K202492 · Teleflex Medical · FMI · Mar 22, 2021 · General Hospital
Device Facts
Record ID
K202492
Device Name
EZ-IO Intraosseous Vascular Access System
Applicant
Teleflex Medical
Product Code
FMI · General Hospital
Decision Date
Mar 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours. Insertion sites: ADULTS (≥22 years old) - Proximal humerus - Proximal tibia - · Distal tibia PEDIATRICS (<21 years old) - Proximal humerus - Proximal tibia - · Distal tibia - · Distal femur For patients ≥12 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
Device Story
The EZ-IO Intraosseous Vascular Access System provides vascular access by inserting a needle set (stylet and catheter) into the bone marrow medullary space. The system consists of a reusable cordless battery-powered driver and single-use needle sets (15mm, 25mm, 45mm) made of 304 stainless steel with polycarbonate hubs. Used in pre-hospital, hospital, and acute care settings by clinicians. The clinician identifies anatomical landmarks, attaches the needle set to the driver, and drills through the bone cortex until a tactile 'pop' indicates entry into the medullary space. The driver is detached, the stylet is removed, and the catheter is connected to standard IV tubing via a Luer lock for fluid/medication administration. The device facilitates rapid vascular access when traditional IV access is difficult, potentially life-saving in emergent scenarios.
Clinical Evidence
Bench testing only. MRI compatibility testing was conducted per ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 to evaluate RF-induced heating in 1.5T and 3T scanners, supporting the MR Conditional labeling.
Technological Characteristics
Needle sets: 304 stainless steel cannula/stylet, polycarbonate hubs, 15G (1.8mm) diameter, 15/25/45mm lengths. Reusable cordless driver powered by lithium batteries. Biocompatible materials per ISO 10993-1. Sterilization: Ethylene Oxide. Depth control via visual markers and tactile feedback. MR Conditional status.
Indications for Use
Indicated for adult and pediatric patients requiring vascular access when difficult to obtain in emergent, urgent, or medically necessary cases. Contraindications: fracture in target bone; previous significant orthopedic procedure, prosthetic limb, or joint at site; IO access in targeted bone within past 48 hours; infection at insertion site; excessive tissue/obesity preventing identification of anatomical landmarks.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
EZ-IO Intraosseous Vascular Access System (K180395)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 22, 2021
Teleflex Medical Rachel Rehl Regulatory Affairs Specialist 3015 Carrington Mill Blyd. Morrisville, North Carolina 27560
Re: K202492
Trade/Device Name: EZ-IO Intraosseous Vascular Access System Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: February 19, 2021 Received: February 23, 2021
Dear Rachel Rehl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202492
#### Device Name
EZ-IO Intraosseous Vascular Access System
Indications for Use (Describe)
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours.
Insertion sites:
ADULTS (≥22 years old)
- · Proximal humerus
- Proximal tibia
- · Distal tibia
PEDIATRICS (<21 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
- · Distal femur
For patients ≥12 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K202492 510(k) SUMMARY
### EZ-IO Intraosseous Vascular Access System MR Conditional Safety Status Labeling
### 1. Submitter Information
| Name: | Teleflex Medical |
|-------------------|---------------------------------------------------------|
| Address: | 3015 Carrington Mill Blvd.<br>Morrisville, NC 27560 USA |
| Contact Person: | Rachel Rehl |
| Telephone Number: | (919)433-2588 |
| Email: | rachel.rehl@teleflex.com |
Date Prepared: August 28, 2020
## 2. Device Name
Device Trade Name: EZ-IO Intraosseous Vascular Access System Common Name: Intraosseous Infusion System Classification Name: Needle, Hypodermic, Single Lumen (Class II, FMI, 21 CFR 880.5570)
### 3. Predicate Device
EZ-IO Intraosseous Vascular Access System (K180395)
### 4. Device Description and Clinical Operation
#### Device Description
The EZ-IO Intraosseous Vascular Access System previously cleared with K180395 is designed to allow the user to insert a needle set consisting of a stylet and catheter through the cortex of the bone to a desired depth within the bone marrow to facilitate intraosseous infusion of desired fluids and medications for vascular access. All system materials are biocompatible. Needle sets are single-use and composed of 304 stainless steel with polycarbonate hubs and available in 15 mm (for patients 3-39 kg); 25 mm (for patients 3 kg or over) and 45 mm (for patients 40 kg or over) lengths. Black lines on the needle set serve as depth markers. The reusable cordless driver/drill is powered by lithium batteries with a battery-power indicator light. An extension tubing set accessory, the EZ-Connect, is included with every needle set. The EZ-Connect contains a needleless connector system and Luer lock adapter. An optional dressing, the EZ-Stabilizer, is an accessory to the EZ-IO Vascular Access System. It is designed as a securement device with an adhesive backing that is placed over an EZ-IO Needle to
{4}------------------------------------------------
keep the needle securely anchored to the patient; and is recommended in the Instructions for use (IFU).
## Clinical Operation
Clinicians locate anatomical landmarks and clean the insertion site. Using the cordless driver with needle set attached, the needle set is pressed through the soft tissue to the outer cortex of the bone. At least one depth marker on the cannula must be visible prior to powering the driver to ensure adequate needle length for proper placement within the medullary space. Clinicians then squeeze the driver trigger and apply moderate, steady pressure. The trigger is released when a sudden "give" or "pop" is felt, which indicates entry into the medullary space; the needle set will not always be inserted to the hub. After insertion of the needle set, the driver unit is detached from the needle set, leaving the stylet and cannula firmly seated in the bone. The stylet is then separated and removed from the cannula by turning the stylet hub counterclockwise leaving the catheter with a standard Luer lock hub securely seated in the bone. The cannula Luer lock permits attachment of the provided EZ-Connect, standard syringes or other IV tubing for administration of medications and fluids.
### 5. Indications for Use
For intraosseous access anytime in which vascular access is difficult to obtain in emergent, urgent or medically necessary cases for up to 24 hours.
Insertion sites:
ADULTS (≥22 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
# PEDIATRICS (≤21 years old)
- Proximal humerus
- Proximal tibia
- · Distal tibia
- · Distal femur
For patients 2 years old, use of the device may be extended for up to 48 hours when alternate intravenous access is not available or reliably established.
- 6. Technological Characteristics and SubstantialEquivalence
The proposed device is identical to the predicate device described in K180395 in design, materials of construction, functional performance, principles of operation, manufacturing, packaging, sterilization, and shelf life.
{5}------------------------------------------------
The only change is the labeled MR Conditional Safety Status. Table 5-1 below provides a comparison of the proposed and predicate devices.
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Indications<br>for Use | For intraosseous access anytime in<br>which vascular access is difficult to<br>obtain in emergent, urgent or<br>medically necessary cases for up to 24<br>hours.<br>Insertion sites:<br>ADULTS (≥22 years old)<br>• Proximal humerus<br>• Proximal tibia<br>• Distal tibia<br><br>PEDIATRICS (≤21 years old)<br>• Proximal humerus<br>• Proximal tibia<br>• Distal tibia<br>• Distal femur<br><br>For patients ≥12 years old, use of the<br>device may be extended for up to 48<br>hours when alternate intravenous<br>access is not available or reliably<br>established. | For intraosseous access anytime in<br>which vascular access is difficult to<br>obtain in emergent, urgent, or<br>medically necessary cases for up to 24<br>hours. Insertion sites:<br>ADULTS (≥22 years old): proximal<br>humerus, proximal tibia, distal tibia<br>PEDIATRICS (≤21 years old): proximal<br>humerus, proximal tibia, distal tibia,<br>distal femur<br><br>Use of the device may be extended for<br>up to 48 hours when alternate<br>intravenous access is not available or<br>reliably established. Insertion sites:<br>ADULTS (≥22 years): proximal<br>humerus, proximal tibia, distal tibia<br>PEDIATRICS (≥12 years through 21<br>years old): proximal humerus,<br>proximal tibia, distal tibia, distal femur | Wording<br>updated for<br>clarity but the<br>indications for<br>use are the<br>same. |
| Contraindicati<br>ons | • Fracture in target bone.<br>• Previous, significant orthopedic<br>procedure at the site,<br>prosthetic limb or joint.<br>• IO access (or attempted IO<br>access) in targeted bone within<br>past 48 hours.<br>• Infection at the area of<br>insertion.<br>• Excessive tissue (severe<br>obesity) and/or absence of<br>adequate anatomical<br>landmarks. | • Fracture in target bone.<br>• Previous, significant orthopedic<br>procedure at the site,<br>prosthetic limb or joint.<br>• IO access (or attempted IO<br>access) in targeted bone within<br>past 48 hours.<br>• Infection at the area of<br>insertion.<br>• Excessive tissue (severe<br>obesity) and/or absence of<br>adequate anatomical<br>landmarks. | Same |
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
| Target<br>Population | Adult and pediatric patients who<br>are in need of vascular access. | Adult and pediatric patients who<br>are in need of vascular access. | Same |
| Where Used | Pre-hospital, In hospital, Acute care | Pre-hospital, In hospital, Acute care | Same |
| Anatomical<br>Sites Used | Proximal Tibia Proximal Humerus Distal Tibia Distal Femur in pediatric<br>population | Proximal Tibia Proximal Humerus Distal Tibia Distal Femur in pediatric<br>population | Same |
| Needle/Cannula<br>Design | Sterile, single use Hubs: polycarbonate and color<br>additive Stylet/catheter: Stainless Steel Faceted tip Standard Luer connection 15 mm;<br>25 mm; 45 mm 15 gauge (0.071", 1.8 mm) Needle Cover: Polypropylene | Sterile, single use Hubs: polycarbonate and color<br>additive Stylet/catheter: Stainless Steel Faceted tip Standard Luer connection 15 mm;<br>25 mm; 45 mm 15 gauge (0.071", 1.8 mm) Needle Cover: Polypropylene | Same |
| Needle Set<br>Guidelines | Available Needle Sets: 15 mm: 3-39 kg 25 mm: 3 kg or over 45 mm: 40 kg or over | Available Needle Sets: 15 mm: 3-39 kg 25 mm: 3 kg or over 45 mm: 40 kg or over | Same |
| Depth Control | Positioning marks at 5 mm and 10<br>cm apart to provide visual reference<br>points<br><br>Tactile feedback for change of<br>pressure | Positioning marks at 5 mm and 10<br>mm apart to provide visual reference<br>points<br><br>Tactile feedback for change of<br>pressure | Same |
| Sterile single<br>use components<br>and accessories | EZ-IO Needle Sets EZ-Connect Extension Set EZ-IO Patient Wristband NeedleVISE 1-port Sharps Block EZ-Stabilizer Dressing | EZ-IO Needle Sets EZ-Connect Extension Set EZ-IO Patient Wristband NeedleVISE 1-port Sharps Block EZ-Stabilizer Dressing | Same |
| Sterility of<br>single use<br>components<br>and accessories | Ethylene Oxide | Ethylene Oxide | Same |
| Shelf life of<br>single use<br>components<br>and accessories | 4 Years | 4 Years | Same |
| | Proposed Device<br>EZ-IO Intraosseous Vascular<br>Access System<br>(MR Conditional Safety<br>Status Labeling) | Predicate Device K180395<br>EZ-IO Intraosseous Vascular<br>Access System | Comparison |
| Biocompatibility | Biocompatible materials used (per<br>ISO 10993-1 prolonged contact<br>duration). | Biocompatible materials used (per<br>ISO 10993-1 prolonged contact<br>duration). | Same |
| MR Safety<br>Status | MR Conditional | MR Unsafe | Update to MR<br>Safety Status<br>as described in<br>submission. |
Table 5-1: Substantial Equivalence Comparison to Predicate
{6}------------------------------------------------
{7}------------------------------------------------
# 7. Non-clinical Testing
Non-clinical testing was conducted as the basis of the MR Conditional safety status claims that are proposed in this 510(k). The testing is listed in Table 5-2 below. Summaries of the testing are provided as well. The test article was the 45 mm needle.
| Testing | Summary |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MRI Testing,<br>Exponent, 2020 | This study evaluated the MRI compatibility of the EZ-IO Needle Set in regard to radiofrequency (RF)-induced heating in 1.5 T and 3 T clinical scanners. The evaluation was conducted using methodologies prescribed in ASTM F2052, ASTM F2213, ASTM F2182, and ASTM F2119 as a guide.<br><br>The evaluation concluded that the EZ-IO Needle Set should be labeled as MR Conditional and included labeling guidelines in the report. |
# Table 5-2: Non-Clinical Testing
# 8. Conclusions
The conclusions drawn from the non-clinical tests inform the MRI labeling that is proposed in this 510(k). As there is no change to the device itself, we can conclude that the device is substantially equivalent to the predicate 510(k) K180395.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.