The SuperSonic Imagine AIXPLORER® MACH / SUPERSONIC MACH range ultrasound diagnostic systems and transducers are intended for general purpose pulse echo ultrasound imaging, soft tissue viscoelasticity imaging, doppler fluid flow analysis of the human body. The SuperSonic Imagine AIXPLORER® MACH / SUPERSONIC MACH ultrasound diagnostic systems are indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculosketal, Vascular, Intraoperative, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac. In addition, the SuperSonic Imagine AIXPLORER® MACH ultrasound diagnostic systems and associated transducers are intended for: Measurements of abdominal anatomical structures, Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen Measurements of brightness ratio between liver and kidney, Visualization of abdominal vascularization, microvascularization and perfusion, Quantification of abdominal vascularization and perfusion. The shearwave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion of vascularization and perfusion may be used as an aid to clinical management of adult and pediatric patients with liver disease. The supported clinical applications for contrast enhancement imaging does not constitute permission to do such imaging beyond the scope of the contrast agent.
Device Story
Cart-based ultrasound diagnostic system; utilizes various linear, curved, microconvex, and phased array transducers. Inputs: ultrasonic pulse-echo signals. Processing: scan converter transforms signals into images displayed on LCD monitor; includes B-mode, M-mode, Color/Power Doppler, 3D imaging, CEUS, and elastography (ShearWave/Strain). Output: diagnostic images, measurements, and reports. Used in clinical settings by healthcare professionals. Aids clinical management of liver disease via stiffness/perfusion quantification. Benefits: non-invasive visualization and measurement of anatomical structures and tissue properties.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), acoustic output (NEMA UD-2/UD-3), and biocompatibility (ISO 10993-1).
Technological Characteristics
Cart-based ultrasound system with LCD monitor and control panel. Transducers: linear, curved, microconvex, motorized linear, phased array. Imaging modes: B-mode (harmonic/fundamental), M-mode, Color/Power Doppler, CW/PW Doppler, 3D, CEUS, ShearWave/Strain elastography. Connectivity: digital memory, external printer. Safety: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37. Biocompatibility: ISO 10993-1.
Indications for Use
Indicated for adult and pediatric patients for general purpose pulse echo ultrasound imaging, soft tissue viscoelasticity imaging, and Doppler fluid flow analysis. Applications include abdominal, small organs, musculoskeletal, superficial musculoskeletal, vascular, peripheral vascular, intraoperative, OB-GYN, pelvic, pediatric, urology, trans-rectal, trans-vaginal, neonatal/adult cephalic, and non-invasive cardiac imaging. Specific liver disease management includes measurements of shear wave speed, tissue stiffness, brightness ratio, and visualization/quantification of vascularization and perfusion.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
AIXPLORER® MACH 20, AIXPLORER® MACH 30 Ultrasound Diagnostic Systems (K191007)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
December 29, 2020
SuperSonic Imagine % Jacques Souquet Chief Innovation Officer Les Jardins de la Duranne - Bât. E&F 510. rue René Descartes 13857 Aix-en-Provence Cedex FRANCE
Re: K202455
Trade/Device Name: AIXPLORER® MACH 20, AIXPLORER® MACH 30, SUPERSONIC MACH40, SUPERSONIC MACH30 & SUPERSONIC MACH20 Ultrasound Diagnostic Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 12, 2020 Received: December 18, 2020
Dear Jacques Souquet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202455
### Device Name
AIXPLORER® MACH 20, AIXPLORER® MACH 30, SUPERSONIC MACH40, SUPERSONIC MACH30 & SUPERSONIC MACH20 Ultrasound Diagnostic Systems
### Indications for Use (Describe)
The SuperSonic Imagine AIXPLORER® MACH / SUPERSONIC MACH range ultrasound diagnostic systems and transducers are intended for general purpose pulse echo ultrasound imaging, soft tissue viscoelasticity imaging, doppler fluid flow analysis of the human body.
The SuperSonic Imagine AIXPLORER® MACH / SUPERSONIC MACH ultrasound diagnostic systems are indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculosketal, Vascular, Intraoperative, OB-GYN, Pelvic, Pediatric, Urology, Trans-rectal, Trans-vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac.
In addition, the SuperSonic Imagine AIXPLORER® MACH ultrasound diagnostic systems and associated transducers are intended for:
Measurements of abdominal anatomical structures,
Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen Measurements of brightness ratio between liver and kidney,
Visualization of abdominal vascularization, microvascularization and perfusion,
Quantification of abdominal vascularization and perfusion.
The shearwave speed, beam attenuation, viscosity and stiffness measurements, the brightness ratio, the visualization of vascularization, microvascularization and perfusion of vascularization and perfusion may be used as an aid to clinical management of adult and pediatric patients with liver disease.
The supported clinical applications for contrast enhancement imaging does not constitute permission to do such imaging beyond the scope of the contrast agent.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Device Name: AIXPLORER® MACH 30 (SSIP95030) , AIXPLORER® MACH 20 (SSIP95020), SUPERSONIC MACH30 (SSIP95030-HOLX), SUPERSONIC MACH20 (SSIP95020-HOLX) & SUPERSONIC MACH40 (SSIP95040) Ultrasound Diagnostic Systems
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | | | | | | | |
|---------------------------|---------------------------------------------------------------------|---|---|-----|-----|------------------|----------------------------|------------------------------------------------------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | P | | P | | P : 5, 6, 10 |
| Fetal Imaging &<br>Other | Fetal | P | P | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8, 9, 10, 11,13,<br>14, 15, 16 - N: 21,22 |
| | Abdominal (including urolology): Liver,<br>Kidney, Spleen... | P | P | P | | P | P : 1, 3, 4 | P: 5, 6, 8, 9 |
| | Intra-operative (Specify) vascular,<br>abdominal, small organs | P | | P | | P | P : 1, 3, 4 | P: 5, 6, 8, 9 |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8, 9, 10, 11, 13, 15,<br>16 - N: 14, 18, 19, 21, 20 |
| | Small Organ (Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | | P | | P | P : 1, 2, 3, 4,12 | P: 5, 6, 7, 8, 9, 10, 14, 15, 18,<br>19, 20, 21<br>N : 13 |
| | Neonatal Cephalic | P | | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 9 |
| | Adult Cephalic | P | | P | | P | P : 1, 3, 4 | P : 5, 6 |
| | Trans-rectal | P | | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8<br>N : 13 |
| | Trans-vaginal | P | P | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8, 11<br>N : 13 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8, 9, 10, 14, 15,19,<br>20<br>N : 13 |
| | Musculo-skeletal (Superficial) | P | | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8, 9, 10, 14, 15,19,<br>20<br>N : 13 |
| | Intravascular | | | | | | | |
| | GYN | P | P | P | | P | P : 1, 2, 3, 4 | P : 5, 6, 7, 8 11, 14<br>N : 13 |
| | Pelvic | P | P | P | | P | P : 1, 2, 3, 4 | P: 5, 6, 7, 8,11, 14<br>N : 13 |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | P | P : 1, 3, 4, 11, 17,<br>18 | P : 5 |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral | Peripheral vessel | P | | P | | P | P : 1, 3, 4 | P : 5, 6, 8, 9, 10, 15<br>N : 13 |
| Vessel | Other (Specify) | P | | P | | P | P : 1, 3, 4 | P : 5, 6, 8, 9, 10, 15<br>N : 13 |
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow
3: Combined modes include: B+ Pulsed Wave
5: Harmonic Imaging 6: Spatial Compounding 7: ShearWave™ Elastography
9: Panoramic Imaging
8: Imaging Guidance for Biopsies
2: Combined modes include: B+ ShearWave™ Elastography
4: Combined modes include: B+ Pulsed Wave + Color Flow
10: 3D Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
TM Elastography 12: Combined modes include: B Mode + Color flow + Shearwave
12. Combined modes in
Enhancement | UltraSound
Enhancement Ultrasound
13: CEUS (Contrast)
Enhancement UltraSound)
13: CEUS (Contrast)
14: Angio PL.U.S (Color Doppler improvement) 15: Needle
14: Angio PL.U.S (Color Doppler improvement) 15: Needle
16: Brightnes
PL.U.S
10-B-14
17: Combined mode include: B+ Continuous Wave 18: Combined mode
include: B+ M mode + Color flow 19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE 21: Vi PLUS
22: Att PLUS & SSp PLUS
20. Combined mode: B Mode + Strain + SWE 21.
21. AU-DIUC & CO-DIUC
22: Att PLUS & SSp PLUS
{4}------------------------------------------------
Device Name: L10-2 transducer (linear transducer, SSIP95103)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
- 4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
- 6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
- 18: Combined mode include: B+ M mode + Color flow
- 19: Strain Elastography
- 20: Combined mode: B Mode + Strain + SWE
- 21: Vi PLUS
22: Att PLUS & SSp PLUS
Clinical Application Mode of Operation в M |PWD|CWD General Specific Color Combined Other* (Specify) (Track 1 Only) (Tracks 1 & 3) Doppler (Specify) Ophthalmic Ophthalmic Fetal Fetal Imaging & Other P Abdominal (including urolology): Liver, P P P: 1, 2, 3, 4 P: 5, 6, 7, 8, 9, 13,14, 15, Kidney, Spleen ... 16 - N: 21,22 Intra-operative (Specify) Intra-operative (Neuro) Laparoscopic Pediatric P P P P: 1, 2, 3, 4 P: 5, 6, 7, 8, 9,14, 15, 16, 21, 22 N : 13 Small Organ (for example Breast, Thyroid, P P P P: 1, 2, 3, 4, P: 5, 6, 7, 8, 9, 14, 15,20, Testicle, Prostate, penis, etc ... ) 21, 22 12 N : 13 P P P Neonatal Cephalic N, 1, 2, 3, 4 N, 5, 6, 7, 9 Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) P P P Musculo-skeletal (Conventional) P: 1, 2, 3, 4 P: 5, 6, 7, 8, 9, 14,15,20 N : 13 P Musculo-skeletal (Superficial) P P P: 1, 2, 3, 4 P: 5, 6, 7, 8, 9, 14, 15,20 N : 13 Intravascular GYN Pelvic Other (Specify) Cardiac Adult Cardiac Cardiac Pediatric Intravascular (Cardiac) Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral Peripheral vessel P P P P: 1, 3, 4 P: 5, 6, 8, 9, 15 N : 13 P Other (Specify) P P P: 5, 6, 8, 9, 15 Vessel P: 1, 3, 4 N : 13
{5}------------------------------------------------
510(k) Number (if known):
Device Name: C6-1X transducer (curved array transducer, SSIP95101)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|---------------------------------------------------------------------------------|-------------------|-------------------|-----|-----|------------------|-----------------------|-----------------------------------------|-------------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | P | P | P | | P | P: 1, 3, 4, 11 | P: 5, 6 | |
| | Abdominal (including urolology), Liver,<br>Kidney, Spleen. | P | P | P | | P | P: 1, 2, 3, 4, 11 | P: 5, 6, 7, 8, 9, 13, 14, 16,<br>21, 22 | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 13, 14, 16,<br>21, 22 | |
| | Small Organ (Breast, Thyroid, Testicle,<br>Prostate, penis, etc...) | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8<br>N: 13 | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | GYN | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8,11,14<br>N: 13 | |
| | Pelvic | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 11,14<br>N: 13 | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral | Peripheral vessel | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8<br>N: 13 | |
| Vessel | Other (Specify) | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8<br>N: 13 | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | P | P | P | | P | P: 1, 3, 4, 11 | P: 5, 6 |
| | Abdominal (including urolology): Liver, Kidney,<br>Spleen... | | P | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8<br>N : 13 |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8<br>N : 13 |
| | Trans-vaginal | | P | P | P | | P | P: 1, 2, 3, 4, 11 | P: 5, 6, 7, 8<br>N : 13 |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | GYN | | P | P | P | P | P | P: 1, 2, 3, 4, 11 | P: 5, 6, 7, 8<br>N : 13 |
| | Pelvic | | P | P | P | P | P | P: 1, 2, 3, 4, 11 | P: 5, 6, 7, 8<br>N : 13 |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (Specify) | | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8<br>N : 13 |
N = new indication; P = Previously cleared by FDA (K191007)
Additional Comments:
- 1: Combined modes include: B+ Color Flow
- 2: Combined modes include: B+ ShearWave™ Elastography
- 3: Combined modes include: B+ Pulsed Wave
- 4: Combined modes include: B+ Pulsed Wave + Color Flow
- 5: Harmonic Imaging
- 6: Spatial Compounding
- 7: ShearWave™ Elastography
- 8: Imaging Guidance for Biopsies
- 9: Panoramic Imaging
- 10: 3D Imaging
- 11: Combined modes include: B+ M mode
- 12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
- 13: CEUS (Contrast Enhancement UltraSound)
- 14: Angio PL.U.S (Color Doppler improvement)
- 15: Needle PL.U.S
- 16: Brightness ratio
- 17: Combined mode include: B+ Continuous Wave
- 18: Combined mode include: B+ M mode + Color flow
- 19: Strain Elastography
- 20: Combined mode: B Mode + Strain + SWE
- 21: Vi PLUS
- 22: Att PLUS & SSp PLUS
{6}------------------------------------------------
510(k) Number (if known):
Device Name: E12-3 transducer (endocavitary transducer, SSIP95102)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow 2: Combined modes include: B+ ShearWave™ Elastography 12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
3: Combined modes include: B+ Pulsed Wave
- 4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
- 14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
- 20: Combined mode: B Mode + Strain + SWE
- 21: Vi PLUS
- 22: Att PLUS & SSp PLUS
{7}------------------------------------------------
510(k) Number (if known):
Device Name: LV16-5 transducer (motorized linear transducer, SSIP95108)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|---------------------------|------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|----------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal (including urolology):<br>Liver, Kidney, Spleen... | P | | P | P | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 10 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 10 |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, penis,<br>etc...) | P | P | P | | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 10 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 10 |
| | Musculo-skeletal (Superficial) | P | P | P | | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 10 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | P: 1, 3, 4 | P: 5, 6, 8, 9, 10 |
| | Other (Specify) | | | | | | | |
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE
21: Vi PLUS
22: Att PLUS & SSp PLUS
{8}------------------------------------------------
510(k) Number (if known):
Device Name: MC12-3 transducer (micro-curved transducer, SSIP95106)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|----------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal (including urolology): Liver,<br>Kidney, Spleen... | P | | P | P | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 13 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 13 |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | | P | | | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9 |
| | Neonatal Cephalic | P | | P | P | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9 |
| | Musculo-skeletal (Superficial) | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | P | P | P | | P | P: 1,3,4,11 | P: 5, 6 |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8, 9 |
| | Other (Specify) | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8, 9 |
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE
21: Vi PLUS
22: Att PLUS & SSp PLUS
{9}------------------------------------------------
510(k) Number (if known):
#### Device Name: P5-1X transducer (Phased Array transducer, SSIP95107)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|---------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal (including urolology): Liver,<br>Kidney, Spleen... | P | P | P | P | P | P: 1, 3, 4, 11,<br>17 | P: 5, 6, 16 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | P | | P | | P: 1, 3, 4 | P: 5, 6 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | P | P | | P | | P: 1, 3, 4 | P: 5, 6 |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | P | P : 1, 3, 4, 11,<br>17, 18 | P: 5, 6 |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P: 1, 3, 4, 17 | P: 5, 6 |
| | Other (Specify) | P | P | P | P | P | P: 1, 3, 4, 17 | P: 5, 6 |
N = new indication; P = Previously cleared by FDA (K191007)
1: Combined modes include: B+ Color Flow
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE
21: Vi PLUS
22: Att PLUS & SSp PLUS
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging 6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
{10}------------------------------------------------
510(k) Number (if known):
Device Name: LH20-6 transducer (linear transducer, SSIP95104)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------|---------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|-------------------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal (including urolology): Liver,<br>Kidney, Spleen... | | | | | | | |
| | Intra-operative (Specify) Vascular,<br>abdominal, small organs | P | P | P | | P | P: 1, 3, 4 | P: 5, 6, 9, 15 |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9, 15 |
| | Small Organ (for example Breast, Thyroid,<br>Testicle, Prostate, penis, etc...) | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9, 15 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9, 15 |
| | Musculo-skeletal (Superficial) | P | P | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9, 15 |
| | Intravascular | | | | | | | |
| | GYN | | | | | | | |
| | Pelvic | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P: 1, 3, 4 | P: 5, 6, 9, 15 |
| | Other (Specify) | P | P | P | | P | P: 1, 3, 4 | P: 5, 6, 9, 15 |
N = new indication; P = Previously cleared by FDA (K191007)
2: Combined modes include: B+ ShearWave™ Elastography
4: Combined modes include: B+ Pulsed Wave + Color Flow
1: Combined modes include: B+ Color Flow
3: Combined modes include: B+ Pulsed Wave
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE
21: Vi PLUS
22: Att PLUS & SSp PLUS
9: Panoramic Imaging
10: 3D Imaging
5: Harmonic Imaging
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
11: Combined modes include: B+ M mode
4. Angio P.L.U.S (Color Doppler Im
5: Needle P.L.U.S
{11}------------------------------------------------
510(k) Number (if known):
#### Device Name: L18-5 transducer Linear Array transducer, SSIP95100)
Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | | |
|---------------------------|-------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|----------------------------------------|--|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | | | |
| | Abdominal (including urology):<br>Liver, Kidney, Spleen... | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 14, 15 | | |
| | Intra-operative (Specify) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9, 14, 15 | | |
| | Small Organ (for example Breast,<br>Thyroid, Testicle, Prostate, Penis) | P | | P | | P | P: 1, 2, 3, 4, 12 | P: 5, 6, 7, 8, 9, 14 – N: 19,<br>20 | | |
| | Neonatal Cephalic | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 9 | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9,14, 15 – N:<br>19, 20 | | |
| | Musculo-skeletal (Superficial) | P | | P | | P | P: 1, 2, 3, 4 | P: 5, 6, 7, 8, 9,14, 15 – N:<br>19, 20 | | |
| | Intravascular | | | | | | | | | |
| | GYN | | | | | | | | | |
| | Pelvic | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8, 9, 15 | | |
| | Other (Specify) | P | | P | | P | P: 1, 3, 4 | P: 5, 6, 8, 9, 15 | | |
N = new indication; P = Previously cleared by FDA (K180572)
1: Combined modes include: B+ Color Flow
2: Combined modes include: B+ ShearWave™ Elastography
3: Combined modes include: B+ Pulsed Wave
- 4: Combined modes include: B+ Pulsed Wave + Color Flow
5: Harmonic Imaging
6: Spatial Compounding
7: ShearWave™ Elastography
8: Imaging Guidance for Biopsies
9: Panoramic Imaging
10: 3D Imaging
11: Combined modes include: B+ M mode
12: Combined modes include: B Mode + Color flow + Shearwave™ Elastography
13: CEUS (Contrast Enhancement UltraSound)
14: Angio PL.U.S (Color Doppler improvement)
15: Needle PL.U.S
16: Brightness ratio
17: Combined mode include: B+ Continuous Wave
18: Combined mode include: B+ M mode + Color flow
19: Strain Elastography
20: Combined mode: B Mode + Strain + SWE
21: Vi PLUS
- 22: Att PLUS & SSp PLUS
{12}------------------------------------------------
510(k) Number (if known):
Device Name: C9-2X transducer Curved Array transducer, SSIP95105) Intended Use: Diagnostic ultrasound imaging, soft tissue elasticity imaging, fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------------------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | P | P | P | | P | P: 1, 3, 4, 11 | P: 5, 6 |
| Other | Abdominal (including urology):<br>Liver, Kidney, Spleen... | P | P | P | | P | P: 1, 2, 3, 4, 11 | P: 5, 6, 7, 8, 9, 13, 14, 16 -<br>N: 21, 22 |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.