The OsteoSync™ Ti Dental Mesh is intended for stabilization and support of bone grafts in dento-alveolar bony defect sites.
Device Story
OsteoSync™ Ti Dental Mesh is a system of dental membranes used for permanent fixation to aid in reconstruction and augmentation of alveolar ridges of maxilla and mandible. Device consists of Grade 2 titanium mesh (ASTM F67) available in various shapes and sizes. Mesh is contoured to accommodate patient-specific anatomy and used to envelop bone grafts, with optional fastener attachment. Used in clinical settings by dental professionals. Provides structural support for bone grafts; helps maintain graft volume during healing. Benefits include reduced need for intraoperative manipulation of generic mesh due to preformed shapes and insert diameters. Sterilized via gamma radiation.
Clinical Evidence
No human clinical data. Substantial equivalence supported by non-clinical bench testing (mechanical testing per ASTM F564, sterilization validation per ISO 11137, shelf-life validation, bacterial endotoxin testing per ANSI/AAMI ST72) and preclinical animal studies in a canine model (leveraged from an FDA-authorized master file).
Technological Characteristics
Material: Grade 2 titanium (ASTM F67). Design: Dental mesh, various shapes/sizes. Thickness: 0.41-0.52mm. Pore diameter: 0.43-0.66mm. Surface: Smooth/smooth or smooth/rough. Stabilization: Envelopment of bone graft with optional fastener. Sterilization: Gamma radiation (SAL 10^-6).
Indications for Use
Indicated for stabilization and support of bone grafts in dento-alveolar bony defect sites in adults.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
Sites Medical, LLC % Karen Warden President BackRoads Consulting Po Box 566 Chesterland, Ohio 44026
July 14, 2021
Re: K202918
Trade/Device Name: OsteoSync™ Ti Dental Mesh Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: June 15, 2021 Received: June 16, 2021
Dear Karen Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name OsteoSync™ Ti Dental Mesh
#### Indications for Use (Describe)
The OsteoSync™ Ti Dental Mesh is intended for stabilization and support of bone grafts in dento-alveolar bony defect sites.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date: | 14 July 2021 |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Sites Medical, LLC<br>5865 E State Road 14<br>Columbia City, IN 46725<br>Office: 260.625.3347 |
| Sponsor Contact: | Greg Stalcup, President/CEO |
| 510(k) Contact: | Karen E. Warden, PhD<br>BackRoads Consulting Inc.<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457 |
| Trade Name: | OsteoSync™ Ti Dental Mesh |
| Common Name: | Bone plate |
| Regulatory Class: | Class II |
| Regulation Name,<br>Regulation, Product<br>Code: | Bone plate, 872.4760, JEY |
| Device Description: | The OsteoSync™ Ti Dental Mesh is a system of dental membranes intended<br>for permanent fixation in adults and designed to aid in the reconstruction and<br>augmentation of the alveolar ridges of the maxilla and mandible. They are<br>available in a variety of sizes and shapes and can be contoured to<br>accommodate the individual anatomic and clinical circumstances of each<br>patient. The implants are sold sterile. |
| Indications for Use: | The OsteoSync™ Ti Dental Mesh is intended for stabilization and support of<br>bone grafts in dento-alveolar bony defect sites. |
| Materials: | The OsteoSync™ Ti Dental Mesh implants are manufactured from Grade 2<br>titanium as described by ASTM F67. |
| Primary Predicate: | Neo Titanium mesh and CTi-mem (neobiotech Co., Ltd – K111761) |
| Reference Devices: | SMARTbuilder System (OSSTEM Implant Co., Ltd – K130840), TraumaOne<br>(Biomet Microfixation - the K081067), BoniPlus Dental G-Mesh System<br>(MontJade Engineering Co., Ltd – K182759) |
| Non-clinical Testing: | Mechanical testing - worst case devices were evaluated per ASTM F564.<br>Biocompatibility evaluation was leveraged from an FDA authorized master file.<br>Sterilization validation was performed using gamma radiation and Method<br>VDmax, in accordance with ISO 11137-1 and ISO 11137-2.<br>Shelf life validation - sterile integrity was verified by visual, seal strength and<br>seal integrity evaluations in accordance with ASTM F1886, ASTM F88, ASTM<br>F1929 and ASTM F2096.<br>Pyrogenicity - Bacterial endotoxin testing (specifically Limulus amebocyte<br>lysate, LAL) was performed per ANSI/AAMI ST72.<br>Preclinical animal studies in a canine model were relied upon to establish<br>substantial equivalence. |
| Technological<br>Characteristics: | The OsteoSync™ Ti Dental Mesh has many of the same technological<br>characteristics as the predicate devices. These include anatomic location,<br>material of manufacture, basic design, method of stabilization, sterile<br>condition, variety of components offered and their general dimensional<br>characteristics. The similarities and differences are provided in the following<br>tables. |
{4}------------------------------------------------
| Overall system | Subject Device | Primary<br>Predicate | Reference Devices | | |
|----------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Features<br>System → | OsteoSync™ Ti<br>Dental Mesh | Neo Titanium<br>mesh & CTI-<br>mem | SMARTbuilder<br>System | BoniPlus Dental<br>G-Mesh System | TraumaOne |
| 510(k) | K202918 | K111761 | K130840 | K182759 | K081067 |
| Manufacturer: | Sites Medical<br>LLC | neobiotech Co.,<br>Ltd. | OSSTEM<br>Implant Co.,<br>Ltd. | MontJade<br>Engineering<br>Co., Ltd. | Biomet<br>Microfixation,<br>Inc. |
| Technological<br>characteristics | | | | | |
| Basic design: | Dental mesh,<br>various shapes | Dental mesh,<br>various shapes | Dental mesh,<br>various shapes | Dental mesh,<br>various shapes | Dental mesh<br>and plates,<br>various shapes |
| Material of<br>manufacture: | ASTM F67, Gr2 | ASTM F67, Gr2 | ASTM F67, Gr2 | ASTM F67, Gr2 | ASTM F67, Gr2<br>& Gr4 |
| Method of<br>stabilization: | Envelopment of<br>bone graft with<br>optional<br>fastener<br>attachment | Envelopment of<br>bone graft with<br>optional<br>fastener<br>attachment | Envelopment of<br>bone graft with<br>optional<br>fastener<br>attachment | Envelopment of<br>bone graft with<br>optional<br>fastener<br>attachment | Envelopment of<br>bone graft with<br>optional<br>fastener<br>attachment |
| Condition when<br>used: | Sterile | Sterile | Sterile | Sterile | Sterile |
| System<br>characteristics: | | | | | |
| Thickness: | 0.41 to 0.52mm | 0.07 &<br>0.085mm | 0.1mm | 0.1 to 0.3mm | 0.2 to 0.6mm |
| Surface: | Smooth/smooth<br>or<br>smooth/rough<br>sides | Smooth on both<br>sides | Smooth on both<br>sides | Smooth on both<br>sides | Smooth on both<br>sides |
| Pore Diameter: | 0.43 to 0.66mm | 0.4 to 0.8mm | 0.6 to 1.0mm | Unknown | Unknown |
Comments:
The thickness of the subject device is greater than that of the Neo Titanium & CT-mem and the . SMARTbuilder systems but only fractionally larger than the largest BoniPlus implants. The thickness of the subject device is within the range offered by TraumaOne.
The pore diameter size of the subject devices lies within those specified by the predicates. .
The surface texture of the subject device and cited predicates is smooth on both sides. However the . subject device additionally offers a smooth gingival side / rough bone side option. The roughened side is intended to assist in securing the mesh to bone. Preclinical animal studies in a canine model provided support for this design. The preclinical animal studies were leveraged from an FDA-authorized master file.
{5}------------------------------------------------
| Oval/round<br>Features↓<br>System → | Subject Device | Primary Predicate | Reference Device |
|-------------------------------------|----------------------------------|------------------------------------|----------------------------|
| 510(k) | K202918 | K111761 | K130840 |
| Manufacturer: | Sites Medical LLC | neobiotech Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Overall dimensions | Image: OsteoSync™ Ti Dental Mesh | Image: Neo Titanium mesh & CTi-mem | Image: SMARTbuilder System |
| Insert diameter | 2 to 6mm | 2.5mm | ~3mm (approximated) |
| Proximal width | 5 to 12mm | 4 to 15mm | 4 to 12mm |
| Buccal width | 4 to 8mm | 8 to 15mm | 7 to 12mm |
| Buccal length | 5 to 8mm | 6 to 20mm | 3 to 9mm |
Comments:
- . The insert diameter of the subject device encompasses the ranqe offered by the Neo Titanium & CTi-mem and the SMARTbuilder systems but extends to a larger range to broaden the system functionality for the end user by minimizing the need for manipulation of generic mesh.
- The buccal width of the subject device overlaps those provided in the Neo Titanium & CTi-. mem and the SMARTbuilder systems but extends narrower to accommodate smaller defects such that the end user does not have to trim a generic mesh to obtain the implant form.
| Posterior Graft | Subject Device | Primary<br>Predicate | Reference Devices | |
|-----------------------|----------------------------------|------------------------------------|--------------------------------------|-------------------------------|
| Features<br>System → | OsteoSync™ Ti<br>Dental Mesh | Neo Titanium<br>mesh & CTi-mem | BoniPlus Dental<br>G-Mesh System | TraumaOne |
| 510(k) | K202918 | K111761 | K182759 | K081067 |
| Manufacturer: | Sites Medical<br>LLC | neobiotech Co.,<br>Ltd. | MontJade<br>Engineering Co.,<br>Ltd. | Biomet<br>Microfixation, Inc. |
| Overall<br>dimensions | Image: OsteoSync™ Ti Dental Mesh | Image: Neo Titanium mesh & CTi-mem | Image: BoniPlus Dental G-Mesh System | Image: TraumaOne |
| Insert diameter | 2 to 6mm | NA | NA | NA |
| Proximal width | 15 to 34mm | 12 to 35mm | 25 to 60mm | 80 to 85mm |
| Buccal width | 10 to 20mm | 12 to 35mm | 17 to 60mm | 80 to 85mm |
None of the dimensional differences in the individual mesh shapes raise new questions of substantial equivalence.
{6}------------------------------------------------
| Posterior Graft | Subject Device | Primary<br>Predicate | Reference Devices |
|--------------------------|------------------------------|--------------------------------|--------------------------------------------|
| Features<br>System → | OsteoSync™ Ti<br>Dental Mesh | Neo Titanium<br>mesh & CTi-mem | BoniPlus Dental<br>G-Mesh System TraumaOne |
| Buccal length<br>(total) | 15 to 24mm | 15 to 30mm | NA 50 to 53mm |
Comments:
- None of the predicate devices have a preformed insert diameter slot to accommodate a . dental implant. However, all of the mesh shapes can be trimmed to form and perforated to accommodate a screw. The option broadens the functionality for the end user by minimizing the need for manipulation of generic mesh.
- . The proximal width, buccal width and buccal length dimensions of the subject device lie within the range offered by the Neo Titanium & CTi-mem, BoniPlus G-mesh and the TraumaOne mesh systems.
None of the dimensional differences in the individual mesh shapes raise new questions of substantial equivalence.
| Socket Graft | Subject Device | Primary<br>Predicate | Reference Devices | |
|-----------------------|---------------------------------|------------------------------------|-----------------------------|--------------------------------------|
| Features<br>System → | OsteoSync™ Ti<br>Dental Mesh | Neo Titanium<br>mesh & CTi-mem | SMARTbuilder<br>System | BoniPlus Dental<br>G-Mesh System |
| 510(k) | K202918 | K111761 | K130840 | K182759 |
| Manufacturer: | Sites Medical<br>LLC | neobiotech Co.,<br>Ltd. | OSSTEM Implant<br>Co., Ltd. | MontJade<br>Engineering Co.,<br>Ltd. |
| Overall<br>dimensions | Image: OsteoSync Ti Dental Mesh | Image: Neo Titanium mesh & CTi-mem | Image: SMARTbuilder System | Image: BoniPlus Dental G-Mesh System |
| Insert diameter | 2 to 6mm | 2.5mm | 3.5-4mm | ~3mm<br>(approximated) |
| Proximal width | 5 to 9.3mm | 4 to 12mm | 4 to 12mm | ~5mm<br>(approximated) |
| Buccal Length | 10 to 13.5mm | 6 to 10mm | 7 & 9mm | 9.5mm |
| Buccal Width | 7 to 29mm | 8 to 12mm | 8 to 12mm | 9mm |
Comments:
- The insert diameter of the subject device encompasses the range offered by the Neo . Titanium & CTi-mem and the SMARTbuilder systems but extends to a larger range to broaden the system functionality for the end user by minimizing the need for manipulation of generic mesh.
- . The buccal width and length of the subject device overlap but extend longer than the range offered by the cited predicates to accommodate larger defects as necessary.
- . The proximal width of the subject device lies within the range offered by the Neo Titanium & CTi-mem system.
- . None of the dimensional differences in the individual mesh shapes raise new questions of substantial equivalence.
{7}------------------------------------------------
| Conclusion: | The OsteoSync™ Ti Dental Mesh possesses the same intended use and<br>technological characteristics as the predicate devices.<br>Nonclinical performance testing demonstrated substantial equivalence.<br>Specifically, |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | A biological risk assessment showed the submitted information and the<br>preclinical animal studies provided in an FDA authorized master file<br>mitigated risks with respect to biocompatibility. |
- The sterilization validation achieved the predetermined acceptance . criteria for a sterility assurance level (SAL) of 10-6.
- The shelf-life validation met the predetermined acceptance criteria for a . five year shelf-life.
- The bacterial endotoxin testing results met the predetermined pyrogen ● limit of 20EU/device.
Therefore the OsteoSync™ Ti Dental Mesh is substantially equivalent for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.