Powder Free NR Latex Surgical Glove (Sterile) with Protein content labelling claims 50ug/dm2 or less per glove of extractable Protein
K202401 · Eco Medi Glove Sdn. Bhd. · KGO · Feb 17, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202401
Device Name
Powder Free NR Latex Surgical Glove (Sterile) with Protein content labelling claims 50ug/dm2 or less per glove of extractable Protein
Applicant
Eco Medi Glove Sdn. Bhd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Feb 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
Powder Free NR Latex Surgical Glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Disposable, sterile, powder-free natural rubber latex surgical glove; intended for use by operating room personnel to protect surgical wounds from contamination. Device is yellow, anatomic, and provided in various sizes. Sterilized via gamma irradiation to achieve a sterility assurance level of 10^-6; packaged to ensure a 3-year shelf life. Device acts as a physical barrier between the clinician and the patient's surgical site. Performance is verified through physical property testing (tensile strength, elongation), pinhole leak testing, and biocompatibility assessments.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ASTM D3577-09 (dimensions, physical properties), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder residue), and ISO 10993-10 (biocompatibility: primary skin irritation and dermal sensitization). All tests passed.
Indicated for operating room personnel to protect surgical wounds from contamination. Device is a sterile, powder-free, natural rubber latex surgical glove.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Latex surgeon's Glove Powder Free with Protein content labeling claim of 50ug/dm2 or less per glove of extractable protein (K190632)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 17, 2021
ECO Medi Glove Sdn. Bhd Suresh Kumar OA Manager Lot 26910, Lorong Perusahaan 3/3, Kawasan Perusahaan, Kamunting Raya Taiping, Perak 34600 Malaysia
#### Re: K202401
Trade/Device Name: Powder Free NR Latex Surgical Glove (Sterile) with Protein content labelling claims 50ug/dm2 or less per glove of extractable Protein Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: August 1, 2020 Received: August 21, 2020
Dear Suresh Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ryan Ortega, PhD. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
## 510(k) Number (if known)
K 202401
#### Device Name
Powder Free NR Latex Surgical Glove (Sterile)with protein content labeling claims 50ug/dm2 or less per glove of extractable protein.
Indications for Use (Describe)
Powder Free NR Latex Surgical Glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# 510(K) Summary Powder Free NR Latex Surgical Glove (Sterile) with Protein content labelling claim 50ug/dm² or less per glove of extractable Protein
#### 1.0 Submitter:
#### 4.0 Predicate De vice Information:
Latex Surgeon's Gloves Powder Free with Protein content labelling claims 50ug/dm² or less per glove of extractable Protein, 79KGO, which meets all the requirement of ASTM D 3577-09 and FDA 21 CFR 878.4460. It is equivalent to K190632, Latex surgeon's Glove Powder Free with Protein content labeling claim of 50ug/dm² or less per glove of extractable protein
#### 5.0 Indication For Use
Powder Free NR Latex Surgical Glove is a device made of Natural Rubber Intended to be worn by operating room personal to protect a surgical wound from contamination
{4}------------------------------------------------
## 6.0 Device Description
The subject device in this 510(k) Notification is Latex Surgeon's Glove Powder Free with Protein content labelling claims 50ug/dm2 or less per glove of extractable Protein. The subject device is a Surgeon's glove made from compounded primarily Natural Latex, Yellow, powder free and sterile (Per 21 CFR 878.4460, class I. The device meets all the specifications in ASTM 3577-09, Standard specification for Rubber Surgical Glove. Additionally, the gloves have been tested for biocompatibility.
Latex Surgeon's Gloves Powder Free with Protein content labelling claims 50ug/dm² or less per glove of extractable Protein is classification as Type 1: Glove compounded primarily form Natural Rubber Latex
The Proposed device is sterilized using Gamma irradiation method to achieve the sterility Assurance Level (SAL) of 10E and place in a sterility maintenance package to ensure a shelf life of 3 years
Powder Free NR Latex Surgical Glove (Sterile), is a disposable device intended for medical purposes that is worn by operating room personal to protect a surgical wound from contamination (product Code KGO).
## 7.0 Specification for Rubber Surgical gloves:
| Dimension/Size | 5.5 | 6.0 | 6.5 | 7.0 | 7.5 | 8.0 | 8.5 | 9.0 |
|----------------------------------------|------|------|------|------|------|------|------|------|
| Overall Length (mm) (Minimum) | 265 | 265 | 265 | 265 | 265 | 265 | 265 | 265 |
| Width (± 6mm) | 70 | 76 | 83 | 89 | 95 | 102 | 108 | 114 |
| Thickness at Palm (mm) (Minimum) | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 |
| Thickness at Finger Tip (mm) (Minimum) | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 |
| Thickness at cuff (mm) (Minimum) | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 |
#### 7.1 Dimension and Thickness of Gloves
#### 7.2 Gloves Physical Properties and Holes
| Measurement | Before Ageing | After Aging at 70°C for<br>168 hrs. @ 100°C for 22<br>hrs. |
|-------------------------|---------------------------------|------------------------------------------------------------|
| Tensile Strength (MPa) | 24 MPa minimum | 18 MPa minimum |
| Ultimate Elongation (%) | 750% minimum | 560% minimum |
| Pin-hole Level | AQL 1.5<br>Inspection Level G-1 | AQL 1.5<br>Inspection Level G-1 |
{5}------------------------------------------------
# 8.0 Technological Characteristics Comparison
| Characteristics | Powder Free NR Latex Surgical Glove<br>(Sterile) with Protein content labelling<br>claims 50ug/dm² or less per glove of<br>extractable Protein<br>K202401 (Subject Device)<br>Eco Medi Glove Sdn. Bhd.<br>KGO | Latex surgeon's Glove Powder Free with<br>Protein content labeling claim of 50ug/dm² or<br>less per glove of extractable protein<br>K190632 (Predicate Device)<br>Lenora Glove Pvt. Ltd<br>KGO | Comparison |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended use | Powder Free NR latex Surgical Glove is a<br>device made of Natural Rubber intended to be<br>worn by operating room personal to protect a<br>surgical wound from contamination | Surgeons glove is a device made of natural rubber<br>intended to be worn by operating room personal<br>to protect a surgical wound from contamination | Same |
| Description | Sterile Powder free, surgical<br>gloves are made of natural<br>rubber latex. The gloves are<br>provided in Sizes 5.5 6.0, 6.5, 7.0, 7.5,<br>8.0, 8.5 and 9.0 | Sterile Powder free, surgical<br>gloves are made of natural<br>rubber latex. The gloves are<br>provided in Sizes 5.5,6.0, 6.5, 7.0, 7.5, 8.0,<br>8.5 and 9.0 | Same |
| Presentation | Sterile gloves are provided in pouches | Sterile gloves are provided in pouches | Same |
| Anatomic | Yes | Yes | Same |
| Material use | Natural Rubber Latex | Natural Rubber Latex | Same |
| Colour | Yellow | Natural (No color is added) | Different |
| Sterility | Sterile | Sterile | Same |
| Single used | Single used | Single used | Same |
| Non-Sterile or Sterile | Sterile | Sterile | Same |
| Dimensions | Meets ASTM D3577 - 09 | Meets ASTM D3577 - 09 | Similar |
| Physical properties | Meets ASTM D3577 -09 and<br>ASTM D412, D573. | Meets ASTM D3577 -09 and<br>ASTM D412, D573. | Similar |
| Freedom from pinholes | Meets ASTM D 5151 -06<br>and ASTM D3577- 09 | Meets ASTM D 5151 -06<br>and ASTM D3577- 09 | Same |
| Characteristics | Powder Free NR Latex Surgical Glove<br>(Sterile) with Protein content labelling<br>claims 50ug/dm² or less per glove of<br>extractable Protein,K202401 | Latex surgeon's Glove Powder Free with<br>Protein content labeling claim of 50ug/dm² or<br>less per glove of extractable protein, K190632<br>(Predicate Device) | Comparison |
| Residual Powder | Meets ASTM<br>D6124-06, (Reapproved<br>2017)<br>Powder content < 2 mg/Glove | Meets ASTM<br>D6124-06, (Reapproved<br>2017)<br>Powder content < 2 mg/Glove | Same |
| Protein Content | Meets ASTM D 5712, ASTM D 6499-12 &<br>ASTM D3577 | Meets ASTM D 5712, ASTM D 6499-12 &<br>ASTM D3577 | Same |
| Mode of<br>Sterilization | Irradiation | EO Sterilization/<br>Irradiation | Different |
| Biological Evaluation on<br>Medical Device<br>-Primary Skin Irritation<br>Test | Under the conditions of this study, the test<br>article was a non-irritant. | Under the conditions of this study, the test article<br>was a non-irritant. | Same |
| Biological Evaluation on<br>Medical Device-<br>Dermal Sensitization Assay | Under the conditions of this study, the test<br>article was a non-sensitizer. | Under the conditions of this study, the test article<br>was a non-sensitizer. | Similar |
| Acute Systemic Test | Non Toxic | Non Toxic | Same |
| Labelling | * Powder free<br>* Latex Surgical Glove<br>* Sterile<br>* Single use<br>* Anatomic<br>* Manufactured for<br>* Lot No<br>* Intended use<br>* Quantity<br>* Country of origin | * Powder free<br>* Latex Surgical Glove<br>* Sterile<br>* Single use Anatomic<br>* Manufactured for<br>* Lot No<br>* Intended use<br>* Quantity<br>* Country of origin | similar |
{6}------------------------------------------------
{7}------------------------------------------------
The above table shows similarities and differences of the performance between the subject device and the predicate device. There are no major differences in technological characteristics of the subject device compare with the predicate device. There are only minor differences between the subject device and the predicate device. One difference is that the subject device is sterilized by Irradiation whereas the predicate device is sterilized by EO sterilization. These minor differences do not impact the intended use, safety and performance of the device. The gloves are made from Natural latex compound, Latex Surgeon's Gloves (Sterile).The gloves met all the specifications in ASTM D3577-09 Standard specification for Rubber Surgical Glove as well Biological Evaluation on medical device.
#### 9.0 SUMMARY OF NON-CLINICAL TESTING
Gloves are made from Latex compound, which is Sterile and Powder free. The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below,
| Characteristics | Standards | Performance of<br>Sterile Latex<br>Surgical<br>Gloves, Powder free | Characteristics | Test | Test standard | Sampling<br>plan<br>/Inspection<br>level/AQL | Sterile Latex<br>Surgical<br>Gloves,<br>Powder free | Result |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|---------------------------|-----------------------------------|-----------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------------------------|--------|
| Freedom from Holes | ASTM D3577-09 / ASTM D5151-06 | Meets | Freedom from<br>Pin holes | FDA 1000 ml<br>water leak<br>test | ASTM D5151 -06<br>(Re-approved<br>2011) | ISO 2859-1 /<br>G1/AQL 1.5 | Meet AQL<br>1.5 | PASS |
| Dimensions | ASTM D3577-09 | Meets | | Length | ASTM D3577 -09 | ISO 2859-1 /<br>S2/AQL 4.0 | Min 265 | PASS |
| Physical Properties | ASTM D3577-09 / ASTM D412-06 | Meets | Dimensions | Width | ASTM D3577-09 | ISO 2859-1 /<br>S2/AQL 4.0 | 70±6 mm to<br>114±6<br>mm (sizes 5.5<br>to 9.0) | PASS |
| Powder-free<br>residue | ASTM D6124-06 | Meets | | Thickness | ASTM D3577-09 | ISO 2859-1 /<br>S2/AQL 4.0 | >0.10mm<br>(cuff, palm<br>,finger) | PASS |
| Powder Content | ASTM D 5712, ASTM D 6499-12 | Meets | Physical<br>properties | Before aging | ASTM D3577-09<br>and ASTM D412-<br>06 | ISO 2859-1 /<br>S2/AQL 4.0 | Tensile<br>strength :<br>> 24 Mpa<br>Ultimate<br>Elongation<br>: >750% | PASS |
| Bio-compatibility | Primary skin irritation - ISO 10993-10 | Non-irritant | | After<br>Accelerated<br>aging | ASTM D3577-09<br>and ASTM D412-<br>06 | ISO 2859-1 /<br>S2/AQL 4.0 | Tensile<br>strength :<br>> 18 Mpa<br>Ultimate<br>Elongation :<br>> 560% | PASS |
| | Skin Sensitization - ISO 10993-10 | Non-sensitizer | Powder-free<br>residue | Powder-free<br>residue | ASTM D3577-09<br>and<br>ASTM D6124-06 | N=5 | Less than 2<br>mg per<br>glove | PASS |
| | Acute Systemic Test | Non Toxic | Powder Content | Powder<br>Residue | ASTM D 5712,<br>ASTM D 6499-12 | N=5 | Less than 2<br>mg per<br>glove | PASS |
| Expiration dating/<br>Shelf life | ASTM D7160-05 | Three years | Biocompatibility | Primary skin<br>irritation | ISO 10993 -10 | Under the conditions of the<br>study the device is not an<br>irritant | | PASS |
| Sterilization | ISO 11737-Sterilization of medical<br>device-Microbiological method - part 1 | Meets | | Skin<br>Sensitization | ISO 10993 -10 | Under the conditions of the<br>study the device is not a<br>sensitizer | | PASS |
| Sterility | ISO 11737 -2<br>Sterilization of Medical Devices-<br>Microbiological<br>methods Part 2: Test of sterility<br>performed in the<br>definition, validation and maintenance<br>of sterilization process | Sterile | Sterility | Sterility | ISO-11737-2 | Sterile | | PASS |
{8}------------------------------------------------
# Performance data of gloves based on ASTM D3577-09 and FDA 1000ml water leak test
{9}------------------------------------------------
## 10.0 Conclusion
Based on intended uses, technological characteristics and non - clinical performance data, the Latex Surgeon's Gloves Powder Free with Protein content labelling claims 50ug/dm2 or less per glove of extractable Protein is as safe, as effective, and performs as well as the legally marketed predicate device, Latex surgeon's Glove Powder Free with Protein content labeling claim of 50ug/dm2 or less per glove of extractable protein, K190632
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.