K202379 · Invacare Corporation · ITI · Mar 19, 2021 · Physical Medicine
Device Facts
Record ID
K202379
Device Name
Invacare AVIVA Storm RX Power Chair
Applicant
Invacare Corporation
Product Code
ITI · Physical Medicine
Decision Date
Mar 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Invacare AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.
Device Story
The Invacare AVIVA Storm RX is a 24V DC motor-driven power wheelchair designed for mobility and positioning. It features a rigid, non-folding base with rearward drive wheel placement, two front casters, and two rear anti-tippers. The device is powered by two 12V sealed VRLA gel batteries and utilizes the LiNX control system, which includes a microprocessor-based architecture, CAN bus interface, and Bluetooth LE for diagnostics and programming. The system supports various driver controls, including joysticks, sip-and-puff, and attendant controls. It incorporates a four-bar linkage independent suspension system and electro-mechanical friction brakes. The device is operated by the user or an attendant to navigate indoor/outdoor environments. Output consists of physical movement and positioning adjustments (tilt/recline). The device benefits patients by providing independent mobility and postural support. It is substantially equivalent to the Invacare AVIVA FX (K192216), differing primarily in drive wheel placement (rear vs. front) and associated mechanical modifications.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence supported by performance testing per ISO 7176 standards (stability, braking, energy consumption, dimensions, speed, fatigue, climatic, obstacle climbing, EMC, and wireless coexistence) and IEC 62304 software life cycle compliance.
Technological Characteristics
24V DC system; 4-pole single-stage drive motors; (2) 12V sealed VRLA gel batteries; LiNX microprocessor-controlled system; CAN bus interface; Bluetooth LE connectivity; four-bar linkage independent suspension; electro-mechanical friction brakes; rigid non-folding base; rear-wheel drive configuration. Complies with ISO 7176 series and IEC 62304 standards.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility and positioning assistance. Prescription use only.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below.
March 19, 2021
Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035
Re: K202379
Trade/Device Name: Invacare AVIVA Storm RX Power Chair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 10, 2021 Received: February 16, 2021
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202379
Device Name Invacare AVIVA Storm RX Power Wheelchair
Indications for Use (Describe)
The Invacare AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Submitter Information per 21 CFR 807.92(a)(1)
| SPONSOR: | Invacare Corporation<br>One Invacare Way<br>Elyria, OH 44035 |
|----------|--------------------------------------------------------------|
|----------|--------------------------------------------------------------|
| PRIMARY CONTACT: | Elijah Wreh |
|------------------|----------------------------|
| | Regulatory Affairs Manager |
| | Phone: (440) 329-6840 |
| | Email: ewreh@invacare.com |
| MANUFACTURER: | Invacare Corporation<br>1200 Taylor Street<br>Elyria, Ohio 44035 |
|---------------|------------------------------------------------------------------|
|---------------|------------------------------------------------------------------|
| DATE PREPARED per 21 CFR 807.92(a)(1): | March 19, 2021 |
|----------------------------------------|----------------|
|----------------------------------------|----------------|
Device Information per 21 CFR 807.92(a)(2)
| NAME OF SUBJECT DEVICE: | Invacare® AVIVA Storm RX Power Wheelchair |
|-------------------------|----------------------------------------------------------------------------------------------------------------|
| COMMON/USUAL NAME: | Power Wheelchair |
| CLASSIFICATION NAME: | Powered Wheelchair [21 CFR §890.3860] |
| REGULATORY CLASS: | 2 |
| PRODUCT CODE: | ITI: Wheelchair, Mechanical |
| PREDICATE DEVICES: | Primary: Invacare AVIVA FX Power Wheelchair<br>(K192216)<br>No reference devices were used in this submission. |
{4}------------------------------------------------
## Device Description per 21 CFR 808.92(a)(4)
The subject device is a rearward drive wheel version substantially equivalent to the previously cleared forward drive wheel Invacare® AVIVA FX Power Wheelchair (K192216) with LiNX Electronics and Ultra Low Maxx Seating System. The subject device version consists of the following changes:
- . Rearward drive wheel placement
- . Modification to suspension system
- . Modification to battery box design and seat interface brackets
- Minor changes to mechanical components such as new rims and colors .
The Invacare® AVIVA Storm RX Power Wheelchair is a 24V DC system, motor-driven wheelchair, utilizing the predicate device LiNX® control system. The subject device consists of a rigid or "non-folding" type power wheelchair base with rearward drive wheel placement with two (2) casters in the front and two anti-tippers in the rear. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.
## Intended Use per 21 CFR 807.92(A)(5)
The intended use of the device is to provide mobility and positioning to persons limited to a sitting position.
## Indications for Use per FORM FDA 3881
The Invacare® AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.
## Indications for Use Characteristics Comparison
Both the subject and predicate device share the same Indications for Use and Intended use.
# Technological Characteristics Comparison with the predicate device per 21 CFR 807.92(a)(6)
The technological characteristics comparison demonstrates that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device.
{5}------------------------------------------------
### COMPARISON of TECHNOLOGICAL CHARACTERISTICS with the PREDICATE DEVICE per 21 CFR 807.92(a)(6)
The device comparison shows that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device. The subject and predicate devices are intended to provide mobility and positioning to persons limited to a sitting position. The subject device components met the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise different questions of safety and effectiveness. The different technological characteristics do not raise different questions of safety and effectiveness as compared to the predicate device.
#### Basis of Substantial Equivalence per 21 CFR 807.100(b)(2)(ii)(A)
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A). The design verification testing and device comparison demonstrates that the subject device is substantially equivalent to the predicate device. The data generated from the subject device design verification test reports support a finding of substantial equivalence regarding device comparison, device specifications, and design characteristics and to provide mobility and positioning to persons limited to a sitting position.
## Summary of the Nonclinical Testing used for Substantial Equivalence
The performance testing, device comparison, and dimensional analysis demonstrate that the subject device components and features are the same or substantially equivalent to the predicate device regarding the following:
- . Static Stability
- Dynamic Stability of Electric Wheelchairs ●
- . Effectiveness of Brakes
- . Energy Consumption of Electric Wheelchairs and Scooters for Determination of Theoretical Distance Range
- . Dimensions Mass and Maneuvering Space
- . Maximum Speed Acceleration and Deceleration of Electric Wheelchairs
- . Seating and Wheel Dimensions
- Methods for Static Impact and Fatigue Strengths .
{6}------------------------------------------------
- Climatic Tests for Electric Wheelchairs ●
- Climbing Ability of Electrically Powered Wheelchairs ●
- Power and Control Systems for Electrically Powered Wheelchairs and Scooters ●
- Information Disclosure Documentation and Labeling ●
- Wheeled Mobility Devices for Use as Seats in Motor Vehicles
- Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and ● Scooters and Battery Chargers
- Batteries and chargers for powered wheelchairs ●
- Software Life Cycle ●
- Flammability Testing .
- Assessment of the Ignitability of Upholstered Furniture Ignition Source Smoldering ● Cigarette
- Wireless Coexistence
The data generated from the subject Invacare® AVIVA Storm RX Power Wheelchair design verification test reports support a finding of substantial equivalence regarding the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility and positioning to persons limited to a sitting position.
### Indications for Use Comparison Table
| Device | Subject Device | Predicate Device |
|------------------------|---------------------------------------------------------------------------------|-------------------------------------------------|
| | Invacare® AVIVA Storm RX Power Wheelchair | Invacare AVIVA FX Power Wheelchair<br>(K192216) |
| Indications for<br>Use | To provide mobility and positioning to persons<br>limited to a sitting position | Same |
| Intended Use | To provide mobility and positioning to persons<br>limited to a sitting position | Same |
| Type of Use | Prescription (RX Only) | Same |
{7}------------------------------------------------
| Component | Description | Invacare® AVIVA FX Power Wheelchair (K192216) Predicate Device | Invacare® AVIVA Storm RX Power Wheelchair Subject Device |
|-----------------------------|-----------------------------------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------|
| SEATING<br>UltraLow<br>Maxx | Powered Positioning Configurations | Fixed, Combinations of Tilt/Recline/Elevate | Combinations of Tilt/Recline |
| | Seat Widths | 16" to 24" | Same |
| | Seat Depths | 16" to 23" | Same |
| | Back Heights | 18" to 25" (tilt) or 20" to 27" (tilt and recline) | Same |
| | Upholstery | Meshtex, Startex, Spacetex, O-Vinyl, Polyester | Same |
| | Tilt Range | 50° | Same |
| | Recline Range | 168° | Same |
| | Patient Support Surfaces | Ref subject device 'Essential Equivalence' section | Same |
| | System Name | LiNX | Same |
| CONTROL<br>SYSTEM<br>LiNX | Cables | Variable cable lengths<br>A range of standard cable lengths available | Same |
| | System Architecture | Microprocessor Controlled | Same |
| | Non-Expandable Options | Yes | Same |
| | Expandable Options | Yes | Same |
| | Wireless Devices | Bluetooth | Same |
| | Power Source | 24V nominal | Same |
| | Bus Interface | CAN | Same |
| BASE | Base Configuration | Forward Drive Wheel Placement | Rearward Drive Wheel Placement |
| | Base Width | 24.3" | 24.8" |
| | Base Length | 42.7" | 33.5" |
| | Ground Clearance | 2.6" | 2.8" |
| | Batteries | (2) 12V Sealed VRLA Gel Batteries | Same |
| | Braking System | Electro-mechanical Friction Brake | Same |
| Drive Wheel Size | 14" x 3" | Same | |
| Incline Capability | 9° | Same | |
| Maximum Speed | 5.8mph | Same | |
| Motors | 4-Pole SSD | Same | |
| Occupant Weight<br>Cap | Up to 300lbs. | Same | |
| Suspension | Four Bar Linkage Suspension System | Four Bar Linkage Suspension System | |
| | Four Bar Linkage Suspension System<br>(Front Drive Wheel) | Four Bar Linkage Suspension System<br>(Rear Drive Wheel) | |
# Design and Technological Characteristics Comparison – Finished Device
{8}------------------------------------------------
{9}------------------------------------------------
| Description | Subject Device<br>Invacare® A VIVA StormRX Power Wheelchair | Predicate Device<br>Invacare A VIVA FX Power Wheelchair (K192216) |
|------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------|
| Base Configuration | Rear Wheel Drive | Front Wheel Drive |
| Base Width | 24.8" | 24.3" |
| Base Length | 33.5" | 42.7" |
| Seat-to-Floor Heights | 16.75"<br>17.25"<br>18.25"<br>19.25" | 16.75"<br>17.75"<br>18.75"<br>19.75" |
| Ground Clearance | 2.8" | 2.6" |
| Batteries | (2) 12V Sealed VRLA Gel Batteries | (2) 12V Sealed VRLA Gel Batteries |
| Braking System | Electro-mechanical Friction Brake | Electro-mechanical Friction Brake |
| Castor Size | 8" x 2.5" | 8" x 2.5" |
| Control System | LiNX | LiNX |
| Drive Wheel Size | 14" x 3" | 14" x 3" |
| Incline Capability | 9° | 9° |
| Maximum Speed | 5.8 mph | 5.8 mph |
| Motors | 4-Pole SSD | 4-Pole SSD |
| Maximum Weight<br>Bearing Capacity | 300lbs. | 300lbs. |
| Range | >12 miles | >12 miles |
| Suspension | Four Bar Linkage Independent Suspension System<br>(Rear Drive Wheel) | Four Bar Linkage Independent Suspension System<br>(Front Drive Wheel) |
# Performance Characteristics Comparison – Finished Device
{10}------------------------------------------------
| LiNX Control System- Modules | Description | AVIVA Storm RX | AVIVA FX | Wireless Technology | Impact to Safety and Effectiveness |
|------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|----------|----------------------------------------------------|---------------------------------------|
| Primary Driver Controls | Drive Only Remote | Yes | Yes | Bluetooth LE Wireless: Diagnostics | The versions are the same. |
| | Remotes with LED Display | Yes | Yes | Bluetooth LE Wireless: Diagnostics | The versions are the same. |
| | Rehabilitation Remote with Colour LCD Display and Touch | Yes | Yes | Bluetooth LE Wireless: Mouse Mover and Diagnostics | The versions are the same. |
| Alternative Driver Controls (LiNX) | Compact Remote | Yes | Yes | Bluetooth LE Wireless: Mouse Mover and Diagnostics | The versions are the same. |
| | Sip 'N' Puff (SNP) | Yes | Yes | No Wireless | |
| | Attendant Driver Control | Yes | Yes | No Wireless | |
| Power Modules | 120A, Dual Motor, Power Controller with Lights and 2 Actuator Channels | Yes | Yes | No Wireless | The versions are the same. |
| Actuator Modules | Actuator Controls only modules: 4 Channel | Yes | Yes | | |
| Stability Control | G-Trac / Gyro | Yes | Yes | No Wireless | The versions are the same. |
| Programming tool | LAK: LiNX Access Key | Yes | Yes | Bluetooth LE Wireless: Remote Programming | |
| Switch Input Module | Module used to add separate switched for Power on/off when used with REM 215/216 | Yes - The difference is that the switch input module adds wired remote on/off capability | No | No Wireless | *Differences |
| LiNX Control System- Modules | Description | AVIVA Storm RX | AVIVA FX | Wireless Technology | Impact to Safety and<br>Effectiveness |
| Environmental Control Module | Environmental Control module used for<br>off-chair control of various devices | Yes | Yes | No Wireless | The versions are the<br>same. |
| Switch Interface Module | Switch Interface Module | Yes – The<br>difference is the<br>switch module is a<br>junction box. | No | No Wireless | *Differences |
| Light Module | Interface to LED lights on StormRX<br>chair | Yes – The<br>difference is low<br>voltage power<br>supply for LED<br>light. | No | No Wireless | *Differences |
| USB Charge Port | Allows charging of Smart pone via<br>USB | Yes | Yes | No Wireless | The versions are the<br>same. |
| Enhanced DLX-IN500 Input<br>Module | Sip 'N' Puff (SNP) | Yes – Same as the<br>regular DLX-<br>IN500 input<br>module, but the<br>difference is that<br>the firmware was<br>updated. | Yes | No Wireless | *Differences |
# LiNX® Components and Accessories Specifications
{11}------------------------------------------------
*Differences: Differences between the subject device features and the previously cleared predicate device do not raise new issues of safety or effectiveness.
{12}------------------------------------------------
## Performance Data
## Non-Clinical Test per 21 CFR 807.92(b)(1)
International Organization of Standardization (ISO) testing, California Technical (CAL), American National Standards Institute (ANSI) and European (EN) standards testing were performed to demonstrate that the subject Invacare® AVIVA Storm RX Power Wheelchair meet the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise questions of safety and effectiveness.
{13}------------------------------------------------
| Test Standard | Test Description |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 7176-1:2014 | Wheelchairs Part 1: Determination of Static Stability |
| ISO 7176-2:2017 | Wheelchairs Part 2: Determination of Dynamic Stability of Electrically Powered Wheelchairs |
| ISO 7176-3:2012 | Wheelchairs Part 3: Determination of Effectiveness of Brakes |
| ISO 7176-4:2008 | Wheelchairs Part 4: Energy Consumption of Electrical Wheelchairs and Scooters for Determination of Theoretical Distance Range |
| ISO 7176-5:2008 | Wheelchairs Part 5: Determination of Dimensions, Mass and Maneuvering Space |
| ISO 7176-6:2018 | Wheelchairs Part 6: Determination of Maximum Speed, Acceleration and Deceleration of Electric Wheelchairs |
| ISO 7176-7:1998 | Wheelchairs Part 7: Measurement of Seating and Wheel Dimensions |
| ISO 7176-8:2014 | Wheelchairs Part 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths |
| ISO 7176-9:2009 | Wheelchairs Part 9: Climatic Tests for Electric Wheelchairs |
| ISO 7176-10:2008 | Wheelchairs Part 10: Determination of Obstacle Climbing Ability of Electrically Powered Wheelchairs |
| ISO 7176-11:2012 | Wheelchairs Part 11: Test Dummies |
| ISO 7176-13:1989 | Wheelchairs Part 13: Determination of Coefficient of Friction of Test Surface |
| ISO 7176-14:2008 | Wheelchairs Part 14: Power and Control Systems for Electrically Powered Wheelchairs and Scooters – Requirements and Test Methods |
| ISO 7176-15:1996 | Wheelchairs Part 15: Requirements for Information Disclosure, Documentation and Labeling |
| ISO 7176-16:2012 | Wheelchairs Part 16: Resistance to Ignition of Postural Support Devices |
| ISO 7176-19 | Wheelchairs Part 19: Wheeled Mobility Devices for use as Seats in Motor Vehicles |
| ISO 7176-21:2008 | Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers |
| ISO 7176-22:2014 | Wheelchairs Part 22: Set-up Procedures |
| ISO 7176-25:2013 | Wheelchairs Part 25: Batteries and chargers for powered wheelchairs |
| IEC 62304:2006 | Medical Device Software – Software Life Cycle |
| ANSI C63.27:2017 | American National Standard for Evaluation of Wireless Coexistence |
{14}------------------------------------------------
## Performance Data Conclusions per 21 CFR 807.92(b)(3)
The subject device utilizes the same intended use, same material composition, and similar technological characteristics as the predicate device. The non-clinical laboratory data support a finding of substantial equivalence of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.