
Dear Allison Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202927
Device Name EYE-SYNC
#### Indications for Use (Describe)
The EYE-SYNC® is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects.
The EYE-SYNC® is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI), within three days of sport-related head injury in patients 17-24 years of age in conjunction with a standard neurological assessment, for use by medical professionals qualified to interpret the results of a concussion assessment examination.
A negative EYE-SYNC® classification corresponds to eye movements that are consistent with a lack of concussion.
A positive EYE-SYNC® classification corresponds to eye movements that may be present in patients with concussion.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">✖</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span>❍</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This summary is being submitted in accordance with the requirements of 21 CFR Part 807.92.
## 510(k) Owner Information:
SyncThink, Inc. 2172 Staunton Court Palo Alto, CA 94306 Tel 508-446-4287 Contact: John Ferreira, VP Engineering jferreira@syncthink.com
### Submission Correspondent:
Allison Kumar Arina Consulting, LLC allison@arinaconsulting.com
#### Device Information:
Trade Name: EYE-SYNC Device: Traumatic Brain Injury Eye Movement Assessment Aid Regulation: 882.1455 Classification Panel: Neurology Device Class: 2 Product Code: QEA
## Predicate Device:
Oculogica EveBOX (DEN170091)
## Reference Devices:
SyncThink EYE-SYNC (K152915), GN Otometric 1068 A/S Type 1085 ICS Impulse (K151504)
## Intended Use/Indications for Use:
The EYE-SYNC® is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects.
The EYE-SYNC® is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI), within three days of sport-related head injury in
{4}------------------------------------------------
patients 17 -24 years of age in conjunction with a standard neurological assessment, for use by medical professionals qualified to interpret the results of a concussion assessment examination.
A negative EYE-SYNC® classification corresponds to eye movements that are consistent with a lack of concussion.
A positive EYE-SYNC® classification corresponds to eye movements that may be present in patients with concussion.
# Device Description:
The SyncThink EYE-SYNC device is a portable, fully enclosed eye tracking environment with three primary components:
- 1. Eye Tracker (head-mounted device) with eye tracking sensor
- 2. Eye tracking display
- 3. Android Tablet
The eye tracker is a modified Samsung GearVR provided by SensoMotoric Instruments (SMI). It is mounted to the subject's face and held either by hands placed on the side or using a strap. The eye tracking sensor includes two high-speed infrared cameras (for each eye) connected to a visual display for battery, computation, and display. Camera lighting is provided by 12 high-quality Light-emitting Diodes (LEDs) centered at 850 nanometers. Eye gaze tracking is performed using a proprietary implementation of the pupil- corneal reaction method. The eye tracker display is a nonnetworked mobile device that fits within the eye tracking sensor and connects over USB. The display receives eye tracking sensor information for post-processing, manages sensor calibration, provides binocular visual display to the subject, and interfaces with the Android tablet over Bluetooth. Eye gaze tracking and visual display is combined to provide several assessment paradigms to characterize subject eye tracking performance:
- Smooth Pursuit
- · Saccades
- · Vestibular-Ocular Reflex (VOR)
- · VOR Cancellation (VORx)
An EYE-SYNC software app on the display device manages these functions.
The Android tablet is a standard off-the-shelf 9.7" mobile tablet from Samsung with Verizon 4G cellular connectivity. A second EYE-SYNC software app designed to provides an integrative platform for data collected on the HMD eye tracker:
- · Patient, administrator, and records management
- · Eye tracker assessment Bluetooth control
- · Assessment Vestibular/Ocular-Motor Screening (VOMS) self-report tools
- · Eye tracker assessment real-time Bluetooth monitoring
{5}------------------------------------------------
- Eye tracker data log Bluetooth transfer
- Eye tracker assessment analysis using visual synchronization metrics
- · Analysis report generation with visualizations
- · Background diagnostics to verify device health
- · Cloud connectivity for data synchronization
Internal batteries from the eye tracking display and Android tablet provide power for remote use, away from power source. Each device has an EYE-SYNC software app installed to provide the described functionality. EYE-SYNC is provided as a complete system and both apps are managed a single software project with identical version numbers. EYE-SYNC is provided in a standalone carry-case with user manual, strap, cleaning, and charging accessories.
## Comparison of Technological Characteristics with Predicate Device:
The claim of substantial equivalence of EYE-SYNC to the predicate device is based on a detailed comparison of several factors. This compairson is included in the table below:
| | Subject Device<br>(EYE-SYNC) | Predicate Device<br>(Eye-BOX) | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission<br>Number | K202927 | DEN170091 | Comparison |
| Product Code | QEA | QEA | SAME |
| Intended Use | Measure and<br>analyze eye<br>movement as an aid<br>in the diagnosis of<br>concussion | Measure and<br>analyze eye<br>movement as an aid<br>in the diagnosis of<br>concussion | SAME |
| Indications for<br>Use | The EYE-SYNC® is<br>intended for<br>recording, viewing,<br>and analyzing eye<br>movements in<br>support of identifying<br>visual tracking<br>impairment in human<br>subjects.<br>The EYE-SYNC® is<br>intended to record,<br>measure, and<br>analyze eye<br>movements as an | The EyeBOX is<br>intended to measure<br>and analyze eye<br>movements as an<br>aid in the diagnosis<br>of concussion within<br>one week of head<br>injury in patients 5<br>through 67 years of<br>age in conjunction<br>with a standard<br>neurological<br>assessment of | SIMILAR<br>Differences between the<br>subject device and predicate<br>include:<br>• Age range of population for<br>subject device is a small<br>subset of predicate.<br>• Population of subject device<br>is specific to sports-related<br>injury.<br>• Use of test is a shorter time<br>period since injury compared<br>to predicate.<br>The differences do not raise |
| | aid in the diagnosis<br>of concussion within<br>three days of sport-<br>related head injury<br>in patients 17 -24<br>years of age in<br>conjunction with a<br>standard<br>neurological<br>assessment, for use<br>by medical<br>professionals<br>qualified to interpret<br>the results of a<br>concussion<br>assessment<br>examination and aid<br>in the management<br>of concussion.<br><br>A negative EYE-<br>SYNC®<br>classification<br>corresponds to eye<br>movements that<br>are consistent with<br>a lack of<br>concussion.<br><br>A positive EYE-<br>SYNC®<br>classification<br>corresponds to eye<br>movements that<br>may be present in<br>patients with<br>concussion. | concussion.<br><br>A negative EyeBOX<br>classification may<br>correspond to eye<br>movement that is<br>consistent with a<br>lack of concussion.<br><br>A positive EyeBOX<br>classification<br>corresponds to eye<br>movement that may<br>be present in both<br>patients with or<br>without concussion. | new questions of safety or<br>effectiveness. |
| Rx Use Only | Yes | Yes | SAME |
| Hardware<br>Components | VR Headset<br>Tablet | Stand with<br>integrated Computer | DIFFERENT<br>Subject device uses VR<br>headset to capture gaze<br>measurements. Analysis is<br>done via a software app on the<br>accompanying tablet. This<br>method has been validated<br>with performance and clinical<br>testing to verify it provides the<br>same information as the |
| | | | |
| | | | predicate device and therefore<br>does not raise new questions<br>of safety and effectiveness. |
| Wireless<br>Communication | Bluetooth<br>connection between<br>VR Headset and<br>Tablet. Tablet has<br>4G cellular<br>connection Cloud. | Single, non-<br>networked device | DIFFERENT<br>Subject eye tracking device<br>communicates over Bluetooth<br>wireless to an administrative<br>tablet for control and data<br>exchange. The predicate<br>device achieves the same<br>functionality within a single<br>computer system.<br>Cellular connectivity to the<br>cloud is a convenience feature<br>and does not contribute to the<br>IFU. Bluetooth connectivity is a<br>common systems interface.<br>This method has been<br>validated with performance<br>and clinical testing to verify itprovides the same functionality<br>as the predicate device and<br>therefore does not raise new<br>questions of safety and<br>effectiveness. |
| | | | |
| Eye Tracking | Gaze | Gaze | SAME |
| Camera Speed | 60 Hz | 500 Hz | DIFFERENT<br>Eye tracking cameras can be<br>oversampled with respect to<br>the signal of interest. The<br>subject device has elected to<br>go with a lower sample rate,<br>but has validated this decision<br>with performance and clinical<br>testing to verify it provides the<br>same functionality as the<br>predicate device and therefore<br>does not raise new questions<br>of safety and effectiveness. |
| | | | therefore does not raise new<br>questions of safety and<br>effectiveness. |
| Software | Algorithm provides<br>analysis and results | Algorithm provides<br>analysis and results | SAME |
| Patient<br>Contacting<br>Materials | Yes - Foam cushion<br>on face mask. | N/A | Different<br>Patient contacting material is<br>biocompatible. Does not raise<br>new questions of safety or<br>effectiveness. |
| Sterility | Non-Sterile | Non-Sterile | SAME |
| EMC and<br>Electrical<br>Safety | IEC 60601-1:2012<br>IEC 60601-1-2:2014 | IEC 60601-1:2012<br>IEC 60601-1-2:2014 | SAME |
| Performance<br>Testing - Bench | Light Safety | Light Safety | SAME |
| Performance<br>Testing -<br>Clinical | 82% sensitivity<br>(74%, 89%)<br>93% specificity<br>(91%, 94%)<br>56% PPV***<br>(48%, 64%)<br>98% NPV***<br>(97%, 99%) | 80.4% sensitivity<br>(66.1%, 91.9%)<br>66.1% specificity<br>(59.7%, 72.1%)<br>31.6% % PPV<br>(23.3%, 40.9%)<br>94.5% NPV<br>(89.9%, 97.5%) | Similar<br>Results from clinical testing*<br>are substantially equivalent to<br>predicate device** |
| Special<br>Controls | Yes | Yes | SAME |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
*Clinical performance analysis did not include assessments with poor eye-tracking quality. In these cases, the device does not provide an assessment result.
**95% Confidence intervals noted for the subject and predicate device are derived from Clopper-Pearson method. 95% Confidence intervals for the subject device derived using the Wilson Score method for sensitivity are (74%, 88%) and (91%, 94%) for specificity.
*** PPV and NPV was obtained from observed study data without adjustment of the prevalence (10%).
## Performance Data
The following performance data were provided in support of the substantial equivalence determination.
{9}------------------------------------------------
#### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the EYE-SYNC device. The system complied with the IEC 60601-1:2012 Ed. 3.0 standard for operator and patient safety and IEC 60601-1-2:2014 Ed 4.0 standard for EMC in the intended use environment. The system complied with IEC 60601-2-57:2011 Ed1.0 Medical electrical equipment – Part 2 for safety of non-laser light source equipment intended for diagnostic use.
#### Software Verification and Validation Testing and Cybersecurity
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator, either directly or indirectly through incorrect or delayed information or through the action of a care provider.
Cybersecurity was assessed and documented according to the FDA Guidance, "Off-the-shelf Software - Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software."
## Performance Testing - Bench
Testing was performed to demonstrate conformance with the following standards:
- EN62471 Photobiological Safety of Lamps and Lamp Systems
## Biocompatibility Testing
The biocompatibility evaluation of the EYE-SYNC device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, "" May 1, 1005, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity .
- Sensitization ●
- Irritation
#### Sterility
The device is provided non-sterile. Cleaning instructions are provided in
{10}------------------------------------------------
the user manual.
## Summary of Clinical Testing:
#### Clinical performance
Clinical data was submitted to support substantial equivalence and conform to special controls. Validation data analysis was performed on 1,069 subjects. The validation study population included subjects ages 17-24 years actively engaged in competitive athletics. Of these 1,069 subjects, 107 subjects met the SCAT-5 clinical reference standard definition of concussion, where the evaluations were conducted by a healthcare practitioner blinded to EYE-SYNC device output +3 days from injury.
Retrospective analysis determined 82% (74%, 89%) sensitivity and 93% (91%, 94%) specificity of the EYE-SYNC test. Negative predictive value made by the classifying algorithm was 98% (97%, 99%) and positive predicative value was 56% (48%, 64%). For PPV and NPV reference, the study prevalence of clinical classification of concussion (mTBI) was 10% (107/1069). The analysis validates the clinical utility of the eye-tracking assessment as an aid in diagnosis of concussion within three days of sport-related head injury.
#### Test-retest reliability
The test-retest reliability of the EYE-SYNC metrics were reported through an independent study of 150 athletes from 11 sports with a mean (SD) age of 20 (1.3) years and 55% female population. The Intra-class correlation (ICC) was calculated with its 95%CI for each EYE-SYNC metric:
- SD tangential error: 0.86 (0.82, 0.90) .
- SD radial error: 0.78 (0.71, 0.84) .
- Mean phase error: 0.83 (0.77, 0.87) .
The ICC describes the level of correlation between two or more observations made on the same student athlete; a higher ICC (closer to 1) indicates greater reliability of the measure.
#### Conclusion:
The non-clinical and clinical data support the safety of the device and the software verification and validation demonstrate that SyncThink's EYE-SYNC device should perform as intended in the specified use conditions, and which are comparable to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.