# PRE-MILLED Abutment (K202269)

_Mit Global Co., Ltd. · NHA · Nov 25, 2020 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202269

## Device Facts

- **Applicant:** Mit Global Co., Ltd.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Nov 25, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The PRE-MILLED Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. It is compatible with the following systems: Zimmer 3.1mmD Dental Implant System (K142082) Screw Vent® and Tapered Screw Vent® (K013227) All digitally designed abutments for use with the PRE-MILLED Abutments must be designed and milled through the 3Shape CAD/CAM System and are intended to be sent to a validated milling center for manufacture.

## Device Story

PRE-MILLED Abutment is a titanium alloy (Ti-6Al-4V ELI) component used to fabricate patient-specific dental abutments. It features a pre-manufactured connection interface for endosseous dental implants. Clinicians use CAD/CAM systems (specifically 3Shape) to design the patient-specific abutment based on intra-oral anatomy; the design is then sent to a validated milling center for final manufacturing. The device is placed over the implant shoulder and secured with a screw. It functions as a support for dental prostheses, restoring chewing function and aesthetics. Used in dental clinics; operated by dental professionals. Benefits include customized fit for individual patient anatomy while maintaining mechanical integrity of the implant-abutment connection.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including fatigue testing per ISO 14801:2016, biocompatibility testing per ISO 10993, and sterilization validation per ISO 17665/ANSI/AAMI ST79. Dimensional analysis and reverse engineering confirmed compatibility with reference implant systems.

## Technological Characteristics

Material: Titanium grade Ti-6Al-4V ELI (ASTM F-136). Connection: Internal hex/conical. Dimensions: Various diameters (2.9mm to 5.7mm platform). Manufacturing: CAD/CAM (3Shape system). Sterilization: Non-sterile (end-user steam sterilization).

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Medentika CAD/CAM Abutment ([K150203](/device/K150203.md))

## Reference Devices

- Zimmer 3.1mmD Dental Implant System ([K142082](/device/K142082.md))
- Screw Vent® and Tapered Screw Vent® ([K013227](/device/K013227.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

November 25, 2020

MIT Global Co., Ltd. Lim Won Regulatory Affairs Manager 474 Dunchon-daero, Jungwon-gu Seongnam-si, Gyeonggido 13229 SOUTH KOREA

Re: K202269

Trade/Device Name: PRE-MILLED Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 15, 2020 Received: October 26, 2020

## Dear Lim Won:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202269

Device Name PRE-MILLED Abutment

#### Indications for Use (Describe)

The PRE-MILLED Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

It is compatible with the following systems:

- · Zimmer 3.1mmD Dental Implant System (K142082)
- · Screw Vent® and Tapered Screw Vent® (K013227)

All digitally designed abutments for use with the PRE-MILLED Abutments must be designed and milled through the 3Shape CAD/CAM System and are intended to be sent to a validated milling center for manufacture.

The PRE-MILLED Abutment is compatible with the following devices:

Zimmer 3.1mmD Dental Implant System (K142082) Implant Body Diameter 3.1/ Implant Platform Diameter 2.9 / Internal Hex

Screw Vent® and Tapered Screw Vent® (K013227) Implant Body Diameter 3.7/ Implant Platform Diameter 3.5 / Internal Hex Implant Body Diameter 4.1/ Implant Platform Diameter 3.5 / Internal Hex Implant Body Diameter 4.7/ Implant Platform Diameter 4.5 / Internal Hex Implant Body Diameter 6.0/ Implant Platform Diameter 5.7 / Internal Hex

Type of Use (Select one or both, as applicable)

| <input type="checkbox"/> For research and/or ID | <input type="checkbox"/> For Quality Control or Quality Assurance |
|-------------------------------------------------|-------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------|

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MIT, which stands for Medical Implant Technology. The letters "MIT" are in red and are slanted to the right. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are written in a smaller, gray font.

## 510(k) Summary

## K202269

#### Applicant

| Name:    | MIT Global Co., Ltd.                                                            |
|----------|---------------------------------------------------------------------------------|
| Address: | 474 Dunchon-daero #501 Jungwong-gu, eongnam-si<br>Gyeonggido, 13229 South Korea |
| Phone:   | +82 31 745 2838                                                                 |
| Contact: | Lim, Rokwon                                                                     |
| Email:   | rokwonlim@mit-global.com                                                        |

11/20/2020

# Date Prepared:

## Subject Device

| Trade Name:         | PRE-MILLED Abutment                   |
|---------------------|---------------------------------------|
| Common Name:        | Endosseous dental implant abutment    |
| Classification Name | Abutment, Implant, Dental, Endosseous |
| Product Code:       | NHA                                   |
| Panel:              | Dental                                |
| Regulation Number:  | 21 CFR 872.3630                       |
| Device Class:       | Class II                              |

## Primary Predicate

| Trade Name:         | Medentika CAD/CAM Abutment (K150203)  |
|---------------------|---------------------------------------|
| Common Name:        | Endosseous dental implant abutment    |
| Classification Name | Abutment, Implant, Dental, Endosseous |
| Product Code:       | NHA                                   |
| Panel:              | Dental                                |
| Regulation Number:  | 21 CFR 872.3630                       |
| Device Class:       | Class II                              |

## Reference Device

Trade Name:

Common Name: Classification Name Product Code: Panel: Regulation Number: Device Class:

Zimmer 3.1mmD Dental Implant System (K142082) Endosseous dental implant abutment Abutment, Implant, Dental, Endosseous NHA Dental 21 CFR 872.3630 Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MIT Medical Implant Technology. The letters "MIT" are in large, red, bold, sans-serif font. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are in a smaller, gray, sans-serif font. The logo is simple and modern.

# Reference Device

Trade Name:

Common Name: Classification Name Product Code: Panel: Regulation Number: Device Class:

Screw Vent® and Tapered Screw Vent® (K013227) Endosseous dental implant abutment Abutment, Implant, Dental, Endosseous NHA Dental 21 CFR 872.3630 Class II

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MIT Medical Implant Technology. The letters "MIT" are in large, bold, red font and are slanted to the right. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are in a smaller, gray font.

## Device Description

The PRE-MILLED Abutment is used in fabricating a patient-specific abutment in titanium alloy. It has a pre-manufactured connection interface that fits directly to an endosseous dental implant. It is made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F-136).

The abutment is placed over the implant shoulder and mounted into the implant with the provided screw. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system.

Mechanical resistance of the implant-abutment connection is essential to ensure the correct longterm functional performance of the complete dental restoration. Dimensional compatibility and mechanical performance of abutments and screws together with the underlying implant are of primary importance. These concepts are the basis upon which the system design characteristics and functional performance are established.

The PRE-MILLED is a device that can only be sold, distributed, or used upon the order of an authorized healthcare provider, generally referred to as prescription (Rx) devices.

The proposed abutments are available in internal connection, and are compatible with Zimmer 3.1mmD Dental Implant System (K142082) and Screw Vent® and Tapered Screw Vent® (K013227).

| No. | Implant System        | Diameters<br>(Ø) | Implant Platform (mm) | Type of Implant-<br>Abutment Connection |
|-----|-----------------------|------------------|-----------------------|-----------------------------------------|
| 1   | Zimmer 3.1mmD Implant | 3.1              | 2.9                   | Internal Hex<br>Conical Connection      |
| 2   | Zimmer SV/TSV         | 3.7              | 3.5                   |                                         |
| 3   | Zimmer SV/TSV         | 4.1              | 3.5                   |                                         |
| 4   | Zimmer SV/TSV         | 4.7              | 4.5                   |                                         |
| 5   | Zimmer SV/TSV         | 6.0              | 5.7                   |                                         |

The PRE-MILLED Abutments are compatible with the following OEM devices:

#### Indication for Use

The PRE-MILLED Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

It is compatible with the following systems:

- · Zimmer 3.1mmD Dental Implant System (K142082)
- · Screw Vent® and Tapered Screw Vent® (K013227)

All digitally designed abutments for use with the PRE-MILLED Abutments must be designed and milled through the 3Shape CAD/CAM System and are intended to be sent to a validated milling center for manufacture.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for MIT Medical Implant Technology. The logo is red and consists of the letters "MIT" in a bold, stylized font. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are written in a smaller, gray font. The logo is simple and modern, and the use of red and gray gives it a professional look.

The PRE-MILLED Abutment is compatible with the following devices:

Zimmer 3.1mmD Dental Implant System (K142082) Implant Body Diameter 3.1/ Implant Platform Diameter 2.9 / Internal Hex

Screw Vent® and Tapered Screw Vent® (K013227)

Implant Body Diameter 3.7/ Implant Platform Diameter 3.5 / Internal Hex Implant Body Diameter 4.1/ Implant Platform Diameter 3.5 / Internal Hex Implant Body Diameter 4.7/ Implant Platform Diameter 4.5 / Internal Hex Implant Body Diameter 6.0/ Implant Platform Diameter 5.7 / Internal Hex

#### Summary of Technological Characteristics

The subject device, the primary predicate have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device, the primary predicate encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the primary predicate listed above.

#### Non-clinical Testing

The subject device was tested to evaluate its substantial equivalence according to the following standards.

- · Fatigue Test according to ISO 14801:2016
- · End User Steam Sterilization Test according to ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010.
- · Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

Non-clinical test data was used to evaluate the proposed device's substantial equivalence compared to the primary predicate. The results of the above tests have met the criteria of the standard and demonstrated the substantial equivalence with the primary predicate.

Non-clinical testing was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", and it consisted of testing finished assembled implant/abutment systems of the worstcase scenario through fatigue testing.

Dimensional analysis and reverse engineering of the implant-to-abutment connection platform were performed, including an assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible implant body as well as the OEM implant body, the OEM implant abutment, and the OEM abutment screw. The testing aided implant to abutment compatibility and has established substantial equivalency of the proposed device with reference devices.

Clinical testing was not necessary to establish substantial equivalency of the device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Medical Implant Technology (MIT). The letters "MIT" are in large, red, bold, and italicized font. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are written in a smaller, gray, sans-serif font.

# Primary Predicate / Reference devices:

The subject device is substantially equivalent to the following primary predicate and reference devices:

- Primary Predicate
	- Medentika CAD/CAM Abutment (K150203) -
- Reference devices
	- Zimmer 3.1mmD Dental Implant System (K142082) -
	- Screw Vent® and Tapered Screw Vent® (K013227) -

## Comparison between Primary predicate

| Feature                           | Proposed Device<br>PRE-MILLED Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary predicate<br>Medentika CAD/CAM<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SE discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant                         | MIT Global Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Medentika GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Trade Name                        | PRE-MILLED Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Medentika CAD/CAM Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 510(K) No.                        | K202269                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K150203                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification<br>Name            | Endosseous Dental Implant Abutments<br>(872.3630)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Endosseous Dental Implant<br>Abutments (872.3630)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                      | NHA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | NHA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Screw and<br>Abutment<br>Material | Ti-6Al-4V ELI<br>(ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Ti-6Al-4V ELI<br>(ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications<br>For Use            | The PRE-MILLED Abutments are<br>intended for use with dental implants as<br>a support for single or multiple tooth<br>prostheses in the maxilla or mandible of<br>a partially or fully edentulous patient.<br>It is compatible with the following<br>systems:<br>• Zimmer 3.1mmD Dental Implant<br>System (K142082)<br>• Screw Vent® and Tapered Screw<br>Vent® (K013227)<br>All digitally designed abutments for use<br>with the PRE-MILLED Abutments must<br>be designed and milled through the<br>3Shape CAD/CAM System and are<br>intended to be sent to a validated milling<br>center for manufacture.<br>The PRE-MILLED Abutment is<br>compatible with the following devices: | (Medentika TiBase CAD/CAM<br>Abutments not applicable to this<br>submission)<br>Medentika Preface CAD/CAM<br>Abutments are intended for use with<br>dental implants as a support for single<br>or multiple tooth prostheses in the<br>maxilla or mandible of a partially or<br>fully edentulous patient.<br><br>Image: Table of Implant System Compatibility<br><br>Medentika Preface is intended for use<br>with the Straumann CARES System.<br>All digitally designed abutments for<br>use with Medentika CAD/CAM<br>Abutments are intended to be<br>manufactured at a Straumann CARES<br>validated milling center. | The subject device is substantially<br>equivalent in indications and design<br>principles to the primary predicate<br>device listed above. Provided tables<br>are comparing the Indications for<br>Use Statements and the<br>technological characteristics of the<br>subject device, the primary<br>predicate device. The Indications<br>for Use Statement (IFUS) for the<br>subject device is substantially<br>equivalent in intended use to the<br>primary predicate device. All are<br>intended for use with endosseous<br>dental implants in the maxilla and<br>mandible to provide functional and<br>esthetic rehabilitation of the<br>edentulous maxilla and mandible.<br>Slight differences in the language of<br>the subject device and primary<br>predicate is Indications for Use<br>statements do not affect the<br>intended use as an endosseous<br>dental implant abutment for support<br>of a prosthesis to restore chewing<br>function. The minor differences<br>between the subject device and the<br>primary predicate device are related |
| Feature                           | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Primary predicate<br>Medentika CAD/CAM<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SE discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                   | PRE-MILLED Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                   | Zimmer 3.1mmD Dental Implant System<br>(K142082) Implant Body Diameter 3.1/<br>Implant Platform Diameter 2.9 / Internal<br>Hex<br>Screw Vent® and Tapered Screw Vent®<br>(K013227)<br>Implant Body Diameter 3.7/ Implant<br>Platform Diameter 3.5 / Internal Hex<br>Implant Body Diameter 4.1/ Implant<br>Platform Diameter 3.5 / Internal Hex<br>Implant Body Diameter 4.7/ Implant<br>Platform Diameter 4.5 / Internal Hex<br>Implant Body Diameter 6.0/ Implant<br>Platform Diameter 5.7 / Internal Hex                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | to the compatible OEM implant<br>lines and the implant platform<br>diameter.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Platform<br>Diameters             | 2.9/3.5/4.5/5.7mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 3.0 - 7.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The difference between the<br>two products in the platform<br>diameter are the same except<br>for the minimum diameter.<br>The minimum diameter of<br>the product compatible with<br>the subject device is smaller<br>than the primary predicate<br>device (K150203). Since the<br>platform diameter of the<br>subject device is included in<br>the range of primary<br>predicate's diameter,<br>therefore, it doesn't impact<br>the product's substantial<br>equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CAD Design<br>Limits              | Maximum Angulation : 0~25°<br>Maximum Cuff Height : 0.5~5mm<br>Minimum Diameter : Ø 4.0~ Ø 8.0mm<br>Minimum Wall Thickness : 0.4mm<br>Minimum Post Height : 4mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Maximum Angulation : 0~30°<br>Maximum Cuff Height : Not identified<br>Minimum Diameter : Ø3.0~ Ø7.0mm<br>Minimum Wall Thickness : Not identified<br>Minimum Post Height : Not identified                                                                                                                                                                                                                                                                                                                                                                                                                              | The minor difference between the<br>two products in the design<br>parameters are the minimum<br>maximum diameter and angulation.<br>The minimum diameter of the<br>product compatible with the subject<br>device is larger than the primary<br>predicate device. The diameter of<br>the primary predicate device is<br>Ø3.0mm~ Ø7.0mm, while the<br>subject device can be designed from<br>Ø 4.0mm up to Ø8.0mm. The<br>maximum angle of the product<br>compatible with the subject device<br>is smaller than the primary<br>predicate device. The angle of the<br>primary predicate device is 30°,<br>while the subject device can be<br>designed up to 25°.                                                                                                                                                                                                                                                                                                                                                                                              |
| Feature                           | Proposed Device<br>PRE-MILLED Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary predicate<br>Medentika CAD/CAM<br>Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | SE discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Surface<br>Treatment              | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterile                           | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for "MIT MEDICAL IMPLANT TECHNOLOGY". The logo is red and the text below the logo is gray. The logo is a stylized version of the letters "MIT".

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MIT Medical Implant Technology. The logo is red and features the letters "MIT" in a bold, slanted font. Below the letters, the words "MEDICAL IMPLANT TECHNOLOGY" are written in a smaller, gray font. The logo is simple and modern, and the use of red and gray gives it a professional look.

## Substantial Equivalence Discussion

PRE-MILLED Abutment incorporates the same material, indications for use, dimension, design, abutment seat, screw seat, anatomical site, connection, type of retention and technological characteristics as the primary predicate.

The Indications for Use of the subject and primary predicate are identical other than the compatible implant bodies. This difference is mitigated by fatigue testing, reverse engineering, dimensional analysis, and identification of reference predicate for compatible implant bodies. Both the predicate and subject devices are intended to be milled into patient-specific abutments using CAD/CAM technology under the manufacturing control of the sponsor.

Any differences in technology characteristics are accompanied by information that demonstrated the device is substantially equivalent as the predicate and do not raise different questions of safety and effectiveness than the predicate.

PRE-MILLED Abutments are compatible with reference devices (K142082 and K013227). Each PRE-MILLED abutment platform has a precision implant/abutment interface corresponding to the implant system predicate for that platform.

#### Conclusion

The PRE-MILLED Abutment constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its primary predicate. Therefore, PRE-MILLED Abutment and its predicate are substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K202269](https://fda-staging.innolitics.com/device/K202269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).