LYKOS Assisted Reproduction Laser with Dynamic Targeting System (DTS)
K202241 · Hamilton Thorne, Inc. · MRX · Jul 2, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K202241
Device Name
LYKOS Assisted Reproduction Laser with Dynamic Targeting System (DTS)
Applicant
Hamilton Thorne, Inc.
Product Code
MRX · Obstetrics/Gynecology
Decision Date
Jul 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6200
Device Class
Class 2
Indications for Use
The LYKOS Assisted Reproduction Laser with Dynamic Target System (DTS) is intended for in vitro fertilization (IVF) laboratory use in assisted reproduction procedures to ablate a small, tangential hole or thin the zona pellucida of select IVF embryos to facilitate assisted reproduction hatching procedures and to biopsy trophectoderm cells from blastocyst stage embryos for purposes of preimplantation genetic diagnosis or screening.
Device Story
LYKOS with DTS is an assisted reproduction laser system for IVF laboratories; operated by embryologists. System inputs include live video feed from an inverted microscope; outputs include laser ablation of zona pellucida or trophectoderm biopsy. Hardware includes laser control unit, 40x objective, mirror/motor modules, and computer interface. Operation modes: Fixed Direction (manual embryo guidance) or DTS (computer-controlled laser path navigation). Laser is infrared (1460 nm) with concentric red LED (RED-i) for visual targeting. System records images/videos of procedures. Clinical benefits include precise, user-defined ablation paths for embryo hatching and biopsy, facilitating genetic screening.
Clinical Evidence
Bench testing only. Performance validation included laser accuracy/precision for single and multiple-pulse DTS modes, RED-i alignment verification, electrical safety (IEC 61010-1, IEC 60825-1), EMC (IEC 61326-1), software validation (moderate level of concern), cybersecurity, and reprocessing validation.
Technological Characteristics
Class II laser system. Infrared laser (1460 nm); power 3-300 mW; pulse length 0.1-3.0 ms. Components: laser control unit, 40x objective, mirror/motor modules, computer interface. Connectivity: USB/communication cable to PC. Standards: IEC 61010-1, IEC 60825-1, IEC 61326-1. Software: moderate level of concern.
Indications for Use
Indicated for IVF laboratory use in assisted reproduction procedures for zona pellucida ablation or thinning to facilitate hatching, and for trophectoderm biopsy of blastocyst stage embryos for preimplantation genetic diagnosis or screening.
Regulatory Classification
Identification
The assisted reproduction laser system is a device that images, targets, and controls the power and pulse duration of a laser beam used to ablate a small tangential hole in, or to thin, the zona pellucida of an embryo for assisted hatching or other assisted reproduction procedures.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Assisted Reproduction Laser Systems.” See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font below.
July 2, 2021
Hamilton Thorne, Inc. Donald Fournier Director, Regulatory Affairs & Quality Assurance 100 Cummings Center, Suite 465E Beverly, MA 01915
Re: K202241
> Trade/Device Name: LYKOS™ Assisted Reproduction Laser with Dynamic Targeting System (DTS) Regulation Number: 21 CFR§ 884.6200 Regulation Name: Assisted Reproduction Laser System Regulatory Class: II Product Code: MRX Dated: June 3, 2021 Received: June 4, 2021
Dear Donald Fournier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202241
Device Name
LYKOS™ Assisted Reproduction Laser with Dynamic Targeting System (DTS)
Indications for Use (Describe)
The LYKOS Assisted Reproduction Laser with Dynamic Target System (DTS) is intended for in vitro fertilization (IVF) laboratory use in assisted reproduction procedures to ablate a small, tangential hole or thin the zona pellucida of select IVF embryos to facilitate assisted reproduction hatching procedures and to biopsy trophectodern cells from blastocyst stage embryos for purposes of preimplantation genetic diagnosis or screening.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K202241
### LYKOS™ Assisted Reproduction Laser with Dynamic Targeting System (DTS)
| Submitter | Hamilton Thorne, Inc.<br>100 Cummings Center, Suite 465E<br>Beverly, MA 01915<br>Tel: (978) 921-2050<br>Fax: (978) 921-0250 |
|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Donald Fournier<br>Director, Regulatory Affairs & Quality Assurance<br>100 Cummings Center, Suite 465E<br>Beverly MA 01915<br>Tel: (978) 921-2050 Ext. 1726<br>Fax: (978) 921-0250<br>dfournier@hamiltonthorne.com |
| Date Prepared | June 30, 2021 |
| Subject Device | |
| Trade Name:<br>Common Name:<br>Regulation Name:<br>Regulation Number:<br>Product Code:<br>Regulatory Class: | LYKOS™ Assisted Reproduction Laser with Dynamic Targeting System (DTS)<br>Assisted Reproduction Laser System<br>Assisted Reproduction Laser System<br>21 CFR § 884.6200<br>MRX (System, Assisted Reproduction Laser)<br>Class II |
## Predicate Device
| 510(k) Number: | K141434 |
|----------------|------------------------------------|
| Device Name: | Saturn 5 <sup>™</sup> Laser System |
| Manufacturer: | Research Instruments Ltd |
The predicate device has not been subject to a design-related recall.
#### Device Description
The LYKOS with DTS is comprised of hardware, software and firmware. Hardware components include a laser control unit, communication cable, optional foot pedal, microscope mount(s) and a 40x objective. Fully integrated within the objective is the laser, a rotating mirror frame and motors that rotate the mirror and control the direction of the laser beam. The objective can be mounted on most commercially available inverted microscopes. The laser hardware interfaces with a computer and camera, providing a live image of the objective's field of view on the computer monitor. The system also records images and videos of the laser procedures.
{4}------------------------------------------------
The system includes three laser modes: Clinical, Multipulse, and Validation. The Clinical Mode includes laser settings recommended for zona thinning and embryo hatching procedures. The Multipulse Mode allows the use of a series of laser pulses and is intended for trophectoderm biopsy procedures. The Validation Mode is intended to validate the proper operation of the laser including RED-i and laser alignment.
Targeting of the laser beam (i.e., either aiming its direction or plotting its ablation path) may be performed in two modes: fixed direction or DTS. In the fixed direction mode of operation, the direction of the beam remains fixed and the embryo is manually guided to the laser by the user. In the DTS Mode, the embryo remains fixed and the direction of the laser is computer-controlled, navigating a user defined plot as drawn on the computer monitor. The system can be programmed to aim and fire the laser anywhere in the field of view or along a user-defined ablation path.
In both Manual and DTS modes, because the laser operates in the invisible infrared wavelength (i) = 1460 mm), to aim and visually verify the position of the laser beam at all times prior to firing, the laser is concentrically aligned with a visible red LED beam (RED-i). Through an automated mapping and verification system using the X-Y coordinates of the computer monitor, the laser can be aimed and an ablation path plotted and verified by the user prior to each use.
# Indications for Use
The LYKOS Assisted Reproduction Laser with Dynamic Target System (DTS) is intended for in vitro fertilization (IVF) laboratory use in assisted reproduction procedures to ablate a small, tangential hole or thin the zona pellucida of select IVF embryos to facilitate assisted reproduction hatching procedures and to biopsy trophectoderm cells from blastocyst stage embryos for purposes of preimplantation genetic diagnosis or screening.
#### Substantial Equivalence Comparison
A comparison of the indications for use and technological characteristics of the subject and predicate devices are summarized in the table below:
| Parameter | Subject Device | Predicate Device | Comparison |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | LYKOS DTS | SATURN 5 – K141434 | |
| Indications for Use | The LYKOS Assisted<br>Reproduction Laser with<br>Dynamic Target System (DTS) is<br>intended for in vitro fertilization<br>(IVF) laboratory use in assisted<br>reproduction procedures to ablate<br>a small, tangential hole or thin the<br>zona pellucida of select IVF<br>embryos to facilitate assisted<br>reproduction hatching procedures<br>and to biopsy trophectoderm cells<br>from blastocyst stage embryos for<br>purposes of preimplantation<br>genetic diagnosis or screening. | For use in assisted reproduction<br>procedures to ablate or thin the<br>zona pellucida of an oocyte or<br>embryo to facilitate assisted<br>hatching or recovery of cells for<br>pre-implantation genetic<br>diagnosis (blastomeres). The<br>device can also be used on<br>blastocyst stage embryos for<br>biopsy of trophectoderm cells<br>for preimplantation diagnosis<br>procedures, and blastocyst<br>collapse prior to vitrification<br>procedures. | Different: The indications for<br>use statements for the subject<br>and predicate devices are<br>different. Both the subject and<br>predicate devices are for use in<br>ablating or thinning the zona<br>pellucida (ZP) to facilitate<br>assisted hatching of embryos<br>and can also be used for<br>trophectoderm biopsy.<br>However, the predicate device<br>is indicated for additional uses<br>beyond those of the subject<br>device, including<br>ablation/thinning of the ZP of<br>oocytes and blastocyst collapse<br>prior to vitrification, These<br>differences represent a more<br>limited use of the subject<br>device and do not represent a<br>new intended use. Therefore,<br>the indented uses are the same. |
| | | | |
| System Configuration | • Laser control unit<br>• 40X objective<br>• Microscope adapters<br>• Mirror module<br>• Motor module<br>• Laser application software<br>• Computer user-interface<br>• Camera<br>• Foot switch (optional) | • Laser control unit<br>• 40X objective<br>• Microscope adapters<br>• Mirror module<br>• Motor module (Active<br>version only)<br>• Laser application software<br>• Computer user-interface<br>• Camera<br>• Foot switch (optional) | Different: The subject and<br>predicate are comprised of<br>different components. The<br>differences in device<br>components do not raise<br>different questions of safety<br>and effectiveness (S&E). |
| Operating Modes | Fixed Direction – The direction of<br>the laser beam is fixed; user<br>manually guides embryo. | Fixed Mode – The direction of<br>the laser beam is fixed; user<br>manually guides embryo. | Similar |
| | DTS – The direction and path of<br>the laser beam are computer-<br>controlled; embryo is stationary. | Active Mode – The direction<br>and path of the laser beam is<br>computer-controlled; embryo is<br>stationary. | |
| Laser Modes | Clinical Mode: Allows a single<br>ablation of the target (Fixed Mode<br>and DTS Mode) or a series of<br>ablations (DTS Mode) | Single Pulse: Allows a single<br>ablation of the target (Fixed<br>Mode and Active Mode) | Different: The subject and<br>predicate devices have<br>different laser modes. The<br>differences do not raise<br>different questions of S&E. |
| | Multipulse Mode: Allows the use<br>of a series of laser pulses for<br>biopsy procedures (Fixed Mode<br>only) | Biopsy Mode: Allows to ablate<br>a series of holes for biopsy<br>procedures (Active Mode only) | |
| | Validation Mode: For validation<br>of proper operation of the laser<br>(Fixed Mode and DTS Mode) | | |
| Laser Specifications | $λ$ = 1460 nm<br>Power: 3-300 mW<br>Pulse length: 0.1 to 3.0 ms | $λ$ = 1480 nm<br>Power 400 mW<br>Pulse length 5-2000 μs | Different: The subject and<br>predicate devices have<br>differences in laser<br>wavelength, power and pulse<br>lengths. These differences do<br>not raise different questions of<br>S&E. |
| Alignment Indicator | Yes - RED-i | Yes -Pilot laser | Similar |
| Laser Classification<br>(U.S CFR 1040.10) | Class I | Class I | Same |
| Microscope Compatibility | Compatible with various models<br>of inverted microscopes. | Compatible with various models<br>of inverted microscopes. | Same |
| Laser Firing Mechanism | Mouse; foot pedal | Mouse; foot pedal | Same |
| Hole Size Indicator | Yes | Yes | Same |
{5}------------------------------------------------
As noted in the table above, there are differences in the indications for use statements and technological features of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not represent a new intended use and the differences in technological features do not raise different questions of safety and effectiveness.
### Summary of Performance Testing
The following performance data was provided in support of the substantial equivalence determination:
- Laser Performance Testing in accordance with the 2004 FDA guidance document "Assisted Reproduction ● Laser Systems - Class II Special Controls Guidance Document for Industry and FDA Staff." Additional testing included the following:
{6}------------------------------------------------
- Validation of the RED-i target alignment feature -
- -Validation of the accuracy and precision of DTS single-pulse across the field of view
- -Validation of the accuracy and precision of DTS multiple-pulse (linear or curved plots) across the field of view
- Electrical Safety and Electromagnetic Compatibility (EMC) Testing in accordance with: ●
- Electrical Safety Testing: IEC 61010-1:20117; IEC 60825-1:2014 -
- -EMC Testing: IEC 61326-1:2013
- Software and Cybersecurity: ●
- Software documentation in accordance with the 2005 FDA guidance document "Guidance for the -Content of Premarket Submissions for Software Contained in Medical Devices" to support device software with a moderate level of concern.
- Cybersecurity information in accordance with the 2014 FDA guidance document "Content of -Premarket Submissions for Management of Cybersecurity in Medical Devices."
- Reprocessing: Validation testing conducted in accordance with the 2015 FDA guidance document, ● "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
#### Conclusion
The results of the testing described above demonstrate that the substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.