← Product Code [GEI](/productcode/GEI) · K202415

# VirtueRF (K202415)

_ShenB Co., Ltd. · GEI · Jan 22, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202415

## Device Facts

- **Applicant:** ShenB Co., Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jan 22, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

1MHz Indications for use: Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF Electrosurgical System is intended for use with Skin Types I-V. 2MHz Indications for use: Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The 2mHz frequency is not intended to treat wrinkles.

## Device Story

VirtueRF is a radiofrequency (RF) electrosurgical system delivering 1MHz or 2MHz energy to human tissue via a sterile disposable 36-needle bipolar array. System components include main console, Smart RF handpiece cable, foot switch, and power cord. Used in clinical settings by healthcare professionals for electrocoagulation, hemostasis, and wrinkle treatment. Energy delivery is controlled by the operator via the handpiece; thermal effects on tissue facilitate the intended surgical outcomes. Device provides controlled RF energy to target tissue, aiding in dermatologic and general surgical procedures.

## Clinical Evidence

Bench testing only. Testing included IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (high-frequency surgical equipment), ISO 10993-1 (biocompatibility), and thermal effect tests on liver, kidney, muscle, and skin tissues per FDA guidance.

## Technological Characteristics

Bipolar fractional RF electrosurgical system; 1MHz and 2MHz frequencies; 36-needle bipolar array; max power 35.9W (1MHz) and 25W (2MHz); RF duration 100-800ms. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and ISO 10993-1 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Vivace Electrosurgical System ([K150409](/device/K150409.md))

## Reference Devices

- Secret RF ([K182355](/device/K182355.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

January 22, 2021

ShenB Co Ltd % Connie Hoy Consultant Hoy and Associates 3916 N. Potsdam Ave #4676 Sioux Falls, South Dakota 57104

Re: K202415

Trade/Device Name: VirtueRF Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 29, 2020 Received: December 1, 2020

Dear Connie Hoy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202415

#### Device Name

## VirtueRF

#### Indications for Use (Describe)

The VirtueRF Electrosurgical System has separate indications for use for each of its functionalities: 1MHz and 2MHz.

## 1MHz Indications for use

Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis, and the percutaneous treatment of facial wrinkles. The VirtueRF Electrosurgical System is intended for use with Skin Types I-V.

## 2MHz Indications for use

Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The 2mHz frequency is not intended to treat wrinkles.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

{ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

{3}------------------------------------------------

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

# 510K Summary K202415

This 510(K) Summary of safety and effectiveness for the VirtueRF is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

| Applicant:                         | ShenB Co Ltd.                                                                                                                                                                                                                                                                                           |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                           | Shenb Bldg 148<br>Seongsui-ro<br>Seongdong-Gu, Seoul , KR 04796                                                                                                                                                                                                                                         |
| Contact Person:                    | Mr. Oskar Lee                                                                                                                                                                                                                                                                                           |
| Telephone:                         | 82 70.4912.2702 - phone<br>oskar@shenb.com                                                                                                                                                                                                                                                              |
| Preparation Date:                  | January 22, 2021                                                                                                                                                                                                                                                                                        |
| Device Trade Name:                 | VirtueRF                                                                                                                                                                                                                                                                                                |
| Common Name:                       | Electrosurgical, cutting & coagulation & accessories                                                                                                                                                                                                                                                    |
| Regulation Name:                   | Electrosurgical cutting and coagulation device and<br>accessories                                                                                                                                                                                                                                       |
| Regulation Number:                 | 21 CFR 878.4400 (Product Code: GEI)                                                                                                                                                                                                                                                                     |
| Legally Marketed Predicate Device: | Vivace Electrosurgical System<br>K150409                                                                                                                                                                                                                                                                |
| Legally Marketed Reference Device: | Secret RF<br>K182355                                                                                                                                                                                                                                                                                    |
| Regulatory Class:                  | Class II Prescription Use                                                                                                                                                                                                                                                                               |
| Description of the VirtueRF :      | The VirtueRF device is a radio frequency output device<br>using 1MHz & 2Mhz to deliver radio frequency energy to the<br>human body with an applicator attached to the connecting<br>cable. The energy is applied to the human body using a<br>sterile disposable needle array consisting of 36 needles. |
|                                    | This product consists of main body, Needle Bipolar RF<br>(Smart RF) handpiece cable, foot switch and power cord.                                                                                                                                                                                        |
| Intended use of VirtueRF :         | The VirtueRF Electrosurgical System has separate<br>indications for use for each of its functionalities:<br>1MHz and 2MHz.                                                                                                                                                                              |
|                                    | 1MHz Indications for use: Intended for use in<br>dermatologic and general surgical procedures for<br>electrocoagulation and hemostasis, and the<br>percutaneous treatment of facial wrinkles. The<br>VirtueRF Electrosurgical System is intended for use<br>with Skin Types I-V.                        |

{5}------------------------------------------------

## 510K Summary K202415

2MHz Indications for use: Intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis. The 2mHz frequency is not intended to treat wrinkles.

Performance Testing: The following performance data was provided in support of the substantial equivalence determination:

> IEC 60601-1 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance;

> IEC 60601-1-2 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility

IEC 60601-2-2 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

IEC 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

Thermal Effect Tests on 4 types of tissues (liver, kidney, muscle and skin) per FDA guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery. Test shows that the VirtueRF is substantially equivalent to the predicate device.

Technical Specifications / Indications for Use Comparison:

| Comparison for 1MHz Functionality |                                       |                                                                     |            |
|-----------------------------------|---------------------------------------|---------------------------------------------------------------------|------------|
|                                   | Proposed Device<br>ShenB VirtueRF     | Predicate device<br>for 1MHz<br>Vivace<br>Electrosurgical<br>System | Comparison |
| Source of<br>energy/Modality      | Bipolar Fractional RF                 | Bipolar Fractional RF                                               | Same       |
| Radiofrequency                    | 1MHz                                  | 1MHz                                                                | Same       |
| Output energy type                | High Frequency                        | High Frequency                                                      | Same       |
| Electrode type                    | Bipolar microneedle                   | Bipolar microneedle                                                 | Same       |
| Max Power                         | 35.9W                                 | 61 w                                                                | Same       |
| RF Duration                       | 100ms-800ms, with<br>100ms increments | 100ms-800ms, with<br>100ms increments                               | Same       |

{6}------------------------------------------------

## 510K Summary K202415

|                              | Comparison for 2MHz Functionality     |                                           |            |
|------------------------------|---------------------------------------|-------------------------------------------|------------|
|                              | Proposed Device<br>ShenB VirtueRF     | Reference device<br>for 2MHz<br>Secret RF | Comparison |
| Source of<br>Energy/Modality | Bipolar Fractional RF                 | Bipolar Fractional RF                     | Same       |
| Radiofrequency               | 2MHz                                  | 2MHz                                      | Same       |
| Output energy type           | High Frequency                        | High Frequency                            | Same       |
| Electrode type               | Bipolar microneedle                   | Bipolar microneedle                       | Same       |
| Max Power                    | 25 W                                  | 25W                                       | Same       |
| RF Duration                  | 100ms-800ms, with<br>100ms increments | 50ms - 950ms                              | Similar    |

Conclusion: The VirtueRF and the predicate device relevant to 1MHz functionality, the Vivace Electrosurgical System, have the same indications for use and the same technological characteristics. The VirtueRF and the reference device relevant to 2MHz functionality, the Secret RF, have the same indications for use and nearly identical technological characteristics. The RF duration for the proposed VirtueRF (100ms-800ms) varies from the Secret RF only in that it is a subset of the Secret RF Duration. Therefore, VirtueRF substantially equivalent to both the predicate device and the reference device and there are no new questions of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K202415](https://fda-staging.innolitics.com/device/K202415)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
