← Product Code [PON](/productcode/PON) · K202427

# Aqueduct 100 Plus Cervical Dilation Balloon Catheter (K202427)

_Gtimd, LLC · PON · Nov 6, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202427

## Device Facts

- **Applicant:** Gtimd, LLC
- **Product Code:** [PON](/productcode/PON.md)
- **Decision Date:** Nov 6, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.4260
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.

## Device Story

The Aqueduct 100 Plus is a balloon cervical dilation catheter used to mechanically dilate the cervical canal. The device consists of a 2-lumen polycarbonate shaft and a single cylindrical dilation balloon. In clinical use, a physician inserts the catheter through the vagina into the cervix, positioning the balloon at the internal and external cervical os. The balloon is inflated with 12.5 ml of saline to provide gradual mechanical dilation. An infusion lumen allows for saline injection to confirm dilation completion. The procedure typically achieves 8-9 mm dilation within 5 minutes, with total procedure time of 6-7 minutes. The device is intended for use in clinical settings by healthcare professionals. It benefits patients by providing a controlled, mechanical method for cervical dilation to facilitate various gynecological procedures.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11135, ISO 10993-7), and mechanical testing per ISO 10555-1 and ISO 10555-4. Accelerated aging studies confirmed 15-month shelf-life stability for dimensional verification, balloon burst pressure, fatigue, inflation/deflation time, bond strength, and pushability.

## Technological Characteristics

2-lumen polycarbonate shaft; Pebax dilation balloon. Inflated diameter 10 mm; burst pressure > 8 atm. Sterilized via ethylene oxide (EO) per ANSI/AAMI/ISO 11135:2014. Complies with ISO 10555-1:2013 and ISO 10555-4:2013 for balloon dilation catheters. Biocompatibility per ISO 10993-1.

## Regulatory Identification

A hygroscopic Laminaria cervical dilator is a device designed to dilate (stretch open) the cervical os by cervical insertion of a conical and expansible material made from the root of a seaweed (Laminaria digitata or Laminaria japonica). The device is used to induce abortion.

## Predicate Devices

- Aqueduct 100 Cervical Dilation Balloon Catheter ([K160664](/device/K160664.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 6, 2020

GTIMD LLC Eran Levit Official Correspondent 6 Columbia Drive Amherst, NH 03031

Re: K202427

> Trade/Device Name: Aqueduct 100 Plus Cervical Dilation Balloon Catheter Regulation Number: 21 CFR § 884.4260 Regulation Name: Hygroscopic Laminaria Cervical Dilator Regulatory Class: II Product Code: PON Dated: September 30, 2020 Received: October 9, 2020

Dear Eran Levit:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202427

#### Device Name

Aqueduct 100 Plus Cervical Dilation Balloon Catheter

Indications for Use (Describe)

The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.

Type of Use (Select one or both, as applicable)

| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in a dark gray.

www.gtimd.com

K202427

# 510(K) SUMMARY - K202427

| I. | Submitter            |                                     |
|----|----------------------|-------------------------------------|
|    | Submitter's Name:    | GTIMD LLC                           |
|    | Address:             | 6 Columbia Drive, Amherst, NH 03031 |
|    | Phone:               | (603) 809-3089                      |
|    | Contact Person:      | Eran Levit                          |
|    | Contact Email:       | elevit@gtimd.com                    |
|    | Date of Preparation: | November 5, 2020                    |

- II. Device

| Name of Device:    | Aqueduct 100 Plus Cervical Dilation Balloon Catheter    |
|--------------------|---------------------------------------------------------|
| Common Name:       | Catheter, Balloon, Dilation of the Cervical Canal       |
| Regulation Number: | 21 CFR 884.4260                                         |
| Regulation Name:   | Hygroscopic Laminaria Cervical Dilator                  |
| Regulatory Class:  | II                                                      |
| Product Code:      | PON (Catheter, Balloon, Dilation of the Cervical Canal) |

- III. Predicate Devices
### Primary Predicate Device:

Aqueduct 100 Cervical Dilation Balloon Catheter (K160664) manufactured by GTIMD LLC

This predicate device has not been subject to a design-related recall.

- IV. Device Description
The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is a balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The subject device is an updated version of the Aqueduct 100 Cervical Dilation Balloon Catheter cleared under K160664 that merges two separate dilation balloons into one balloon, removes the anchor balloon, replaces the four-way hub with a three-way hub, and removes the stylet from the shaft.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, bold, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in gray.

The catheter consists of a 2-lumen shaft. One lumen inflates a cylindrical dilation balloon. A second lumen is for infusion of saline solution. In use, the catheter is inserted through the vagina and cervical canal and into the uterus. The catheter is positioned inside the cervix. The user aligns the cylindrical balloon at the internal cervical os and external cervical os. The balloon is inflated with 12.5 ml of saline, providing gradual mechanical dilation of the cervix.

After 3 minutes of dilation of the internal and external orifices of the uterus, a controlled injection of 1-2.5 ml of saline may be made through the catheter infusion lumen which exits on the distal end of the catheter. Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed exiting through the external opening of the cervix. Optimal dilation of 8-9 mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes.

The deflation of the balloon is conducted by attaching a luer lock syringe to the swabable valve and removing saline from the balloon.

The subject device is packaged in a mylar/Tyvek pouch and ethylene oxide (EO) sterilized to a SAL 10-6. The packaged device has a shelf-life of 15 months.

- V. Indications For Use
The Aqueduct 100 Plus Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in gray.

- VI. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below provides a comparison of the intended use and technological characteristics of the subject and predicate device:

|                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                           | Aqueduct 100 Plus Cervical<br>Dilation Balloon Catheter<br>(K202427)                                                                                                                                                                                                                                                                                                                            | Aqueduct 100 Cervical<br>Dilation Balloon Catheter<br>(K160664)                                                                                                                                                                                                                                                                                                                         |
| Manufacturer              | GTIMD LLC                                                                                                                                                                                                                                                                                                                                                                                       | GTIMD LLC                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use    | The Aqueduct 100 Plus<br>Cervical Dilation Balloon<br>Catheter is intended to be<br>used whenever cervical<br>softening and dilation is<br>desired. Some examples are:<br>treatment of cervical stenosis,<br>IUD placement and removal,<br>Radium placement, drainage<br>of uterine cavity, endometrial<br>biopsy, uterine curettage,<br>suction cannula aspiration,<br>operative hysteroscopy. | The Aqueduct 100 Cervical<br>Dilation Balloon Catheter is<br>intended to be used whenever<br>cervical softening and dilation<br>is desired. Some examples are:<br>treatment of cervical stenosis,<br>IUD placement and removal,<br>Radium placement, drainage<br>of uterine cavity, endometrial<br>biopsy, uterine curettage,<br>suction cannula aspiration,<br>operative hysteroscopy. |
| Balloon<br>volume(s)      | Dilation Balloon: 12.5 mL                                                                                                                                                                                                                                                                                                                                                                       | Anchor Balloon: 2.5 mL<br>Dilation Balloons: 7.5 mL<br>(combined)                                                                                                                                                                                                                                                                                                                       |
| Balloon burst<br>pressure | Dilation Balloon: > 8 atm                                                                                                                                                                                                                                                                                                                                                                       | Anchor Balloon: > 2 atm<br>Dilation Balloon: > 2 atm                                                                                                                                                                                                                                                                                                                                    |
| Usable Length             | 25 cm                                                                                                                                                                                                                                                                                                                                                                                           | 25 cm                                                                                                                                                                                                                                                                                                                                                                                   |
| Outer diameter            | 3.3 mm deflated and 10 mm<br>inflated                                                                                                                                                                                                                                                                                                                                                           | 3.3 mm deflated and 12 mm<br>inflated                                                                                                                                                                                                                                                                                                                                                   |
| Shaft type                | 2-lumen polycarbonate shaft                                                                                                                                                                                                                                                                                                                                                                     | 3 lumen Pebax shaft                                                                                                                                                                                                                                                                                                                                                                     |
| Materials                 | Pebax balloons and<br>polycarbonate catheter                                                                                                                                                                                                                                                                                                                                                    | Pebax balloons and Pebax<br>catheter                                                                                                                                                                                                                                                                                                                                                    |
| Dilation<br>Mechanism     | Fixed diameter balloons that<br>expand to 10 mm                                                                                                                                                                                                                                                                                                                                                 | Fixed diameter balloons that<br>expand to 12 mm                                                                                                                                                                                                                                                                                                                                         |
| Stylet                    | No                                                                                                                                                                                                                                                                                                                                                                                              | Yes, unremovable                                                                                                                                                                                                                                                                                                                                                                        |
| Shelf Life                | 15 months                                                                                                                                                                                                                                                                                                                                                                                       | 2 years                                                                                                                                                                                                                                                                                                                                                                                 |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, bold, sans-serif font and is colored in a light blue. Below "GTIMD" is the text "CATHETER SOLUTIONS" in a smaller, sans-serif font and is colored in gray.

The Indications for Use statement for the subject device is the same as the predicate device. Therefore, the intended use of the subject and predicate devices are the same.

The following technological differences exist between the subject and predicate devices:

- The balloon volume and burst pressure specifications are increased for the subject device compared to the predicate device.
- The subject device balloon dimensions are different than the predicate device. The inflated diameter of the dilation balloon is less than the predicate device.
- . The subject device includes one dilation balloon while the predicate device contains two separate dilation balloons and an anchor balloon.
- . The number of lumens and component materials are different between the subject and predicate device.
- . The subject device does not contain a stylet while the predicate device does.

These differences in technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device. Nonclinical performance data were used to address the differences related to design and materials to demonstrate substantial equivalence to the predicate device.

- VII. Performance Data
To support the modifications to the subject device, the following design verification and validation activities were performed:

# Biocompatibility Testing

Biocompatibility tests were conducted on the Aqueduct 100 Plus Cervical Dilation Balloon Catheter according to the 2020 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" as follows:

- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010) ●
- Irritation (ISO 10993-10:2010)

The results of this testing demonstrated that the subject device is noncytotoxic, non-irritating, and non-sensitizing.

# Sterilization Validation

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for "GTIMD CATHETER SOLUTIONS". The text "GTIMD" is in a large, sans-serif font and is colored in a light blue. Below this, the text "CATHETER SOLUTIONS" is in a smaller, sans-serif font and is colored in gray.

The subject device utilizes the same sterilization process as the predicate device and uses the same sterilization chamber and cycle. The ethylene oxide sterilization cycle was validated using the overkill method, in accordance with ANSI/AAMI/ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices. Ethylene oxide residuals were evaluated according to ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.

## Bench Performance Testing

Performance tests were performed on the Aqueduct 100 Plus Cervical Dilation Balloon Catheter according to the requirements of ISO 10555-1:2013 and ISO 10555-4:2013 Intravascular catheters — Sterile and single-use catheters — Part 1: General requirements and Part 4: Balloon dilation catheters, and ISO 11737-1:2018 Sterilization of health care products — Microbiological methods Part 1: Determination of a population of microorganisms on products.

Performance testing was also conducted on the Aqueduct 100 Plus Cervical Dilation Balloon Catheter following an accelerated aging study to simulate a 15 month shelf life, including:

- . Dimensional Verification
- Balloon Burst Pressure ●
- Balloon Fatigue ●
- Balloon Inflation/Deflation Time ●
- Catheter Bond Strength ●
- Catheter Pushability

#### VIII. Conclusion

The Aqueduct 100 Plus Cervical Dilation Balloon Catheter has the same intended use as the predicate device. The subject device has different technological characteristics than the predicate device, but these differences do not raise different questions of safety and effectiveness. Finally, the submitted performance testing demonstrates that the subject device is as safe and effective as the predicate device. Therefore, the Aqueduct 100 Plus Cervical Dilation Balloon Catheter is substantially equivalent to the cleared predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K202427](https://fda-staging.innolitics.com/device/K202427)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
