ThermaCor 1200 Disposable Sets for the ThermaCor 1200 Rapid Thermal Infusion System
K202461 · Smisson-Cartledge Biomedical, LLC · LGZ · Aug 12, 2022 · General Hospital
Device Facts
Record ID
K202461
Device Name
ThermaCor 1200 Disposable Sets for the ThermaCor 1200 Rapid Thermal Infusion System
Applicant
Smisson-Cartledge Biomedical, LLC
Product Code
LGZ · General Hospital
Decision Date
Aug 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System is indicated for use over a full range of flow rates from slow feed to rapid, high flow infusion of: crystalloid, or blood product, including packed red blood cells [not in additive solutions and stored up to 21 days with anticoagulants], as volume replacement for patients suffering from blood loss due to trauma or surgery; warmed fluid to rewarm patients after surgery or for hypothermia; warmed fluid for irrigation in urology procedures
Device Story
Disposable set for ThermaCor® 1200 Rapid Thermal Infusion System; includes cassette with heat exchanger, tubing, and patient/supply lines. Heat exchanger features Parylene coating to reduce aluminum ion leaching. System uses volumetric pump for continuous infusion (up to 100L per cassette) or bolus delivery (up to 1L). Used in hospital ER, OR, ICU, and L&D; operated by clinicians. Device warms fluids to body temperature via AC power; battery backup available for transport (heating disabled). Clinicians use system to manage fluid/blood delivery and patient temperature; improves safety by mitigating potential ion leaching.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993, ASTM F756), bond strength (ISO 8536-4), coating adhesion (ASTM D3359), particulate matter (USP <788>), aluminum ion leaching verification, and heat transfer performance verification compared to non-coated heat exchangers.
Technological Characteristics
Disposable set includes cassette with Parylene-coated heat exchanger. Materials meet ISO 10993 biocompatibility standards. System uses volumetric pump; AC power for heating, lithium-ion battery for transport. Sterilized via Ethylene Oxide (EO).
Indications for Use
Indicated for patients suffering from blood loss due to trauma or surgery requiring volume replacement (crystalloid, colloid, or blood products), patients requiring rewarming after surgery or for hypothermia, and patients requiring warmed irrigation fluid for urology procedures.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Smisson-Cartledge Biomedical TIS-1200 Thermal Infusion System (K052055)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 12, 2022
Smisson-Cartledge Biomedical LLC % Julie Stephens President/Consultant Regulatory Resources Group, Inc. 111 Laurel Ridge Dr. Alpharetta, Georgia 30004
Re: K202461
Trade/Device Name: ThermaCor® 1200 Disposable Sets for the ThermaCor® 1200 Rapid Thermal Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LGZ, FRN, FPA Dated: July 19, 2022 Received: July 20, 2022
Dear Julie Stephens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark J. Antonino, M.S. Acting Division Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202461
#### Device Name
ThermaCor® 1200 Disposable Sets for the ThermaCor® 1200 Rapid Thermal Infusion System
Indications for Use (Describe)
The Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System is indicated for use over a full range of flow rates from slow feed to rapid, high flow infusion of: crystalloid, or blood product, including packed red blood cells [not in additive solutions and stored up to 21 days with anticoagulants], as volume replacement for patients suffering from blood loss due to trauma or surgery; warmed fluid to rewarm patients after surgery or for hypothermia; warmed fluid for irrigation in urology procedures
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| <div><span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Smisson-Cartledge Biomedical LLC Traditional 510(k) # K202461 - ThermaCor® 1200 Disposable Sets
# 510(k) Summarv
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements under 21 CFR 807.92.
| Submitted By: | Smisson-Cartledge Biomedical LLC<br>487 Cherry Street - Third Street Tower<br>Macon, GA 31201<br>Phone: (478) 744-9992 |
|-----------------|------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Julie Stephens, President/Consultant<br>Regulatory Resources Group, Inc. - Phone: (678) 513-0693 |
August 9, 2022 Date Submitted:
Device Name and Classification:
| Trade/Proprietary Name: | ThermaCor® 1200 Disposable Sets for the<br>ThermaCor® 1200 Rapid Thermal Infusion System |
|-------------------------|------------------------------------------------------------------------------------------|
| Common Name: | Rapid thermal infusion system |
| Regulation Number: | 21 CFR 880.5725 |
| Regulation Name: | Infusion Pump |
| Regulatory Class: | II |
| Product Code: | LGZ, FRN, FPA |
- Note: ThermaCor® 1200 Rapid Thermal Infusion System is the current "trade/proprietary name" for Smisson-Cartledge Biomedical TIS-1200 Thermal Infusion System, which was the name used within the prior 510(k) # K052055.
#### Legally Marketed Predicate Device:
Smisson-Cartledge Biomedical TIS-1200 Thermal Infusion System - 510(k) # K052055
#### Reason for 510(k) Submission:
To obtain clearance for a component modification within the ThermaCor® 1200 Disposable Sets. The component modification is the addition of a Parylene coating to the heat exchanger. The heat exchanger is a component of the cassette within the ThermaCor® 1200 Disposable Sets used with the ThermaCor® 1200 Rapid Thermal Infusion System.
#### Device Description:
The ThermaCor® 1200 Disposable Sets are intended for use only with the Smisson-Cartledge Biomedical (SCB) ThermaCor® 1200 Rapid Thermal Infusion System. The Disposable Sets consist of configurations that include cassettes with tubing connectors, various patient lines, and supply lines capable of interfacing with intravenous (IV) bags or optional-use reservoir equipment. Use of the ThermaCor® 1200 Disposable Sets ensures the intended use of the ThermaCor® 1200 Rapid Thermal Infusion System is met. The ThermaCor® 1200 Disposable Sets are provided sterile, non-pyrogenic and are single patient use only. They are sterilized by Ethylene Oxide (EO) sterilization method.
The Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System is a portable or pole-mounted device intended for use in the hospital emergency room (ER), operating room (OR), intensive care unit (ICU), and Labor and Delivery (L&D) environments.
{4}------------------------------------------------
### Smisson-Cartledge Biomedical LLC Traditional 510(k) # K202461 - ThermaCor® 1200 Disposable Sets
# 510(k) Summarv
The system consists of an infusion device and a compatible, single patient use sterile disposable set with supply lines capable of interfacing with intravenous (IV) bags or an optional reservoir. The ThermaCor® 1200 Infuser may be used with a Footswitch (optional accessory) to allow hands-free, user-controlled delivery.
The ThermaCor® 1200 Infuser can deliver flow rates from 10 mL per Hour to 1200 mL per Minute selectable in 10mL/hr in Slow mode and 20mL/min in Rapid mode at normothermic temperatures and is intended for continuous operation. The ThermaCor® 1200 Rapid Thermal Infusion System is driven by a volumetric pump capable of variable-rate continuous infusion up to approximately 100 Liters total volume per Cassette. The system is also capable of delivering discrete bolus infusions. The user can select Bolus Mode to deliver a fixed, predetermined bolus of up to 1 Liter at a default or adjustable rate. When connected to alternating current (AC) power, the ThermaCor® 1200 Infuser can deliver fluids warmed to body temperature under most conditions. The Infuser can also run on battery power (heating capabilities will be disabled) to allow transport of the patient. A lithium-ion battery pack provides the power backup.
### Indications for Use:
The Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System is indicated for use over a full range of flow rates from slow feed to rapid, high flow infusion of:
- crystalloid, colloid, or blood product, including packed red blood cells [not in additive . solutions and stored up to 21 days with anticoagulants], as volume replacement for patients suffering from blood loss due to trauma or surgery
- warmed fluid to rewarm patients after surgery or for hypothermia .
- warmed fluid for irrigation in urology procedures ●
## Discussion of Indications for Use for ThermaCor® 1200 Disposable Sets:
The Indications for Use as proposed for the ThermaCor® 1200 Disposable Sets are same as the Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System. The ThermaCor® 1200 Disposable Sets are necessary to enable the ThermaCor® 1200 Rapid Thermal Infusion System to meet its Indications for Use.
The Indications for Use statement for the Smisson-Cartledge Biomedical ThermaCor® 1200 Rapid Thermal Infusion System did not change from the cleared 510(k) # K052055.
## Discussion of Technological Characteristics:
The ThermaCor® 1200 Disposable Sets did not change technological characteristics from the cleared 510(k) # K052055 except for the addition of Parylene coating to the heat exchanger. The heat exchanger transfers heat to the fluids and is a component within the cassette. The cassette is then part of the Disposable Set that mounts onto the infusion pump unit. The addition of the Parylene coating did not change the performance of the Disposable Sets within the ThermaCor® 1200 Rapid Thermal Infusion System and demonstrated an improvement in the safety margin for potential aluminum ion leaching. The Disposable Sets are substantially equivalent to the Disposable Sets within the cleared 510(k) # K052055.
{5}------------------------------------------------
### Smisson-Cartledge Biomedical LLC Traditional 510(k) # K202461 - ThermaCor® 1200 Disposable Sets
# 510(k) Summarv
### Summary of Testing:
The biocompatibility risk assessment was completed as directed by FDA quidance under ISO 10993 biocompatibility requirements. Biocompatibility testing completed included Cytotoxicity. Acute Systemic Toxicity, Intracutaneous Reactivity Test, Material Mediated Pyrogenicity Test, Sensitization, and Blood Contact testing per ASTM F756 and ISO 10993-6 and additional FDA requests. Based on risk assessment, performance and safety testing was conducted on the ThermaCor® 1200 Disposable Sets with Parvlene Coated Heat Exchangers for bond strength tests per ISO 8536-4, mechanical adhesion of the Parylene coating per ASTM D3359, particulate matter testing per USP <788>, and aluminum ion leaching verification testing. Performance verification testing was conducted with the ThermaCor® 1200 Rapid Thermal Infusion System to demonstrate maximum heat transfer equivalency of the Parylene Coated Heat Exchangers to previous versions of Heat Exchangers without Parylene coating.
#### Substantial Equivalence Conclusions:
The modification to the Smisson-Cartledge Biomedical ThermaCor® 1200 Disposable Sets [proposed] is substantially equivalent to the Disposable Cassette Sets within the cleared 510(k) # K052055 [predicate] in terms of indications for Use/Intended Use, technical characteristics, principles of performance and safety.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.