ClearRT Helical kVCT for the Radixact Treatment Delivery System
K202412 · Accuray Incorporated · IYE · Dec 18, 2020 · Radiology
Device Facts
Record ID
K202412
Device Name
ClearRT Helical kVCT for the Radixact Treatment Delivery System
Applicant
Accuray Incorporated
Product Code
IYE · Radiology
Decision Date
Dec 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The kVCT Imaging Feature is an option within the intended use of the Radixact Treatment Delivery System. The Radixact Treatment Delivery System is intended to be used for the delivery of radiation therapy, stereotactic radiotherapy or stereotactic radiosurgery to tumors or other targeted tissues. The megavoltage x-ray radiation is delivered using rotational, non-rotational, intensity modulated (IMRT), or non-modulated (non-IMRT/three dimensional conformal) treatment techniques and using image-guided (IGRT) or non-image-guided workflows in accordance with the physician approved plan.
Device Story
The kVCT Imaging Feature is an add-on for the Radixact Treatment Delivery System, a radiation therapy device. It enables the acquisition of kV x-ray images to provide Image Guided Radiation Therapy (IGRT). The system uses a kV x-ray source and a flat panel detector to acquire 2D images, which are then reconstructed into 3D CT images. The device is operated by clinical staff in a treatment room (vault). The output consists of 3D CT images used for patient positioning and treatment verification. It does not diagnose disease, recommend treatment, or quantify effectiveness. The feature improves tissue contrast and image quality compared to the predicate's megavoltage CT (MVCT) imaging. The system uses dynamic collimation and filtering to manage the imaging beam. The output assists clinicians in verifying target alignment before radiation delivery, potentially increasing the precision of therapy.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including verification and validation of design specifications, performance testing of imaging characteristics (spatial resolution, contrast, uniformity, noise), and compliance with applicable FDA-recognized consensus safety standards for radiation therapy equipment.
Technological Characteristics
The system includes a kV x-ray source (40-150 kV) and a CsI:Tl flat panel detector. It features dynamic IEC-Xb/Yb collimating blades and a dynamically selectable copper filter (0.2, 0.3, 0.5 mm). The gantry is a CT-style design with an 85 cm bore. The system is a stand-alone radiation delivery unit. Software controls include an Embedded Controls Subsystem (ECS) modified to manage kV triggering, collimation, and panel actuation. The system is designed for helical acquisition.
Indications for Use
Indicated for the delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues anywhere in the body under the direction of a licensed medical practitioner.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Motion Tracking And Compensation Feature For The Radixact Treatment Delivery System (K182687)
Submission Summary (Full Text)
{0}------------------------------------------------
December 18, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Accuray Incorporated % Keith Picker Senior Regulatory Affairs Specialist 1240 Deming Way MADISON WI 53717
## Re: K202412
Trade/Device Name: kVCT Imaging Feature for the Radixact Treatment Delivery System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: November 24, 2020 Received: November 25, 2020
Dear Keith Picker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202412
#### Device Name
kVCT Imaging Feature for the Radixact Treatment Delivery System
Indications for Use (Describe)
The kVCT Imaging Feature is an option within the indications for use of the Radixact Treatment Delivery System. The Radixact Treatment Delivery System is indicated for the delivery of radiation therapy, stereotactic radiotherapy or stereotactic radiosurgery to tumors or other targeted tissues anywhere in the direction of a licensed medical practitioner.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M0 2a2 2 0 0 1 2-2h12a2 2 0 0 1 2 2v12a2 2 0 0 1-2 2H2a2 2 0 0 1-2-2V2zm1.5 0a.5.5 0 0 0-.5.5v11a.5.5 0 0 0 .5.5h11a.5.5 0 0 0 .5-.5v-11a.5.5 0 0 0-.5-.5h-11z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Accuray logo. The logo consists of a stylized atom symbol in teal, green, and purple, followed by the word "ACCURAY" in teal, with a registered trademark symbol next to it. The logo is clean and modern.
#### Section 8 510(k) Summary
K202412
#### Submitter
Accuray Incorporated 1240 Deming Way Madison, WI 53717 Phone: 608-824-3069 Fax: 608-824-2996
| Contact: | Keith Picker |
|----------------|-------------------|
| Date Prepared: | November 24, 2020 |
### Device Identification
Device Name: kVCT Imaging Feature for the Radixact Treatment Deliverv System Trade & Brand Names: ClearRT Helical kVCT for the Radixact Treatment Delivery System Common Name: Radiation Therapy System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE
### Predicate Device
Radixact Treatment Delivery System (K161146)
#### Reference Device:
Motion Tracking And Compensation Feature For The Radixact Treatment Delivery System (K182687)
#### Device Description
The kVCT Imaging Feature is designed for use with the predicate Radixact Treatment Delivery System last cleared on K161146. The primary functions of the kVCT Imaging Feature are: 1) acquisition of kV x-ray images according to input protocols in concert with other subsystems, 2) correcting 2D images and 3) returning reconstructed 3D images.
The Radixact Treatment Delivery System is a radiation therapy delivery system that provides Image Guided Radiation Therapy (IGRT) using integral megavoltage CT
{4}------------------------------------------------
imaging capabilities and delivers helical (rotational) and fixed-angle (non-rotational) radiation therapy to tumors and other targeted tissues.
Neither the Radixact Treatment Delivery System nor the kVCT Imaging Feature diagnose disease, recommend treatment regimens or quantify treatment effectiveness. Accordingly, they are not intended for diagnostic use.
### Intended Use
The kVCT Imaging Feature is an option within the intended use of the Radixact Treatment Delivery System. The Radixact Treatment Delivery System is intended to be used for the delivery of radiation therapy, stereotactic radiotherapy or stereotactic radiosurgery to tumors or other targeted tissues. The megavoltage x-ray radiation is delivered using rotational, non-rotational, intensity modulated (IMRT), or non-modulated (non-IMRT/three dimensional conformal) treatment techniques and using image-guided (IGRT) or non-image-guided workflows in accordance with the physician approved plan.
### Indications for Use
The kVCT Imaging Feature is an option within the indications for use of the Radixact Treatment Delivery System. The Radixact Treatment Delivery System is indicated for the delivery of radiation therapy, stereotactic radiotherapy or stereotactic radiosurgery to tumors or other targeted tissues anywhere in the body under the direction of a licensed medical practitioner.
The intended use and indications for use statements for the kVCT Imaging Feature (shown above) are the same as for the Radixact Treatment Delivery System (last cleared on K161146), except for the addition of the introductory sentence as kVCT Imaging is a type of Image Guided Radiotherapy (IGRT).
## Technological Characteristics
The Radixact Treatment Delivery System with the use of the kVCT Imaging Feature has imaging and treatment capabilities equivalent to those of the predicate Radixact Treatment Delivery System. It also has a similar functionally-equivalent CT style gantry and patient couch.
The intended use and indications for use of the kVCT Imaging Feature fit within those of the predicate Radixact Treatment Delivery System. Additionally, the predicate and subject devices have substantially equivalent performance specifications and technological characteristics. Further, the Radixact Treatment Delivery System with the use of the kVCT Imaging Feature and the predicate device employ the same fundamental scientific principles and have substantially equivalent principles of operation. The main difference between the predicate and the subject device is the use of the kVCT Imaging Feature. Where there are technological differences between the subject and predicate devices, those differences do not raise different questions of safety or effectiveness.
{5}------------------------------------------------
A substantial equivalence table comparing the similarities and differences between the Radixact Treatment Delivery System with the use of the kVCT Imaging Feature and the predicate device is presented below in Table 5.5.1. The minor differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
A table comparing the similarities and differences between the Radixact Treatment Delivery System with the use of the kVCT Imaging Feature and the reference device (K182687) is provided in Table 5.5.2 below. The minor differences between the subject and predicate devices do not raise different questions of safety or effectiveness. The kV Imaging subsystem provides three-dimensional CT images with equivalent or better performance than the predicate device (MVCT). The kVCT energy range offer improved intrinsic tissue contrast and image quality characteristics compared to that of MVCT imaging.
{6}------------------------------------------------
# Table 5.5.1 Device Comparison Table: Radixact Treatment Delivery System with kVCT Imaging Feature
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System Description | | | |
| Intended Use | The Radixact Treatment<br>Delivery System is<br>intended to be used for<br>the delivery of radiation<br>therapy, stereotactic<br>radiotherapy or<br>stereotactic radiosurgery<br>to tumors or other<br>targeted tissues. The<br>megavoltage x-ray<br>radiation is delivered<br>using rotational, non-<br>rotational, intensity<br>modulated (IMRT), or<br>non-modulated (non-<br>IMRT/three dimensional<br>conformal) treatment<br>techniques and using<br>image-guided (IGRT) or<br>non-image-guided<br>workflows in accordance<br>with the physician<br>approved plan. | The kVCT Imaging<br>Feature is an option<br>within the intended use<br>of the Radixact<br>Treatment Delivery<br>System. The Radixact<br>Treatment Delivery<br>System is intended to<br>be used for the delivery<br>of radiation therapy,<br>stereotactic<br>radiotherapy or<br>stereotactic<br>radiosurgery to tumors<br>or other targeted<br>tissues. The<br>megavoltage x-ray<br>radiation is delivered<br>using rotational, non-<br>rotational, intensity<br>modulated (IMRT), or<br>non-modulated (non-<br>IMRT/three<br>dimensional conformal)<br>treatment techniques<br>and using image-guided<br>(IGRT) or non-image-<br>guided workflows in<br>accordance with the<br>physician approved<br>plan. | Identical |
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
| Indications for<br>Use | The Radixact Treatment<br>Delivery System is<br>indicated for the delivery<br>of radiation therapy,<br>stereotactic radiotherapy<br>or stereotactic<br>radiosurgery to tumors or<br>other targeted tissues<br>anywhere in the body<br>under the direction of a<br>licensed medical<br>practitioner. | The kVCT Imaging<br>Feature is an option<br>within the indications<br>for use of the Radixact<br>Treatment Delivery<br>System. The Radixact<br>Treatment Delivery<br>System is indicated for<br>the delivery of<br>radiation therapy,<br>stereotactic<br>radiotherapy or<br>stereotactic<br>radiosurgery to tumors<br>or other targeted tissues<br>anywhere in the body<br>under the direction of a<br>licensed medical<br>practitioner. | Identical |
| Classification | §892.5050 Medical<br>charged-particle radiation<br>therapy system, class II | Same as predicate | Identical |
| System<br>Configuration | Stand-alone radiation<br>delivery system (does not<br>include data management<br>system or planning<br>system) | Stand-alone radiation<br>delivery system with<br>kV imaging added<br>(does not include data<br>management system or<br>planning system) | Substantially<br>equivalent. The<br>introduction of<br>the kVCT<br>imaging feature<br>does not raise<br>different issues<br>of safety or<br>effectiveness.<br>The risk<br>acceptability<br>determination<br>was not<br>impacted by the<br>modified design. |
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
| Vault (Treatment Room) | | | |
| Minimum Room<br>Dimensions<br>(H*W*L) | 274 x 462 x 602 cm | 274.3 x 463 x 602 cm | Substantially<br>equivalent. Minor<br>differences are<br>negligible. |
| Device<br>Dimensions<br>(gantry + couch)<br>(H*W*L) | 255 x 280 x 473 cm | 255 x 280 x 470.5 cm | Substantially<br>equivalent. Minor<br>differences are<br>negligible. |
| Device Mass<br>(kg) | 6580 kg | 7000 kg (approximate) | Substantially<br>equivalent. The<br>added mass due to<br>the kVCT imaging<br>subsystem does not<br>result in different<br>questions of safety<br>or effectiveness. |
| Line Voltage | 380 - 480 Vac 3-<br>Phase | 380 - 480 Vac 3-Phase | Identical |
| Ambient Room<br>Temperature | 20 - 24 °C | 20 - 24 °C | Identical |
| Relative<br>Humidity<br>(non-<br>condensing) | 30 - 60 % | 30 - 60 % | Identical |
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
| Gantry Mechanical Features | | | |
| Bore Diameter | 85 cm | 85 cm | Identical |
| Degrees of<br>Rotation | Continuous rotation<br>around Y- axis<br>(axes per IEC 61217) | Continuous rotation<br>around Y- axis<br>(axes per IEC 61217) | Identical |
| Direction of<br>Rotation | Clockwise | Clockwise | Identical |
| Rotational<br>Speed<br>(Treatment) | 1 - 5 RPM | 1 - 5.08 RPM | Substantially<br>equivalent. Minor<br>differences are<br>negligible and do<br>not result in<br>different questions<br>of safety or<br>effectiveness. |
| Rotational<br>Speed (Imaging) | 10 RPM | Up to 10 RPM | Substantially<br>equivalent.<br>Rotational speed is<br>unchanged for<br>MVCT imaging.<br>Slower speeds are<br>used for some<br>kVCT imaging<br>applications. |
| Couch Support<br>in Bore | Provided | Provided | Identical |
| Radiation Delivery Modes | | | |
| Description | Helical, Direct | Helical, Direct | Identical |
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
| Photon Beam | | | |
| Accelerator<br>Type | Standing wave | Standing wave | Identical |
| RF Source | Magnetron | Magnetron | Identical |
| Nominal Energy | 6 MV | 6 MV | Identical |
| Fixed Field Size | 1.0 cm x 40 cm<br>2.5 cm x 40 cm<br>5.0 cm x 40 cm | 1.0 cm x 40 cm<br>2.5 cm x 40 cm<br>5.0 cm x 40 cm | Identical |
| Dynamic Field<br>Size | 1.0 – 2.5 cm x 40 cm<br>1.0 – 5.0 cm x 40 cm | 1.0 – 2.5 cm x 40 cm<br>1.0 – 5.0 cm x 40 cm | Identical |
| Dose Rate | 850 cGy/min standard<br>1000 cGy/min<br>optional | 850 cGy/min standard<br>1000 cGy/min optional | Identical |
| MV Beam Collimation | | | |
| Description | Primary collimation,<br>jaws and multi-leaf<br>collimator (MLC) | Primary collimation,<br>jaws and multi-leaf<br>collimator (MLC) | Substantially equivalent. kVCT<br>does not change<br>the behavior of the<br>jaws. |
| Device<br>Characteristic | Predicate Device<br>Radixact Treatment<br>Delivery System<br>(K161146) | Subject Device<br>Radixact Treatment<br>Delivery System with<br>kVCT Imaging<br>Feature | Analysis |
| MVCT Imaging | | | |
| Source | Megavoltage<br>Computed<br>Tomography | Megavoltage<br>Computed<br>Tomography | Identical |
| Field of View<br>(FOV) | 39 cm diameter | 39 cm diameter | Identical |
| Scan Length<br>[Specification<br>not detailed in<br>K161146] | 135 cm | 135 cm | Identical |
| Imaging Dose | 0.5 - 3.0 cGy<br>(imaging dose near<br>center of Tomo-<br>Phantom) | 1.1 - 3.4 cGy<br>(CTDIvol, Head)<br>0.8 - 2.5 cGy<br>(CTDIvol, Body) | Identical.<br>Differences shown<br>are due to<br>additional detail<br>provided and dose<br>characterized in<br>CTDIvol |
| Slice Spacing | 1, 2, 3, 4 and 6 mm<br>reconstruction<br>Intervals…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.