The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, clinics and similar care environment for clinical diagnosis of patients.
Device Story
HS30 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers (linear, curved, endocavity, phased array) and displays images in multiple modes: 2D, M, Color/Power/PW/CW Doppler, Tissue Doppler, and 3D/4D. Used in hospitals and clinics by trained healthcare professionals for clinical diagnosis. The system provides real-time acoustic output (MI/TI) and includes analysis packages for anatomical measurements. Healthcare providers use these images and measurements to inform clinical decision-making. Benefits include non-invasive diagnostic imaging and fluid analysis.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electromagnetic, and mechanical safety testing in accordance with recognized standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, NEMA UD 2).
Technological Characteristics
Mobile cart-based ultrasound system with LED monitor. Supports linear, curved, endocavity, and phased array transducers (1.0-20.0 MHz). Modes: 2D, M, Doppler (Color, Power, PW, CW, Tissue), 3D/4D. Connectivity: DICOM. Safety: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2. Software-controlled with automated analysis features (Auto IMT+, Needle Mate+, Strain+, etc.).
Indications for Use
Indicated for fetal/obstetrics, abdominal, gynecology, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional/superficial), urology, cardiac (adult/pediatric), and peripheral vessel imaging in patients requiring diagnostic ultrasound.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2020
Samsung Medison Co., LTD % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA
Re: K202556
Trade/Device Name: HS30 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 31, 2020 Received: September 3, 2020
Dear Ji Yea Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202556
Device Name HS30 Diagnostic Ultrasound System
### Indications for Use (Describe)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Premarket Notification - Traditional
## 5. 510(K) Summarv: K202556
In accordance with 21 CFR 807.92 the following summary of information is provided:
- Date Prepared August 31, 2020 1.
- 2. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
- 3. Primary Contact Person Ji Yea Lee Requlatory Affairs Specialist Phone: +82.2.2.2194.1594 Fax: +82.31.8017.9573 Email: jiyea722.lee@samsungmedison.com
- 4. Secondary Contact Person Ninad Gujar Vice President Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com
- Proposed Device 5.
- Common/Usual Name: Diagnostic Ultrasound System and Accessories
- Proprietary Name: HS30 Diagnostic Ultrasound System -
- Common Name: Diagnostic Ultrasound System ।
- Classification Names: system, imaqing, pulsed doppler, ultrasonic ।
- । Product Code: IYN, IYO, ITX
- -Regulation: 21 CFR 892.1550, 892.1560, 892.1570
### 6. Predicates
- HS30 Diagnostic Ultrasound System (K182632) Primary -
- HS40 Diagnostic Ultrasound System (K200339) -
- 7. Device Description
The HS30 is a general purpose, mobile, software controlled, diagnostic ultrasound system, Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, Elastoscan Mode or as a combination of these modes. The HS30 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS30 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
- 8. Intended Use
The diagnostic ultrasound system and probes are designed to obtain ultrasound images
{4}------------------------------------------------
510(k) Premarket Notification – Traditional
and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode.
- 9. Technology
The HS30 employs the same fundamental scientific technology as the predicates.
- 10. Determination of Substantial Equivalence
The HS30 is substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
| Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel |
| Scanhead<br>Types: | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes |
| Scanhead Frequency | 1.0 ~ 20.0 MHz | 1.0 ~ 20.0 MHz | 1.0 ~ 20.0 MHz |
| Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) |
| Acoustic<br>Output Display<br>& FDA Limits: | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display |
| Modes of<br>Operation: | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad)<br>- 3D/4D mode | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad) | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad)<br>- 3D/4D mode |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
| System<br>Characteristics: | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz |
| Product Safety<br>Certification | - IEC 60601-1<br>- IEC60601-2-37 | - IEC 60601-1<br>- IEC60601-2-37 | - IEC 60601-1<br>- IEC60601-2-37 |
| EMC<br>Compliance | IEC60601-1-2 | IEC60601-1-2 | IEC60601-1-2 |
| Functionality | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) |
| Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) |
| | 3D MagiCut<br>3D XI(MSV, Oblique<br>View, XI VOCAL)<br>XI STIC<br>2D NT<br>BiometryAssist<br>LaborAssist | 3D MagiCut | 3D MagiCut<br>3D XI (MSV, Oblique<br>View, XI VOCAL)<br>XI STIC<br>2D NT<br>BiometryAssist<br>LaborAssist |
| Transducers | [Linear array] | [Linear array] | [Linear array] |
| | -LN5-12<br>-L5-12/50 | -LN5-12<br>-L5-12/50 | -LA3-16AD<br>-LN5-12<br>-L5-12/50<br>-L4-7<br>-LS6-15 |
| | [Curved array]<br>-CF4-9<br>-C2-8<br>-C2-5<br>-CA2-6BM | [Curved array]<br>-CF4-9<br>-C2-8<br>-C2-5 | [Curved array]<br>CA2-8AD<br>-CF4-9<br>-C2-8<br>-C2-5<br>-CA2-6BM |
| | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 |
| | [Phased array]<br>-PN2-4<br>-SP3-8 | [Phased array]<br>PN2-4 | [Phased array]<br>-PN2-4<br>-SP3-8 |
| | [3D]<br>-VN4-8<br>-EV2-10A | | [3D]<br>-VN4-8<br>-V5-9<br>-EV2-10A |
| | [CW]<br>DP2B | [CW]<br>DP2B | [CW]<br>DP2B<br>-DP8B |
| Biopsy Guides | BP-KIT-024 | BP-KIT-024 | - BP-KIT-024<br>- BP-KIT-029 |
| | - BP-KIT-035 [BP-KIT-<br>035-NG]<br>- BP-KIT-040 [BP-KIT-<br>040-NG]<br>- BP-KIT-045 [BP-KIT-<br>045-NG]<br>- BP-KIT-047 [BP-KIT-<br>047-NG]<br>- BP-KIT-049 [BP-KIT-<br>049-NG]<br>- BP-KIT-061<br>- BP-KIT-085<br>- BP-KIT-086 | - BP-KIT-035 [BP-KIT-<br>035-NG]<br>- BP-KIT-040 [BP-KIT-<br>040-NG]<br>- BP-KIT-045 [BP-KIT-<br>045-NG]<br>- BP-KIT-047 [BP-KIT-<br>047-NG]<br>- BP-KIT-061 | - BP-KIT-054 [BP-KIT-<br>054-NG]<br>- BP-KIT-055 [BP-KIT-<br>055-NG]<br>- BP-KIT-049 [BP-KIT-<br>049-NG]<br>- BP-KIT-060<br>- BP-KIT-061<br>- BP-KIT-068 [BP-KIT-<br>068-NG]<br>- BP-KIT-043<br>- BP-KIT-085<br>- BP-KIT-086 |
| Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) |
| optional<br>devices | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) |
| Accessories | - Foot Switch<br>- ECG<br>Gel Warmer | - Foot Switch<br>- ECG<br>Gel Warmer | - Foot Switch<br>- ECG<br>Gel Warmer |
{5}------------------------------------------------
# SAMSUNG
510(k) Premarket Notification – Traditional
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "SAMSUNG" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
510(k) Premarket Notification – Traditional
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "SAMSUNG" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The word is horizontally oriented and centered.
510(k) Premarket Notification - Traditional
## 11. Summary of Non-Clinical Test
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The HS30 and its applications comply with voluntary standards.
| Reference No. | Title |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R)2012 and A2:2010 /(R)2012<br>Medical Electrical Equipment - Part 1: General Requirements for<br>basic safety and essential performance. |
| IEC 60601-1-2 | IEC60601-1-2: 2014(4th Edition), Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential<br>performance - EMC |
| IEC 60601-2-37 | IEC60601-2-37:2007 + A1:2015, Particular requirements for the<br>safety of ultrasonic medical diagnostic and monitoring equipment |
| ISO10993-1 | ISO 10993-1, Biological evaluation of medical devices -- Part 1:<br>Evaluation and testing within a risk management process. |
| ISO14971 | ISO 14971:2007, Medical devices - Application of risk management<br>to medical devices |
| NEMA UD 2-<br>2004 | NEMA UD 2-2004 (R2009)<br>Acoustic Output Measurement Standard for Diagnostic Ultrasound<br>Equipment Revision 3 |
- 12. Summary of Clinical Tests
The subject of this premarket submission, HS30 is not required clinical studies to support substantial equivalence.
- 13. Conclusion
Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the HS30 to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
### END of 510(K) Summary i
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.