Disposable Surgical Mask

K202580 · Qingdao Likangyuan Medical Device Co., Ltd. · FXX · Jun 7, 2021 · General, Plastic Surgery

Device Facts

Record IDK202580
Device NameDisposable Surgical Mask
ApplicantQingdao Likangyuan Medical Device Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJun 7, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile.

Device Story

Disposable 3-ply surgical mask; flat-pleated design with ear loops. Constructed of outer/inner spunbond polypropylene layers and middle melt-blown polypropylene filter layer. Includes nose piece for facial fit. Used in clinical settings for infection control; worn by healthcare personnel and patients to provide barrier protection against blood, body fluids, and particulates. Single-use, non-sterile device.

Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F1862 (synthetic blood penetration), ASTM F2299 (particulate filtration efficiency ≥99%), ASTM F2101 (bacterial filtration efficiency ≥99%), MIL-M-36954C (differential pressure 3.7 mmH2O/cm²), and 16 CFR 1610 (flammability Class 1). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

Technological Characteristics

3-ply flat-pleated mask. Materials: Spunbond polypropylene (inner/outer), melt-blown polypropylene (filter), spandex/nylon ear loops, galvanized wire nose piece with PE coating. ASTM F2100 Level 3 compliant. Non-sterile. Dimensions: 9.5 cm x 17.5 cm.

Indications for Use

Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue. June 7, 2021 Qingdao Likangyuan Medical Device Co., Ltd. % Helen Nan General Manager Wenzhou Cytech Information Service Co., Ltd. Room302. Building3, Hangqian Mansion, Hangqian Street, Lucheng District Wenzhou, Zhejiang 325000 China Re: K202580 Trade/Device Name: Disposable Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 16, 2021 Received: May 6, 2021 Dear Helen Nan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Bifeng Qian -S For Clarence W. Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) k202580 Device Name Disposable Surgical Mask #### Indications for Use (Describe) The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary for K202580 This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. #### 1.0 Submitter Information | Company: | Qingdao Likangyuan Medical Device Co., | |----------------------|------------------------------------------| | | Ltd. | | Address: | No.257 Leigongshan Road,Huangdao | | | District,Qingdao,Shandong, 266423, China | | Phone: | 086-532-89052813 | | Contact Name: | Ying Xu | | Title: | Quality Control Manager | | E-mail: | likangyuan@hotmail.com | | Date of Preparation: | June 4, 2021 | ## 2.0 Submission Correspondent: | Company: | Wenzhou Cytech Information Service Co., Ltd. | |------------------|-----------------------------------------------------------------------------------------------------------------| | Mailing Address: | Room302, Building3, Hangqian Mansion,<br>Hangqian Street, Lucheng District, Wenzhou,<br>Zhejiang, 325000, CHINA | | Contact Name: | Helen Nan | | Title: | General Manager | | Email: | ctc_registration@hotmail.com | # 3.0 Subject Device Information | Trade/Device Name: | Disposable Surgical Mask | |-------------------------|--------------------------| | Model: | Ear-loop Type | | Common Name: | Surgical Face Mask | | Regulation Description: | Surgical apparel | | Device: | Mask, Surgical | | Review Panel: | General Hospital | | Product Code: | FXX | | Regulation Number: | 21 CFR 878.4040 | | Device Class: | Class II | # 4.0 Predicate Device Information | Submitter: | DemeTECH Corporation | |-------------------------|----------------------| | 510k Number: | K201479 | | Trade/Device Name: | DemeMASK | | Common Name: | Surgical Face Mask | | Regulation Description: | Surgical apparel | | Device: | Mask, Surgical | | Review Panel: | General Hospital | {4}------------------------------------------------ | Product Code: | FXX | |--------------------|-----------------| | Regulation Number: | 21 CFR 878.4040 | | Device Class: | Class II | # 5.0 Device Description The Disposable Surgical Masks (non-sterile) are 3-ply, ear loop, flat-pleated surgical masks. The devices are manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. There surgical masks to be secured on users via ear loops made of spandex and nylon. Each mask contains a nose piece to firmly fit over the nose. The devices are sold non-sterile and are intended to be single use and disposable. This device is not made from any natural rubber latex. # 6.0 Indications for Use The Disposable Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. This face mask is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device provided non-sterile. | Device | Predicate Device | Proposed Device | Comparison | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Manufacturer | DemeTECH Corporation | Qingdao Likangyuan<br>Medical Device Co., Ltd. | N/A | | 510K Number | K201479 | K202580 | N/A | | Trade/Device<br>Name | DemeMASK | Disposable Surgical<br>Mask (non-sterile) | N/A | | Product Code | FXX | FXX | Same | | Classification | Class II<br>(21 CFR 878.4040) | Class II<br>(21 CFR 878.4040) | Same | | Intended Use | The Disposable Surgical<br>Face Mask is intended to<br>be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body<br>fluids and particulate<br>material.<br>This face mask is<br>intended for use in<br>infection control practices | The Disposable Surgical<br>Masks (non-sterile) are<br>intended to be worn to<br>protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids and particulate<br>material.<br>These face masks are<br>intended for use in | Same | # 7.0 Comparison of Technological Characteristics Table 1 - General Comparison {5}------------------------------------------------ | | to reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device provided non-<br>sterile. | infection control<br>practices to reduce the<br>potential exposure to<br>blood and body fluids.<br>This is a single use,<br>disposable device<br>provided non-sterile. | | |-----------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | | Material | | | | Outer Facing<br>Layer | Spunbond Polypropylene | Spunbond Polypropylene | Same | | | Melt Blown | Melt Blown | | | Middle Layer | Polypropylene<br>Filter | Polypropylene<br>Filter | Same | | Inner Facing<br>Layer | Spunbond Polypropylene | Spunbond Polypropylene | Same | | Nose Piece | Single Galvanize Wire,<br>Coated By PE | Single Galvanize Wire,<br>Coated By PE | Same | | Mask Style | 3 Ply, Ear Loops, Flat-<br>Pleated Style | 3 Ply, Ear Loops, Flat-<br>Pleated Style | Same | | Ear loops | Spandex and Nylon | Spandex and Nylon | Same | | Color | White | Blue (Outer layer)<br>White (Inner layer) | Similar | | Dimension<br>(Width) | $9.5 cm \pm 1 cm$ | $9.5 cm$ | Similar | | Dimension<br>(Length) | $17.5 cm \pm 1 cm$ | $17.5 cm$ | Similar | | OTC Use | Yes | Yes | Same | | Sterility | Non-Sterile | Non-Sterile | Same | | Shelf life | No Shelf Life | 2 yrs | Different | | Use | Single Use | Single Use | Same | | ASTM F2100<br>Level | Level 3 | Level 3 | Same | | | Table 2 - Performance Testing | | |--|-------------------------------|--| |--|-------------------------------|--| | Device | Predicate Device<br>DemeMASK | Proposed Device<br>Disposable Surgical<br>Mask | Comparison | |--------|------------------------------|------------------------------------------------|------------| |--------|------------------------------|------------------------------------------------|------------| {6}------------------------------------------------ | Fluid Resistance<br>Performance<br>ASTM F1862 | Pass at 160 mmHg<br>(Level 3 Fluid<br>Resistance) | Pass at 160 mmHg<br>(Level 3 Fluid<br>Resistance) | Same | |-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------| | Particulate<br>Filtration<br>Efficiency<br>ASTM F2299 | Pass at ≥99% | Pass at ≥99% | Same | | Bacterial Filtration<br>Efficiency<br>ASTMF2101 | Pass at ≥99% | Pass at ≥99% | Same | | Differential<br>Pressure (Delta P)<br>MIL-M-36954C | Average 3.6<br>mmH2O/cm² | Average 3.7<br>mmH2O/cm² | Similar | | Flammability<br>16 CFR 1610 | Class 1 | Class 1 | Same | | Biocompatibility<br>ISO 10993-5 and<br>ISO 10993-10 | Under the conditions of<br>the studies employed,<br>the device is non-<br>cytotoxic, non-<br>sensitizing, and non-<br>irritating. | Under the conditions of<br>the studies employed,<br>the device is non-<br>cytotoxic, non-<br>sensitizing, and non-<br>irritating. | Same | 8.0 Summary of Non-clinical Tests Performed on the Proposed Device The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004. Table of Conformity to Standards | Standards | Name | |-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASTM<br>F1862/F1862M-17 | Standard Test Method for Resistance of Medical Face Masks to<br>Penetration by Synthetic Blood (Horizontal Projection of Fixed<br>Volume at a Known Velocity) | | ASTM F2100-19 | Standard Specification for Performance of Materials Used in<br>Medical Face Masks | | ASTM F2101-19 | Standard Test Method for Evaluating the Bacterial Filtration<br>Efficiency (BFE) of Medical Face Mask Materials Using a<br>Biological Aerosol of Staphylococcus aureus | | MIL-M-36954C | Differential Pressure (Delta-P) | | ASTM<br>F2299/F2299M-03 | Standard Test Method for Determining the Initial Efficiency of<br>Materials Used in Medical Face Masks to Penetration by<br>Particulates Using Latex Spheres | | 16 CFR PART 1610 | Standard for the flammability of clothing textiles | {7}------------------------------------------------ | ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in<br>vitro cytotoxicity | |--------------|----------------------------------------------------------------------------------------------------| | ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for<br>Irritation and skin sensitization | {8}------------------------------------------------ | Standards | Disposable Surgical<br>Mask<br>Ear Loop | Acceptance Criteria | Result | |-----------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------|--------| | Resistance to<br>penetration by<br>synthetic blood<br>ASTM F1862 | 32 out of 32 passed in<br>160 mm Hg | 29 out of 32 passed in<br>160 mm Hg | Pass | | Sub-micron<br>particulate filtration<br>efficiency at<br>0.1 micron<br>ASTM F2299 | ≥99% | ≥98% | Pass | | Bacterial filtration<br>efficiency<br>ASTM F2101-19 | ≥99% | ≥98% | Pass | | Differential pressure<br>MIL-M-36954 | Average 3.7<br>mmH₂O/cm² | < 6.0<br>mmH₂O/cm² | Pass | | Flame spread<br>16 CFR 1610 | Class 1<br>Non-Flammable | Class 1<br>Non-Flammable | Pass | ## 9.0 Table of BiocompatibilityTesting | Standards | Proposed Device | Result | |-----------------------------------------|-------------------------------------------------------------------|--------| | Cytotoxicity<br>ISO 10993-5 | Under the conditions of the study, the device is non-cytotoxic. | Pass | | Skin Sensitization Test<br>ISO 10993-10 | Under the conditions of the study, the device is non-sensitizing. | Pass | | Skin Irritation Test<br>ISO 10993-10 | Under the conditions of the study, the device is non-irritating. | Pass | ## 9.0 Summary of Clinical Performance Test No clinical study is included in this submission. # 10.0 Conclusion: The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K202580, the Disposable Surgical Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared underK201479.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...