The ASAHI Corsair Armet 18 is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet 18 is also intended to assist in the delivery of contract media into the peripheral vasculature. This device should not be used in coronary vasculature or neurovasculature.
Device Story
Microcatheter used in peripheral vasculature to support guidewire placement, facilitate guidewire exchange, and deliver contrast media. Operated by physicians in clinical settings. Device consists of distal tip, shaft tube, protector, and connector. Distal tip and distal shaft feature hydrophilic coating; radiopaque tip aids visualization. Physician advances catheter over guidewire to target site; contrast media injected through lumen. Provides mechanical support for guidewire navigation through peripheral vessels. Benefits include improved guidewire tracking and procedural efficiency during peripheral interventions.
Clinical Evidence
Bench testing only. Performance evaluated via air leakage, burst pressure, corrosion resistance, kink resistance, liquid leakage, radio-detectability, slide durability, tensile strength, torque durability, and torque transmission. All samples met acceptance criteria based on ISO 10555-1 or validated in-house procedures. No clinical data required.
Technological Characteristics
Percutaneous catheter; polyamide-elastomer shaft; hydrophilic coating on distal tip/shaft; radiopaque tip. Dimensions: 600-1350cm length; 0.73mm distal/0.87mm proximal OD. Sterilized via Ethylene Oxide (SAL 10^-6). Complies with ISO 10555-1.
Indications for Use
Indicated for patients requiring guidewire support or exchange and contrast media delivery within the peripheral vasculature. Contraindicated for use in coronary or neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
September 8, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Asahi Intecc Co., Ltd. Cynthia Valenzuela Director, Regulatory Affairs 3002 Dow Avenue, Suite 212 Tustin, California 92780
Re: K202252
Trade/Device Name: ASAHI Corsair Armet 18 Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 31, 2020 Received: August 10, 2020
### Dear Cynthia Valenzuela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medical-devices/ device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-andcontinuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202252
Device Name ASAHI Corsair Armet 18
Indications for Use (Describe)
The ASAHI Corsair Armet 18 is intended to provide support to facement of guidewires in the peripheral vasculature, and can be used to exchange one guide wire for another.
The ASAHI Corsair Armet 18 is also intended to assist in the delivery of contract media into the peripheral vasculature.
This device should not be used in coronary vasculature or neurovasculature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) Summary K202252
In accordance with the Safety Medical Devices Act of 1990 (SMDA), a 510(K) Summary of safety and effectiveness information upon which this substantial equivalence determination is based, is provided in this section.
| DATE PREPARED: | 31JUL2020 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | ASAHI INTECC CO., LTD<br>3-100 Akatsuki-cho, Seto<br>Aichi 489-0071, Japan |
| PRIMARY CONTACT: | Mrs. Cynthia Valenzuela<br>Director, Regulatory Affairs<br>ASAHI INTECC USA, INC.<br>3002 Dow Avenue, Suite 212<br>Tustin, California 92780<br>Phone: (714) 442 0575<br>Fax: (949) 377 3255<br>Email: cynthiav@asahi-intecc-us.com |
| TRADE NAME: | ASAHI Corsair Armet 18 |
| DEVICE CLASSIFICATION: | Class II, 21CFR§870.1250 |
| CLASSIFICATION NAME: | Percutaneous Catheter |
| PRODUCT CODE: | DQY, Catheter, Percutaneous |
| PREDICATE DEVICE(S): | ASAHI Corsair Armet (K161362) |
#### Intended Use/Indications for Use
The ASAHI Corsair Armet 18 is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.
The ASAHI Corsair Armet 18 is also intended to assist in the delivery of contract media into the peripheral vasculature.
This device should not be used in coronary vasculature or neurovasculature.
#### Device Description:
This product consists of a distal tip, a shaft tube, a protector and a connector. The distal tip and distal side of shaft are coated with hydrophilic coating on the outer surface. The tip is radiopaque, and the distal end is clearly distinguished by the tip.
{4}------------------------------------------------
### Comparison with Predicate Device:
A comparison of the ASAHI Corsair Armet 18 Microcatheter and predicate device shows that the technological characteristics of the ASAHI Corsair Armet 18 Microcatheter such as components, design, materials, sterilization method, shelf life and operating principle are identical or similar to the currently marketed predicate device.
The intended use/indications of the Subject Device are a subset of the predicate. There are specific design features of the Subject Device that are similar to the predicate, which has demonstrated equivalence for these similar features.
| Name of Device | ASAHI Corsair Armet | ASAHI Corsair Armet |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) | (K161362) | Subject Device / Current<br>Application |
| Intended Use and<br>Indications | The ASAHI Corsair Armet is<br>intended to provide support to<br>facilitate the placement of guide<br>wires in the peripheral<br>vasculature, and can be used to<br>exchange one guide wire for<br>another.<br>The ASAHI Corsair Armet is also<br>intended to assist in the delivery<br>of contract media into the<br>peripheral vasculature. | The ASAHI Corsair Armet 18 is<br>intended to provide support to<br>facilitate the placement of guide<br>wires in the peripheral<br>vasculature, and can be used to<br>exchange one guide wire for<br>another.<br>The ASAHI Corsair Armet 18 is<br>also intended to assist in the<br>delivery of contract media into<br>the peripheral vasculature. |
| | This device should not be used<br>in coronary vasculature or<br>neurovasculature. | This device should not be used<br>in coronary vasculature or<br>neurovasculature. |
| Target Body Location | Peripheral | |
| Hydrophilic Coating | Yes | |
| Effective Length | 600-1500cm | 600-1350cm |
| Nominal Outer Diameter | Distal: 0.75mm<br>Proximal: 0.83mm | Distal: 0.73mm<br>Proximal: 0.87mm |
| Catheter Shaft Material | Polyamide-elastomer | |
| Distal Tip Length | 1.2mm | 1.50mm |
| Single Use | Yes | |
| Sterilization | Ethylene Oxide to SAL-6 | |
| Shelf life | 3 Years | |
{5}------------------------------------------------
## NON-CLINICAL TESTING / PERFORMANCE DATA;
Non Clinical Laboratory testing performed on the ASAHI Corsair Armet 18 Microcatheter to determine substantial equivalence:
- Air Leakage ●
- Appearance / Dimensions .
- Burst Pressure Under Static Condition ●
- Corrosion Resistance ●
- Kink Resistance .
- Liquid Leakage Under Pressure ●
- Radio-Detectability
- . Slide Durability
- Tensile Strength / Force at Break ●
- Torque Durability .
- Torque Transmission ●
In the in vitro bench tests demonstrated that the ASAHI Corsair Armet 18 Microcatheter met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrates that the device functions as intended, and is substantially equivalent to the predicate device.
All bench testing of the ASAHI Corsair Armet 18 Microcatheter performed on finished sterilized product. The acceptance criteria is based on ISO10555-1. Tests not conducted per ISO10555-1 were completed using in-house validated procedures with internally developed acceptance criteria.
All ASAHI Corsair Armet 18 Microcatheter test samples met the acceptance criteria for each of the tests listed in this submission. There were no deviations from the acceptance criteria. Testing shows that the ASAHI Corsair Armet 18 Microcatheter is equivalent to the predicate ASAHI Corsair Armet.
#### BIOCOMPATIBILITY:
The Biological Safety test of the ASAHI Corsair Armet 18 Microcatheter was not required. as the subject device did not change in material from its predicate device deeming biocompatibility testing unnecessary.
#### CONCLUSION:
The ASAHI Corsair Armet 18 Microcatheter has substantial identical intended use, the same similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended.
Therefore, the ASAHI Corsair Armet 18 Microcatheter is considered, substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.