K202258 · El.En Electronic Engineering Spa · GEX · Apr 26, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202258
Device Name
ERISE Laser handpiece
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ERISE Laser Handpiece is indicated for use in soft tissue (skin and cutaneous tissue) such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectas; Actinic chelitis; Keloids; Verrucae; Skin tags; Keratoses; Scar revision (including acne scars). With microbeam end piece it is indicated for Skin resurfacing.
Device Story
ERISE is a 2940 nm Erbium:YAG laser handpiece; connects to DEKA LUXEA console (K192539) for power and control. Components include connector, cord, plastic shell, laser cavity, mechanical shutter, optical guide, and interchangeable tips (2, 4, 9mm spots; 9mm microbeam array). Operated by physicians in clinical settings for dermatological and plastic surgery procedures. Device delivers laser radiation to soft tissue; energy/fluence parameters controlled via console. Clinical benefit includes tissue ablation/resurfacing for aesthetic and therapeutic skin conditions. Healthcare providers use output to perform precise tissue removal or resurfacing; clinical decision-making based on visual assessment of skin condition and desired treatment depth.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical performance data, including electrical safety (IEC 60601-2-22, IEC 60825-1) and electromagnetic compatibility (IEC 60601-1-2) testing, alongside software verification and validation.
Technological Characteristics
2940 nm Er:YAG laser handpiece; 3 J max energy; 95-121 J/cm2 max fluence; 0.3-1.5 ms pulse duration; 6 Hz max repetition rate. Features microlens array for microbeam creation (300-600 μm diameter, 65-70 dots, 12% coverage). Connects to DEKA LUXEA console. Complies with IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-2.
Indications for Use
Indicated for soft tissue (skin/cutaneous) procedures including skin resurfacing, wrinkle treatment, epidermal nevi, telangiectasia, spider veins, actinic cheilitis, keloids, verrucae, skin tags, anal tags, keratoses, and scar revision (including acne scars) in dermatology and plastic surgery patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Quanta System 2940 nm Er:YAG laser handpiece (K173002)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 26, 2021
ELEN. Electronic Engineering Spa Paolo Peruzzi Regulatory Affair Manager Via Baldanzese 17 Calenzano, FI 50041 Italy
Re: K202258
Trade/Device Name: ERISE Laser handpiece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 17, 2021 Received: March 23, 2021
Dear Paolo Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202258
Device Name ERISE Laser Handpiece
Indications for Use (Describe)
ERISE Laser Handpiece is indicated for use in soft tissue (skin and cutaneous tissue) such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectas; Actinic chelitis; Keloids; Verrucae; Skin tags; Keratoses; Scar revision (including acne scars). With microbeam end piece it is indicated for Skin resurfacing.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) Summary
### ERISE Laser Handpiece
### Submitter:
El.En. S.p.A.
Via Baldanzese, 17
50041 Calenzano (FI), Italy
### Contact:
Paolo Peruzzi
Regulatory Affairs Manager & Official Correspondent
Phone: +39.055.8826807
E-mail: p.peruzzi@elen.it
### Date Summary Prepared:
April 15, 2021
### Device Trade Name:
ERISE Laser handpiece
### Common Name:
2940 nm Er:YAG laser handpiece
### Classification:
Class II
### Classification Name:
Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology
{4}------------------------------------------------
# Product Code :
GEX
## Classification Number:
21 CFR 878.4810
## Predicate Device:
- K173002 – Quanta System 2940 nm Er:YAG laser handpiece
## Device Description:
ERISE is an Erbium:YAG laser handpiece delivering laser radiation at 2940nm.
It is intended to be used with DEKA LUXEA platform, cleared by FDA with K192539.
The Erise laser handpiece includes:
- . A connector to connect to the console
- A cord where electrical cables and hydraulic tubes pass in
- A plastic shell
- . A Laser cavity
- . A mechanical shutter
- . An optical guide
- . An ending tip which provides different spot sizes (diameter 2, 4, 9mm), including microbeam output (diameter 9mm)
The 2940 nm Erbium:YAG laser handpiece is controlled by the DEKA LUXEA console which it is connected to, through the connector.
## Indications for Use:
ERISE Laser Handpiece is indicated for use in soft tissue (skin and cutaneous tissue) such as, but not limited to: Dermatology and plastic Surgery: Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins; Actinic chelitis; Keloids; Verrucae; Skin tags; Anal tags; Keratoses; Scar revision (including acne scars).
With microbeam end piece it is indicated for Skin resurfacing.
{5}------------------------------------------------
# Technological Characteristics Comparison:
ERISE laser handpiece is as safe, as effective, and performs as well as the legally marketed predicate device Quanta System 2940 nm Er:YAG laser handpiece (K173002):
| | Proposed 510(k) Device | Predicate Device |
|------------------------------------|------------------------------------------------------------|------------------------------------------------------------|
| | | K173002 |
| Device Trade Name | ERISE laser handpiece | Quanta System 2940 nm Er:YAG laser<br>handpiece |
| Product code | GEX | GEX |
| Laser Wavelength | 2940 nm | 2940 nm |
| MAX Energy | 3 J | 3 J |
| | Up to 95 J/cm2 (non microbeam mode); | Up to 95 J/cm2 (non microbeam mode); |
| MAX Fluence | Up to 121 J/cm2 (with stacking pulses -<br>microbeam mode) | Up to 121 J/cm2 (with stacking pulses -<br>microbeam mode) |
| Handpiece Spot | Ø 2, 4, 9mm | Ø 2, 4, 9mm |
| Sizes | Ø 9mm dots array | Ø 9mm dots array |
| Pulse Duration | 0.3 to 1.5 ms | 0.3 to 1.5 ms |
| Pulse Repetition Rate | Up to 6 Hz | Up to 6 Hz |
| Diameter of the microbeams | 300- 600 μm | 300- 600 μm |
| Creation of microbeams | By microlenses array | By microlenses array |
| Number of dots | 65-70 | 65-70 |
| MAX Energy per Microbeam | 180 mJ (stacking mode) | 180 mJ (stacking mode) |
| Coverage of skin in microbeam mode | 12% | 12% |
## Clinical Performance Data:
None
{6}------------------------------------------------
## Non-Clinical Performance Data:
The following performance data are provided in support of the substantial equivalence determination, for the ERISE laser handpiece used in conjunction with DEKA LUXEA platform.
### Safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the DEKA LUXEA platform.
The system complies with the IEC 60601-2-22, IEC 60825-1 standards for safety and the IEC 60601-1-2 standard for EMC.
## Software Verification and Validation Testing
Software of DEKA LUXEA platform successfully underwent verification and validation testing including the use of ERISE Handpiece.
Software documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
## Conclusions:
Based on the comparison of indications for use and the technological characteristics, and on the outcome of non-clinical performance data provided , we can conclude that ERISE Laser handpiece is as safe, as effective, and performs as well as the legally marketed predicate device.
## Additional Information:
None
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.