The intended use of the Gelsoft Plus ERS vascular graft is for the extra-anatomical vascular repair.
Device Story
Gelsoft Plus ERS is an externally reinforced, gelatin-sealed, knitted polyester vascular graft. Device consists of a tubular knitted polyester base fabric, crimped for flexibility, with an external polypropylene support to provide kink resistance and a smooth flow surface for extra-anatomical applications. The polypropylene support is peelable at the ends to facilitate anastomosis. The graft is impregnated with bovine-derived gelatin to seal interstices, eliminating the need for pre-clotting. Used by vascular surgeons in clinical settings for bypass procedures. The device serves as a conduit for blood flow in patients with arterial occlusive or aneurysmal disease, restoring circulation and potentially preventing complications associated with arterial blockages or aneurysms.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical testing including physical bench testing, biocompatibility, chemical characterization, and an animal performance study.
Technological Characteristics
Knitted polyester textile graft; externally reinforced with polypropylene support; impregnated with bovine-derived gelatin sealant. Tubular form with crimping. Peelable polypropylene support at ends. Sterilized device.
Indications for Use
Indicated for extra-anatomical vascular repair, primarily axillo-femoral, bi-femoral, and femoro-popliteal bypass/reconstruction in patients with aneurysmal and occlusive arterial disease. Contraindicated in patients with sensitivity to polyester or bovine-derived materials; not for coronary repair, blood access fistula (hemodialysis), pulmonary shunting, thoracic use, or cardiovascular patching.
Regulatory Classification
Identification
A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services USA logo, and on the right is the FDA logo. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
April 6, 2021
Vascutek Ltd Chirag Merchant Senior Regulatory Affairs Associate Newmains Avenue, Inchinnan Renfrewshire PA4 9RR Scotland. UK
Re: K202703
Trade/Device Name: Gelsoft Plus ERS Vascular Grafts Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular graft prosthesis Regulatory Class: Class II Product Code: DSY Dated: February 12, 2021 Received: February 25, 2021
Dear Chirag Merchant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Rohini Retarekar -S
for Carmen Gacchina Johnson, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202703
Device Name Vascutek Gelsoft Plus ERS Vascular Graft
Indications for Use (Describe)
Indicated for extra anatomical vascular repair, primarily for axillo-femoral bypass and femoro- popliteal reconstruction.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Terumo Aortic logo. The word "TERUMO" is in green, with a red swoosh above it. Below that, the word "Aortic" is in gray. The logo is simple and modern.
Vascutek Ltd Newmains Avenue, Inchinnan Renfrewshire PA4 9RR Scotland, UK
> Tel. +44 (0) 141 812 5555 terumoaortic.com
## 510 (k) Summary
This 510(k) Summary is being submitted in accordance with 21 CFR 807.92.
| Submitter: | Vascutek Ltd<br>Newmains Avenue<br>Inchinnan<br>Renfrewshire<br>PA4 9RR<br>Scotland, UK |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------|
| | Tel: +44 (0) 141 812 555 |
| Contact: | Chirag Merchant<br>Senior Regulatory Affairs Associate<br>Tel: +44 (0) 141 812 555<br>Email: c.merchant@terumoaortic.com |
| Date of Preparation: | April 2, 2021 |
| Trade Name: | Vascutek Gelsoft Plus ERS Vascular Graft |
| Common or Usual Name: | Vascular Graft Prosthesis |
| Classification Name: | Prosthesis, vascular graft, of 6mm and<br>greater diameter |
| Product Code: | DSY |
| Regulation Number: | 21 CFR 870. 3450 |
| Device Class: | II |
Image /page/3/Picture/7 description: This image is a logo for The Queen's Awards for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design attached to it. Above the "e" is a crown, symbolizing the Queen. The text "THE QUEEN'S AWARDS FOR ENTERPRISE: INNOVATION 2016" is written below the "e".
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Terumo Aortic. The word "TERUMO" is in large, green, sans-serif font with a red swoosh above the letters "TE". Below the word "TERUMO" is the word "Aortic" in a smaller, gray, sans-serif font. The logo is simple and modern.
#### Identification of the legally marketed device to which equivalence is being claimed:
Vascutek Ltd. are claiming equivalence to the following legally marketed devices:
- . Vascutek Gelsoft Plus ERS (K034010).
### Device Description:
Gelsoft Plus ERS grafts are an externally reinforced (ERS), gelatin-sealed, knitted polyester grafts.
Vascutek's polyester vascular prosthesis family is based on a polyester textile technology. The starting point for all products is polyester yarn. This is fabricated into tubular form by knitting. This is referred to as the base fabric and exists in a variety of designs to meet particular end uses. Further processing is designed to retain the tubular form of the graft e.g. crimping and addition of an external polypropylene support. In Gelsoft Plus ERS, this external polypropylene support is to provide kink resistance and a smooth flow surface for extra-anatomical applications. The polypropylene support may be peeled where it extends to the ends of the prosthesis, in order to facilitate the fashioning of the anastomosis. Additional branches are also attached in a variety of configurations. The last process is impregnation with a bovine derived gelatin sealant. This process fills the gaps between threads of the polyester base graft and eliminates the need for pre-clotting of the graft prior to implant.
#### Intended Use:
The intended use of the Gelsoft Plus ERS vascular graft is for the extra-anatomical vascular repair.
#### Indications for Use:
Vascutek Gelsoft Plus ERS Graft is indicated for extra anatomical vascular repair, primarily for axillo-femoral/bi-femoral bypass and femoro- popliteal reconstruction.
Gelsoft Plus ERS AX -FEM Vascular Graft: Indicated for vascular repair i.e. primarily for axillo-femoral bypass procedures & femoral- popliteal reconstruction in aneurysmal & occlusive disease of the arteries.
Image /page/4/Picture/12 description: The image is a logo for The Queen's Awards for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design extending from it. Above the "e" is a crown, symbolizing the British monarchy. The text "THE QUEEN'S AWARDS FOR ENTERPRISE: INNOVATION 2016" is written below the "e".
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Terumo Aortic logo. The word "TERUMO" is in green, with a red swoosh above it. Below that, the word "Aortic" is in gray. The logo is simple and clean, and the colors are eye-catching.
Gelsoft Plus ERS FEM Vascular Graft: Indicated for vascular repair i.e. primarily for femoro-femoral bypass procedures in aneurysmal & occlusive disease of the arteries.
Gelsoft Plus ERS AX- BIFEM Vascular Graft: Indicated for vascular repair i.e. primarily for axillo-bi-femoral bypass procedures in aneurysmal & occlusive disease of the arteries.
#### Intended Patient Population:
Patients requiring extra anatomical vascular repair whenever knitted polyester grafts are normally indicated, primarily for axillo-femoral/bi-femoral bypass and femoro-popliteal reconstruction in aneurysmal & occlusive disease of the arteries.
They are contraindicated for use in patients with a sensitivity to polyester or materials of bovine origin. They are contraindicated for coronary vascular repair, blood access fistula (e.g. haemodialysis) and pulmonary shunting. They are contraindicated for thoracic use. They are contraindicated for use in the pulmonary positions, use in arteriovenous shunting or cardiovascular patching.
#### Technological Characteristics:
Equivalency is based on identical design, technology, construction and intended use.
The only change is that the gelatin used to seal the grafts will be purchased from a new supplier.
The nonclinical testing performed, including physical bench, biocompatibility, chemical characterisation and an animal performance study, have demonstrated that the device is substantially equivalent to the predicate device.
Image /page/5/Picture/10 description: The image is a black and white logo for The Queen's Award for Enterprise: Innovation 2016. The logo features a stylized letter "e" with a wing-like design extending from it. Above the "e" is a crown, symbolizing the British monarchy. Below the "e" and crown, the text "THE QUEEN'S AWARD FOR ENTERPRISE: INNOVATION 2016" is arranged in a stacked format.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.