The EndoPilot2 systems are dental devices which combine in a single control unit an endo motor to clean the root canal, a dental obturator to fill and pressurize, an electronic apex locator to locate the file tip in the rot canal and an ultrasonic-handpiece for root-canal cleaning and preparation. The EndoPilot is intended solely for use by trained dental professional health care facilities on patients that need root-canal-treatment.
Device Story
EndoPilot2 is a standalone AC-powered dental control unit with a touch display, designed for use by dental professionals in clinics. It integrates multiple handpieces for root canal procedures: an endo motor for cleaning, an obturator for filling/pressurizing, an electronic apex locator for file tip positioning, and an ultrasonic handpiece for cleaning/preparation. The apex locator functions via electrical impedance changes. The device uses a modular design allowing various handpieces to connect to the central unit. It features a wireless footswitch for activation. The clinician operates the system via the touchscreen to select functions and monitor parameters. By consolidating these tools into one unit, the device streamlines endodontic workflows, potentially improving procedural efficiency and accuracy in root canal treatments.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993), electrical safety (IEC 60601-1, IEC 80601-2-60), EMC (IEC 60601-1-2), reprocessing validation (ISO 17664/17665), mechanical coupling (ISO 3964/14457), software V&V, usability engineering (IEC 62366-1), and risk management (ISO 14971).
Technological Characteristics
Standalone AC-powered control unit with LCD touchscreen. Components: endo motor, obturator, apex locator, ultrasonic handpiece. Materials: stainless steel, POM-C, USP Plastic Class VI, TPR, PVC. Connectivity: Bluetooth 4.1 (2.402-2.480 GHz). Power: 100-240V AC, 50/60Hz; 12V DC. Ultrasonic frequency: 27-33 kHz. Apex locator accuracy: < ±0.5 mm. Sterilization: per ISO 17664/17665-1.
Indications for Use
Indicated for patients requiring root-canal treatment. Intended for use by trained dental professionals in professional health care facilities.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 13, 2021
Schlumbohm GmbH & Co. KG % Oliver Eikenberg, PhD Senior Consultant QA/RA Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K202906
Trade/Device Name: EndoPilot2 Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC, EKX, EKR, LQY Dated: March 2. 2021 Received: March 9, 2021
Dear Oliver Eikenberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav. therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael Adjodha Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202906
Device Name EndoPilot2
Indications for Use (Describe)
The EndoPilote systems are dental devices which combine in a single control unit an endo motor to clean the root canal, a dental obturator to fill and pressurize, an electronic apex locator to locate the file tip in the rot canal and an ultrasonic-handpiece for root-canal cleaning and preparation.
The EndoPilot is intended solely for use by trained dental professional health care facilities on patients that need root-canal-treatment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over The Counter Use (21 CFR 801 Subpart C) |
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K202906
# EndoPilot2
#### 1. Submission Sponsor
Schlumbohm GmbH & Co. KG Klein Floyen 8-10 D-24616 Brokstedt, Schleswig-Holstein , GERMANY Phone: +49-(0)4324-8929-0 Contact: Gerald Schlumbohm , Managing Director E-mail: gs@schlumbohm.de
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Oliver Eikenberg, PhD Title: Senior Consultant, Quality & Regulatory Affairs
#### 3. Date Prepared
April 12, 2021
#### 4. Device Identification
#### Trade/Proprietary Name: EndoPilot2
| Common/Usual Name: | Dental hand instrument |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Ultrasonic Scaler |
| Regulation Number: | 21 CFR 872.4850 |
| Product Code: | ELC Ultrasonic scaler<br>EKX direct drive, AC-powered handpiece<br>EKR endodontic plugger, root canal<br>LQY root apex locator (unclassified), |
| Class: | Class II |
| Classification Panel: | Dental |
{4}------------------------------------------------
#### 5. Legally Marketed Predicate Device(s)
For the Endodontic unit systems (applicable to all EndoPilot²model variants): Primary Predicate
Device name: EMS-200 510(k) number: K153285 Manufacturer: Meta Systems Co., Ltd., Korea
For the Ultrasonic Unit (applicable only to EndoPilot²model variant "ultra" and " ultra plus"): Reference Device
Device name: SUPRASSON P5 NEWTRON 510(k) number: K050895 Manufacturer: SATELEC-ACTEON Group
#### 6. Indication for Use Statement
The EndoPilot² systems are dental devices which combine in a single control unit an endo motor to clean the root canal, a dental obturator to fill and pressurize, an electronic apex locator to locate the file tip in the root canal and an ultrasonic-handpiece for root-canal cleaning and preparation.
The EndoPilot² is intended solely for use by trained dental professionals in professional health care facilities on patients that need root-canal-treatment.
#### 7. Device Description
The EndoPilot2 systems, including the models "comfort", "plus", "ultra plus" and the model variant under brand "CanalPro Jeni" (marketed by Distributor Coltène/Whaledent Inc. and identical for technical aspects to model EndoPilot² comfort) are standalone AC-powered dental control units with a touch display to which multiple hand-held dental handpieces for root canal preparation (Apex measurement, EndoMotor, Ultrasonic handpiece) and/or root-canal filling (DownPack, BackFill handpiece) can be connected.
These multifunctional devices are intended for use by professionals in the dental clinic use environment. Based on the modular concept of these EndoPilot2 systems different handpieces can be combined with the control unit and different device variants (systems) of the EndoPilot²exist. The key hand-held components like handpieces or endodontic tools of the EndoPilot² are medical devices commercially available by themselves and have separate FDA registration or clearance for marketing in the US.
#### 8. Substantial Equivalence Discussion
The following tables compare the EndoPilot2 to the predicate devices EMS-200 and SUPRASSON P5 NEWTRON with respect to indications for use, principles of operation, technological characteristics, components, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate devices.
{5}------------------------------------------------
### Table 5A – Comparison of Characteristics between Subject Device and Predicate Device EMS-200
| Attribute | SUBJECT DEVICE | PREDICATE DEVICE /<br>510(k) HOLDER (K153285) | Device<br>Comparison |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Schlumbohm GmbH & Co. KG | Meta Systems Co., Ltd. | |
| Trade Name | EndoPilot2 | EMS-200 | |
| Regulation Number | 872.4200 Dental Handpiece and Accessories | | Same |
| Regulation Name | EKX, direct drive, ac-powered handpiece | | |
| Product Codes | 872.4565 Dental hand instrument<br>EKR, endodontic plugger, root canal<br>LQY, root apex locator (unclassified)<br>872.4850 Ultrasonic scaler<br>ELC Ultrasonic scaler | NA | See second<br>predicate in<br>Table 5B |
| Indications for Use | The EndoPilot2 systems are dental devices<br>which combine in a single control unit an<br>endo motor to clean the root canal, a<br>dental obturator to fill and pressurize, an<br>electronic apex locator to assist the<br>operator to locate the file tip in the root<br>canal and an ultrasonic-handpiece for root-<br>canal cleaning and preparation.<br>The EndoPilot2 is intended solely for use<br>by trained dental professionals in<br>professional health care facilities on<br>patients that need root-canal-treatment. | The EMS-200 is a dental device<br>which combines in a single LCD<br>unit an endo motor which ablates<br>the tooth to expand the root<br>canal, a dental obturator to fill<br>and pressurize various shaped<br>packaging elements and an<br>electronic apex locator which<br>assists the operator the location<br>of the front tip in the root canal,<br>for use by trained dental<br>professionals. | Similar<br>The difference in<br>indications for<br>use result out of<br>the different<br>design and<br>modules used<br>for both devices.<br>The intended<br>use is the same. |
| Professional Use | Dental professionals | | Same |
| Location of Use | Dental practice | | Same |
| Mode of Action | Endo-Motor: ablates the tooth to clean the root canal (rotating endo-files)<br>Obturation Unit: to fill and pressurize various shaped packing elements<br>(Warming up by a resistance-wire)<br>Apex locator: to ensure the location of the front tip in root canal through<br>changes of electric resistance value into one unit (Electrical impedance)<br>Display: displayed through a single touch screen | | Same |
| AC/DC<br>Power<br>supply | AC: 100-240 V, 50/60 Hz<br>DC: 12 V, 1.5 A | AC: 100-240 V, 50/60 Hz<br>DC: 12 V, 5.0 A | Similar<br>Small DC<br>difference results<br>out of the lower<br>power<br>consumption for<br>EndoPilot2system<br>power adapters |
| Battery Operated | Yes, Li-Ion battery, 7.2 V,<br>power output 48 Wh | no | Different<br>Battery tested for<br>IEC 62133 andan<br>UN 38.3 Test |
| Attribute | SUBJECT DEVICE | PREDICATE DEVICE /<br>510(k) HOLDER (K153285) | K202906<br>Device<br>Comparison |
| Manufacturer | Schlumbohm GmbH & Co. KG | Meta Systems Co., Ltd. | |
| Trade Name | EndoPilot2 | EMS-200 | |
| Components<br>for<br>systems | 1) Control unit with display panel (touch<br>screen), 5 connecting sockets, a microSD<br>slot, power supply and wireless foot switch<br>2) Apex cable set (Lip-clip, Cap for Lip-clip,<br>Cable for file clamp, File clamp, Retainer<br>for apex cable, measuring cable)<br>3) Contra-angle for apex measurement<br>4) Endodontic Motor with apex measuring<br>contact<br>5) DownPack (D-Pack) handpiece with LED<br>indicator<br>6) BackFill handpiece (K042828)<br>7) Ultrasonic handpiece, Ultrasonic<br>Module (K050895) | 1) Control unit with display panel<br>(touch screen), 4 connecting<br>sockets, power supply<br>2) Electronic apex locator and<br>accessories (lip holder, cable<br>assembly and 2 kinds of file<br>holder).<br>3) Contra-angle for apex<br>measurement<br>4) Endodontic-motor Handpiece<br>with apex measuring contact<br>5) Obturation Unit (K031664)<br>6) BackFill handpiece included in<br>Obturation Unit (see 5)<br>7) NA | Similar<br>The<br>small<br>differences<br>in<br>design do not<br>raise<br>new<br>questions<br>of<br>substantial<br>equivalence |
| Processing<br>(reuse of components<br>sterilized by user) | To be reprocessed in the dental practice before re-use. | | Similar<br>Manuals include<br>reprocessing<br>instruction<br>based on ISO<br>17664/17665-1 |
| Electrical Safety | IEC 60601-1 | IEC 60601-1 | Same |
| Protection type and<br>level against electric<br>shock | Class II equipment, Type BF applied part | Class I equipment, Type B applied<br>part | Similar<br>Equivalent<br>technology is<br>used, |
| Electromagnetic<br>Compatibility<br>Electric Safety Tests | IEC 60601-1-2<br>IEC 61000-3 series<br>IEC 61000-4 series (2,3,4,5,6,8,11)<br>IEC 80601-2-60 | IEC 60601-1-2<br>IEC 61000-4 series (2,3,4,5,6,8,11) | Similar |
| | Technological comparison of System Component Functions | | |
| | Control Unit and display, Apex cable set and foot switch | | |
| Display<br>Device<br>Function | LCD Touchscreen for display of working<br>components | LCD Touchscreen for display of<br>working components, stand | Similar |
| Functional<br>Specification | All main function can be selected directly at the start screen<br>The device switches off after a longer period of non-operation | | Same |
| Accessory<br>Foot switch Function | Single wireless footswitch or optional<br>Twin wireless foot switch | NA, Function is started by pressing<br>a switch by finger | Different<br>The foot switch<br>provides |
| Wireless<br>connection/Bluetooth | 2.402-2.480 GHz, TX Power: +7 dBm | NA | different<br>method to start<br>the device |
| Attribute | SUBJECT DEVICE | PREDICATE DEVICE /<br>510(k) HOLDER (K153285) | K202906<br>Device<br>Comparison |
| Manufacturer | Schlumbohm GmbH & Co. KG | Meta Systems Co., Ltd. | |
| Trade Name | EndoPilot2 | EMS-200 | |
| Apex locator (Apex cable set) | | | |
| Function | Apex location of the front tip (of the endo-file) in the root canal through<br>changes of electric resistance value | | Same |
| Apex locator | Electronic apex locator (Schlumbohm) | Electronic apex locator | Same |
| Apex cable set<br>components<br>(material) | Lip clip (Stainless-Steel)<br>cap for lip-clip plug socket (POM-C)<br>58 mm 2mm | Lip holder (Stainless steel)<br>64 mm 2 mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.