ENCORE Latex Textured Surgical Gloves, Powder Free with Protein Content Labeling Claim (50 micrograms or less) and Tested for use with Chemotherapy drugs
K202765 · Careplus (M) Sdn Bhd · KGO · May 27, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202765
Device Name
ENCORE Latex Textured Surgical Gloves, Powder Free with Protein Content Labeling Claim (50 micrograms or less) and Tested for use with Chemotherapy drugs
Applicant
Careplus (M) Sdn Bhd
Product Code
KGO · General, Plastic Surgery
Decision Date
May 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Sterile, powder-free latex surgical gloves; textured surface; manufactured via on-line polymer and anti-tack coating. Used by operating room personnel to protect surgical wounds from contamination. Tested for resistance to chemotherapy drug permeation per ASTM D6978. Provides barrier protection; reduces risk of wound contamination during surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including ASTM D3577 (physical properties), ASTM D5151 (freedom from holes), ASTM D6124 (residual powder), ASTM D6978 (chemotherapy permeation), and ISO 10993 (biocompatibility).
Technological Characteristics
Natural rubber latex surgical gloves; powder-free with polymer/anti-tack coating. Titanium dioxide pigment. Sterilized via ISO 11137-2. Complies with ASTM D3577, D5151, D6124, D6978, D5712, and D7160. Protein content ≤ 50 µg/g.
Indications for Use
Indicated for operating room personnel to protect surgical wounds from contamination. Sterile, disposable, powder-free latex surgical gloves.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Encore Mark IV Powder Free Surgical Gloves (K983489)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 27, 2021
Careplus (M) SDN BHD Lim Shyan CEO/Managing Director Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate Seremban, Negeri Sembilan 70450 Malaysia
Re: K202765
Trade/Device Name: ENCORE Latex Textured Surgical Gloves, Powder Free with Protein Content Labeling Claim (50 micrograms or less) and Tested for use with Chemotherapy drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: April 13, 2021 Received: April 16, 2021
Dear Lim Shyan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202765
#### Device Name
ENCORE® Latex Textured Surgical Gloves, Powder free with Protein Content Labeling Claim (50 micrograms or less) and Tested for use with Chemotherapy drugs, Non-Pyrogenic
#### Indications for Use (Describe)
Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect a surgical wound from contamination.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy:
| Test Chemotherapy Drug (Concentration) | Minimum Breakthrough Detection Time in minutes |
|----------------------------------------|------------------------------------------------|
| Fluorouracil (50.0mg/ml) | >240 |
| Etoposide (20.0mg/ml) | >240 |
| Cyclophosphamide (20.0mg/ml) | >240 |
| Carmustine (3.3mg/ml) | 13.2 |
| Thiotepa (10.0mg/ml) | 12.0 |
| Paclitaxel (6.0mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0mg/ml) | >240 |
| Methotrexate (25.0mg/ml) | >240 |
| Vincristine Sulfate (1.0mg/ml) | >240 |
"WARNING: Do not use Carmustine and Thiotepa"
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K202765
1.0
### 510(K) SUMMARY
- CAREPLUS (M) SDN BHD Applicant: Address: Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia. Phone No. 60-6-6772781 Fax No. 60-6-6772780 Contact Person Lim Kwee Shyan 2.0 Date of Summary 11th May, 2021 3.0 Device Information ENCORE® Latex Textured Surgical Gloves, Powder free with Protein Content Device Name: Labeling Claim (50 micrograms or less) and Tested for use with Chemotherapy drugs, Non-Pyrogenic Common Name: Surgical Gloves Product Code: KGO Subsequent Product Code: LZC Classification Name: Surgeon's Gloves
### 4.0 Device Description
It is the powder-free variation of the class I sterile latex surgical gloves made by on-line polymer and anti-tack coating on inner and outer surface. The process modifies the surface istics and causes it to remain tack-free without the use of any dusting or donning powder. The device size comes in different sizes - 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0. Gloves meet the specification of ASTM D3577 with dimension as below:
Patient Examination Gloves Specialty (21 CFR 88.6250, LZC)
| Size | 5.5 | 6.0 | 6.5 | 7.0 | 7.5 | 8.0 | 8.5 | 9.0 |
|--------------------|-----|-----|-----|-----|------|-----|-----|-----|
| Length, min. mm | 245 | 265 | | | | | | |
| Thickness, min. mm | | | | | 0.10 | | | |
| Width, ± 6 mm | 70 | 76 | 83 | 89 | 95 | 102 | 108 | 114 |
# 5.0 Predicate Device
K983489, Encore Mark IV Powder Free Surgical Gloves (Protein Labeling Claim), Ansell Perry. The device is a class I, Surgeon's Gloves, KGO, LZC powder-free..
#### 6.0 Indication for Use
Powder Free Surgical gloves are sterile disposable devices intended to be worn by operating room personnel to protect aa surgical wound from contamination.
{4}------------------------------------------------
7.0 Technological Comparison Table:
| 7.0 Technological Comparison Table: | | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Characteristics | ASTM Specification | Standard References | Subject Glove K202765<br>(ENCORE® Latex Textured Surgical Gloves, Powder free with Protein Content Labeling Claim (50 micrograms or less and Tested for use with Chemotherapy drugs, Non-Pyrogenic) | Predicate Device (Encore Mark IV Powder Free Surgical Gloves, Protein Labeling Claim, K983489) | Discussion |
| Freedom from holes | ASTM D3577<br>ASTM D5151 | G-I, AQL 1.5 | Meets ASTM D3577<br>Meets ASTM D5151 | Meets ASTM D3577<br>Meets ASTM D5151 | Same |
| Dimension | ASTM D3577 | | | | Same |
| Length (size: 5.5), mm | | 245 min | min. 295 | Meet 245mm min | |
| Length (size: 6.0), mm | | 265 min | min. 295 | Meet 265mm min | |
| Length (size: 6.5), mm | | 265 min | min. 296 | Meet 265mm min | |
| Length (size: 7.0), mm | | 265 min | min. 295 | Meet 265mm min | |
| Length (size: 7.5), mm | | 265 min | min. 296 | Meet 265mm min | |
| Length (size: 8.0), mm | | 265 min | min. 296 | Meet 265mm min | |
| Length (size: 8.5), mm | | 265 min | min. 297 | Meet 265mm min | |
| Length (size: 9.0), mm | | 265 min | min. 295 | Meet 265mm min | |
| Thickness (cuff), mm | | 0.10 min | min. 0.158 | Meet 0.10mm min | |
| Thickness (palm), mm | | 0.10 min | min. 0.220 | Meet 0.10mm min | |
| Thickness (finger), mm | | 0.10 min | min. 0.225 | Meet 0.10mm min | |
| Width (size: 5.5), mm | | 70 ± 6 | 73-75 mm | Meet 70 ± 6 mm | |
| Width (size: 6.0), mm | | 76 ± 6 | 78-79 mm | Meet 76 ± 6 mm | |
| Width (size: 6.5), mm | | 83 ± 6 | 83-84 mm | Meet 83 ± 6 mm | |
| Width (size: 7.0), mm | | 89 ± 6 | 90-91 mm | Meet 89 ± 6 mm | |
| Width (size: 7.5), mm | | 95 ± 6 | 96-97 mm | Meet 95 ± 6 mm | |
| Width (size: 8.0), mm | | 102 ± 6 | 102-103 mm | Meet 102 ± 6 mm | |
| Width (size: 8.5), mm | | 108 ± 6 | 107-109 mm | Meet 108 ± 6 mm | |
| Width (size: 9.0), mm | | 114 ± 6 | 114-115 mm | Meet 114 ± 6 mm | |
| Physical Properties<br>(Before Ageing)<br>i) Tensile Strength (Mpa)<br>ii) Ultimate Elongation (%)<br>iii) Stress at 500% Elongation<br>(After Ageing)<br>i) Tensile Strength (Mpa)<br>ii) Ultimate Elongation (%) | ASTM D3577 | Min. 24<br>Min. 750<br>Max. 5.5<br><br>Min. 18<br>Min. 560 | min. 24<br>min. 780<br>max. 2.0<br><br>min. 20<br>min. 880 | Meets ASTM D3577 | Same |
| Powder Content | ASTM D3577<br>ASTM D6124 | Max. 2 mg/glove | Max. 0.57 mg/glove | Meets ASTM D3577<br>Meets ASTM D6124 | Same |
| K202765 | | | | | |
| Biocompatibility Test<br>i) Primary Skin Irritation<br>Test | ISO 10993-10 | No Animal<br>Irritation | Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>cause and irritant<br>response. | Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>cause and irritant<br>response. | Same |
| ii) Dermal Sensitization<br>Test | ISO 10993-10 | No Animal<br>Irritation | Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>produce a skin<br>sensitization effect. | Conclusion: Under the<br>conditions of this study,<br>the test material did not<br>produce a skin<br>sensitization effect. | Same |
| iii) In vitro Cytotoxicity Test | ISO 10993-5 | | Conclusions: Under the<br>conditions of this study,<br>the test material is<br>cytotoxic (grade 4) at<br>undiluted, 1:2, 1:4, 1:8<br>dilutions; and<br>Non-cytotoxic, grade 2<br>at 1:16 dilution, grade 0<br>at 1:32 and 1:64<br>dilutions | iii) No data available | Different |
| iv) Acute Systemic Toxicity | ISO 10993-11 | | Conclusion: Under the<br>conditions of this study,<br>the test material both<br>inner and outer surface<br>did not reveal systemic<br>toxicity. | iv) No data available | Different |
| Protein Label Claim | | | Contains 50<br>micrograms or less of<br>total water extractable<br>protein per gram. | Contains 50<br>micrograms or less of<br>total water extractable<br>protein per gram. | Same |
| Chemo Drugs Claim | ASTM D6978 | - | Chemo Claim | No Chemo Claim | Different |
| Non- Pyrogenic Claim | ISO 10993-12 | - | Non-Pyrogenic | No Non-Pyrogenic<br>claim | Different |
| Color | | - | Natural | Natural | Same |
| White Pigment | | - | Titanium Dioxide | Titanium Dioxide | Same |
| Intended Use | | - | Powder Free Surgical<br>gloves are sterile<br>disposable devices<br>intended to be worn by<br>operating room<br>personnel to protect a<br>surgical wound from<br>contamination. | Powder Free Surgical<br>gloves are sterile<br>disposable devices<br>intended to be worn by<br>operating room<br>personnel to protect a<br>surgical wound from<br>contamination. | Same |
{5}------------------------------------------------
{6}------------------------------------------------
| Chemotherapy Drug | ASTM D6978-05 | Subject Glove<br>(ENCORE® Latex Textured Surgical Gloves, Powder free with<br>Protein Content Labeling Claim (50 micrograms or less) and Tested<br>for use with Chemotherapy drugs, Non-Pyrogenic) |
|-------------------------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test Chemotherapy<br>Drug | Concentration | Minimum Breakthrough Detection Time (min) |
| Fluorouracil (Adrucil) | 50.0mg/ml | > 240 |
| Etoposide (Toposar) | 20.0mg/ml | > 240 |
| Cyclophosphamide<br>(Cytoxan) | 20.0mg/ml | > 240 |
| *Carmustine (BCNU) | 3.3mg/ml | 13.2 |
| *Thiotepa | 10.0mg/ml | 12.0 |
| Paclitaxel (Taxol) | 6.0mg/ml | > 240 |
| Doxorubicin<br>Hydrochloride | 2.0mg/ml | > 240 |
| Methotrexate | 25.0mg/ml | > 240 |
| Vincristine Sulfate | 1.0mg/ml | > 240 |
#### 8.0 Summary of Non-Clinical Testing
The performance test data of the non-clinical test that support a determination of substantial equivalence are the same as mentioned immediately above (ASTM requirement).
Following standards were used for the testing:
- ASTM D3577-09 Standard Specification for Rubber Surgical Gloves
- ASTM D6124-06 Standard Test Method for Residual Powder on Medical Gloves
- ASTM D5151-06 Standard Test Method for Detection of Holes in Medical Gloves
- ASTM D5712-15 Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method
- ASTM D7160-16 Standard Practice for Determination of Expiration Dating for Medical Gloves
- ASTM D6978-16 Standard Assessment of Resistance of Medical Devices to Permeation
- ISO 11137-2:2013 Sterilization of Healthcare Products- Establishing the sterilization dose
#### 9.0 Summary of Clinical Testing
Clinical data is not needed for medical gloves or of most devices cleared by the 510(k) process.
- 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.