APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are intended for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size. APTUS® K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants.
Device Story
System comprises cannulated compression screws (headless and headed) and K-wire fixation system. Headless screws utilize differential thread pitch to generate compression; headed screws utilize conventional head-to-thread compression. Features include triangular SpeedTip® geometry for improved insertion torque and internal hexalobular instrument interface. Used by orthopedic surgeons in clinical settings for bone fixation, osteotomies, and arthrodesis. Screws are cannulated for over-the-wire placement using compatible K-wires. Provides mechanical stabilization of bone segments to facilitate healing. Supplied non-sterile; requires user-performed sterilization.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including mechanical testing per ASTM F543 and biocompatibility/sterilization data referenced from previously cleared predicate devices.
Indicated for patients requiring treatment of bone fractures, osteotomies, and arthrodesis, or bone reconstruction, and for use as guide pins for other implants.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
November 6, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Medartis AG % Lauren M. Wessell RA Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K202589
Trade/Device Name: APTUS Cannulated Compression Screws, APTUS headed Cannulated Compression Screws, APTUS K-Wire System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HTY, HTN Dated: September 8, 2020 Received: September 8, 2020
Dear Lauren M. Wessell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202589
Device Name
APTUS® Cannulated Compression Screws and APTUS® headed Cannulated Compression Screws
Indications for Use (Describe)
APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are intended for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202589
Device Name APTUS® K-Wire System
Indications for Use (Describe)
APTUS® K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
## APTUS® headed Cannulated Compression Screws
# APTUS® Cannulated Compression Screws
# APTUS® K-Wire System
#### Medartis AG
#### November 6, 2020
#### ADMINISTRATIVE INFORMATION Manufacturer Name Medartis AG Hochbergerstrasse 60E CH-4057 Basel, Switzerland Telephone: Official Contact Andrea Kiefer-Schweizer Head of Quality Management and Regulatory Affairs Lauren M. Wessell, RAC Representative/Consultant Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1 858-792-1235 +1 858-792-1236 Fax: Email: lwessell(@paxmed.com kthomas(@paxmed.com; flarson(@paxmed.com DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | APTUS® headed Cannulated Compression Screws |
|----------------------|-------------------------------------------------------------------|
| | APTUS® Cannulated Compression Screws |
| | APTUS® K-Wire System |
| Common Name | Screw, Fixation, Bone (Primary) |
| | Pin, Fixation, Smooth |
| | Washer, Bolt Nut |
| Regulation Number | 21 CFR 888.3040 (Primary) |
| | 21 CFR 888.3030 |
| Regulation Name | Smooth or threaded metallic bone fixation fastener (Primary) |
| Regulatory Class | Class II |
| Product Code | HWC (Primary), HTY, HTN |
| Classification Panel | Orthopedic |
| Reviewing Division | Office of Orthopedic Devices (OHT6) |
| | Division of Restorative, Repair and Trauma Devices (DHT6C) |
| | Stereotaxic, Bone Growth Stimulators and Fracture Fixation Device |
| | Team |
{5}------------------------------------------------
#### PREDICATE DEVICE INFORMATION
Primary Predicate
K133460, APTUS® Cannulated Compression Screws, Medartis AG
Additional Predicates
K110658, APTUS® Cannulated Compression Screws, Medartis AG K092038, APTUS® K-Wire System, Medartis AG K050681, Omnitech System & Easy Lock Osteosystem with Xtremities Plates, TriMed, Inc. K963192, Synthes Sterile 3.5 mm and 4.0 mm Cannulated Screws, Synthes (USA)
Reference Devices
K193633, APTUS Ankle Trauma System 2.8/3.5, Medartis AG K191848, APTUS Wrist Spanning Plates 2.5, Medartis AG
### INDICATIONS FOR USE STATEMENT
APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are intended for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size.
APTUS® K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants.
## SUBJECT DEVICE DESCRIPTION
The subject device includes APTUS Cannulated Compression Screws in thread diameters 1.7 mm, 2.2 mm, 3.0 mm, 5.0 mm, and APTUS headed Cannulated Compression Screws in diameters 2.2 mm, 3.0 mm, 5.0 mm, and 7.0 mm. APTUS Cannulated Compression Screws are headless screws that incorporate threads with different pitch on the head and tip of the screws; this difference in pitch provides compression as the screw is inserted. APTUS headed Cannulated Compression Screws have conventional bone screw heads that apply compression between the threads and the head. Both types of screws provide compression of the bone segments upon insertion of the screw. Additionally, all subject device screws have a triangular SpeedTip® shape designed to improve cutting and insertion torque and an internal hexalobular instrument face.
The APTUS Cannulated Compression Screws and APTUS headed Cannulated Compression Screws come in partially threaded and fully threaded designs. The 1.7 mm diameter screws are provided in overall lengths ranging from 8 mm to 20 mm. The 2.2 mm diameter screw are provided in overall lengths ranging from 10 mm to 40 mm. The 3.0 mm diameter screws are provided in overall lengths ranging from 10 mm to 40 mm. The 4.0 mm diameter screws are provided in overall lengths ranging from 16 mm to 60 mm. The 5.0 mm diameter screws are provided in overall lengths ranging from 24 mm to 70 mm. The 7.0 mm diameter screws are provided in overall lengths ranging from 30 mm to 140 mm.
The subject device 1.7 mm, 2.2 mm, 4.0 mm, 5.0 mm, and 7.0 mm thread diameter APTUS Cannulated Compression Screws and APTUS headed Cannulated Compression screws are used with the corresponding subject device K-wires (0.6, 0.8, 1.1, 1.25, 1.6, and 2.2 mm, respectively). The subject device screws also are compatible with the Medartis K-wires cleared under K092038. Similarly, the subject device K-wires are compatible with the APTUS Cannulated Compression Screws cleared under K133460 and K110658.
{6}------------------------------------------------
This submission also includes the corresponding washers for APTUS headed Cannulated Compression Screws.
The subject device cannulated compression screws and washers are manufactured from titanium allov conforming to ASTM F136 Standard Specification for Wrought Titanium-61 anadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The subject device K-wires are manufactured from stainless steel conforming to ASTM F138 Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673).
All subject devices are provided non-sterile to the end user. The subject devices are single-use only.
### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility referenced from K133460, K110658, K092038, and K193633; moist heat sterilization (to be performed by the end user) also referenced from K133460, K110658, and K193633; X-ray beam sterilization, packaging, and sterile barrier shelf life referenced from K191848 and K 193633; and mechanical testing according to ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws. Clinical data were not provided in this submission.
### EQUIVALENCE TO MARKETED DEVICES
The subject devices are substantially equivalent in indications and design principles to the primary predicate device and the additional predicates devices listed above. Additionally, the subject devices have similar technological characteristics as the reference devices.
The subject device APTUS Cannulated Compression Screws and APTUS headed Cannulated Compression Screws have similar Indications for Use Statements (IFUS) to those of devices previously cleared in K133460, K110658, K050681, and K963192. Differences among the IFUS include specific language not in the subject device IFUS: children and adolescent patient populations (K963192), specific anatomical locations (K050681), and references to "small bones" (K050681 and K963192). These differences do not impact the substantial equivalence because all IFUS express equivalent intended use for the treatment of fractures, osteotomies and arthrodesis of bones.
The primary predicate K133460, and additional predicates K110658 and K092038, are in support of substantial equivalence in terms of comparable cannulated compression screw and K-wire designs, compatible cannulated compression screws and K-wires, and identical materials (titanium alloy conforming to ASTM F136 and stainless steel conforming to ASTM F138).
Additional predicate devices K050681 and K963192 are in support of substantial equivalence in terms of comparable screw designs and for comparative mechanical properties.
The reference device K193633 is in support of substantial equivalence for comparable washer designs and identical material (titanium alloy conforming to ASTM F136). Furthermore, the device-specific accessories and Class I instruments have similar designs and are made of identical materials as the Class II accessories and instruments previously cleared in K193633.
{7}------------------------------------------------
The reference devices K191848 and K193633 are in support of substantial equivalence for terminal sterilization and shelf life for product provided sterile to the end user. Additionally, reference devices K191848 and K193633 are in support of identical materials for Class II device-specific accessories and Class I instruments supplied non-sterile to the end-user
### CONCLUSION
The subject devices, the primary predicate device, and the additional predicate devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.