← Product Code [ODC](/productcode/ODC) · K202237

# Locking device (K202237)

_Hangzhou AGS MedTech Co., Ltd. · ODC · Dec 23, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202237

## Device Facts

- **Applicant:** Hangzhou AGS MedTech Co., Ltd.
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Dec 23, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This device is an accessory to be used with endoscopic biliary devices to lock the guidewire(s) in place during ERCP procedure. The device is supplied sterile and intended for single use only.

## Device Story

One-piece integrated system; secures to endoscope accessory channel; features central access port for wire guides/biliary accessories; includes wire guide lock mechanism; used during ERCP procedures; operated by physician/clinician; allows locking of guidewire for continued ductal access during therapy; benefits patient by maintaining guidewire position during complex endoscopic interventions.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization/reactivity). Sterility validated per ISO 11135.

## Technological Characteristics

Materials: ABS, Silica gel. Form factor: One-piece integrated system with central access port and wire guide lock. Connectivity: None. Sterilization: ISO 11135. Energy source: None (mechanical).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Wilson-Cook USW Cap and Wire Lock ([K040137](/device/K040137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

December 23, 2020

Hangzhou AGS MedTech Co., Ltd. Yanping Fu RA Supervisor Building 5, Building 6, No.597 Kangxin Road Yuhang District Hangzhou, Zhejiang 311106 China

Re: K202237

Trade/Device Name: Locking device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: ODC Dated: November 17, 2020 Received: December 21, 2020

Dear Yanping Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202237

Device Name Locking Device

Indications for Use (Describe)

This device is an accessory to be used with endoscopic biliary devices to lock the guidewire(s) in place during ERCP procedure. The device is supplied sterile and intended for single use only.

Type of Use (Select one or both, as applicable)

| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>                                                                                                                                                                                                       |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a logo with a symbol and text. The symbol is on the left and appears to be three curved shapes in blue and green, intertwined to form a circular design. To the right of the symbol, there is text in two lines. The first line reads "AGS MEDTECH" in a sans-serif font, and the second line is in Chinese characters, which translates to "An Jie Si Medical".

## 510(k) Summary

We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.

| 5.1 Submitter                         |                                                                                                                                   |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Submitted by/Owner:                   | Hangzhou AGS MedTech Co., Ltd.<br>Building 5, Building 6, NO.597 Kangxin Road Yuhang<br>District, Hangzhou, Zhejiang 311106 China |
| Establishment<br>Registration Number: | 3010288205                                                                                                                        |
| Registration Status:                  | Active                                                                                                                            |
| Contact Person:                       | Yanping Fu<br>Phone: 0086-15958493282<br>Fax: 0086-0571-87671225<br>Email: fuyp@bioags.com                                        |
| Date Prepared:                        | July 30, 2019                                                                                                                     |

## 5.2 Proposed Device

| Trade Name:             | Locking device              |
|-------------------------|-----------------------------|
| Device Name:            | Locking device              |
| Common Name:            | Locking device              |
| Regulation class:       | Class II                    |
| Regulation Number:      | 876.1500                    |
| Regulation Description: | Endoscope and accessories.  |
| Review Panel:           | Gastroenterology/Urology    |
| Product Code:           | ODC                         |
| Product Code Name:      | Endoscope Channel Accessory |

## 5.3 Predicate Device

| Trade Name:             | Wilson-Cook USW Cap and Wire Lock              |
|-------------------------|------------------------------------------------|
| Device Name:            | Wilson-Cook USW Cap and Wire Lock              |
| Common Name:            | Cap and Wire Lock                              |
| 510(k) Number:          | K040137                                        |
| Regulation class:       | Class II                                       |
| Regulation Number:      | 876.1500                                       |
| Regulation Description: | Endoscope and accessories.                     |
| Review Panel:           | Gastroenterology/Urology                       |
| Product Code:           | OCY                                            |
| Product Code Name:      | Endoscopic Guidewire, Gastroenterology-Urology |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle around a green center. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters below it. The logo appears to be for a medical technology company.

K202237 Page 2 of 3

### 5.4 Device Description

The proposed Locking Device is one-piece integrated system that secures onto the accessary channel of an endoscope. The access port in the center of the cap allows access of wire guides and other biliary accessories. The wire guide lock allows the practitioner to lock the wire guide in place for continued ductal access while proceeding with other ERCP therapies. This device consists of locking device and biopsy cap.

#### 5.5 Indication for use statement

This device is an accessory to be used with endoscopic biliary devices to lock the guidewire(s) in place during ERCP procedure. The device is supplied sterile and intended for single use only.

#### 5.6 Comparison of Technology Characteristics

Our proposed device Locking Device is substantially equivalent to the predicate devices. The differences between the Locking Device and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below.

| Item                                                                                                      | Proposed device                                                                                                                                      | Predicate device | Comparison                                                                                                                                                                                                                                                                                                    |
|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Recommended Wire Guide<br>Diameter inch                                                                   | 0.025inch<br>0.035inch<br>0.038inch                                                                                                                  | Unknown          | Different.<br>We have conducted guide wire<br>fixing force performance for our<br>proposed device, demonstrate that<br>our Locking device is suitable for<br>guide wire with<br>0.025inch/0.035inch/0.038inch<br>diameter. For details, please refer<br>to Appendix 18-1 Bench<br>performance Testing report. |
| Biological<br>Materials or<br>substances in<br>contact with the<br>same human<br>tissue or body<br>fluids | ABS, Silica gel                                                                                                                                      | Unknown          | Different.<br>Biocompatibility tests have been<br>done for the difference. Biological<br>risks are acceptable.                                                                                                                                                                                                |
| Biocompatibility                                                                                          | In Vitro Cytotoxicity<br>Test: ISO 10993-5:<br>2009;<br>Skin sensitization<br>Test: ISO 10993-10:<br>2010;<br>Intracutaneous<br>Reactivity Test: ISO | Unknown          |                                                                                                                                                                                                                                                                                                               |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo with a circular design on the left and text on the right. The circular design features three curved shapes in blue, forming a triangle in the center, which is filled with green. To the right of the logo, the text "AGS MEDTECH" is written in a smaller, sans-serif font, followed by Chinese characters in a larger font size. The overall design appears to represent a medical technology company.

| Item | Proposed device | Predicate device | Comparison |
|------|-----------------|------------------|------------|
|      | 10993-10: 2010; |                  |            |

### 5.7 Applicable Guidance Document

NA

## 5.8 Performance Data

The Locking Device meets all design specifications and medical device standards for biocompatibility (ISO 10993) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.

## 5.9 Clinical Test

No Clinical test is included in this submission.

## 5.10 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS MedTech Co., Ltd has demonstrated that proposed device Locking Device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K202237](https://fda-staging.innolitics.com/device/K202237)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
