Medical Face Masks

K202627 · Guilin Hbm Sanitary Protections, Inc. · FXX · Feb 3, 2021 · General, Plastic Surgery

Device Facts

Record IDK202627
Device NameMedical Face Masks
ApplicantGuilin Hbm Sanitary Protections, Inc.
Product CodeFXX · General, Plastic Surgery
Decision DateFeb 3, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

Medical Face Masks are flat-pleated, three-layer, single-use, disposable devices; worn via ear loops; include a nose piece for facial fit. Construction: outer and inner layers of spun-bonded polypropylene; middle layer of melt-blown polypropylene filter. Used in clinical or general settings for infection control; protects wearer and patient from transfer of microorganisms, body fluids, and particulates. Device is non-sterile; provides physical barrier; reduces exposure to blood and body fluids.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862 (fluid resistance 120mmHg), ASTM F2299 (99.82% PFE at 0.1μm), ASTM F2101 (99.78% BFE), EN 14683 (3.3 mmH2O/cm² differential pressure), and 16 CFR 1610 (Class I flammability). Biocompatibility confirmed per ISO 10993-5 and 10993-10.

Technological Characteristics

Flat-pleated, 3-layer non-woven polypropylene mask. Layers: spun-bonded PP (outer/inner), melt-blown PP (filter). Ear loops: 85% spandex/15% polyester. Nose piece: iron wire covered in polypropylene. Dimensions: 17.5cm x 9.5cm. Non-sterile, single-use. Biocompatibility per ISO 10993 series.

Indications for Use

Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material in infection control settings.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 3, 2021 Guilin HBM Sanitary Protections, Inc % Shelley Li Director Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705, Baohua International Plaza, West Guangzhong Road 555, Jingan Shanghai, 200071 China Re: K202627 Trade/Device Name: Medical Face Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 9, 2020 Received: November 13, 2020 Dear Shelley Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K202627 Device Name Medical Face Masks #### Indications for Use (Describe) The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | Type of Use (Select one or both, as applicable) | | | |-------------------------------------------------|--|--| | | | | | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary: K202627 #### I. Submitter Guilin HBM Sanitary Protections, Inc No.1-2, Shuijing East Road, Economic and Technological Development Area, Guilin 541805, China Contact person: Liu Xin Position: General Manager Assistant Tel.: +86-13761713489 E-mail: 724095871@qq.com Preparation date: Jan.30, 2021 ### Submission Correspondent Ms. Shelley Li Tel.: +86-021-80317637 Shanghai Landlink Medical Information Technology Co., Ltd. E-mail: shelley.li@landlink-healthcare.com ### II. Proposed Device | Trade Name of Device: | Medical Face Masks | |-----------------------|--------------------| | Common name: | Surgical Mask | | Regulation Number: | 21 CFR 878.4040 | | Regulatory Class: | Class II | | Product code: | FXX | | Review Panel | General Hospital | ### III. Predicate Devices | 510(k) Number: | K153496 | |-----------------|-----------------------------------------| | Trade name: | Disposable Surgical Face Mask | | Common name: | Surgical Mask | | Classification: | Class II | | Product Code: | FXX | | Manufacturer | Xiantao Rayxin Medical Product Co., Ltd | ### IV. Device description {4}------------------------------------------------ The Medical Face Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Medical Face Masks are manufactured with three layers. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric. The Medical Face Masks are single use, disposable device, provided non-sterile. # V. Indication for use The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. VI. Comparison of technological characteristics with the predicate devices | Item | Proposed device<br>(K202627) | Predicate device<br>(K153496) | Comparison | | |----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Product name | Medial Face Masks | Disposable Surgical<br>Face Mask | Similar | | | Product Code | FXX | FXX | Same | | | Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | | Class | Class II | Class II | Same | | | Mask style | Flat-pleated, ear loop,<br>3 layers | Flat-pleated, ear loop,<br>tie-on, 3 Layers | Similar | | | Indication for use | The Medical Face<br>Masks are intended<br>to be worn to protect<br>both the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms,<br>body fluids and<br>particulate material. | The Disposable<br>Surgical Face Masks<br>are intended to be<br>worn to protect both<br>the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms,<br>body fluids and | Same | | | | | These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. | | | Materials | Inner layer | Spun-bond polypropylene | Spun-bond polypropylene | Same | | | Middle layer | Melt blown polypropylene filter | Melt blown polypropylene filter | Same | | | Outer layer | Spun-bond polypropylene | Spun-bond polypropylene | Same | | | Ear loop | 85% Spandex + 15% Polyester | Polyester | Different1 | | | Nose piece | Iron wire covered polypropylene | Malleable aluminum wire | Different1 | | Color | | Blue | Blue | Same | | Length | | 17.5cm | 17.5±1cm | Same | | Width | | 9.5cm | 9.5±1cm | Same | | OTC use | | Yes | Yes | Same | | Sterile | | Non-sterile | Non-sterile | Same | | Single for use | | Yes | Yes | Same | | Biocompatibility | | Confirm to the requirements of ISO 10993 series standards | Confirm to the requirements of ISO 10993 series standards | Same | | Fluid Resistance Performance<br>ASTM E1862 | | Pass at 120mmHg | pass at 120mmHg | Similar | | Particulate<br>Filtration Efficiency<br>ASTM F2299 | Average 99.82% at<br>0.1μm | Average 99.74% at<br>0.1μm | | | | Bacterial Filtration<br>Efficiency<br>ASTM F2101 | Average 99.78% | Average 99.4% | | | | Differential<br>Pressure | Average 3.3<br>mmH2O/cm² | Average 2.7<br>mmH2O/cm² | | | | Flammability<br>16 CFR 1610 | Class I<br>Non Flammable | Class 1<br>Non Flammable | | | | In Vitro<br>Cytotoxicity<br>ISO 10993-5 | Under the conditions<br>of this study the<br>device is<br>non-cytotoxic | Under the conditions<br>of this study the<br>device is<br>non-cytotoxic | Same | | | Skin Irritation<br>ISO 10993-10 | Under the conditions<br>of this study the<br>device is non-irritating | Under the conditions<br>of this study the<br>device is non-irritating | Same | | | Skin Sensitization<br>ISO 10993-10 | Under the conditions<br>of this study the<br>device is<br>non-sensitizing | Under the conditions<br>of this study the<br>device is<br>non-sensitizing | Same | | Table 1 Comparison of characteristics {5}------------------------------------------------ {6}------------------------------------------------ 1 The difference in the materials does not raise additional questions for safety and effectiveness of the device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device which includes all construction materials and color additives. The test results shows pass the requirements. # VII. Summary of Non-Clinical Testing Non-clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of test article, which demonstrated that the proposed device complies with the related standards: | Methodology/<br>Standard | Purpose | Acceptance<br>Criteria | Results | |--------------------------|---------------------------------|---------------------------------|-------------------------| | ASTM<br>F1862M-17 | Fluid Resistance<br>Performance | 29 out of 32 pass at<br>120mmHg | 29 out of 32<br>pass at | {7}------------------------------------------------ | | | | 120mmHg | |--------------------------|-----------------------------------|--------------------------|-------------------------------| | ASTM F2299 | Particulate Filtration Efficiency | ≥98% | Average<br>99.82% at<br>0.1μm | | ASTM<br>F2101-19 | Bacterial Filtration Efficiency | ≥98% | Average<br>99.78% | | EN 14683:2019<br>Annex C | Differential Pressure | <6mmH2O/cm² | Average 3.3<br>mmH2O/cm² | | 16 CFR 1610 | Flammability | Class I<br>Non Flammable | Meet Class I | # VIII. Clinical Testing No clinical study is included in this submission. ### IX. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe as effective, and performs as well as or better than the legally marketed predicate deviceK153496.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...