K202346 · Veran Medical Technologies, Inc. · EOQ · Feb 19, 2021 · Ear, Nose, Throat
Device Facts
Record ID
K202346
Device Name
SPiN Vision Video Bronchoscope System
Applicant
Veran Medical Technologies, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Feb 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Veran SPIN Vision™ Single-Use Flexible Bronchoscope is intended to be used with the SPiN Vision™ Digital Video Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree. The Veran SPiN Vision™ Single-Use Flexible Bronchoscopes and SPiN Vision™ Video Processor are for use in hospitals, clinics, and/or urgent care centers by trained physicians.
Device Story
System comprises single-use flexible video bronchoscope and reusable video processor; used for airway/tracheobronchial tree endoscopy. Bronchoscope features distal CMOS sensor and LED light source; captures reflected light to generate video signal. Processor receives signal via encoding circuit; performs brightness/image correction; outputs to external monitor. Operated by trained physicians in hospitals/clinics/urgent care. Provides real-time visualization to facilitate diagnostic and therapeutic procedures. Benefits include sterile single-use design reducing cross-contamination risk; maneuverable tip and working channel allow accessory insertion for clinical interventions.
Clinical Evidence
Bench testing only. Testing included ISO 8600-3/4 compliance, mechanical performance (leaking, bending, articulation, endurance), biocompatibility, aging, sterile packaging integrity, electrical safety (IEC 60601-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), color performance, image intensity uniformity, and software verification/validation. All tests passed.
Technological Characteristics
Single-use flexible bronchoscope; reusable video processor. Distal CMOS sensor; LED light source. Sterilized via Ethylene Oxide. Complies with ISO 8600-3, ISO 8600-4, IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Features maneuverable tip with locking function, working channel, and suction capability.
Indications for Use
Indicated for use in hospitals, clinics, and urgent care centers by trained physicians for endoscopy within the airways and tracheobronchial tree.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
February 19, 2021
Veran Medical Technologies, Inc. B. Nathan Hunt Vice President, Quality Assurance and Regulatory Affairs 1908 Innerbelt Business Center Dr St. Louis, Missouri 63114
Re: K202346
Trade/Device Name: Veran SPiN Vision™ Bronchoscope System Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ Dated: January 15, 2021 Received: January 19, 2021
Dear B. Nathan Hunt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shu-Chen Peng, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202346
Device Name Veran SPiN VisionTM Bronchoscope System
#### Indications for Use (Describe)
The Veran SPIN VisionTM Single-Use Flexible Bronchoscope is intended to be used with the SPiN Vision™ Digital Video Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Veran SPiN Vision™ Single-Use Flexible Bronchoscopes and SPiN Vision™ Video Processor are for use in hospitals, clinics, and/or urgent care centers by trained physicians.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size: 20px;">☐</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Veran Medical Technologies. The logo features a green and blue circular design on the left, followed by the word "VERAN" in large, bold, black letters. Below the word "VERAN" are the words "MEDICAL TECHNOLOGIES" in smaller, light blue letters.
1908 INNERBELT BUSINESS CENTER DRIVE ST. LOUIS, MISSOURI 63114 PHONE 314.659.8500 FAX 314.659.8560 www.veranmedical.com
K202346
# 510(k) Summary Veran Medical Technologies, Inc. SPiN Vision™ Bronchoscope System
In accordance with 21 CFR 807.92 the following summary of information is provided, on this date,
January 14th, 2021:
# 1. 510(k) SUBMITTER
Veran Medical Technologies 1908 Innerbelt Business Center Drive St. Louis, MO 63114 Tel: 314-659-8500 Fax: 314-659-8560
## Contact Person:
B. Nathan Hunt Vice President, Quality Assurance and Requlatory Affairs Veran Medical technologies 1908 Innerbelt Business Center Drive St. Louis. MO 63114 Tel: 314-659-8500 Fax: 314-659-8560
## Date Prepared:
January 14th, 2021
## 2. DEVICE
## Trade Name of Device:
Veran SPiN Vision™ Bronchoscope System Models: SYS-5100, SPiN Vision™ Video Processor INS-7100, SPiN Vision™ Scope, 4.0mm OD, 2.0mm WC INS-7130, SPiN Vision™ Scope, 6.0mm OD, 3.0mm WC
### Common or Usual Name:
Bronchoscope (flexible or rigid) and accessories
### Classification Names:
Bronchoscope (flexible or rigid) and accessories (21 CFR 874.4680)
Regulation Class: II
Classification Code: EOQ
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Veran Medical Technologies. The logo consists of a green and blue geometric shape resembling a stylized flower or starburst, followed by the word "VERAN" in a bold, black serif font. Below "VERAN" are the words "MEDICAL TECHNOLOGIES" in a smaller, light blue sans-serif font.
1908 INNERBELT BUSINESS CENTER DRIVE ST. LOUIS, MISSOURI 63114 PHONE 314.659.8500 FAX 314.659.8560 www.veranmedical.com
K202346
# 3. PREDICATE DEVICE
### Predicate Device:
Trade Name: Vathin® Video Bronchoscope System
Common Name: Bronchoscope (flexible or rigid) and accessories
510(K) No.: K191828
Device Models:
Vathin®H-SteriScope™ I Single-use flexible Video Bronchoscope BCV1-01 BCV1-02 BCV1-C1 BCV1-C2 BCV1-H1 BCV1-K1 BCV1-K1 BCV1-K2 BCV1-M1 BCV1-M2 BCV1-01 BCV1-02 BCV1-S1 BCV1-S2 BCV1-U1 BCV1-U2 BCV1-W1 BCV1-W2 Vathin®VisionCenterTM I Digital Video Processor DVP-A1
Classification Name: Bronchoscope (flexible or rigid) and accessories (21 CFR 874.4680)
Product Code: EOQ
Manufacturer: Hunan Vathin Medical Instrument Co., Ltd.
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
### 4. DEVICE DESCRIPTION
The Veran SPiN Vision™ Bronchoscope System consists of the Veran SPiN Vision™ Single-Use Flexible Video Bronchoscope accessories (Model Nos. INS-7100 and INS-7130) and the Veran SPiN Vision™ Video Processor (Model: SYS-5100) for clinical image processing. The Veran SPiN Vision™ Single-Use Flexible Bronchoscope is introduced within the airways or tracheobronchial tree during Bronchoscopy. The Veran SPiN Vision™ Video Processor provides power and processes the images for medical electronic endoscope.
The Veran SPiN Vision™ Single-use Flexible Video Bronchoscope is a sterile, single use flexible bronchoscope. The Veran SPiN Vision™ Video Processor is a reusable device.
The light emitted by the LED cold light source of the Veran SPiN Vision™ Single-use Flexible Video Bronchoscope lens is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is imaged on the CMOS (complementary metal oxide semiconductor). The CMOS acquisition image is controlled by the CMOS drive circuit, and the standard color video signal is output to the Veran SPiN Vision™ Video Processor via the encoding circuit. The Veran SPiN Vision™ Video Processor adjusts the brightness of the light source or corrects the image according to the video signal output from the CMOS, and outputs the corrected standard color video signal.
The Veran SPiN Vision™ Single-use Flexible Video Bronchoscope has the following physical and performance characteristics:
- Maneuverable tip controlled by the user
- . Flexible insertion cord
- . Camera and LED light source at the distal tip
- Working channel
- . Sterilized by Ethylene Oxide
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Veran Medical Technologies. The logo features a green and blue circular design to the left of the company name. The company name is written in a bold, sans-serif font, with the word "VERAN" in black and "MEDICAL TECHNOLOGIES" in blue.
- For single use .
The differences between the Veran SPiN Vision™ Single-use Flexible Video Bronchoscope models are as follow:
- . Working channel diameters
- . Insertion tube outer diameter
The Veran SPiN Vision™ Video Processor has the following physical and performance characteristics:
- . Provides images from the Veran SPiN Vision™ Single-use Flexible Video Bronchoscope for observation
- . Can connect to an external monitor
- . Reusable device
### 5. INDICATION FOR USE
The Veran SPiN Vision™ Single-Use Flexible Bronchoscope is intended to be used with the SPiN Vision™ Digital Video Processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Veran SPiN Vision™ Single Bronchoscopes and SPiN Vision™ Video Processor are for use in hospitals, clinics, and/or urgent care centers by trained physicians.
### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Veran SPiN Vision™ Bronchoscope System has the same technological characteristics and fundamental design as the predicate device. The Veran SPiN Vision™ Bronchoscope System and the predicate device are designed to provide real-time images to the physician in order to facilitate diaqnostic and therapeutic procedures in the lung tract.
The differences between The Veran SPiN Vision™ Bronchoscope System and predicate device do not alter suitability of the proposed device for its intended use.
The Veran SPiN Vision™ Bronchoscope System is substantially equivalent to the cleared predicate device (K191828).
The Veran SPiN Vision™ Bronchoscope System has equivalent Indications for Use as the predicate device.
The subject and predicate device are similar in the following technological elements:
- . They are single-use devices delivered sterile.
- . They are flexible endoscopes with a maneuverable tip.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Veran Medical Technologies. The logo features a green and blue circular graphic on the left, followed by the word "VERAN" in large, black, serif font. Below "VERAN" are the words "MEDICAL TECHNOLOGIES" in a smaller, blue, sans-serif font.
K202346
- . They are video endoscopes with a camera located in the distal tip to provide an image on a separate monitor.
- . They use an LED-light source located in the distal tip.
- They have suction functionality.
- . They have a working channel for insertion of endoscopic accessories.
The following technological differences exist between the subject and predicate device:
- The proposed device has different physical specifications.
- . The conical lock/working channel port of the proposed device is compatible with existing biopsv valves.
- . The steering level of the proposed device includes a locking function.
- . The video processor of the proposed device has a different cable connector.
Non-clinical bench testing has been completed ensuring that the differences don't affect the safety and effectiveness of the proposed subject device.
# 7. PERFORMANCE DATA
The following performance data are provided in support of the substantial equivalence determination:
- . Declaration of Conformity with the product specific standards ISO 8600-3, and ISO 8600-4
- Performance tests to test leaking, bending, articulating angle, endurance of the bending . section, radius of the bending section and irrigation tests were performed.
- . Packaging and labeling performance tests were performed
- . Biocompatibility Test
- . Aging Performance Test
- Sterile Packaging Integrity Test •
- Electrical Compatibility according to IEC 60601-1-2
- . Electrical Safety according to IEC 60601-1 and IEC 60601-2-18
- . Color Performance Testing and Image Intensity Uniformity (IIU) Testing
- . Software Verification and Validation Testing
Results: All tests passed.
## 8. CONCLUSION
The Veran SPiN Vision™ Video Bronchoscope System is substantially equivalent to its predicate device. The non-clinical testing demonstrates that the device is as safe and as effective and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.