K202490 · Vision Quest Industries Inc./Dba VQ Orthocare · IPF · Nov 16, 2020 · Physical Medicine
Device Facts
Record ID
K202490
Device Name
Avid CT2
Applicant
Vision Quest Industries Inc./Dba VQ Orthocare
Product Code
IPF · Physical Medicine
Decision Date
Nov 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Avid CT2 Neuromuscular and Interferential Stimulator, Model AV-CT20A. is indicated for use in the following applications: Interferential Stimulation can be used in the following applications: - Symptomatic relief of post-surgical and/or post traumatic acute pain - Symptomatic relief of chronic intractable pain - Relaxation of muscle spasms - Maintain or increase range of motion - Increase local blood circulation Neuromuscular Stimulation can be used in the following applications: - Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis - Prevention or retardation of disuse atrophy - Muscle-re-education
Device Story
Avid CT2 is a combination therapy device providing interferential and neuromuscular electrical stimulation. It delivers low electrical current via lead wires to skin-surface electrodes. Input parameters are set by a clinician; end-users select protocols and adjust amplitude via an LCD touchscreen and keypad. The device uses a microcontroller to gate the interferential waveform on and off, enabling neuromuscular stimulation control. It is self-contained, powered by internal rechargeable lithium-ion batteries or an external power supply. Data can be uploaded via wired or wireless interface. The device allows preset-to-preset transitions without patient interaction to facilitate physician-prescribed protocols. It is intended for use in clinical or home settings under clinician guidance to manage pain, prevent atrophy, and support muscle re-education.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through comparative analysis of electrical output specifications, waveform characteristics (scope traces), and device functionality against predicate devices.
Technological Characteristics
Microcontroller-based stimulator; molded ABS/PC plastic housing; two internal non-removable rechargeable lithium-ion batteries or external power supply; transformer-coupled channel isolation; LCD touchscreen interface; wired/wireless data connectivity. Complies with IEC60601-1 (safety) and IEC60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring symptomatic relief of acute post-surgical/post-traumatic or chronic intractable pain, muscle spasm relaxation, range of motion maintenance/increase, local blood circulation improvement, prevention of venous thrombosis via calf muscle stimulation, prevention/retardation of disuse atrophy, and muscle re-education.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
11/16/2020
Vision Quest Industries Inc. Mohamed Ouerghi Director of QA/RA 1390 Decision Street. Suite A Vista, California 92081
Re: K202490
Trade/Device Name: Avid CT2 Neuromuscular and Interferential Stimulation System Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, LIH Dated: October 2, 2020 Received: October 5, 2020
Dear Mohamed Ouerghi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaorui Tang, Ph.D. Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K202490
#### Device Name
Avid CT2, Neuromuscular and Interferential Stimulation System, Model AV-CT20A
#### Indications for Use (Describe)
Interferential Stimulation can be used in the following applications:
- · Symptomatic relief of post-surgical and/or post traumatic acute pain
- · Symptomatic relief of chronic intractable pain
- Relaxation of muscle spasms
- Maintain or increase range of motion
- Increase local blood circulation
Neuromuscular Stimulation can be used in the following applications:
- · Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
- · Prevention or retardation of disuse atrophy
- · Muscle-re-education
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| 510(k) Owner: | Vision Quest Industries, Inc.<br>18011 Mitchell South,<br>Irvine, CA, 92614 |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mohamed Ouerghi<br>Director of QA/RA<br>Vision Quest Industries, Inc.<br>Phone 760-477-8201<br>Mobile 760-691-0168<br>Fax 760-727-5950<br>mouerghi@vqorthocare.com |
| Date Summary Prepared: | 8/26/2020 |
| Proprietary Name: | Avid CT2 Neuromuscular and<br>Interferential Stimulation System,<br>Model AV-CT20A |
| Device Name and Classification: | Neuromuscular and Interferential Stimulator, Class II,<br>21 CFR 882.5890, Product Code LIH and<br>21CFR 890.5850, Product Code IPF |
| Predicate Devices: | Surgi Stim/T.E.A.R. Tech by Vision Quest Industries, Inc.<br>K982388 and<br>Avid IF2 by Vision Quest Industries, Inc.K183692 |
| Device Description: | The Avid CT2, Model AV-CT20A is a combination therapy<br>device. Like its predicate Avid IF2, it is an Interferential<br>Stimulator that produces a low electrical current that is<br>transmitted via lead wires to electrodes placed on the skin in the<br>area predetermined by a clinician. Operating parameters can be<br>adjusted throughout their range by a trained clinician but the<br>end-user is limited to protocol selection and amplitude. The user<br>interface consists of an LCD display and a keypad. The primary<br>difference between the two devices is the addition of user-<br>adjustable parameters that allow the existing interferential<br>waveform to turn on and off within a small, preselected range to<br>provide necessary control for neuromuscular stimulation. This<br>same method of gating the interferential on and off was used in<br>VQ's previous Surgi Stim stimulator. |
{4}------------------------------------------------
Statement of Intended Use:
The Avid CT2 Neuromuscular and Interferential Stimulator, Model AV-CT20A. is indicated for use in the following applications:
Interferential Stimulation can be used in the following applications:
- · Symptomatic relief of post-surgical and/or post traumatic acute pain
- · Symptomatic relief of chronic intractable pain
- Relaxation of muscle spasms
- · Maintain or increase range of motion
- Increase local blood circulation
Neuromuscular Stimulation can be used in the following applications:
- · Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
- · Prevention or retardation of disuse atrophy
- Muscle-re-education
#### Substantial Equivalence
The Avid CT2 is a minor product enhancement to the Avid IF2. The primary difference between the two devices is the addition of user-adjustable parameters that allow the existing interferential waveform to turn on and off within a small, preselected range to provide necessary control for neuromuscular stimulation. This same method of gating the interferential on and off was used in VQ's previous Surgi Stim stimulator.
The Avid CT2 is equivalent to the Avid IF2 in all areas except these new user-adjustable parameters. For those parameters we are substantially equivalent to VO's Surgi Stim stimulator..
#### Indications for Use
The Avid CT2 has the same indications for use as the Avid IF2 when these are used in IF mode. The Avid CT2 has the additional indications for use as allowed by neuromuscular stimulators.
### Device Functionality Equivalency
- Like the predicate devices, the new device uses a microcontroller and LCD display to create a user friendly interface.
- The Avid IF2 and CT2 Stimulator is self-contained and includes two non-removable, rechargeable lithium ion batteries and an external power supply like the predicates.
- The Avid IF2 and CT2 Stimulator also contain a single output jack for both output channels, and a user interface consisting of a touchscreen LCD for improved user experience. The electrodes used for stimulation are the same used with the predicates. The lead wires are of proprietary design.
- The user is able to select presets on the device for the desired treatment and waveform output based upon prescriptions from the treating clinician. This is accomplished by displayed menu items and selection through the device interface. If desired, the user will be able to upload data stored on the device to Vision Ouest Industries, Inc. (via wired interface or wirelessly). The device has the ability to move from preset to preset without patient interaction. This allows for easy use of physician prescribed protocols.
{5}------------------------------------------------
# Device Characteristics and Output Specifications Equivalency
The Avid IF2 only has the IF mode whereas the Avid CT2 has an additional neuromuscular mode. The second predicate device from Vision Quest Industries, Inc. has three modes of stimulation: High Volt Pulsed Current (HVPC), Interferential (IF), and a Neuromuscular Electrical Stimulation (NMES) mode.
The tables below compare the Avid CT2 to the two predicate devices.
{6}------------------------------------------------
# Device Characteristics Comparison
| 510(k) Number | K982388 | K183692 | Unassigned |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Device Name | Surgi Stim /T.E.A.R. Tech | Avid IF2 | Avid CT2 |
| Manufacturer | Vision Quest Industries, Inc. | Vision Quest Industries, Inc. | Vision Quest Industries, Inc. |
| Power Source | 2 battery packs consisting of<br>4 'AA' alkaline cells each or<br>external power supply | 2 internal, non-removable,<br>rechargeable Lithium-ion<br>batteries or external power<br>supply | 2 internal, non-removable,<br>rechargeable Lithium-ion<br>batteries or external power<br>supply |
| —Method of Line Current<br>Isolation | Use of UL2601-1 approved<br>external power supply | Use of UL2601-1 approved<br>external power supply | Use of UL2601-1 approved<br>external power supply |
| —Patient Leakage Current | | | |
| —Normal Condition<br>(μΑ) | <500 | <500 | <500 |
| —Single fault<br>condition (μΑ) | <500 | <500 | <500 |
| No. of Output Modes | 3 (IF, HVPC, NMES) | 4 (IF) | 1 (IF, NMES) |
| No. of Output Channels | IF Mode – 2<br>IF Mode – 1<br>NMES Mode– 2<br>NMES Mode –1<br>HVPC Mode – 1 | IF Mode – 2<br>IF Mode – 1 | IF Mode – 2<br>IF Mode – 1<br>NMES Mode – 2<br>NMES Mode –1 |
| Synchronous or<br>Alternating | IF — Synchronous<br>NMES — Synchronous<br>HVPC — Synchr. Or Alt. | IF — Synchronous | IF — Synchronous<br>NMES — Synchronous |
| Method of Channel<br>Isolation | IF - Transformer coupled<br>NMES - Transformer<br>coupled<br>HVPC – N/A | IF - Transformer coupled | IF - Transformer coupled<br>NMES - Transformer coupled |
| Reciprocal | IF - No<br>NMES - No<br>HVPC - Yes | IF - No | IF – No<br>NMES - No |
| Regulated Current or<br>Regulated Voltage | IF - Regulated voltage<br>NMES - Regulated voltage<br>HVPC - Regulated voltage | IF - Regulated voltage | IF - Regulated voltage<br>NMES– Regulated voltage |
| Software/Firmware/<br>Microprocessor Control | Microprocessor Control | Microprocessor Control | Microprocessor Control |
| Software Provided | Yes-Embedded Firmware | Yes- Embedded Firmware | Yes- Embedded Firmware |
| Automatic Overload Trip | No | Yes | Yes |
| Automatic No-Load Trip | No | Yes (w/override option) | Yes (w/override option) |
| Automatic Shut Off | Yes | Yes | Yes |
| Patient Override Control | Yes | Yes | Yes |
| Indicator Display: | | | |
| Unit Functioning | Yes | Yes | Yes |
| On/Off Status | Yes | Yes | Yes |
| Low Battery | Yes | Yes | Yes |
| Voltage/Current<br>Level | 4.4V | 5.75V | 6.0V |
| Other | LCD panel displays all<br>parameter settings. | LCD panel displays all<br>parameter settings. | LCD panel displays all<br>parameter settings. |
| 510(k) Number<br>Device Name | K982388<br>Surgi Stim /T.E.A.R. Tech | K183692<br>Avid IF2 | Unassigned<br>Avid CT2 |
| Manufacturer | Vision Quest Industries,<br>Inc. | Vision Quest Industries,<br>Inc. | Vision Quest Industries, Inc. |
| Constant Voltage | NMES- No<br>HVPC- No<br>IF- Yes<br>NMES- Yes<br>HVPC- Yes | NMES - No<br>IF - Yes | NMES - No<br>IF - Yes<br>NMES - Yes |
| Timer Range (minutes) | | | |
| Timer Settings | 10 min to 8 hours or<br>continuous | 1 min to 24 hours or<br>continuous | 1 min to 24 hours or continuous |
| Compliance with<br>voluntary Standards | Standards-AAMI/ANSI NS4<br>1986 | NA | NA |
| Compliance with<br>EN60601-1 (Safety) | Not Tested | Yes | Yes |
| Compliance with<br>IEC60601-1-2 (EMC) | Not Tested | Yes | Yes |
| Compliance with 21<br>CFR 898 (Mandatory<br>05/09/02) | NA | Yes | Yes |
| Weight (with batteries) | 10.6 oz. | 6.8 oz. | 6.8 oz. |
| Dimensions (inches) | 5.7 x 3.0 x 1.5 | 4.9 x 2.85 x 1.0 | 4.9 x 2.85 x 1.0 |
| Housing Materials and<br>Construction | Molded ABS/PC plastic<br>housing | Molded ABS/PC plastic<br>housing | Molded ABS/PC plastic housing |
{7}------------------------------------------------
# Device Characteristics Comparison
{8}------------------------------------------------
# Technical Explanations:
The following device performance description/comparison to predicate devices are provided in accordance with the FDA document "Guidance Document for Powered Muscle Stimulator 510(k)s" Attachment II section 3, issued on June 9, 1999.
The above document also requires some explanations of calculations - these are provided below.
# Interference Pattern
The interference pattern is created using two different frequencies. When the four electrode stimulation is selected two separate frequencies are provided on the two electrode pairs.
Interference occurs at the patient. When two electrodes stimulation is selected the two frequencies are combined inside the device and the interference pattern is delivered via the one electrode pair.
# Current Density
Current density is calculated using 2 different electrode sizes. A 2" round electrode equal to 20.27 sq. cm and a 2" x 1.25" rectangular electrode is equal to 16.13 sq. cm. This second electrode is used in the Limited mode output where amplitude is limited to 60% of full power.
In the IF mode current density is the pulsed current over the electrode area. Each phase is 50% of the pulse thus the average is given as half.
### Power Density
Power density is calculated in a similar manner to current density except that the peak phase power density is the max voltage times the max current.
### Maximum Phase Charge
In the IF mode, charge (Q) can be calculated as follows:
(Peak voltage/load) X duration of pulse
{9}------------------------------------------------
#### Waveform Drawings Explanations
The waveform drawings are provided in accordance with the "Guidance Document for Powered Muscle Stimulator 510(k) s".
## Waveform Drawing 1
This drawing shows the output waveform in the IF stimulation mode. Waveforms are Figure 1 given with purely resistive loads of 500 Ohms, 2000 Ohms, and 10,000 Ohms as required.
Image /page/9/Figure/4 description: Figure 1 shows a graph of amplitude in volts versus time in microseconds. There are three different plots on the graph, representing loads of 500 Ohms, 2000 Ohms, and 10000 Ohms. The amplitude varies between approximately -27 volts and 27 volts for the 500 and 2000 Ohm loads, and between -25 volts and 25 volts for the 10000 Ohm load. The time varies between 0 and 450 microseconds.
### Waveform Drawing 2
Modality = IF Mode = 6/6 Figure 1 This drawing represents the frequency of a series of pulses when the device is in the IF mode with frequency modulation. The modulation parameters are six second ramping between the preset frequencies.
When the device is turned on pulses begin at 60% of the user selected frequency (4000Hz plus beat frequency) over a six second period, ramp up to 160% of the selected frequency. Over the next six second period the frequency ramps down to 60% of the setting again and the cycle starts over.
Figure 2 Modality = IF Mode = 6|6 This drawing represent the frequency of a series of pulses when the device is in the IF mode with frequency modulation. The modulation parameters are six seconds, abruptly changing between the preset frequencies.
{10}------------------------------------------------
When the device is turned on pulses begin at 60% of the user selected frequency (4000Hz plus beat frequency) over a six second period, instantly change to 160% of the selected frequency for six seconds. The frequency then instantly decreases down to 60% of the setting again and the cycle starts over.
Image /page/10/Figure/1 description: The image shows two waveform drawings, labeled as Figure 1 and Figure 2. Figure 1 displays a triangular waveform that oscillates between 60% and 160% of the user setting. Figure 2 shows a square waveform that also oscillates between 60% and 160% of the user setting. The x-axis of both figures represents time, with each hash mark indicating 1 second.
#### Waveform Description
The waveforms from the Avid CT2 are the same as the predicate devices. A description of the waveforms is provided below in table format allowing comparison of measured values. For a visual comparison, scope traces of all three devices are also provided below.
Note that the scope traces of all three devices show a slight improvement with each generation; cleaner wave forms and less voltage variation over load.
{11}------------------------------------------------
| 510(k) Number | K982388 | K183692 | Unassigned |
|------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------|------------------------------------------------------------|
| Device Name | Surgi Stim /T.E.A.R. Tech | Avid IF2 | Avid CT2 |
| Waveform | IF- Sym. Biphasic<br>NMES-Sym. Biphasic<br>HVPC- Twin peak pulsed<br>monophasic | IF - Sym. Biphasic | IF - Sym. Biphasic<br>NMES-Sym. Biphasic |
| Max. Output<br>Current (500<br>Ohm Load) | IF Mode - 50mA ±10%<br>NMES Mode- 50mA ±10%<br>HVPC- .66A ±10% | IF Mode - 50mA+/- 10% | IF Mode - 50mA+/- 10%<br>NMES Mode- 50mA ±10% |
| Max. Output<br>Voltage (500<br>Ohm Load) | IF- 25V ± 10%<br>NMES- 25V ±10%<br>HVPC 330V ±10% | IF-25V +/- 10% | IF-25V +/- 10%<br>NMES-25V +/- 10% |
| Shape | IF- Square or rectangular<br>NMES- Square or rectangular<br>HVPC- Dual exponential spike | IF - Square or rectangular | IF - Square or rectangular<br>NMES - Square or rectangular |
| Symmetry | IF- Symmetrical<br>NMES- Symmetrical<br>HVPC - No | IF - Symmetrical | IF - Symmetrical<br>NMES - Symmetrical |
| Net Phase Charge | IF - 0μC<br>NMES – 0μC<br>HVPC- 8.25μC | IF - 0μC | IF - 0μC<br>NMES – 0μC |
| Peak Phase<br>Current (500<br>Ohm) | IF - 50mA<br>NMES - 50mA<br>HVPC - 0.66A | IF - 50mA | IF - 50mA<br>NMES-50mA |
| Peak Phase<br>Voltage (500<br>Ohm) | IF-25V<br>NMES - 25V<br>HVPC-330V | IF-25V | IF-25V<br>NMES-25V |
| Phase Rise Time<br>(500 Ohm,<br>max.width) | IF - < 2μS<br>NMES- <2μS<br>HVPC- <1μS | IF - < 2μS | IF - < 2μS<br>NMES - < 2μS |
| Phase Decay<br>Time (500 Ohm,<br>max. width) | IF- < 2μS<br>NMES- <2μS<br>HVPC- 27μS | IF- < 2μS | IF- < 2μS<br>NMES- < 2μS |
| Phase Duration<br>Range (at 50%<br>max. width) | IF - 7μS – 125μS<br>NMES - 7μS – 125μS<br>HVPC- 5μS | IF - 7μS – 125μS | IF - 7μS – 125μS<br>NMES - 7μS – 125μS |
| Interphase<br>Interval | IF – 0μS<br>NMES – 0μS<br>HVPC- 100 - 300μS | IF - 0μS | IF - 0μS<br>NMES – 0μS |
| Frequency Range | IF- 4000 Hz – 4240 Hz<br>NMES-4000Hz - 4240 Hz<br>HVPC - 1-200 Hz | IF- 4000 Hz – 4240 Hz | IF- 4000 Hz – 4240 Hz<br>NMES- 4000 Hz - 4240 Hz |
| Interference<br>Pattern | IF – Yes<br>NMES - Yes<br>HVPC - No | IF - Yes | IF – Yes<br>NMES - Yes |
| Beat<br>Frequencies | IF- 1-240 Hz<br>NMES-1-240Hz<br>HVPC-NA | IF- 1-240 Hz | IF- 1-240 Hz<br>NMES- 1-240 Hz |
| Burst Mode | No | No | No |
| Current Density | | | |
| Peak (per sq.<br>cm) (500 Ohm<br>Load) | IF-2.47mA<br>NMES-2.47mA<br>HVPC-65.1mA | IF-2.47mA | IF-2.47mA<br>NMES-2.47mA |
| Ave. (per sq.<br>cm)<br>(500 Ohm<br>Load) | IF-1.235mA<br>NMES-1.235mA<br>HVPC-0.13mA | IF-1.235mA | IF-1.235mA<br>NMES-1.235mA |
| Output Specifications Comparison | | | |
| 510(k) Number | K982388 | K183692 | Unassigned |
| Device Name | Surgi Stim /T.E.A.R. Tech | Avid IF2 | Avid CT2 |
| Power Density | | | |
| Peak (per sq.<br>cm)<br>(500 Ohm<br>Load) | IF – 61.7mW<br>NMES – 61.7mW<br>HVPC – 10.7 W | IF – 61.7mW | IF – 61.7mW<br>NMES – 61.7mW |
| Ave. (per sq.<br>cm)<br>(500 Ohm<br>Load) | IF – 30.85mW<br>NMES – 30.85mW<br>HVPC – 21.4 mW | IF – 30.85mW | IF – 30.85mW<br>NMES – 30.85mW |
| Max. Phase<br>Charge | | | |
| 500 Ohms | IF- 6.25µC<br>NMES – 6.25 µC<br>HVPC- 9.9 µC | IF- 6.25µC | IF- 6.25µC<br>NMES- 6.25µC |
| 2K Ohms | IF- 1.56 µC<br>NMES – 1.56 µC<br>HVPC- 1.65 µC | IF- 1.56 µC | IF- 1.56 µC<br>NMES- 1.56 µC |
| 10K Ohms | IF- 0.33 µC<br>NMES – 0.33 µC<br>HVPC- 0.33 µC | IF- 0.33 µC | IF- 0.33 µC<br>NMES- 0.33 µC |
{12}------------------------------------------------
{13}------------------------------------------------
Image /page/13/Figure/0 description: The image shows a graph of four electrode stimulation with a 500 Ohm resistive load. The x-axis represents time in microseconds, ranging from -250 to 250. The y-axis represents voltage in DC, ranging from -50 to 50. The graph shows a square wave pattern, with the voltage alternating between approximately 20 and -20.
Image /page/13/Figure/2 description: The image shows the text "Four Electrode Stimulation, 500 Ohm Resistive Load – Predicate (K183692)". The text describes a four-electrode stimulation setup with a 500 Ohm resistive load. The term "Predicate" is also included, followed by an identifier in parentheses, "K183692".
Image /page/13/Figure/3 description: The image shows a graph with voltage (V) on the y-axis and time (us) on the x-axis. The graph displays a square wave pattern, alternating between approximately 20V and -20V. The time scale ranges from -250 us to 250 us, indicating the duration of the pulses. The graph also includes the text "Pico Technology www.picotech.com" at the bottom.
Image /page/13/Figure/4 description: The image shows the title of a document or presentation. The title is "Four Electrode Stimulation, 500 Ohm Resistive Load - Predicate (K982388)". The title provides information about the content of the document or presentation, which is likely related to four electrode stimulation with a 500 Ohm resistive load.
Image /page/13/Figure/5 description: The image shows a graph with the x-axis labeled in microseconds from -250 to 250, and the y-axis labeled in volts from -50 to 50. The graph shows a square wave oscillating between approximately 25 volts and -25 volts. The wave has a period of approximately 200 microseconds.
{14}------------------------------------------------
Image /page/14/Figure/0 description: The image shows a graph with voltage on the y-axis and time on the x-axis. The y-axis ranges from -50.0 to 50.0 DC, while the x-axis ranges from -250.0 to 250.0 microseconds. The graph displays a square wave pattern, alternating between approximately 20V and -20V. The text at the bottom indicates that the graph was created using Pico Technology.
Four Electrode Stimulation, 200 Ohm Resistive Load – Avid CT2
Image /page/14/Figure/2 description: The image shows the text "Four Electrode Stimulation, 200 Ohm Resistive Load - Predicate (K183692)". The text describes a four-electrode stimulation setup with a 200 Ohm resistive load. The term "Predicate" is followed by a code, likely an identifier or reference number. The text is presented in a clear, readable font.
Image /page/14/Figure/3 description: The image shows a graph with voltage on the y-axis and time in microseconds on the x-axis. The voltage ranges from -50.0 to 50.0, while the time ranges from -250.0 to 250.0 microseconds. The graph displays a square wave pattern, alternating between approximately 20.0 and -20.0 voltage levels.
#### Four Electrode Stimulation, 200 Ohm Resistive Load - Predicate (K982388)
Image /page/14/Figure/5 description: The image shows a graph with the x-axis labeled in microseconds from -250 to 250 and the y-axis labeled in volts DC from -50 to 50. The graph shows a square wave that oscillates between approximately 22 volts and -22 volts. The graph also contains the text "Pico Technology www.picotech.com".
{15}------------------------------------------------
Image /page/15/Figure/0 description: The image shows a graph titled "Four Electrode Stimulation, 2000 Ohm Resistive Load – Avid CT2". The graph shows the voltage (V) on the y-axis and time (µs) on the x-axis. The graph shows a square wave oscillating between approximately 25V and -25V. The x-axis ranges from -250 µs to 250 µs, and the y-axis ranges from -50V to 50V.
Image /page/15/Figure/2 description: The image shows the text "Four Electrode Stimulation, 2000 Ohm Resistive Load – Predicate (K183692)". The text describes a four-electrode stimulation setup with a 2000 Ohm resistive load. The term "Predicate" is also included, followed by the identifier "(K183692)".
Image /page/15/Figure/3 description: The image shows a graph of voltage versus time. The y-axis represents voltage in DC volts, ranging from -50 to 50. The x-axis represents time in microseconds, ranging from -250 to 250. The graph shows a square wave oscillating between approximately 25 volts and -25 volts.
#### Four Electrode Stimulation, 2000 Ohm Resistive Load – Predicate (K982388)
Image /page/15/Figure/5 description: The image shows a graph of voltage versus time. The y-axis represents voltage in DC volts, ranging from -50 to 50. The x-axis represents time in microseconds, ranging from -250 to 250. The graph shows a square wave oscillating between approximately 25 and -25 volts.
{16}------------------------------------------------
Image /page/16/Figure/0 description: The image shows a graph with voltage on the y-axis and time in microseconds on the x-axis. The graph displays a periodic waveform that alternates between approximately 20V and -20V. The time scale ranges from -250 microseconds to 250 microseconds. The graph also includes the text 'Pico Technology www.picotech.com' at the bottom.
# Two Electrode Stimulation, 500 Ohm Resistive Load – Avid CT2
Two Electrode Stimulation, 500 Ohm Resistive Load – Predicate (K183692)
Image /page/16/Figure/3 description: The image shows a graph of voltage versus time. The y-axis represents voltage in DC, ranging from -50.0 to 50.0. The x-axis represents time in microseconds, ranging from -250.0 to 250.0. The graph shows a periodic square wave with voltage levels alternating between approximately -23 and 23.
Two Electrode Stimulation, 500 Ohm Resistive Load - Predicate (K982388)
Image /page/16/Figure/5 description: The image shows a graph of voltage over time. The y-axis represents voltage in DC volts, ranging from -50 to 50. The x-axis represents time in microseconds, ranging from -250 to 250. The graph shows a square wave pattern, alternating between approximately 25 volts and -25 volts.
{17}------------------------------------------------
# Substantial Equivalence Summary
Based on the data contained in the previous two tables and comparison waveforms we conclude that the Avid CT2 is equivalent to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.