Aulisa Multiple Patient Digital Vital Sign Monitoring System
K202497 · Taiwan Aulisa Medical Devices Technologies, Inc. · MSX · Oct 30, 2020 · Cardiovascular
Device Facts
Record ID
K202497
Device Name
Aulisa Multiple Patient Digital Vital Sign Monitoring System
Applicant
Taiwan Aulisa Medical Devices Technologies, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Oct 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000, is intended to provide remote central monitoring and display of information as recorded by multiple Aulisa single-patient monitoring systems, on a central remote screen. The system can be used in a hospital type and clinic environment.
Device Story
System acts as central display station for multiple Aulisa single-patient monitors; receives physiological parameters and alarm data via standard wired or wireless LAN; software runs on off-the-shelf PC (Windows 10). Used in hospitals/clinics by healthcare staff to view patient status remotely; supports viewing, exporting, and printing of patient history and reports. Does not perform data analysis; duplicates data from individual units. Does not allow remote adjustment of monitoring unit parameters. Benefits include centralized oversight of multiple patients, facilitating timely clinical response to alarms.
Clinical Evidence
Bench testing only. Software verification and validation performed per FDA guidance; risk management assessment conducted per ISO 14971. No clinical data presented.
Technological Characteristics
Stand-alone software running on off-the-shelf PC hardware; Windows 10 OS; connectivity via standard wired or wireless LAN. Visual and auditory alarm notification. No data analysis performed; data duplication only.
Indications for Use
Indicated for remote central monitoring and display of physiological parameters and alarm data from multiple Aulisa single-patient monitoring systems in hospital or clinic environments.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2020
Taiwan Aulisa Medical Devices Technologies, Inc. Don Mizota Consultant 725 Morninghome Road Danville, California 94526
Re: K202497
Trade/Device Name: Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX, MWI Dated: August 25, 2020 Received: August 31, 2020
Dear Don Mizota:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, & Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202497
Device Name
Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000
Indications for Use (Describe)
The Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000, is intended to provide remote central monitoring and display of information as recorded by multiple Aulisa single-patient monitoring systems, on a central remote screen. The system can be used in a hospital type and clinic environment.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92
### 5.1. General Information
| Date of Preparation: | August 25, 2020 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by: | Taiwan Aulisa Medical Devices Technologies, Inc.<br>Rm. 1052, Bldg. H, 10F., No.3-2, YuanQu St.,<br>Nangang Dist.,Taipei City 115 TW<br>Phone: 886-2-2655-7297<br>FAX: 886-2-2655-7260 |
| Contact Person: | Paul Liu<br>Regulatory Affairs Supervisor<br>Taiwan Aulisa Medical Devices Technologies, Inc.<br>Rm. 1052, Bldg. H, 10F., No.3-2, YuanQu St.,<br>Nangang Dist.,Taipei City 115 TW<br>Phone: 886-2-2655-7297<br>FAX: 886-2-2655-7260<br>Email: paul.liu@aulisa.com |
### 5.2. Trade/Device Name
Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000
### 5.3. Regulatory Information
| Regulation Number | Regulation Name | Regulation Class | Product Code |
|-------------------|--------------------------------------------------------------------------------|------------------|--------------|
| 21 CFR 870.2300 | System, Network and<br>Communication,<br>Physiological Monitors | Class II | MSX |
| 21 CFR 870.2300 | Monitor, Physiological,<br>Patient (Without Arrhythmia<br>Detection or Alarms) | Class II | MWI |
### 5.4. Predicate Device
### Predicate
K151006, EarlySense Central Display Station, EarlySense Ltd.
{4}------------------------------------------------
### 5.5. Device Description
The Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000, operates through an Aulisa-developed software installed on a standard personal computer. The system interlinks multiple Aulisa single-patient monitoring systems to display the information as recorded by each single-patient monitoring system on a central remote screen. The system uses standard off-theshelf communication and IT hardware for data transmission.
The information transmitted from single-patient monitoring systems to the MP1000 system includes the physiological parameters and alarm indications. Adjusting alarm settings can only be performed on the Aulisa Display Unit of the single-patient monitoring system. If configured, the MP1000 screen can also be accessed through remote devices, such as other PCs. In addition, patient history data and report files can be viewed, exported, and/or printed.
#### 5.6. Intended Use
The Aulisa Multiple Patient Digital Sign Monitoring System. MP1000, is intended to provide remote central monitoring and display of information as recorded by multiple Aulisa singlepatient monitoring systems, on a central remote screen. The system can be used in a hospital type and clinic environment.
### 5.7. Comparison with Predicate
The subject device has similar intended use and technology characteristics to the predicate device, which is to provide remote central monitoring through the display of physiological parameters as recorded by individual monitoring units on a central remote screen. The physiological parameters displayed on the subject device are similar to the parameters displayed by the predicate device, EarlySense Central Display Station. Both devices classify as stand-alone software and run on an off-the-shelf computer, with the subject device operating on a Windows 10 operating system and the predicate device on a Linux operating system. The communication method used on both devices for data transmission is either via standard wired or via wireless LAN. The subject device provides the accessibility to remotely view the central remote screen from additional PCs, like its predicate device.
Additionally, the MP1000 and its predicate device are used in the same clinical environmentshospitals or hospital type and clinic environment.
Further, audio/visual alarm notification is provided for both subject and predicate devices. Both devices do not perform analysis of data, but only duplicate the data received from the individual monitoring units. The subject device does not allow for the adjustment of the monitoring unit's settable parameters from the remote station, whereas the predicate device does. These differences do not affect the safety and effectiveness of the device when used as labeled.
The comparison table for the subject device versus the predicate device, EarlySense Central Display Station (K151006) is shown in Table 5.1.
{5}------------------------------------------------
| Table 5.1 - Comparison with Predicate | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Subject Device | Predicate Device |
| Item | Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000 | EarlySense Central Display Station, (K151006) |
| Indication for use | The Aulisa Multiple Patient Digital Vital Sign Monitoring System, MP1000, is intended to provide remote central monitoring and display of information as recorded by multiple Aulisa single-patient monitoring systems, on a central remote screen. The system can be used in a hospital type and clinic environment. | The EarlySense Central Display System is intended to provide remote central monitoring and display of information as recorded by multiple EarlySense bedside units, on a central remote screen. The system can be used in hospitals or hospital type and clinic environment. |
| Environment of use | Hospitals or hospital type and clinic environment | Hospitals or hospital type and clinic environment |
| Alarms | Visual and auditory alarms | Visual and auditory alarms |
| Operating System | Windows 10 | Linux Debian |
| Communication Method | LAN and wireless LAN | LAN and wireless LAN |
| Alarm Adjustment Capability | No | Yes |
K202497
### Summary of Performance Testing 5.8.
Yes
Remote Accessibility
Aulisa Multiple Patient Digital Vital Sign Monitoring System is a medical device stand-alone software; all testing was completed and the software has been verified and validated per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Further, risk management assessment on the subject device was performed per ISO 14971, Medical Devices- Risk Management to verify that it does not raise any new safety and effectiveness issues in comparison to its predicate device.
Yes
Test results indicate that the subject device complies with its predetermined specifications and the applicable guidance document and standards.
#### 5.9. Substantially Equivalent Conclusion
Based on the testing summarized in this 510 (k) submission, the results demonstrate that the subject device is substantially equivalent to the predicate device. The differences do not raise any questions of safety or effectiveness when compared to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.