K202518 · Promaxo, Inc. · LNH · Mar 3, 2021 · Radiology
Device Facts
Record ID
K202518
Device Name
Promaxo MRI System
Applicant
Promaxo, Inc.
Product Code
LNH · Radiology
Decision Date
Mar 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Promaxo Magnetic Resonance Imaging (MRI) System is an office-based MRI system for producing images that display the prostate and adjoining tissues. When used by a trained urologist or interventional/urologic radiologist, the system is intended to be used for targeting prostatic lesions under MR guidance in alignment with the current standard of care. Promaxo MR images are not intended to be used for diagnostic purposes, and a 3T MR image acquired without an endorectal coil is a required input for guidance using the Promaxo MRI System.
Device Story
Office-based MRI system; permanent magnet array; captures/reconstructs prostate images for lesion targeting. Operated by urologists/radiologists in clinical settings. Requires external 3T MR image input for guidance. Includes template holder for needle localization; MR guidance interface for template calibration/registration. Provides visual guidance for interventional procedures; improves targeting accuracy for prostate biopsies/interventions. Cryogen-free; fixed installation.
Clinical Evidence
Clinical feasibility study conducted; sample clinical images provided. Testing included registration accuracy and motion studies on human subjects. Bench testing performed per IEC 60601-1, 60601-1-2, 60601-2-33, and NEMA MS series (geometric distortion, noise, slice thickness, SNR, SAR, RF heating).
Technological Characteristics
Permanent magnet MRI; 0.058T-0.074T field strength; cryogen-free. Two-axis electromagnetic coils (x, y) with built-in permanent z-gradient. Integrated RF transmission coil (4 kW, 2.77 MHz); patient-wearable receive coil. DICOM compliant. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS 2, 4, 5, 6, 8, 9, 10, 14, ISO 10993-1.
Indications for Use
Indicated for use by trained urologists or interventional/urologic radiologists for targeting prostatic lesions under MR guidance in alignment with standard of care. Requires 3T MR image input (no endorectal coil) for guidance. Not for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
March 3, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Promaxo, Inc. % Eva Hellman RA/QA Manager 70 Washington St. Suite 407 OAKLAND CA 94607
Re: K202518
Trade/Device Name: Promaxo MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, MOS Dated: January 21, 2021 Received: January 22, 2021
Dear Eva Hellman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202518
Device Name Promaxo MRI System
#### Indications for Use (Describe)
The Promaxo Magnetic Resonance Imaging (MRI) System is an office-based MRI system for producing images that display the prostate and adjoining tissues. When used by a trained urologist or interventional/urologic radiologist, the system is intended to be used for targeting prostatic lesions under MR guidance in alignment with the current standard of care. Promaxo MR images are not intended to be used for diagnostic purposes, and a 3T MR image acquired without an endorectal coil is a required input for guidance using the Promaxo MRI System.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 1em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Promaxo logo at the top, with the text "510(k) Summary" below it. The logo features the word "promaxo" in a combination of gray and blue, with a blue checkmark extending from the "x" in the word. Below the text "510(k) Summary" is the identifier "K202518".
## 510(k) Summary
| Submitter: | Promaxo, Inc.<br>70 Washington St., Suite 407, Oakland, CA 94607<br>510-770-6819 |
|----------------------------|----------------------------------------------------------------------------------|
| Contact Person: | Eva Hellman<br>RA/QA Manager<br>(510) 770-6961<br>ehellman@promaxo.com |
| Date: | 8/31/2020 |
| Trade or Proprietary Name: | Promaxo MRI System |
| Common or Usual Name: | System, Nuclear Magnetic Resonance Imaging |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Product Code: | LNH, MOS |
| Regulation Number: | 892.1000 |
| Classification: | Class II |
Predicate Device:
| Trade Name | Common Name | Class | Product Code | Manufacturer | K-number |
|--------------------------------------------------------------------------------------------|-------------|-------|--------------|----------------------------------------|----------|
| ASM-030PIII<br>(OPENMARK 111),<br>ASM-040P<br>(OPENMARK 4000), ASM-050P<br>(OPENMARK 5000) | MRI Systems | II | LNH | Shenzhen Anke<br>High-Tech Co.<br>Ltd. | K113281 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Promaxo logo, with the word "promaxo" in gray and blue. A blue checkmark is incorporated into the logo. Below the logo, the text "510(k) Summary" is written in a larger font size.
#### Indications for Use
The Promaxo Magnetic Resonance Imaging (MRI) System is an office-based MRI system for producing images that display the prostate and adjoining tissues. When used by a trained urologist or interventional/urologic radiologist, the system is intended to be used for targeting prostatic lesions under MR guidance in alignment with the current standard of care. Promaxo MR images are not intended to be used for diagnostic purposes, and a 3T MR image acquired without an endorectal coil is a required input for guidance using the Promaxo MRI System.
### Description of the Device and Summary of the Technological Characteristics
Promaxo's MRI system is an open configuration MRI system composed of an array of permanent magnets arranged to provide a constant in-plane magnetic field strength and a built-in z-gradient within its field of view. The system utilizes electromagnetic gradient coils, RF coils, and other components such as the spectrometer and signal amplifiers to capture, reconstruct and display magnetic resonance images of objects within its field of view. Promaxo MRI System's technological features are substantially equivalent to its predicate device. Both of them:
- । Are comprised of a magnet, magnet enclosure, electromagnetic gradient coils, RF transmission coil, and RF receiver coil
- Have the main magnet comprised of an array of permanent magnets
- -Measure spatial distribution of protons exhibiting magnetic resonance
- -Are capable of imaging T1, T2, and Diffusion-Weighted Imaging
- -Are cryogen free
- Provide an interactive user interface to operate the device -
Promaxo's MRI system differs in technology in the following ways:
- -The z-gradient is built into the main magnetic field and, as a result, the system does not require an electromagnetic z-gradient coil
- -The device includes a template holder to be used for procedures under MR guidance
- -The device includes an MR guidance user interface workflow such as template calibration and registration with imported MR images
A detailed comparison of the Promaxo MRI System and predicate device is described below:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "promaxo". The word "promaxo" is written in a sans-serif font, with the "pro" in gray and the "maxo" in blue. A blue checkmark extends from the bottom of the "x" and curves upward, resembling an arrow pointing upwards and to the right.
510(k) Summary
| Product | Promaxo MRI System | 510(k) Summary<br>OPENMARK III, OPENMARK 4000<br>and OPENMARK 5000 |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Information | | |
| Manufacturer | Promaxo | Shenzhen Anke High-tech Co., Ltd |
| 510(k) Number | K202518 | K113281 |
| Product Code | LNH, MOS | LNH |
| Regulation No. | 892.1000 | 892.1000 |
| Intended Use | The Promaxo Magnetic Resonance<br>Imaging (MRI) System is an office-<br>based MRI system for producing<br>images that display the prostate<br>and adjoining tissues. When used<br>by a trained urologist or<br>interventional/urologic<br>radiologist, the system is intended<br>to be used for targeting prostatic<br>lesions under MR guidance in<br>alignment with the current<br>standard of care. Promaxo MR<br>images are not intended to be<br>used for diagnostic purposes, and<br>a 3T MR image acquired without<br>an endorectal coil is a required<br>input for guidance using the<br>Promaxo MRI System. | MRI SYSTEMS, including ASM-<br>030PI1 (OPENMARK 1II), ASM-040P<br>(OPENMARK 4000) and ASM-050P<br>(OPENMARK 5000). Are indicated<br>for use as magnetic resonance<br>diagnostic devices (MRDD) that<br>produce transverse, sagittal,<br>coronal and oblique cross-sectional<br>images, and that display the<br>internal structure and/or function<br>of the head, body, or extremities.<br>These images when interpreted by<br>a trained physician yield<br>information that may assist in<br>diagnosis. |
| Magnet | | |
| Type of Magnet | Permanent Magnet | Permanent Magnet |
| Magnetic field<br>strength in FOV | 0.058T-0.074T | 0.3T, 0.4T, 0.5T |
| Patient accessible<br>bore-space | Open configuration | Open configuration |
| Type of installation | Fixed | Fixed |
| Shimming | Passive | Passive |
| Gradient System | | |
| Cooling System | Cryogen free | Cryogen free |
| Gradient System<br>Description | Two-axis electromagnetic coils for<br>x- and y-gradients, built-in<br>permanent z-gradient | Three-axis electromagnetic coils |
| dB/dt (time rate of<br>change of<br>magnetic field) | Always within normal operating<br>mode, under normal use<br>conditions | Always less than 20T/s, under<br>normal using condition |
| Product | Promaxo MRI System | 510(k) Summary<br>OPENMARK III, OPENMARK 4000<br>and OPENMARK 5000 |
| Radiofrequency System | | |
| RF Transmission<br>Coil | Peak Power: 4 kW<br>Center Frequency: 2.77 MHz<br>Transmission Bandwidth: 2.47 to 3.07 MHz<br>Configuration: Integrated | Peak Power: 6 kW<br>Center Frequency: 21.29 MHz<br>Transmission Bandwidth: Unknown<br>Configuration: Integrated |
| RF Receive Coils | Patient wearable receive coil<br>(plug-in type) | Patient wearable receive coil<br>(various configurations, plug-in type) |
| SAR (Specific<br>Absorption Rate)<br>Management | No active SAR management<br>required | No active SAR management<br>required |
| Operation | | |
| Mains Power<br>Requirements | Power Supply: 208V, 3 phase<br>Power Frequency: 60 Hz<br>Input Power: 15 kVA | Power Supply: 380V 3N~<br>Power Frequency: 50 Hz<br>Input Power: 15 kVA |
| Control System | Spectrometer | Spectrometer |
| Image Acquisition and Processing | | |
| Image Resolution | Varies | Varies |
| Slice Thickness | Selectable by technician | Selectable by technician |
| Pulse Sequences | T1, T2, DWI | T1, T2, DWI, DCE |
| Imaging Protocols | T1, T2, DWI of prostate and<br>adjoining tissues<br>Patient wearable receive coil | T1, T2, DWI of head, body and<br>extremities<br>Patient wearable receive coil<br>Exogeneous contrast media may be<br>applied. |
| Image Processing | None | None |
| DICOM Compliance | Yes | Yes |
| Graphical User<br>Interface | Touchscreen keyboard/mouse | Display, keyboard/mouse |
| Other | | |
| Software<br>Applications | User operates the system using a<br>main user interface application | User operates the system using a<br>main user interface application |
| Networking | Network interface for archival and<br>retrieval | Network interface for archival and<br>retrieval |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Promaxo. The word "promaxo" is written in a sans-serif font, with "pro" in gray and "maxo" in blue. A blue checkmark extends from the bottom of the "x" and curves upward, ending in an arrow pointing to the upper right.
# 510(k) Summary
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the "promaxo" logo with the word "promaxo" in gray and blue. The "o" in "promaxo" has a blue checkmark and arrow pointing upwards. Below the logo, the text "510(k) Summary" is written in a larger font.
### Summary of Non-Clinical Testing
Promaxo MRI System has been evaluated to demonstrate substantial equivalence related to medical electrical equipment, risk management, hardware and software verification and validation, and image quality and has been found to conform to the following medical device standards:
- । IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- । IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-33:2010, AMD1:2013, AMD2:2015, Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis
- -NEMA MS 2-2008 (R2014), Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images.
- -NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices
- -NEMA MS 5-2018, Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging.
- -NEMA MS 6-2008 (R2014), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging
- -NEMA MS 8-2016, Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
- -NEMA MS 9-2008 (R2014), Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images
- -NEMA MS 10-2010, Determination of Local Specific Absorption Rate (SAR) in Diagnostic Magnetic Resonance Imaging Systems
- -NEMA MS 14-2019 Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems
- -ISO 10993-1:2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
Navigation accuracy testing on phantoms was performed to support the indications for use and substantial equivalence. The test included users performing template localization and needle localization activities using the Promaxo MRI System.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "promaxo" in a stylized font, with the first four letters in gray and the last three in blue. A blue checkmark is incorporated into the "o" at the end of the word. Below the word "promaxo" is the text "510(k) Summary" in a simple, sans-serif font. The text is centered and appears to be the title or heading of a document or presentation.
## Summary of Clinical Testing
Sample clinical images and testing on sample clinical images were provided to support substantial equivalence. Users performed registration accuracy and motion studies using the Promaxo MRI System on clinical images of human subjects.
An end-to-end clinical feasibility study was conducted.
### Conclusion
The testing demonstrates that the Promaxo MRI System is as safe and effective as the predicate for the intended use of surgical localization of previously diagnosed prostate lesions on a diagnostic 3T MRI. The subject device is substantially equivalent to the legally marketed predicate device and conforms to applicable medical device safety and performance standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.