← Product Code [GZB](/productcode/GZB) · K202274

# Nalu Neurostimulation Kit (lnte:gmted 40 em: Single 8/Dual 8), Nalu Neurostimulation Kit (Ported, 2 em: Single 8/Dual8), Dual 8 Ported Nalu Implantable Pulse Generator with 40 ern Kil, 40 cm/ 60 cm/60 em TrialJExtension Lead Kits, Patient Kits and miscellaneous replacement kits (K202274)

_Boston Scientific Neuromodulation · GZB · Nov 9, 2020 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202274

## Device Facts

- **Applicant:** Boston Scientific Neuromodulation
- **Product Code:** [GZB](/productcode/GZB.md)
- **Decision Date:** Nov 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5880
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Nalu Neurostimulation system is intended for the stimulation of the spinal cord for treatment of chronic, intractable pain.

## Device Story

Nalu Neurostimulation System provides spinal cord stimulation (SCS) for chronic, intractable pain. System components: implantable pulse generator (IPG), leads with eight cylindrical electrodes, lead anchor, and external Therapy Disc (or Trial Therapy Disc). IPG implanted in epidural space; delivers pulsed electrical current to spinal cord nerves to inhibit pain signal transmission. Therapy Disc houses battery and RF electronics; controlled via remote programmer. Physician performs implantation using standard SCS surgical tools. Patient uses external components to manage therapy levels and profiles. Device is MR Conditional for full-body scans. System provides therapeutic relief by modulating spinal cord nerve activity.

## Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing, including design verification/validation, sterilization validation, biocompatibility, and human factors/usability testing. Updated MR Conditional labeling supported by non-clinical testing per ISO/TS 10974, ISO 14708, and ASTM standards (F2052-15, F2213-17, F2129-2013).

## Technological Characteristics

Implantable neurostimulator (IPG) and leads (platinum-iridium 90:10). RF wireless power transmission. Components: IPG (27.7mm x 9.3mm x 4.2mm), leads (40/60cm), Therapy Disc, remote programmer. Connectivity: RF link between external disc and implant. Standards: ISO 14708, ISO 14971, ISO 13485:2016, ISO/TS 10974, ASTM F2052, F2213, F2129.

## Regulatory Identification

An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.

## Predicate Devices

- Nalu Neurostimulation System for Spinal Cord Stimulation ([K183047](/device/K183047.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 9, 2020

Nalu Medical, Inc. Pauline Lieu, Ph.D. Regulatory Affairs Consultant 2320 Faraday Ave. Suite 100 Carlsbad, California 92009

Re: K202274

Trade/Device Name: Nalu Neurostimulation System Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZB Dated: August 10, 2020 Received: August 11, 2020

Dear Dr. Lieu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K202274

#### Device Name

Nalu Neurostimulation System for SCS

#### Indications for Use (Describe)

For Spinal Cord Stimulation

This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral pain. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### K202274

#### 1. Submission Sponsor

Nalu Medical, Incorporated 2320 Faraday Ave., Suite 100 Carlsbad, CA 92009 Phone: (760) 448-2360 Fax: (760) 448-2377 Contact Pauline Lieu, Regulatory Affairs

Date Prepared: Aug. 10th, 2020

### 2. Device Names and Classification

| Predicate Device          |                                                                     |
|---------------------------|---------------------------------------------------------------------|
| Proprietary Name          | Nalu Neurostimulation System                                        |
| Common Names              | Stimulator, spinal-cord implanted (pain relief);                    |
| Class                     | II                                                                  |
| Classification Regulation | 21 CFR 882.5880; Stimulator, Implanted Spinal-cord (Pain<br>Relief) |
| Product Code              | GZB                                                                 |
| Review Panel              | Division of Neurological and Physical Medicine Devices              |

## 3. Predicate Device:

Nalu Neurostimulation System for Spinal Cord Stimulation (K183047)

### 4. Device Description

The Nalu Neurostimulation system has been cleared by the FDA (K183047) for spinal cord stimulation to provide therapeutic relief for chronic, intractable pain of the trunk and/or

{4}------------------------------------------------

limbs including unilateral or bilateral pain. The Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on nerves in the spinal cord to inhibit the transmission of pain signals to the brain. The Nalu System is implanted only following a successful trial period using the Nalu Neurostimulation trial system.

The Nalu Neurostimulation system is consisted of five components. The implantable pulse generator (IPG) provides electrical stimulation pulses that are transmitted through the leads, through the dura, to the desired location of the spinal cord site. The leads are implantable and designed to deliver electrical pulses to the spinal cord in the epidural space via an array of eight cylindrical electrodes at the distal end. The leads may be secured in place with the Nalu Lead Anchor. The Trial Therapy Disc or the Therapy Disc houses the battery and electronics for RF power and controls the IPG for therapy delivery via the remote programmer. Implantation of the Nalu IPG and lead components for Spinal Cord Stimulation (SCS) is performed via standard SCS surgical tools and techniques, as described in (K183047).

The Nalu Neurostimulation System has been previously cleared by the FDA (K183047) with the magnetic resonance imaging (MRI) Conditional Labeling for the leads, anchor and implantable pulse generator which can be scanned safely with the local RF coils, including head, foot/ankle, knee, or wrist, as stated in the instructions for use. In this submission, Nalu performed MRI testing on the standard horizontal MR bore system to support the safety of the RF body coil. Nalu proposes an update to the MR Conditional Labeling with the full body scan as indicated in the proposed labeling update.

#### 5. Indications for Use

#### Spinal Cord Stimulation (SCS)

This system is indicated as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach for chronic, intractable pain of the trunk and/or limbs, including unilateral or bilateral pain.

The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.

{5}------------------------------------------------

# 6. Comparison with the Predicate Device:

| Device                                | Nalu Neurostimulation<br>System (Predicate<br>Device: K183047)                                                                                                                                                                                                                                                                                                                                                                                                 | Nalu Neurostimulation<br>System (Subject<br>Device)                                                                                                                                                                                                                                                                                                                                                                                                            | Analysis of<br>Technological<br>Differences |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Trade Name                            | Nalu Neurostimulation<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                | Nalu Neurostimulation<br>System                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                        |
| Manufacturer                          | Nalu Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Nalu Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                        |
| Intended Use                          | The Nalu<br>Neurostimulation<br>system is intended for<br>the stimulation of the<br>spinal cord for<br>treatment of chronic,<br>intractable pain.                                                                                                                                                                                                                                                                                                              | The Nalu<br>Neurostimulation<br>system is intended for<br>the stimulation of the<br>spinal cord for<br>treatment of chronic,<br>intractable pain.                                                                                                                                                                                                                                                                                                              | Same                                        |
| Indications for<br>Use                | This system is indicated<br>as the sole mitigating<br>agent, or as an adjunct<br>to other modes of<br>therapy used in a<br>multidisciplinary<br>approach for chronic,<br>intractable pain of the<br>trunk and/or limbs,<br>including unilateral or<br>bilateral pain.<br>The trial devices are<br>solely used for trial<br>stimulation (no longer<br>than 30 days) to<br>determine efficacy<br>before recommendation<br>for a permanent (long<br>term) device. | This system is indicated<br>as the sole mitigating<br>agent, or as an adjunct<br>to other modes of<br>therapy used in a<br>multidisciplinary<br>approach for chronic,<br>intractable pain of the<br>trunk and/or limbs,<br>including unilateral or<br>bilateral pain.<br>The trial devices are<br>solely used for trial<br>stimulation (no longer<br>than 30 days) to<br>determine efficacy<br>before recommendation<br>for a permanent (long<br>term) device. | Same                                        |
| Clinical<br>application               | Treatment of chronic,<br>intractable pain of the<br>trunk and/or limbs,<br>including unilateral or<br>bilateral pain                                                                                                                                                                                                                                                                                                                                           | Treatment of chronic,<br>intractable pain of the<br>trunk and/or limbs,<br>including unilateral or<br>bilateral pain                                                                                                                                                                                                                                                                                                                                           | Same                                        |
| Prescription<br>Use                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                        |
| Intended User                         | Physician, Layperson                                                                                                                                                                                                                                                                                                                                                                                                                                           | Physician, Layperson                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                        |
| Principle of<br>Operation             | Stimulation of the spinal<br>cord to provide<br>therapeutic relief for<br>chronic, intractable pain<br>of the trunk and/or<br>limbs including<br>unilateral or bilateral<br>pain                                                                                                                                                                                                                                                                               | Stimulation of the spinal<br>cord to provide<br>therapeutic relief for<br>chronic, intractable pain<br>of the trunk and/or<br>limbs including<br>unilateral or bilateral<br>pain                                                                                                                                                                                                                                                                               | Same                                        |
| Mode of<br>Action                     | RF wireless transmission<br>of energy to deliver<br>stimulation at stimulator<br>electrodes                                                                                                                                                                                                                                                                                                                                                                    | RF wireless transmission<br>of energy to deliver<br>stimulation at stimulator<br>electrodes                                                                                                                                                                                                                                                                                                                                                                    | Same                                        |
| Implant<br>Neurostimulat<br>or        | Nalu IPG 27.7 mm x 9.3<br>mm x 4.2 mm                                                                                                                                                                                                                                                                                                                                                                                                                          | Nalu IPG 27.7 mm x 9.3<br>mm x 4.2 mm                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                                        |
| Lead                                  | 40 cm, 60 cm; Platinum-<br>iridium 90:10;<br>Multilumen tube                                                                                                                                                                                                                                                                                                                                                                                                   | 40 cm, 60 cm; Platinum-<br>iridium 90:10;<br>Multilumen tube                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                        |
| Externally<br>worn devices            | Trial Therapy Disc and<br>Therapy Disc                                                                                                                                                                                                                                                                                                                                                                                                                         | Trial Therapy Disc and<br>Therapy Disc                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                        |
| Electronics                           | A printed circuit board<br>(PCB) that generates RF<br>power with embedded<br>waveform parameter<br>settings and buttons for<br>changing parameter<br>settings                                                                                                                                                                                                                                                                                                  | A printed circuit board<br>(PCB) that generates RF<br>power with embedded<br>waveform parameter<br>settings and buttons for<br>changing parameter<br>settings                                                                                                                                                                                                                                                                                                  | Same                                        |
| Clinician<br>Programmer               | Software to<br>communicate to Trial<br>Therapy or Therapy Disc                                                                                                                                                                                                                                                                                                                                                                                                 | Software to<br>communicate to Trial<br>Therapy or Therapy Disc                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                        |
| Patient<br>Remote<br>Control          | Software to pair with<br>Trial Therapy or Therapy<br>Disc                                                                                                                                                                                                                                                                                                                                                                                                      | Software to pair with<br>Trial Therapy or Therapy<br>Disc                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                        |
| Human<br>Factors                      | Integrated controls and<br>indicators that allows<br>the user to turn the<br>device on/off, increase<br>or decrease therapy<br>levels, select from<br>configured therapy<br>profiles and monitor<br>device status                                                                                                                                                                                                                                              | Integrated controls and<br>indicators that allows<br>the user to turn the<br>device on/off, increase<br>or decrease therapy<br>levels, select from<br>configured therapy<br>                                                                                                                                                                                                                                                                                   | Same                                        |
| Externally<br>Contacting<br>Materials | Clip adhesives applied to<br>skin<br>Biocompatible PC ABS<br>housing                                                                                                                                                                                                                                                                                                                                                                                           | Clip adhesives applied to<br>skin<br>Biocompatible PC ABS<br>housing                                                                                                                                                                                                                                                                                                                                                                                           | Same                                        |

Table 1. Nalu Neurostimulation System for Spinal Cord Stimulation (SCS)

{6}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{7}------------------------------------------------

|          | Textile material of belt                            | Textile material of belt                              |                                                                                             |
|----------|-----------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Labeling | MR Conditional Labeling<br>for Head and Extremities | MR Conditional Labeling<br>for Head and Extremities   | Same                                                                                        |
|          | Full Body<br>Do not use RF transmit<br>body coil    | Full body<br>MR Conditional Labeling<br>for Full Body | Different but<br>does not raise<br>different<br>questions of<br>safety and<br>effectiveness |

### 7. Technological Characteristics

All of the physical and therapeutic attributes for the proposed Nalu Neurostimulation System and the predicate device (K183047) share the same technological characteristics and has no differences that would impact safety or effectiveness.

#### 8. Summary of nonclinical performance testing

Nalu Medical performed a range of testing to gather data supporting the safety and performance of the Nalu Neurostimulation System prior to use. Nalu follows the Design Controls section of 21 CFR 820.30, ISO 14971, and ISO 13485:2016. These procedures ensure that all designs are appropriately evaluated and tested. The system is designed and tested to ensure that it meets all applicable standards and guidance documents. Bench testing includes design verification and validation, sterilization validation, and biocompatibility testing. Human factors and usability testing were performed on the device. Validation and performance testing demonstrate that the device meets user needs as reflected in the functional specification. The subject device of this 510(k) has the same technological and performance criteria which have not changed from the predicate device. Therefore, test results from the predicate device (K183047) except for the updated Magnetic Resonance testing remain applicable to the subject device of this 510(k).

#### Table 2. Applicable Standards and Guidance Documents

The testing for the labeling changes proposed for the Nalu Neurostimulation System includes the following test standards and guidance:

| Standard Number | Title                                                                                                                                    |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------|
| ISO/TS 10974    | Assessment of the safety of magnetic resonance imaging for<br>patients with an active implantable medical device                         |
| ISO 14708       | Implant for surgery – Active implantable medical devices – Part<br>1 and Part 3, General requirements and Implantable<br>neurostimulator |

{8}------------------------------------------------

| ASTM F2052-15   | Standard Test Method for Measurement of Magnetically<br>Induced Displacement Force on Medical Devices in the<br>Magnetic Resonance Environment |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM F2213-17   | Standard Test Method for Measurement of Magnetically<br>Induced Torque on Medical Devices in the Magnetic Resonance<br>Environment             |
| ASTM F2129-2013 | Standard Test Method for Evaluation of MR Image Artifacts<br>from Passive Implants                                                             |

### 9. Clinical performance data

Nalu Medical determined that bench and non-clinical testing are sufficient to demonstrate that the Nalu Neurostimulation System is safe and effective as the predicate device.

## 10. Conclusions

The subject device of this 510(k) is substantially equivalent to the predicate devices as they are identical with regard to indications for use, performance and the technological characteristics. Nalu performed testing to support the full body MR scan and the proposed changes in the instructions for use to ensure the safety and effectiveness of the device.

---

**Source:** [https://fda-staging.innolitics.com/device/K202274](https://fda-staging.innolitics.com/device/K202274)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).