The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap or where a gap closure is achieved by flexion of the extremity.
Device Story
NervAlign Nerve Cuff is a sterile, freeze-dried, acellular collagen membrane matrix derived from porcine pericardium. Implanted by surgeons during peripheral nerve repair procedures; device acts as a protective scaffold around damaged nerves. Scaffold becomes infiltrated by patient cells and remodels into native tissue over time. Provides physical protection to nerve site during healing process. Single-use device available in multiple sizes. Benefits include nerve protection and support for natural tissue regeneration. No electronic components or software.
Clinical Evidence
No human clinical data. Evidence consists of extensive non-clinical bench and animal testing. Bench testing: suture retention, thickness, tensile strength, endotoxin levels, and ISO 10993 biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, genotoxicity). Animal studies: 4-week and 13-week rat subcutaneous implantation (biocompatibility/degradation), 6-month rabbit sciatic nerve wrap study (safety/compatibility), and rat transected sciatic nerve model (performance comparison vs. predicate). Results confirmed biocompatibility, appropriate degradation profile, and nerve healing outcomes similar to predicate.
Technological Characteristics
Collagen matrix derived from porcine pericardium. Sterile, freeze-dried, pre-cut flat sheet. Dimensions: 10x10mm, 20x30mm, 30x40mm. Sterilization: Gamma irradiation. Mechanical properties: Suture retention and tensile strength sufficient for nerve repair. Bioresorbable scaffold. No energy source, connectivity, or software.
Indications for Use
Indicated for repair of peripheral nerve injuries in patients where no gap exists or gap closure is achieved via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 10, 2022
Renerve Ltd % Chris Sloan President Sloan Regulatory Consulting LLC 322 Hart Road Gaithersburg, Maryland 20878
Re: K202234
Trade/Device Name: NervAlign Nerve Cuff Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: January 10, 2022 Received: January 11, 2022
Dear Chris Sloan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202234
Device Name NervAlign® Nerve Cuff
Indications for Use (Describe)
The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap closure is achieved by flexion of the extremity.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary NervAlign® Nerve Cuff K202234
#### Submitter
| Name | ReNerve Ltd |
|----------------|------------------------------------------------------------------------|
| Address | Suite 3, 21 Vale Street<br>North Melbourne, Victoria 3051<br>Australia |
| Contact Person | David Rhodes |
| Phone | +61 (0)418598835 |
| Email Address | drhodes@renerve.com.au |
| Date Prepared | February 2, 2022 |
### Device
| Trade Name | NervAlign® Nerve Cuff |
|-----------------------|-----------------------|
| Common Name | Nerve Cuff |
| Classification Name | Nerve Cuff |
| Classification Number | 21 CFR 882.5275 |
| Product Code | JXI |
| Regulatory Class | II |
### Predicate Device
| Name [510(k) Number] | AxoGuard Nerve Protector [K132660] |
|----------------------|------------------------------------|
|----------------------|------------------------------------|
### Device Description
The NervAlign® Nerve Cuff is a collagen membrane matrix derived from porcine pericardium. It is a sterile, whitish to light beige colored, freeze-dried, pre-cut, flat sheet of acellular collagen. The NervAlign® Nerve Cuff is available in three (3) different sizes: 10x10mm, 20x30mm and 30x40mm.
The collagen material that comprises the Nerve Cuff is derived from the same species as that of the predicate nerve cuff (AxoGuard Nerve Protector; K132660) manufactured by Cook Biotech Incorporated.
Like the predicate, the NervAlign® Nerve Cuff is implanted providing a scaffold which becomes infiltrated by the patient's cells and is remodelled into native tissue. The Nerve Cuff provides protection of the damaged nerve while the nerve heals.
The NervAlign® Nerve Cuff is packaged in a dried state, is for single use and provided sterile.
{4}------------------------------------------------
## Indications for Use
The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap or where a gap closure is achieved by flexion of the extremity.
## Comparison of Technological Characteristics with the Predicate Device
The proposed device is substantially equivalent to the design and materials in the predicate device. The table below summarizes the comparison between the predicate device and the NervAlign® Nerve Cuff.
| Feature | NervAlign® Nerve Cuff<br>[Proposed Device] | AxoGuard Nerve Protector<br>[Predicate Device] |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K202234 | K132660 |
| Device Class | Class II | Class II |
| Classification Name and<br>Number | Nerve Cuff; 21 CFR 882.5275 | Nerve Cuff; 21 CFR 882.5275 |
| Product Code | JXI | JXI |
| Indications for Use | The NervAlign® Nerve Cuff is<br>indicated for the repair of<br>peripheral nerve injuries in which<br>there is no gap or where a gap<br>closure is achieved by flexion of the<br>extremity. | The Nerve Cuff is indicated for<br>the repair of peripheral nerve<br>injuries in which there is no gap<br>or where a gap closure is<br>achieved by flexion of the<br>extremity. The device is provided<br>sterile and intended for one-time<br>use. |
| Principles of Operation | Collagen matrix intended to wrap<br>and repair damaged peripheral<br>nerves | Collagen matrix intended to wrap<br>and repair damaged peripheral<br>nerves |
| Material | Collagen Matrix - Porcine<br>Pericardium | Collagen Matrix – Porcine Small<br>Intestinal Submucosa (SIS) |
| Sterile/Single Use | Sterile; single use | Sterile; single use |
| Sterilization Method | Gamma Irradiation | Ethylene Oxide (EtO) |
# NervAlign® Nerve Cuff Comparison to AxoGuard Nerve Protector
# Performance Data
The following testing was performed on the ReNerve NervAlign® Nerve Cuff.
{5}------------------------------------------------
| Test | Test Method Summary | Results |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Suture retention<br>strength | A suture was placed through aged and<br>unaged devices and the force required to<br>pull free was measured. | Test was completed and met specification.<br>Like predicate the device has sufficient<br>strength for its intended use. |
| Thickness | Direct measurement of the device with<br>micrometer when hydrated according to<br>IFU. | Thickness measurements of device showed<br>substantially equivalent to predicate. |
| Tensile Strength | Direct measurement of the device.<br>Device placed between two grips and the<br>separation force required to reach device<br>failure was measured. | Test was completed and met specification.<br>Like predicate device is suitable for<br>intended use. |
| Cytotoxicity | The device was evaluated for potential<br>cytotoxic effects using a mammalian cell<br>line following ISO 10993-5 guidelines. | The test article extract showed no cytotoxic<br>potential. |
| Sensitization | The device was evaluated for the<br>potential to cause delayed dermal<br>contact sensitization in guinea pigs<br>based on ISO 10993-10. | Extracts of the test article showed no<br>evidence of inducing delayed contact<br>sensitization in the guinea pig. |
| Acute intracutaneous<br>reactivity (Irritation) | The device was evaluated for the<br>potential to cause irritation following<br>intracutaneous injection in rabbits based<br>on ISO 10993-10. | Extracts of the test article show no evidence<br>of irritation. |
| Acute systemic<br>toxicity | The device was evaluated for acute<br>systemic toxicity in mice based on ISO<br>10993-11. | No mortality or evidence of systemic<br>toxicity from both extracts, injected into<br>mice. |
| Pyrogenicity | The device was evaluated for the<br>potential to induce a pyrogenic response<br>following intravenous injection in<br>rabbits. Study conducted according to<br>the United States Pharmacopeia (USP 41<br>- NF36, General Chapter <151>). | The test article met the requirements and is<br>judged non-pyrogenic. |
| Hemolysis | The device was evaluated for hemolytic<br>potential when in contact with blood<br>based on ASTM F756 and requirements<br>of ISO 10993-4. | The direct contact of the test article was<br>slightly hemolytic.<br>The extracts at 25, 12.5 and 6.25% were<br>non-hemolytic.<br>Device is suitable for intended indication. |
| Genotoxicity<br>(AMES) | The device was evaluated for the<br>potential to induce reverse mutations in<br><i>Salmonella typhimurium</i> and | Extracts of the device were considered to be<br>non-mutagenic to tester strains. |
| Test | Test Method Summary | Results |
| Escherichia coli tester strains per ISO<br>10993-3. | | |
| Genotoxicity (mouse<br>lymphoma assay) | The device was evaluated to determine<br>its mutagenic potential using the mouse<br>lymphoma forward gene mutation assay<br>per ISO 10993-3. | Test article considered non-mutagenic. |
| Endotoxin | Bacterial endotoxin testing is conducted<br>per USP 85 and European<br>Pharmacopoeia 2.6.14 | Device is produced and released with<br>endotoxin level <2.15EU/device. |
| Subacute Systemic<br>Toxicity & Local<br>effects of<br>Implantation | The device was surgically implanted in<br>the subcutaneous tissue of rats for 4<br>weeks to evaluate its potential systemic<br>toxicity and local tissue response. | The device showed no evidence of systemic<br>toxicity. Device appeared as a well-<br>integrated scaffold with remodelling<br>responses. Microscopically the test article<br>demonstrated minimal reaction to the tissue. |
| Sub-chronic Systemic<br>Toxicity & Local<br>effects of<br>implantation | The device was surgically implanted in<br>the subcutaneous tissue of rats for 8 and<br>13 weeks to evaluate potential systemic<br>toxicity after 13 weeks and local tissue<br>responses after 8 and 13 weeks. | On subcutaneous implantation the device<br>showed no evidence of systemic toxicity<br>after 13 weeks.<br>Microscopically the test article<br>demonstrated minimal tissue reaction<br>compared to control. Device was<br>bioresorbing at 13 weeks as expected. |
| 6-Month Nerve Wrap<br>Study in New-<br>Zealand White<br>Rabbits | The device was surgically implanted<br>around the sciatic nerve of rabbits for 1,<br>2 & 6 months. The effects and<br>compatibility of the wrapping material<br>on the nerve and surrounding tissues<br>were assessed. | The device was well tolerated with no<br>device related clinical signs, changes in<br>body weight, food consumption,<br>neurological parameters, NCV, clinical<br>pathology, or organ weights up to 6 months<br>post-implantation.<br>No adverse macroscopic or microscopic<br>changes in the nerve and surrounding<br>tissues were observed. No device related<br>axonal degeneration was observed. |
| Rat transected sciatic<br>nerve model | The device was surgically implanted<br>around the transected sciatic nerve of the<br>rat and compared to nerves wrapped<br>with marketed predicate control<br>material. Data were collected at 1, 4 and<br>13 weeks. | At all time points the changes observed in<br>the nerve were similar between device and<br>predicate control material and were typical<br>of changes in a nerve after transection.<br>At all time points the device was considered<br>to elicit no or minimal reaction in<br>comparison to the predicate control |
| Test | Test Method Summary | Results |
| | Local tissue responses and device<br>degradation were evaluated, and motor<br>and sensory neurological assessments<br>were conducted. | The reactivity scores for device and<br>predicate control decreased at later time<br>points indicating healing.<br>The device was markedly or completely<br>degraded at 13 weeks, compared with the<br>control device which was not degraded.<br>Neuromas were not observed in any animal. |
{6}------------------------------------------------
{7}------------------------------------------------
# Conclusions
The NervAlign® Nerve Cuff has the following similarities to the predicate device:
- . Same intended use,
- Same basic design,
- Similar material, ●
- Similar safety and performance characteristics, and ●
- Same operating principles. ●
Based on the similarities of the intended use/indications for use, device design, principles of operation, technological characteristics and the results of the non-clinical performance testing, the subject device, NervAlign® Nerve Cuff, is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.