← Product Code [GEX](/productcode/GEX) · K202503

# Chrome (K202503)

_Quanta System Spa · GEX · Nov 18, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202503

## Device Facts

- **Applicant:** Quanta System Spa
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Nov 18, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Chrome is intended for use in aesthetic, cosmetic and surgical applications requiring incision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows.

## Device Story

Chrome is a multi-source laser and IPL system for aesthetic and surgical soft tissue procedures. It utilizes Q-Switched Nd:YAG (532nm/1064nm) and pulsed Nd:YAG laser sources, plus optional Twain IPL handpieces (400-1200nm range). The device features an articulated arm for laser delivery and interchangeable handpieces for IPL. An integrated skin cooler reduces pain and side effects. Operated by clinicians via a touchscreen and footswitch, the device enables precise tissue interaction (incision, ablation, coagulation). Output is visual/thermal; clinicians use the device to treat vascular/pigmented lesions, remove tattoos/hair, and perform skin resurfacing. Benefits include improved cosmetic outcomes and effective management of dermatological conditions. The system is intended for clinical/professional use.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance established via compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards. Software verification and validation conducted per FDA guidance. Biocompatibility established by comparison to predicate K191842, confirming identical patient-contacting materials, processing, and geometry.

## Technological Characteristics

Laser system (Nd:YAG 532/1064nm) and IPL (400-1200nm). Articulated arm delivery for lasers; handpieces for IPL. Integrated skin cooling system. Touchscreen interface, footswitch-triggered. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, and IEC 60825-1. Biocompatible materials identical to predicate K191842.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Pico Platform ([K191842](/device/K191842.md))

## Reference Devices

- Evo family ([K160368](/device/K160368.md))
- Er:YAG handpieces ([K173002](/device/K173002.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 18, 2020

Quanta System Spa Francesco Dell'Antonio Vice President Regulatory Affairs and QA via Acquedotto, 109 Samarate (Va), 21017 Italy

Re: K202503

Trade/Device Name: Chrome Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 20, 2020 Received: August 31, 2020

Dear Francesco Dell'Antonio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202503

Device Name Chrome

Indications for Use (Describe)

General intended use

Chrome is intended for use in aesthetic, cosmetic and surgical applications requiring incision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows.

Indications for use

1064 & 532 nm (Q-Switched, nanosecond mode)

Chrome is intended for treatment of benign vascular lesions, benign pigmented lesions, and for hair, tattoo removal and the incision, excision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of:

532 nm (Q-Switched, nanosecond mode), including microbeam handpieces: Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos

Treatment of benign vascular lesions including, but not limited to:

- port wine birthmarks
- telangiectasias
- spider angioma
- Cherry angioma
- Spider nevi

Treatment of benign pigmented lesions including, but not limited to:

- cafe-au-Iait birthmarks
- Ephalides, solar lentigines
- senile lentigines
- Becker's nevi
- freckles
- common nevi
- nevus spilus
- Ota Nevus

Treatment of seborrheic keratosis

Treatment of post inflammatory hyperpigmentation

Skin resurfacing procedures for the treatment of acne scars and wrinkles.

1064 nm (Q-Switched, nanosecond mode), including microbeam handpieces: Removal of dark ink (black, blue and brown) tattoos

Removal of benign pigmented lesions including;

- nevus of Ota
- Café au lait spot
- Ephalides, solar lentigo (lentigines)
- Becker Nevus
- Nevus spilus

Treatment of common nevi

Removal or lightening of unwanted hair

Skin resurfacing procedures for the treatment of acne scars and wrinkles

{3}------------------------------------------------

1064 nm (pulsed)

Dermatology/Plastic Surgery:

Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis.

The laser is also intended for the treatment of benign pigmented lessons such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, sebortheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques.

The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. The laser is also indicated for the treatment of facial wrinkles.

Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

It is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

The laser is also indicated for benign pigmented lesion size, for patients with benign lesions that would potentiallybenefit from aggressive treatment, and for patients with benign that have not responded to other laser treatments.

It is indicated for use on all skin types (Fitzpatrick I-VI) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types.

IPL 590-1200nm; 625-1200nm; 650-1200nm

Indicated for permanent hair removal.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime

## IPL 550-1200nm; 570-1200nm

Indicated for photocoagulation of dermatological benign vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions.

IPL 400-1200nm

Indicated for inflammatory acne (mild to moderate acne vulgaris).

Integrated Skin Cooler

The intended use of the integrated cooling system in the laser hand piece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluencies for laser treatments such as hair removal and benign vascular lesion, and to reduce the potential side effects of laser treatments. Any other different use is considered incorrect.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

#### K202503

### 5. 510(K) SUMMARY

| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy 21017                                                                                                                   |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person:                           | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA<br>Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared:                                   | 10th November 2020                                                                                                                                                                         |
| Device Name:                                     | Chrome                                                                                                                                                                                     |
| Classification:                                  | Class II                                                                                                                                                                                   |
| Classification Name:                             | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology.                                                                                                    |
| Regulation Number:                               | 21 CFR 878.4810                                                                                                                                                                            |
| Product Code:                                    | GEX                                                                                                                                                                                        |
| Main Predicate Device                            | Pico Platform (K191842), Quanta System SpA                                                                                                                                                 |
| Reference Device                                 | Evo family (K160368), Quanta System SpA                                                                                                                                                    |

#### Performance Standards:

There are no mandatory performance standards for this device.

### Description of the device:

Chrome is a laser family that includes Q-Switched and/or Pulsed laser sources, emitting at 532 nm and 1064 nm (Nd:YAG laser).

Chrome, through the special universal Twain connector, can be equipped with intense pulsed light handpieces (Twain IPL) emitting at the following wavelengths: 650-1200nm, 625-1200nm, 590-1200nm, 570-1200nm, 550-1200nm, 400-1200nm.

It can also be connected to Er:YAG handpieces cleared under K173002.

{6}------------------------------------------------

Chrome, when operating with Pulsed laser sources and IPL, can be used in combination with optional contact, or air, cooling systems.

The optical delivery system is an articulated arm with fixed or variable handpieces. The optical delivery system for the IPL system is a handpiece (Twain IPL) with fixed or interchangeable light filters at different wavelengths.

Chrome is controlled via a touch screen display housed in the front of the device.

Emission is triggered by means of a footswitch.

### Indication for use General intended use

Chrome is intended for use in aesthetic, cosmetic and surgical applications requiring incision, excision, ablation, vaporization and coagulation of body soft tissues in the medical specialties of dermatology, general, plastic and oral surgery as follows.

#### Indications for use

#### 1064 & 532 nm (Q-Switched, nanosecond mode)

Chrome is intended for treatment of benign vascular lesions, benign pigmented lesions, and for hair, tattoo removal and the incision, ablation, vaporization of soft tissue for General dermatology such as, but not limited to treatment of:

532 nm (Q-Switched, nanosecond mode), including microbeam handpieces: Removal of light ink (red, sky blue, green, tan, purple, and orange) tattoos Treatment of benign vascular lesions including, but not limited to:

- । port wine birthmarks
- telangiectasias
- י spider angioma
- -Cherry angioma
- -Spider nevi

Treatment of benign pigmented lesions including, but not limited to:

- cafe-au-lait birthmarks
- -Ephalides, solar lentigines
- senile lentigines
- -Becker's nevi
- freckles
- common nevi
- nevus spilus
- -Ota Nevus

Treatment of seborrheic keratosis

Treatment of post inflammatory hyperpigmentation

Skin resurfacing procedures for the treatment of acne scars and wrinkles.

1064 nm (Q-Switched, nanosecond mode), including microbeam handpieces:

5-2

{7}------------------------------------------------

Removal of dark ink (black, blue and brown) tattoos Removal of benign pigmented lesions including;

- । nevus of Ota
- -Café au lait spot
- -Ephalides, solar lentigo (lentigines)
- -Becker Nevus
- -Nevus spilus

Treatment of common nevi

Removal or lightening of unwanted hair

Skin resurfacing procedures for the treatment of acne scars and wrinkles

### 1064 nm (pulsed)

### Dermatology/Plastic Surgery:

Intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis.

The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratosis and plaques.

The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

The laser is also indicated for the treatment of facial wrinkles.

Additionally, the laser is indicated for the treatment of pseudofolliculitis barbae (PFB) and for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

lt is indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

The laser is also indicated for benign pigmented lesion size, for patients with benign lesions that would potentially benefit from aggressive treatment, and for patients with benign lesions that have not responded to other laser treatments.

lt is indicated for use on all skin types (Fitzpatrick I-VI) including tanned skin, and the removal and permanent reduction of unwanted hair in Fitzpatrick I-VI, including suntanned skin types.

#### IPL 590-1200nm; 625-1200nm; 650-1200nm

Indicated for permanent hair removal.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime

#### IPL 550-1200nm; 570-1200nm

Indicated for photocoagulation of dermatological benign vascular lesion (i.e. face telangiectasia), photothermolysis of blood vessels (treatment of facial and leg veins), and treatment of benign pigmented lesions.

{8}------------------------------------------------

#### IPL 400-1200nm

Indicated for inflammatory acne (mild to moderate acne vulgaris).

#### Integrated Skin Cooler

The intended use of the integrated cooling system in the laser hand piece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluencies for laser treatments such as hair removal and benign vascular lesion, and to reduce the potential side effects of laser treatments. Any other different use is considered incorrect.

### Technological Characteristics Comparison

| Specification                                                                    | Main predicate device                                                                                                                                                                                                                                                                    | Additional reference device                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                                  |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name<br>(K number)                                                        | Pico Family K191842                                                                                                                                                                                                                                                                      | EVO Platform K160368                                                                                                                                                                                                                                                                | Chrome                                                                                                                                                                                                                                                                                                                                          |
| Submitter                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                   | Quanta System SpA                                                                                                                                                                                                                                                                                                                               |
| Product Code                                                                     | GEX                                                                                                                                                                                                                                                                                      | GEX                                                                                                                                                                                                                                                                                 | GEX                                                                                                                                                                                                                                                                                                                                             |
| Regulation<br>Number                                                             | 21 CFR 878.4810                                                                                                                                                                                                                                                                          | 21 CFR 878.4810                                                                                                                                                                                                                                                                     | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                 |
| Class                                                                            | II                                                                                                                                                                                                                                                                                       | II                                                                                                                                                                                                                                                                                  | II                                                                                                                                                                                                                                                                                                                                              |
| Laser Sources                                                                    | Nd:YAG, and Ruby                                                                                                                                                                                                                                                                         | Nd:YAG, Ruby and Alexandrite                                                                                                                                                                                                                                                        | Nd:YAG                                                                                                                                                                                                                                                                                                                                          |
| Laser<br>Wavelengths<br>[nm]                                                     | ☑ 1064nm Q-Switched<br>(Nanosecond mode)<br>☑ 1064nm (picosecond mode)<br>☐ 1064nm non-Q-Switched<br>(pulsed mode)<br>☑ 532 nm Q-Switched<br>(Nanosecond mode)<br>☑ 532 nm (picosecond mode)<br>☑ 694nm Q-Switched<br>(Nanosecond mode)<br>☑ 694nm non-Q-Switched (free<br>running mode) | ☑ 1064nm Q-Switched<br>(Nanosecond mode)<br>☑ 1064nm non-Q-Switched<br>(pulsed mode)<br>☑ 532 nm Q-Switched<br>(Nanosecond mode)<br>☑ 532nm pulsed<br>☑ 1320nm pulsed<br>☑ 755nm pulsed<br>☑ 694nm Q-Switched<br>(Nanosecond mode)<br>☑ 694nm non-Q-Switched (free<br>running mode) | ☑ 1064nm Q-Switched<br>(Nanosecond mode)<br>☑ 1064nm non-Q-Switched<br>(pulsed mode)<br>☑ 532 nm Q-Switched<br>(Nanosecond mode)<br>☐ 532nm pulsed<br>☐ 1320nm pulsed<br>☐ 755nm pulsed<br>☑ 694nm Q-Switched<br>(Nanosecond mode)<br>☑ 694nm non-Q-Switched (free<br>running mode)<br>☐ 532 nm (picosecond mode)<br>☐ 1064nm (picosecond mode) |
| IPL<br>wavelengths<br>[nm]                                                       | 650-1200nm<br>625-1200nm<br>590-1200nm<br>570-1200nm<br>550-1200nm<br>400-1200nm                                                                                                                                                                                                         | 650-1200nm<br>625-1200nm<br>590-1200nm<br>570-1200nm<br>550-1200nm<br>400-1200nm                                                                                                                                                                                                    | 650-1200nm<br>625-1200nm<br>590-1200nm<br>570-1200nm<br>550-1200nm<br>400-1200nm                                                                                                                                                                                                                                                                |
| 1064nm (Q-Switched, nanosecond mode)                                             |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Homogeneous spot                                                                 |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Pulse width                                                                      | 6 to 12 ns                                                                                                                                                                                                                                                                               | 6 to 12 ns                                                                                                                                                                                                                                                                          | 6 to 12 ns                                                                                                                                                                                                                                                                                                                                      |
| Max Fluence                                                                      | Up to 38J/cm²                                                                                                                                                                                                                                                                            | Up to 50J/cm²                                                                                                                                                                                                                                                                       | Up to 48J/cm²                                                                                                                                                                                                                                                                                                                                   |
| Spot Size                                                                        | 2 to 12 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 7x7 mm²<br>Squared                                                                                                                                                                                                                     | 2 to 12 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 9x9 mm²<br>Squared                                                                                                                                                                                                                | 2 to 8 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 7x7 mm²<br>Squared                                                                                                                                                                                                                                                                             |
| Specification                                                                    | Main predicate device                                                                                                                                                                                                                                                                    | Additional reference device                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                                  |
| Device Name<br>(K number)                                                        | Pico Family K191842                                                                                                                                                                                                                                                                      | EVO Platform K160368                                                                                                                                                                                                                                                                | Chrome                                                                                                                                                                                                                                                                                                                                          |
| Submitter                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                   | Quanta System SpA                                                                                                                                                                                                                                                                                                                               |
| Repetition<br>Rate                                                               | Up to 10 Hz                                                                                                                                                                                                                                                                              | Up to 10 Hz                                                                                                                                                                                                                                                                         | Up to 20 Hz                                                                                                                                                                                                                                                                                                                                     |
| High<br>coverage<br>microbeam<br>Handpiece                                       | Present                                                                                                                                                                                                                                                                                  | Absent                                                                                                                                                                                                                                                                              | Present                                                                                                                                                                                                                                                                                                                                         |
| Microbeam<br>handpiece<br>Fluence per<br>dot                                     | 6 J/cm²                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                 | From 0.2 to 16.5 J/cm²                                                                                                                                                                                                                                                                                                                          |
| Microbeam<br>handpiece<br>spot<br>dimension<br>(mm)                              | 9 mm diameter                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                 | 9 mm diameter                                                                                                                                                                                                                                                                                                                                   |
| Standard<br>microbeam<br>Handpiece                                               | Present                                                                                                                                                                                                                                                                                  | Absent                                                                                                                                                                                                                                                                              | Present                                                                                                                                                                                                                                                                                                                                         |
| Microbeam<br>handpiece<br>Fluence per<br>dot                                     | Up to 38J/cm²                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                 | From 0.03 to 44.4 J/cm2                                                                                                                                                                                                                                                                                                                         |
| Microbeam<br>handpiece<br>spot<br>dimension<br>(mm)                              | 8 mm diameter                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                 | 8 mm diameter                                                                                                                                                                                                                                                                                                                                   |
|                                                                                  | 1064 nm (non Q-Switched - pulsed mode)                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Pulse width                                                                      | N/A                                                                                                                                                                                                                                                                                      | 0.25 to 300 ms                                                                                                                                                                                                                                                                      | 0.3 to 50 ms                                                                                                                                                                                                                                                                                                                                    |
| Max Fluence                                                                      | N/A                                                                                                                                                                                                                                                                                      | 300 J/cm²                                                                                                                                                                                                                                                                           | 300 J/cm²                                                                                                                                                                                                                                                                                                                                       |
| Spot Size                                                                        | N/A                                                                                                                                                                                                                                                                                      | 2 to 24 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 9x9 mm²<br>Squared                                                                                                                                                                                                                | 2 to 8 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 7x7 mm2<br>Squared                                                                                                                                                                                                                                                                             |
| Repetition<br>Rate                                                               | N/A                                                                                                                                                                                                                                                                                      | Up to 10 Hz                                                                                                                                                                                                                                                                         | Up to 10 Hz                                                                                                                                                                                                                                                                                                                                     |
|                                                                                  | 532nm (Q-Switched, nanosecond mode)                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Homogeneous spot                                                                 |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Pulse width                                                                      | 6 to 12 ns                                                                                                                                                                                                                                                                               | 6 to 12 ns                                                                                                                                                                                                                                                                          | 6 to 12 ns                                                                                                                                                                                                                                                                                                                                      |
| Max Fluence                                                                      | Up to 14 J/cm²                                                                                                                                                                                                                                                                           | Up to 19 J/cm²                                                                                                                                                                                                                                                                      | Up to 15 J/cm²                                                                                                                                                                                                                                                                                                                                  |
| Spot Size                                                                        | 2 to 12 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 7x7 mm²<br>Squared                                                                                                                                                                                                                     | 2, 4, 6, 8 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5 mm² Squared                                                                                                                                                                                                                     | 3 to 10.5 mm diameter Round;<br>2x2, 3x3, 4x4, 5x5, 7x7 mm2<br>Squared                                                                                                                                                                                                                                                                          |
| Repetition<br>Rate                                                               | Up to 10 Hz                                                                                                                                                                                                                                                                              | Up to 10 Hz                                                                                                                                                                                                                                                                         | Up to 20 Hz                                                                                                                                                                                                                                                                                                                                     |
| High<br>coverage<br>microbeam<br>Handpiece                                       | Present                                                                                                                                                                                                                                                                                  | Absent                                                                                                                                                                                                                                                                              | Present                                                                                                                                                                                                                                                                                                                                         |
| Microbeam<br>handpiece<br>Fluence per<br>dot                                     | 3.5 J/cm²                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                                                 | from 0.05 to 5.95 J/cm2                                                                                                                                                                                                                                                                                                                         |
| Specification                                                                    | Main predicate device                                                                                                                                                                                                                                                                    | Additional reference device                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                                  |
| Device Name<br>(K number)                                                        | Pico Family K191842                                                                                                                                                                                                                                                                      | EVO Platform K160368                                                                                                                                                                                                                                                                | Chrome                                                                                                                                                                                                                                                                                                                                          |
| Submitter                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                        | Quanta System SpA                                                                                                                                                                                                                                                                   | Quanta System SpA                                                                                                                                                                                                                                                                                                                               |
| Microbeam<br>handpiece<br>spot<br>dimension<br>(mm)                              | 9 mm diameter                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                 | 9 mm diameter                                                                                                                                                                                                                                                                                                                                   |
| Standard<br>microbeam<br>Handpiece                                               | Present                                                                                                                                                                                                                                                                                  | Absent                                                                                                                                                                                                                                                                              | Present                                                                                                                                                                                                                                                                                                                                         |
| Microbeam<br>handpiece<br>Fluence per<br>dot                                     | Up to 14 J/cm²                                                                                                                                                                                                                                                                           | N/A                                                                                                                                                                                                                                                                                 | From 0.3 to 14.9 J/cm2                                                                                                                                                                                                                                                                                                                          |
| Microbeam<br>handpiece<br>spot<br>dimension<br>(mm)                              | 8 mm diameter                                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                                                                                 | 8 mm diameter                                                                                                                                                                                                                                                                                                                                   |
| IPL 650-1200 nm, 625-1200 nm, 590-1200 nm, 570-1200 nm, 550-1200 nm, 400-1200 nm |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                 |
| Pulse width                                                                      | Up to 40ms                                                                                                                                                                                                                                                                               | Up to 40ms                                                                                                                                                                                                                                                                          | Up to 40ms                                                                                                                                                                                                                                                                                                                                      |
| Max Fluence                                                                      | up to 25 J/cm²                                                                                                                                                                                                                                                                           | up to 25 J/cm²                                                                                                                                                                                                                                                                      | up to 25 J/cm²                                                                                                                                                                                                                                                                                                                                  |
| Spot Size                                                                        | 48mm x13 mm<br>25mm x13 mm                                                                                                                                                                                                                                                               | 48mm x13 mm<br>25mm x13 mm                                                                                                                                                                                                                                                          | 48mm x13 mm<br>25mm x13 mm                                                                                                                                                                                                                                                                                                                      |
|                                                                                  | Repetition<br>Rate                                                                                                                                                                                                                                                                       | up to 0,5 Hz                                                                                                                                                                                                                                                                        | up to 0,5 Hz                                                                                                                                                                                                                                                                                                                                    |

{9}------------------------------------------------

{10}------------------------------------------------

#### Performance data:

The subject device was subjected to performance testing in accordance with the following recognized consensus standards:

- IEC 60601-1:2012, ed 3.1, Medical Electrical Equipment Part 1: General Requirements For Basic i Safety And Essential Performance
- i IEC 60601-1-2 Edition 4: 2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests.
- । IEC 60601-2-22: 2012-10 ed 3.1, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment
- । IEC 60825-1 Ed. 3.0 (2014) Safety of laser products – Part 1: Equipment classification and requirements

The following testing were performed on the subject device:

{11}------------------------------------------------

- -Software Verification and Validation Testing: Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
#### Biocompatibility:

Biocompatibility of the subject device was established based on a comparison to the legally-marketed predicate device, as the patient-contacting components of the subject device in its final finished form are identical to that of K191842 in formulation, processing, sterilization, and geometry, and no other chemicals have been added.

#### Conclusions:

The subject device has the same intended use, the same indications for use and the same fundamental scientific technology as its predicates.

The performance specification of the subject device are equivalent to its identified predicate devices and therefore do not present any new concerns of safety and effectiveness.

Therefore, Chrome is as safe, as effective, and performs as well as the legally marketed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K202503](https://fda-staging.innolitics.com/device/K202503)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).