The SternaFuse® Fixation System is indicated for use in the stabilization and fixation of fractures of the anterior chest wall, including sternotomy, sternal fractures, and sternal reconstructive surgical procedures, to promote fusion. The device is for prescription use only.
Device Story
SternaFuse Fixation System consists of 316LVM stainless steel plates, links, and screws for anterior chest wall stabilization. Implants feature centralized saddles for intra-operative contouring, emergency re-entry cutting, and stainless steel wire placement. System includes self-drilling locking screws (3.0mm/3.3mm diameters; 10mm-16mm lengths) and single-use disposable surgical instrumentation. Used by surgeons in clinical settings for sternal fixation following sternotomy, fractures, or reconstructive procedures. Implants are provided sterile, single-use, and customizable to patient anatomy. Device promotes bone fusion by providing rigid fixation of chest wall segments.
Clinical Evidence
No clinical data or animal studies performed. Substantial equivalence supported by bench testing, including static/dynamic tension testing, screw pullout, driving torque, and torsion testing per ASTM F543. Biocompatibility evaluated per ISO 10993-1; sterility validated per ISO 11137 and ISO 11607-1.
Indicated for stabilization and fixation of anterior chest wall fractures, including sternotomy, sternal fractures, and sternal reconstructive procedures to promote fusion in patients requiring surgical chest wall repair.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Biomet Microfixation Sternal Closure System (K111908)
{0}------------------------------------------------
January 26, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health and Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION".
Fusion Innovations, LLC % Katelyn Jessup Regulatory & Quality Specialist Kapstone Medical, LLC 520 Elliot Street Charlotte, North Carolina 28202
Re: K202914
Trade/Device Name: SternaFuse Fixation System Regulation Number: 21 CFR 888.3030, 888.3040 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 20, 2022 Received: January 21, 2022
Dear Katelyn Jessup:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202914
Device Name SternaFuse® Fixation System
Indications for Use (Describe)
The SternaFuse® Fixation System is indicated for use in the stabilization of fractures of the anterior chest wall, including stemal fixation following sternal fractures, and sternal reconstructive surgical procedures, to promote fusion. The device is for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR 807.92, the following summary of information is provided.
## 1. Date of Preparation:
January 26, 2022
## 2. Applicant
Fusion Innovations, LLC 341 N. Maitland Ave. Ste. 270 Maitland, FL 32751 USA
### 3. Submitter/Contact Person
Katelyn Jessup Regulatory and Quality Specialist Kapstone Medical LLC 520 Elliot St. Charlotte, NC 28202 (856) 577-3432
### 4. Device Name
Trade Name: SternaFuse® Fixation System Common Name: Plate, Fixation, Bone (primary), and Screw Fixation, Bone Classification Name: Single/multiple component metallic bone fixation appliances and accessories (primary), Smooth or Threaded Metallic Bone Fixation Fastener Regulation Number: 888.3030 (primary), 888.3040 Product Code: HRS (primary), HWC Classification: Class II Panel: Orthopedic
### 5. Predicate Devices
Biomet Microfixation Sternal Closure System (K111908) - Primary Ethi-pack Surgical Stainless Steel Suture (K931271) - Reference
{4}------------------------------------------------
## 6. Indications for Use (Intended Purpose and Conditions of Use):
The SternaFuse® Fixation System is indicated for use in the stabilization and fixation of fractures of the anterior chest wall, including sternal fixation following sternotomy, sternal fractures, and sternal reconstructive surgical procedures, to promote fusion. The device is for prescription use only.
## 7. Device Description
The SternaFuse® Fixation System implants are composed of 316 LVM implant quality stainless steel plates, links, and screws intended to stabilize and fixate fractures of the anterior chest wall. The components include various sizes to facilitate customization according to the requirements of the anterior chest wall repair. Self-drilling locking screws come in two diameters, 3.0mm and 3.3mm, and lengths of 10mm, 13mm, and 16mm. Multiple plates may be used in one anterior chest wall repair. Variable Assemblies can be configured, and the Links angled for optimum fixation. Implants are designed with centralized saddles to aid in intra-operative contouring, to facilitate cutting during postoperative emergent re-entry, and to provide a location for stainless steel wire.
The implants are individually packaged and sterilized, allowing the surgeon to customize the desired implants for the specific needs of the patient. All implants are provided sterile with a five-year shelf-life. The implants should never be reused under any circumstance.
### 8. Comparison of Technological Characteristics with the Predicate Devices
As was established in this submission, the subject device SternaFuse® Fixation System is substantially equivalent to the predicate device, cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate device through comparison in areas including design, intended use, material composition, function, and sterility.
### 9. Performance Data
There are no clinical data generated and held by the manufacturer, i.e., no premarketing or post-market clinical studies or animal studies have been performed. The following information is provided in support of substantial equivalence.
{5}------------------------------------------------
#### Biocompatibility 9.1
The SternaFuse® Fixation System implants are classified as an implant device with tissue/bone contact and permanent contact. Therefore, according to ISO 10993-1, the biological evaluation was assessed for potential effects. The SternaFuse® Fixation System devices are manufactured from 316LVM Stainless Steel conforming to ASTM standard F138. The evaluation was based on a risk assessment of the device considering materials, manufacturing process, and finished device testing. The results show that implants made of 316LVM Stainless Steel have a high demonstratable biological safety. No concerns arose that would preclude clinical use of the SternaFuse® Fixation System. The requirements of the ISO 10993 standard are fulfilled.
#### 9.2 Pyrogenicity / Endotoxin Testing
The bacterial endotoxin test was performed utilizing worst case subject implants to verify that the subject implants meet the acceptance criteria of ≤ 20 EU/device. Testing was successfully performed, and it was confirmed that the SternaFuse® Fixation System meets the acceptance criteria of ≤ 20 EU/device according to USP. General Chapter <85>, Bacterial Endotoxins Test, and EP Ch 2.6.14 Bacterial Endotoxins as recommended in ISO 10993-11:2009 Biological Evaluation of Medical Devices, Part 11: Tests for Systemic Toxicity. Endotoxin testing will be conducted on every lot.
#### 9.3 Packaging/Sterility
The SternaFuse® Fixation System implants are supplied sterile to the end user, packaged in a container within a pouch. The pouch serves as the sterile barrier and is intended to maintain sterility of the packaged medical device. The pouch has undergone validation, in accordance with ISO 11607-1, to demonstrate the sterile barrier ability after gamma irradiation and after the claimed shelf-life. A transport simulation was performed according to ASTM D4169. Sterilization was performed according to ISO 11137.
The SternaFuse® Fixation System also includes single-use, disposable surgical instrumentation to assist with implantation. Driver bits are provided sterile, packaged in the same container and pouch described above.
#### 9.4 Mechanical Testing
The following testing was performed to demonstrate equivalence:
- Static Tension Testing ●
- Dynamic Tension Testing ●
{6}------------------------------------------------
- Screw Pullout Testing per ASTM F543 ●
- Driving Torque Testing per ASTM F543 ●
- Torsion Testing per ASTM F543 ●
#### MR Safety Testing 9.5
The SternaFuse® Fixation System implants are manufactured using non-ferromagnetic 316 Stainless Steel. The implants have not been evaluated for safety, heating, migration, or compatibility in the MR environment. The SternaFuse® Fixation System has not been tested for heating, migration, or image artefact in the MR environment. The safety of the SternaFuse® Fixation System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.
### 10. Conclusion
The SternaFuse® Fixation System and the predicate device Biomet Microfixation Sternal Closure System (K111908) have the same "Indications for Use," and are available by prescription only. Compared to the reference device, the SternaFuse® Fixation System is manufactured from the same material and biocompatible. Any technical differences, which were identified, do not result in new questions of safety or effectiveness.
Through assessment of technological characteristics, indications for use and performance data, it can be concluded that SternaFuse® Fixation System is both a safe and effective device and is substantially equivalent to the Biomet Microfixation Sternal Closure System and Ethi-pack Surgical Stainless Steel Suture.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.