The Diagnostic Ultrasound System Aplio a550 Model CUS-AA550,Aplio a450 Model CUS-AA450 and Aplio a Model CUS-AA000 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intraoperative (abdominal), pediatric, small organs (including thyroid, breast, testicle), trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural. This system provides high-quality ultrasound images in the following modes of operation: B (2D), M, Color Doppler (blood-flow imaging), Doppler (PWD, CWD, Power) (blood-flow spectrum), Combined (B/M; B/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD), Precision Imaging, Apli Pure, Micro Pure, BEAM, TDI, Shear wave, Elastography, SMI (ADF), 2D Wall Motion Tracking, Smart Sensor3D, 3D Color (Volume color), 4D, STIC, STIC Color, Fusion, Smart Navigation, ATI, CHI (Per FDA approved contrast agent prescribing information), Shadow Glass. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
Device Story
Mobile diagnostic ultrasound system; uses various probes (2-20 MHz) to acquire ultrasound signals; transforms signals into B-mode, M-mode, Doppler, and advanced imaging modes (e.g., Shear wave, Elastography, 4D, Fusion). Operated by physicians/trained personnel in clinical settings. Provides real-time visualization and diagnostic information. Includes workflow support tools (LI-RADS, BI-RADS, TI-RADS checklists, Workflow Navigator) and automated optimization features (Smart Optimization). Output displayed on system monitor for clinical assessment; aids in diagnosis, biopsy guidance, and procedural planning. Benefits include improved image quality, standardized reporting, and streamlined clinical workflows.
Clinical Evidence
Bench testing and volunteer studies conducted. Verification/validation testing confirmed new and improved features met specifications. No clinical trials required for substantial equivalence determination.
Technological Characteristics
Mobile ultrasound system; flat/convex/sector array probes (2-20 MHz). Conforms to AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, IEC 62359, ISO 10993-1. Software V5.1. Features include B-mode, Doppler, Elastography, 3D/4D, and automated reporting/workflow tools.
Indications for Use
Indicated for visualization of structures and dynamic processes in fetal, abdominal, intraoperative, pediatric, small organs (thyroid, breast, testicle), trans-rectal, neonatal/adult cephalic, cardiac, peripheral vascular, transesophageal, musculo-skeletal, laparoscopic, and thoracic/pleural applications. Suitable for use by physicians or trained qualified persons in hospital/clinical settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Aplio a550 and a450, Software V4.0 Diagnostic Ultrasound System (K191663)
Reference Devices
Aplio i900/i800/i700, Software V4.0 Diagnostic Ultrasound System (K191467)
{0}------------------------------------------------
October 15, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in smaller blue letters below that.
Canon Medical Systems Corporation % Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780
## Re: K202364
Trade/Device Name: Aplio a550, Aplio a, Diagnostic Ultrasound System, Software V5.1 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 17, 2020 Received: August 19, 2020
## Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K202364
### Device Name
Aplio a550, Aplio a450, and Aplio a, Diagnostic Ultrasound System, Software V5.1
### Indications for Use (Describe)
The Diagnostic Ultrasound System Aplio a550 Model CUS-AA550,Aplio a450 Model CUS-AA450 and Aplio a Model CUS-AA000 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intraoperative (abdominal), pediatric, small organs (including thyroid, breast, testicle), trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural.
This system provides high-quality ultrasound images in the following modes of operation: B (2D), M, Color Doppler (blood-flow imaging), Doppler (PWD, CWD, Power) (blood-flow spectrum), Combined (B/M; B/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD), Precision Imaging, Apli Pure, Micro Pure, BEAM, TDI, Shear wave, Elastography, SMI (ADF), 2D Wall Motion Tracking, Smart Sensor3D, 3D Color (Volume color), 4D, STIC, STIC Color, Fusion, Smart Navigation, ATI, CHI (Per FDA approved contrast agent prescribing information), Shadow Glass. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K202364 510(k) SUMMARY
- 1. SUBMITTER'S NAME: Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi-ken, Japan 324-8550
- 2. OFFICIAL CORRESPONDENT Hideo Saeki
- 3. ESTABLISHMENT REGISTRATION: 9614698
#### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
- 5. DATE PREPARED:
August 17, 2020
#### 6. DEVICE NAME: Aplio a550, Aplio a450, and Aplio a, Diagnostic Ultrasound System, Software V5.1
#### 7. TRADE NAME(S):
Diagnostic Ultrasound System Aplio a550 Model CUS-AA550 Diagnostic Ultrasound System Aplio a450 Model CUS-AA450 Diagnostic Ultrasound System Aplio a Model CUS-AA000
#### 8. COMMON NAME:
System, Diagnostic Ultrasound
#### 9. DEVICE CLASSIFICATION:
Class II
Ultrasonic Pulsed Doppler Imaging System – Product Code: 90-IYN [per 21 CFR 892.1550] Ultrasonic Pulsed Echo Imaging System – Product Code: 90-IYO [per 21 CFR 892.1560] Diagnostic Ultrasonic Transducer – Product Code: 90-ITX [per 21 CFR 892.1570]
{4}------------------------------------------------
### 10. PREDICATE DEVICE:
| Product | Marketed by | 510(k) Number | Clearance Date |
|--------------------------------------------------------------------------------------------------|--------------------------------|---------------|--------------------|
| Aplio a550 and a450, Software V4.0 Diagnostic Ultrasound System<br>(Primary Predicate Device) | Canon Medical Systems USA, Inc | K191663 | September 18, 2019 |
| Aplio i900/i800/i700, Software V4.0 Diagnostic Ultrasound System<br>(Reference Predicate Device) | Canon Medical Systems USA, Inc | K191467 | July 19, 2019 |
| Versana Premier<br>(Reference Predicate Device) | GE Healthcare | K182277 | November 21, 2018 |
| AIXPLORER® MACH Ultrasound Diagnostic Systems<br>(Reference Predicate Device) | Super Imagine | K180572 | May 29, 2018 |
| Epiq 7<br>(Reference Predicate Device) | Philips Medical Systems | K181485 | July 27, 2018 |
## 11. DEVICE DESCRIPTION:
The Aplio a550 Model CUS-AA550, Aplio a450 Model CUS-AA450 and Aplio a Model CUS-AA000 are mobile diagnostic ultrasound systems. These systems are Track 3 devices that employ a wide array of probes including flat linear array, convex linear array, and sector array with frequency ranges between approximately 2 MHz to 20 MHz.
### 12. INDICATIONS FOR USE:
The Diagnostic Ultrasound Systems Aplio a550 Model CUS-AA550, Aplio a450 Model CUS-AA450 and Aplio a Model CUS-AA000 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, transvaginal, trans-rectal, neonatal cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and thoracic/pleural. This system provides high-quality ultrasound images in the following modes of operation: B (2D), M, Color Doppler (blood-flow imaging), Doppler (PWD, CWD, Power) (blood-flow spectrum), Combined (B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD), Precision Imaging, Apli Pure, Micro Pure, BEAM, TDI, Shear wave, Elastography, SMI (ADF), 2D Wall Motion Tracking, Smart 3D, Smart Sensor3D, 3D Color (Volume color), 4D, STIC, STIC Color, Fusion, Smart Navigation, ATI, CHI (Per FDA approved contrast agent prescribing information), Shadow Glass. This system is suitable for use in hospital and clinical settings by physicians or legally qualified persons who have received the appropriate training.
{5}------------------------------------------------
## 13. SUBSTANTIAL EQUIVALENCE:
The purpose of this submission is the modification of cleared Aplio a550, and Aplio a, Diagnostic Ultrasound System, V4.0, 510(k) cleared per K191663, marketed by Canon Medical Systems USA, Inc. to which the subject device is substantially equivalent. The Aplio a450, and Aplio a, V5.1 employ the same fundamental scientific technology as the predicate devices, and functions in a manner similar to and is intended for the same use as the predicate devices. The subject device includes modifications to the cleared device which improves upon existing features.
This submission also includes details regarding new features determined to be substantially equivalent to features cleared under the predicate devices referenced within this submission. The Aplio a550, Aplio a450 and Aplio a, V5.1 and predicate Aplio a550, Aplio a450, and Aplio a,
- . V4.0 have the same clinical intended use and the same imaging modes, except that Thoracic/Pleural is now available on several transducers on Aplio a550, and Aplio a, V5.1, and Shear Wave Elastography mode was expanded to include musculo-skeletal clinical use.
- . The transducers supported in the subject Aplio a550, and Aplio a and predicate Aplio a550, Aplio a450 and Aplio a are identical, except for the addition of the new transducer PST-65BT which was previously cleared with reference predicate Aplio i900/i800/i700, V4.0, K191467 and shares the same fundamental technological and biocompatibility characteristics as existing transducers supported by the subject system.
- The software features supported in subject Aplio a550, and Aplio a and predicate Aplio a550, Aplio a450, and Aplio a are identical except the following main, new features:
- o LI-RADS Checklist (Contrast Enhanced Ultrasound (CEUS) Liver Imaging Reporting and Data System), a new feature predicated on the RADS kit feature, which includes the BI-RADS Checklist (Breast Imaging Reporting and Data System) and TI-RADS Checklist (Thyroid lmaging Reporting and Data System), which is also introduced on Aplio a550, Aplio a450, and Aplio a, V5.1 and was previously cleared on reference predicate device Aplio i900/i800/i700, V4.0 (K191467). LI-RADS Checklist supports the standardization of liver characterization based on American College of Radiology (ACR) guidelines, BI-RADS Checklist supports the standardization of breast mass characterization according to ACR guidelines, and TI-RADS Checklist supports the standardization of thyroid mass characterization according to ACR guidelines.
- o Workflow Navigator, a new feature predicated on existing workflow functionality available on predicate Aplio a550, Aplio a450, and Aplio a which provides workflow and reporting support based on American Society of Echocardiography (ASE) guidelines for specific medical conditions.
- Smart Optimization, a new feature similar to "Whizz" on the reference predicate device O Versana Premier by GE Healthcare (K182277) which is intended to improve image quality, targeting the abdominal region.
- O Prostate Biopsy Report, a new feature which adds functionality existing on predicate
{6}------------------------------------------------
Made For life
Aplio a550, Aplio a450, and Aplio a, to provide a workflow panel for biopsy procedures and 3D modeling for visualization during these procedures.
- In addition to these new features, several existing features have been modified for improved functionality on the subject device:
- Smart Body Mark, an improvement to manual breast scan workflow by incorporation of O previously cleared functionality
- MSK Protocol movie, to support the display of musculo-skeletal (MSK) instructional movies O with existing functionality
- Wide View support, an expansion of the scan range available with convex transducers O
- Sensor 3D, modified to support improved resolution O
- Obstetric Image quality, modified to support improved visualization O
- The following features were migrated to the subject Aplio a550, aplio a450, and Aplio a, from reference predicate, Aplio iSeries (K191467)
- O Multi Parametric Report, allows the review of results of up to five applications on a single worksheet
- 2D Wall Motion Tracking Fetal kit, enables the analysis of second trimester fetal hearts O
### 14. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the AAMI/ANSI ES60601-1:2012, IEC 60601-1-2 (2014), IEC 60601-2-37 (2015), IEC 62304 (2015), IEC 62359 (2017) and ISO 10993-1(2009) standards.
### 15. TESTING
Risk Analysis and Verification/Validation testing conducted through the bench testing which is included in this submission demonstrate that the requirements for the new and improved features have been met. Additional performance testing, using test data and volunteer studies, were conducted to assess improvements to existing features. Results of all these studies demonstrate that the features included in this submission meet specifications and perform as intended. FDA guidance document "Marketing Clearance of Diagnostic Ultrasound Systems and Transducer", issued June 27, 2019 was referenced for this submission, along with Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005. Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 18, 2018, was also referenced for this submission.
Additionally, testing of this device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and UL systems.
{7}------------------------------------------------
## 16. CONCLUSION
The Aplio a550 Model CUS-AA550, Aplio a450 Model CUS-AA450 and Aplio a Model CUS-AA000, V5.1 are substantially equivalent to the predicate devices. The subject devices function in a manner similar to and are intended for the same use as the predicate devices, as described in the labeling. Based upon the bench testing, clinical evaluation, successful completion of software validation, and application of risk management and design controls, it is concluded that this device is safe and effective for its intended use and performance, and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.