K202926 · Shanghai Heartcare Medical Technology Co., Ltd. · QJP · Feb 10, 2022 · Cardiovascular
Device Facts
Record ID
K202926
Device Name
Micro Catheter
Applicant
Shanghai Heartcare Medical Technology Co., Ltd.
Product Code
QJP · Cardiovascular
Decision Date
Feb 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Micro Catheter is intended for selective delivery of therapeutic devices and infusion of contrast media into the peripheral and neuro vasculature.
Device Story
Single-lumen intravascular catheter; introduced over steerable guidewire into peripheral and neuro vasculature. Features proximal Luer adapter, lubricious distal coating, and distal radiopaque marker for fluoroscopic visualization. Used in clinical settings by physicians for selective delivery of therapeutic devices and contrast media infusion. Output is visual placement of device/media under fluoroscopy; assists in diagnostic/therapeutic vascular procedures. Benefits patient by enabling targeted access to complex vascular anatomy.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including surface evaluation, radiopacity, dimensional verification, liquid/air leakage, burst pressure, tensile force, torque strength/ability, kink resistance, trackability, particulate testing, coating integrity, tip flexibility, corrosion resistance, connector performance, and compatibility testing. Biocompatibility testing performed per ISO 10993 series.
Technological Characteristics
Single-lumen catheter; materials include PTFE (inner layer), Pebax (outer layer), Platinum-iridium (marker), Polycarbonate (handle), and Dymax 204-CTH-F (adhesive). Lubricious PVP coating. Sterilized via Ethylene Oxide (ISO 11135). Dimensions: 0.021/0.027 inch ID; 2.55/2.85 Fr distal OD. Complies with ISO 10555-1, ISO 594-1/2, ISO 80369-7, and ASTM standards for mechanical/packaging performance.
Indications for Use
Indicated for selective delivery of therapeutic devices and infusion of contrast media into the peripheral and neuro vasculature in patients requiring such diagnostic or therapeutic procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
February 10, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai Heartcare Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 China
Re: K202926
Trade/Device Name: Micro Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: January 8, 2022 Received: January 11, 2022
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K202926
Device Name Micro Catheter
Indications for Use (Describe)
The Micro Catheter is intended for selective delivery of therapeutic devices and infusion of contrast media into the peripheral and neuro vasculature.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of 21 CFR 807.92.
510(k) Number: __ K202926
Date of Preparation: 02/10/2022
#### 1. Sponsor Identification
### Shanghai Heartcare Medical Technology Co., Ltd.
590 Ruiqing Rd, Building 4, Suite 201, East Zhangjing High-Tech Park, Shanghai, P.R. China
Establishment Registration Number: Not registered
Contact Person: Zongyu Xue Position: RA Director Tel: +86-18621683501 Fax: +86-21-68798512 Email: zyxue@strokemedical.com
#### 2. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
#### 3. Identification of Proposed Device
Trade Name: Micro Catheter Common Name: Intravascular Catheter
Regulatory Information
{4}------------------------------------------------
Primary Product Code Classification Name: Catheter, Percutaneous, Neurovasculature Classification: II Product Code: QJP Regulation Number: 21CFR 870.1250 Review Panel: Neurology
Secondary Product Code Classification Name: Catheter, Percutaneous Classification: II Product Code: DQY Regulation Number: 21CFR 870.1250 Review Panel: Cardiovascular
#### Identification of Predicate Device and Reference Devices 4.
Predicate Device 510(k) Number: K192122 Product Name: Trevo Trak 21 Microcatheter
Reference Device 1 510(k) Number: K993672 Product Name: Rebar Micro Catheter
Reference Device 2 510(k) Number: K001966 Product Name: Rebar Micro Catheter
#### રં. Device Description
The proposed device Micro Catheter is a single-lumen catheter designed to be introduced over a steerable guidewire into the peripheral and neuro vasculature. The proximal end of the catheter incorporates standard Luer adapter to facilitate the attachment of accessories. The outer surface of the catheter has a lubricious coating at the distal end of the Micro Catheter has a radiopaque marker at the distal end to facilitate fluoroscopic visualization. The device is intended for single use and is provided sterile.
{5}------------------------------------------------
#### 6. Indications for Use
The Micro Catheter is intended for selective delivery of therapeutic devices and into the peripheral and neuro vasulature.
#### 7. Comparison of Technological Characteristics
| Item | Subject Device | Predicate Device K192122 | Reference Device 1 K993672 | Reference Device 2 K001966 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | II | II | II | II |
| Product Code | QJP, DQY | DQO, DQY | KRA | KRA |
| Regulation Number | 21 CFR 870.1250 | 21 CFR 870.1200<br>21 CFR 870.1250 | 21 CFR 870.1210 | 21 CFR 870.1210 |
| Indication for Use | The Micro Catheter is<br>intended for selective<br>delivery of therapeutic<br>devices and infusion of<br>contrast media into the<br>peripheral and neuro<br>vasculature. | The Microcatheter is indicated<br>for use in the selective<br>placement of devices and/or<br>fluids, such as contrast media,<br>into the peripheral, coronary,<br>and neuro vasculature during<br>diagnostic and/or therapeutic<br>procedures. | The Rebar Micro Catheter is<br>intended for the controlled<br>selective infusion of physician-<br>specified therapeutic agents or<br>contrast media into the<br>vasculature of the peripheral<br>and neuro anatomy. | The Rebar Micro Catheter is<br>intended for the controlled<br>selective infusion of physician-<br>specified therapeutic agents or<br>contrast media into the<br>vasculature of the peripheral<br>and neuro anatomy. |
| Component | Ruhr Handle<br>Outer Stress Tube<br>Inner Stress Tube<br>Coil Tube | Outer Jacket<br>Shaft Braid<br>Strain Relief<br>Inner Layer<br>Catheter Hub | Ruhr Handle<br>Outer Stress Tube<br>Inner Stress Tube<br>Coil Tube | Ruhr Handle<br>Outer Stress Tube<br>Inner Stress Tube<br>Coil Tube |
| Inner Diameter | 0.021, 0.027 inch | 0.021 inch | 0.027 inch | 0.021 inch |
| Proximal Outer<br>Diameter | 2.85 Fr (0.95 mm) | 2.85 Fr (0.95 mm) | 2.8 Fr (0.93 mm) | 2.8 Fr (0.93 mm) |
| Distal Outer Diameter | 2.55 Fr (0.85 mm),<br>2.85 Fr (0.95 mm) | 2.4 Fr (0.80 mm),<br>2.0 Fr (0.67 mm) | 2.8 Fr (0.93 mm) | 2.3 Fr (0.76 mm) |
| Effective length | 110, 130, 145, 153 cm | 162 cm | 110, 130, 145 cm | 110, 130, 153 cm |
| | | | | |
| Maximum Guidewire | 0.010 inch | 0.018 inch | 0.021 inch | 0.018 inch |
| Radiopaque Marker | Yes | Yes | Yes | Yes |
| Single Use | Yes | Yes | Yes | Yes |
| Materials | | | | |
| Inner Layer of Coil<br>Tube | Polytetrafluoroethylene | PTFE | Polytetrafluoroethylene | Polytetrafluoroethylene |
| Outer Layer of Coil<br>Tube | Pebax | Polyolefin | Pebax | Pebax |
| Coating | Polyvinylpyrrolidone (PVP) | Hydrophilic Coating | Hyaluronic Acid | Hyaluronic Acid |
| Radiopaque Marker | Platinum-iridium alloy | Platinum/iridium | Platinum-iridium alloy | Platinum-iridium alloy |
| Ruhr Handle | Polycarbonate (PC) | Polyurethane | unknown | unknown |
| Adhesive | Dymax 204-CTH-F | Acrylic (Acrylated Urethane) | unknown | unknown |
Table 1 Comparison of Technological Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
#### 8. Non-Clinical Performance Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and was substantially equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-3: 2014 Biological evaluation of medical devices-Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- > ISO 10993-4: 2017 Biological evaluation of medical devices-Part 4: Selection of tests for interactions with blood
- > ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitrocytotoxicity
- > ISO 10993-7: 2008 Biological evaluation of medical devices-Part 7: Ethylene oxide sterilization residuals
- > ISO 10993-10: 2010 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
- > ISO 10993-11: 2017 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity
- > USP <85> Bacterial Endotoxins Test
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- > ASTM F88/88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- > ISO 594-1: 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
- > ISO 594-2: 1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
- > ISO 10555-1: 2013 Intravascular catheters - Sterile and single-use intravascular catheters - Part 1: General requirements
- > ISO 80369-7: 2106 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
The results of verification and validation testing conducted on the Micro Catheter demonstrate that it performs as designed and is suitable for its intended use. A summary of the tests performed is provided in the table below:
| Test | Test Method Summary | Results and Conclusion |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Surface Evaluation | Observe the catheter by eye or under microscope at X 2.5 magnification for any structural or mechanical damage. | Surface met acceptance criteria. |
| Radiopacity | The radiopaque marker on the catheter tip should be visible under X - ray. | The radiopaque marker on the catheter tip is visible under X-ray. |
| Dimensional<br>verification | Verify dimensions using specified<br>measurement tools. Record measurements. | Size verification met acceptance<br>criteria. |
| Liquid leakage | The proposed device was evaluated per ISO<br>10555-1 to demonstrate that the device<br>meets the liquid leakage under pressure<br>requirements. | No liquid leakage. |
| Static burst pressure | Burst pressure tests were performed at<br>pressures greater than the manual syringe<br>injection pressures. | Burst pressure met acceptance<br>criteria. |
| Peak tensile force | Use a tensile test machine to apply a tensile<br>load to the sample and determine whether the<br>maximum tensile force meets the acceptance<br>criteria. | Peak tensile force met acceptance<br>criteria. |
| Torque Strength | Fix the distal end of the catheter and rotate<br>the proximal end until the failure. Record the<br>number of rotation and the failure mode. | Torque strength met acceptance<br>criteria. |
| Torqueability | Rotate the proximal end of the catheter 90<br>degrees to observe the distal end of the<br>catheter and calculate a proximal-to-distal<br>rotational ratio for the device. | Torqueability met acceptance<br>criteria. |
| Kink resistance | The proposed device was evaluated per<br>FDA guidance “Non-Clinical Engineering<br>Tests and Recommended Labeling for<br>Intravascular Stents and Associated<br>Delivery Systems", section IV, C.9. | Kink resistance met acceptance<br>criteria. |
| Trackability | The proposed device was evaluated per FDA<br>guidance "Non-Clinical Engineering Tests<br>and Recommended Labeling for<br>Intravascular Stents and Associated<br>Delivery Systems", section IV, C.2. | Trackability met acceptance<br>criteria. |
| Particulate testing | After simulated use testing with compatible<br>devices, determine the quantity and size of<br>the particles generated. | The number and size of the<br>particles were similar to that of<br>the predicate device. |
| Coating integrity | After simulated use testing with compatible<br>devices, the coating is dyed and checked for<br>coating defects at X 500 magnification. | Coating integrity met<br>acceptance criteria. |
| Tip flexibility | The proposed device was evaluated per FDA guidance “Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling”, section IV, G.15. | Tip flexibility met acceptance criteria. |
| Air leakage | The proposed device was evaluated per ISO 10555-1 to demonstrate that the device meets the hub aspiration air leakage requirements. | No air leakage. |
| Corrosion resistance | The proposed device was evaluated per ISO 10555-1 to demonstrate that the device is corrosion resistant. | Corrosion resistance met acceptance criteria. |
| Connector performance | The proposed device was evaluated per ISO 594-1, 594-2 and 80369-7 to demonstrate that the device meets the requirements for small bore connectors. | Connector performance met acceptance criteria. |
| Compatibility test | Simulated use testing with compatible devices in a vascular model was performed. | The device can be used as intended. |
{8}------------------------------------------------
{9}------------------------------------------------
## Biocompatibility
The device is categorized as Externally Communicating Device, Circulating Blood, Limited Contact (< 24 hours), per ISO 10993-1, the following testing was conducted:
| Test | Reference Standard | Test Summary | Conclusion |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Cytotoxicity | Tested in accordance with ISO<br>10993-5:2009, Biological<br>Evaluation of Medical Devices –<br>Part 5: Tests for in vitro toxicity. | Verify the viability, if<br>viability is reduced to < 70%<br>of the blank, it has a<br>cytotoxic potential. | Non-cytotoxic |
| Irritation | Tested in accordance with ISO<br>10993-10:2010, Biological<br>Evaluation of Medical Devices –<br>Part 10: Tests for Irritation and<br>Skin Sensitization. | Study animals tested with the<br>subject device were observed<br>for dermal sensitization. | No irritation |
| Sensitization | Tested in accordance with ISO<br>10993-10, Biological Evaluation<br>of Medical Devices – Part 10<br>Tests for Irritation and Skin<br>Sensitization. | Study animals tested with the<br>subject device were observed<br>for dermal sensitization. | No sensitization |
| Systemic Toxicity | Tested in accordance with ISO<br>10993-11:2017,<br>Biological<br>Evaluation of Medical Devices –<br>Part 11: Tests for Systemic<br>Toxicity. | Study animals tested with the<br>subject device were observed<br>for abnormal clinical signs<br>indicative of toxicity during<br>the 72-hour test period. | No systemic toxicity |
| Hemolysis | Tested in accordance with ASTM<br>F756-17, Standard Practice for<br>Assessment of Hemolytic<br>Properties of Materials. | The difference between the<br>hemolytic indexes of the<br>subject device and the<br>negative control was<br>evaluated. | Non-hemolytic |
| Material Mediated<br>Pyrogenicity | Tested in accordance with ISO<br>10993-11:2017,<br>Biological<br>Evaluation of Medical Devices –<br>Part 11: Tests for Systemic<br>Toxicity and USP 42 <151><br>Pyrogen Test. | Study animals were<br>observed for temperature<br>rise. | Nonpyrogenic |
| In<br>Vivo<br>Thromboresistance | Tested in accordance with ISO<br>10993-4:2017,<br>Biological<br>Evaluation of Medical Devices –<br>Part 4: Selection of Tests for<br>Interactions with Blood. | Study animals with subject<br>device were observed for<br>thrombogenic potentials and<br>test results demonstrated<br>similar thromboresistance<br>characteristics with the<br>control device. | Met the<br>predetermined<br>acceptance criteria,<br>the test score is 0,<br>i.e., No thrombosis. |
| Complement<br>Activation | Tested in accordance with ISO<br>10993-4, Biological Evaluation<br>of Medical Devices – Part<br>4:2017: Selection of Tests for<br>Interactions with Blood, SC5b-9<br>Complement Activation. | Comparison of the subject<br>device SC5b-9 value to the<br>predicate device for all<br>exposure times was<br>performed. | No statistical<br>difference from the<br>predicate device |
| Partial<br>Thromboplastin Time | Tested in accordance with ISO 10993-4:2017, Biological<br>Evaluation of Medical Devices –<br>Part 4: Selection of Tests for<br>Interactions with Blood, and<br>ASTM F2382-2018, Standard<br>Test Method for Assessment of<br>Intravascular Medical Device<br>Materials on Partial<br>Thromboplastin Time. | The clotting time was<br>observed for both the subject<br>device and the predicate. | No statistical<br>difference from the<br>predicate device. |
| Bacterial Reverse<br>Mutation | Tested in accordance with ISO<br>10993-3:2014 Biological<br>evaluation of medical devices-<br>Part 3: Tests for genotoxicity,<br>carcinogenicity and reproductive<br>toxicity. | The difference between the<br>number of revertant colonies<br>of the subject device and the<br>negative control was<br>evaluated. | No backward<br>mutation in<br>salmonella<br>typhimurium. |
| In Vitro Mammalian<br>Cell Gene Mutation | Tested in accordance with ISO<br>10993-3:2014 Biological<br>evaluation of medical devices-<br>Part 3: Tests for genotoxicity,<br>carcinogenicity and reproductive<br>toxicity. | The difference between the<br>mutation frequency of the<br>subject device and the<br>negative control was<br>evaluated. | Non-mutagenic to<br>TK+/--3.7.2C<br>Subline of L5178Y<br>cell. |
{10}------------------------------------------------
{11}------------------------------------------------
## Sterilization and Shelf Life
The Micro Catheter sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10th. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 2.15 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 2 year shelf-life.
#### 9. Clinical Performance Testing
No clinical studies were necessary to demonstrate substantial equivalence.
### 10. Conclusion
The subject device has similar technological characteristics and intended use as the predicate device. The differences do not raise new questions of safety and effectiveness. The biocompatibility, sterility and bench performance testing demonstrate that the Micro Catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.