K202292 · Ebm Technologies, Inc. · LLZ · Oct 5, 2020 · Radiology
Device Facts
Record ID
K202292
Device Name
UDE
Applicant
Ebm Technologies, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
UDE software is intended to display images from CT, MR, CR, US, XA and SC for the trained physician 's diagnosis or referring purpose. UDE provides wireless and portable access to medical images. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography diagnosis.
Device Story
UDE is a software application for Apple iPad Pro; enables wireless, portable access to medical images stored on an existing UDE server. Operates by allowing trained physicians to query, retrieve, and display images (CT, MR, CR, US, XA, SC) for diagnostic or referral purposes. Features include distance calculation, window/level adjustment, pan, zoom, cine loop, and mammography hanging protocols. Used in clinical environments as a secondary viewing tool when primary workstations are inaccessible. Output is visual image presentation on the iPad Pro screen; assists physicians in clinical review and decision-making. Benefits include increased mobility and timely access to patient data.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including software verification/validation per IEC 62304 and display performance testing per AAPM (2005) standards.
Technological Characteristics
Software-based PACS viewer for Apple iPad Pro (iOS 9, 11, 12, 13). Connectivity via wireless network to UDE server. Features include image processing (window/level, zoom, pan, distance measurement), user authentication, and FIFO image management. Display performance evaluated against AAPM 2005 standards. Software developed per IEC 62304.
Indications for Use
Indicated for trained physicians to display and review medical images (CT, MR, CR, US, XA, SC) for diagnostic or referring purposes when a primary workstation is unavailable. Contraindicated for mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
October 5, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
EBM Technologies, Inc. % Mr. John Su Regulatory Affairs Manager 5F., No. 516, Sec. 1, Neihu Rd. Taipei. Taiwan 11493 REPUBLIC OF CHINA
Re: K202292
Trade/Device Name: UDE Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 27, 2020 Received: August 12, 2020
Dear Mr. Su:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K202292
Device Name UDE
#### Indications for Use (Describe)
UDE software is intended to display images from CT, MR, CR, US, XA and SC for the trained physician 's diagnosis or referring purpose. UDE provides wireless and portable access to medical images. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography diagnosis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
K202292
## 5.1 Device Submitter
EBM Technologies Incorporated 5F., No. 516, Sec. 1, Neihu Rd., Taipei, Taiwan, Republic of China Phone: 886 2 8751 4567 Fax: 886 2 8751 3300 Contact Person: John Su Regulation Affairs Manager Tel: 808. 397.6809 john@ebmtech.com
### 5.2 Device Name
Device Trade Name: UDE Device Class : Class 2 Product Code : LLZ Regulation name and number: Picture Archiving and Communications System , 21 CFR 892.2050
## 5.3 Substantially Equivalent (predicate) device(s)
Device Trade Name: EBM iDO Viewer 510(k) Number: K140399 Device Class : Class 2 Product Code : LLZ Regulation name and number: Picture Archiving and Communications System , 21 CFR 892.2050 Manufacturer: EBM Technologies Incorporated
{4}------------------------------------------------
# 5.4 Device Description:
UDE is a software device that can be installed on Apple iPad Pro. Through wireless network, user can login, query and display the images which are stored in heir existing UDE server. The device can be installed in iOS 9, 11, 12 and 13 version platform such as iPad /iPhone, but can't be installed in platforms other than iOS 9, 11, 12 and 13 version . The image display quality on iPad /iPhone will be almost the same as on iPad Pro when it is used for diagnosis purpose of CT, MR, US, XA and SC. However, if it is used for CR diagnosis purpose, we will strongly suggest that users should adopt iPad Pro. because the screen size of iPad Pro is larger than those of Apple® iPad/iPhone.
Main features of UDE are listed below
- Receive, Store, Retrieve, Display, and Process Digital Images(CT, MR, US, CR, Full- Field . Mammography, XA,SC etc.)
- · The displayed CT, MR, CR,, US,XA and SC images can be diagnosed by the trained physicians.
- . The displayed CT, MR, CR, Full- Field Mammography, US, XA and SC images can be referred by the trained physicians. Mammography images are not for diagnostic use
- · Communication log file
- · Auto delete old images (FIFO)
- · Overlay labels
- · User Authentication
- Display of Clinical Patient Data .
- Distance Calculation
- · layout adjustment (1×1, 2×1, 1×2, 2×2)
- · Pan
- Zoom
- Window Level
- · Cine Loop
- Mammography hanging protocol
## 5.5 Indication for Use:
UDE software is intended to display images from CT, MR, CR, XA and SC for the trained physician 's diagnosis or referring purpose. UDE provides wireless and portable access to medical images. It is not intended to be used as, or to replace, a full diagnostic workstation or system and should be used only when there is no access to a workstation. This device is not to be used for mammography diagnosis.
{5}------------------------------------------------
# 5.6 Technical characteristics
UDE is a software device that can be installed on Apple iPad Pro. Through wireless network, user can login, query and display the images which are stored in heir existing UDE server. It has functions related to the medical image presentation and processing. These functions can help the trained physician to perform the medical images review and diagnosis if environment lighting condition has been evaluated and in opportune setting.
| Device | Subject device | Predicate device | Comparison and<br>Explanation of<br>Differences |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Topic | UDE | EBM iDO Viewer | |
| 510k number | N/A | K140399 | |
| Intended Use /<br>Indications for Use | UDE software is intended<br>to display images from<br>CT, MR , CR, US, XA<br>and SC for the trained<br>physician 's diagnosis or<br>referring purpose. UDE<br>provides wireless and<br>portable access to<br>medical images. It is not<br>intended to be used as, or<br>to replace, a full<br>diagnostic workstation or<br>system and should be<br>used only when there is<br>no access to a<br>workstation. This device<br>is not to be used for<br>mammography diagnosis. | EBM iDO Viewer software is<br>intended to display images from<br>CT/MR for the trained physician 's<br>diagnosis or referring purpose.<br>EBM iDO Viewer provides<br>wireless and portable access to<br>medical images. It is not intended<br>to be used as, or to replace, a full<br>diagnostic workstation or system<br>and should be used only when there<br>is no access to a workstation. This<br>device is not to be used for<br>mammography. | The images displayed by<br>subject device have extended<br>to CR and Full-Field<br>Mammography , US, XA and<br>SC. It leads to the change of<br>intended use. It raises no new<br>issues of safety or<br>effectiveness |
| Receive, Store,<br>Retrieve, Display,<br>and Process Digital<br>Medical Images | Yes | Yes | No difference |
| Display of Clinical<br>Patient Data When<br>No Access to a<br>Workstation | Yes | Yes | No difference |
| Distance Calculation | Yes | Yes | No difference |
| layout adjustment (1<br>×1,2×1,1×2,2<br>×2) | Yes | Yes | No difference |
| Window / Level | Yes | Yes | No difference |
| Zoom, Pan | Yes | Yes | No difference |
| User Authentication | Yes | Yes | No difference |
| Modality images for<br>diagnosis | CT, MRI, CR, US,<br>XA,SC | CT, MRI | The images displayed by<br>subject device has extended to<br>CR, US, XA, and SC. It raises<br>no new issues of safety or<br>effectiveness |
| Modality images for<br>reference.<br>Mammography<br>images are not for<br>diagnostic use. | CT, MRI, CR, Full-Field<br>Digital<br>Mammography, US,<br>XA,SC | CT, MRI ,US, CR,XA,SC | The modalities connected<br>with subject device has<br>extended to Full-Field<br>Mammography. It raises no<br>new issues of safety or<br>effectiveness |
| Mammography<br>hanging protocol | Yes | No | Subject device has added a<br>new image process function<br>which raise no new issues of<br>safety or effectiveness |
| Communication log<br>file | Yes | No | Subject device has added a<br>new image process function<br>which raise no new issues of<br>safety or effectiveness |
| Auto delete old<br>images (FIFO) | Yes | No | Subject device has added a<br>new image process function<br>which raise no new issues of<br>safety or effectiveness |
| Overlay labels | Yes | No | Subject device has added a<br>new image process function<br>which raise no new issues of<br>safety or effectiveness |
| Remote Handheld<br>Viewing | Yes | Yes | No difference |
| Operating Platform | Apple® iOS | Apple® iOS | No difference |
| Hardware<br>Requirements | Apple® iPad Pro | Apple® iPad/iPhone | The hardware to be installed is<br>different. It raises no new<br>issues of safety or<br>effectiveness |
| Screen technology | 12.9"/11"TFT Color<br>LCD Panel | 9.7"TFT Color LCD Panel<br>(Apple® iPad)<br>3.5"TFT Color LCD Panel<br>(Apple® iPhone) | The screen size of iPad Pro is<br>larger than those of Apple®<br>iPad/iPhone. It raises no new<br>issues of safety or<br>effectiveness |
| Screen resolution in<br>pixels/square inch | 264 ppi | 264 ppi (Apple® iPad)<br>326 ppi (Apple® iPhone) | The screen of iPad Pro has the<br>same resolution as that of<br>Apple® iPad, but is different<br>that of iPhone. It raises no new<br>issues of safety or<br>effectiveness |
Device Comparison between subject and predicate
{6}------------------------------------------------
{7}------------------------------------------------
# 5.7 Performance data from non-clinical Testing
The software verification testing and validation testing based on IEC 62304 workflows have been performed by designed engineer or professional personnel. These testing include unit, integration and system tests. The test results of software verification and validation had all met and passed the acceptance criteria referred to medical image software quality request.
The non-clinical performance of display had been conducted according to the description and requirements described in the AAPM Assessment of Display Performance for Medical Imaging Devices (2005) document by a third party to ensure high quality laboratory results. All tests had passed successfully.
# 5.8 Performance data from clinical Testing
Given the differences from the predicate device iDO Viewer (K140399), no clinical testing of the UDE software is necessary for this submission. The subject software is based on the cleared predicate. Moreover, bench testing results are showing adequate performance of the iPad Pro display used in combination with the UDE software.
# 5.9 Conclusion
Based on all above evidence, UDE described in this 510(K) is, in our opinion, substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.