K202222 · Irrimax Corporation · FQH · Dec 14, 2020 · General Hospital
Device Facts
Record ID
K202222
Device Name
Irrisept Wound Debridement and Cleansing System
Applicant
Irrimax Corporation
Product Code
FQH · General Hospital
Decision Date
Dec 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.
Device Story
Irrisept is a single-use, manual, self-contained wound irrigation system. It consists of a 450mL bottle containing 0.05% Chlorhexidine Gluconate (CHG) in 99.95% sterile water and a multi-port applicator (Irriprobe). The device is used by clinicians to cleanse wounds; the operator manually compresses the Blow-Fill-Seal (BFS) bottle to deliver the solution through the applicator tip. The mechanical force of the fluid stream loosens and removes debris from the wound bed. The CHG acts as a preservative. The device is provided sterile and is intended for single use to assist in wound management, potentially reducing microbial load through mechanical irrigation.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including USP <51> antimicrobial effectiveness, ISO 13408 aseptic processing, ISO 11135/10993-7 sterilization validation, USP <85>/<151>/<161> endotoxin/pyrogen testing, ISO 11607 packaging validation, ASTM F1980 accelerated aging, and ISO 10993 biocompatibility testing.
Technological Characteristics
Manual, single-use irrigation system. Components: Polypropylene bottle, polycarbonate applicator. Solution: 0.05% CHG in 99.95% sterile water. Sterilization: EO (applicator and bottle) and aseptic BFS processing (solution). Biocompatibility: ISO 10993-1 compliant. Packaging: Tyvek header pouch and CSR wraps. Shelf-life: 2 years.
Indications for Use
Indicated for wound cleansing and debridement via mechanical action to loosen and remove wound debris. Prescription use only.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
Predicate Devices
Irrisept Wound Debridement and Cleansing System (K080779)
Submission Summary (Full Text)
{0}------------------------------------------------
December 14, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Irrimax Corporation Christy Coleman VP Regulatory 1665 Lakes Parkway - Suite 102 Lawrenceville, Georgia 30043
Re: K202222
Trade/Device Name: Irrisept Wound Debridement and Cleansing System Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: Class II, unclassified Product Code: FOH, FRO Dated: November 12, 2020 Received: November 13, 2020
#### Dear Christy Coleman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Kimberly Ferlin, Ph.D. Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202222
Device Name
Irrisept Wound Debridement and Cleansing System
#### Indications for Use (Describe)
The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| 510(K) NUMBER: | K202222 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PREPARATION DATE: | December 11, 2020 |
| APPLICANT: | Irrimax® Corporation<br>1665 Lakes Parkway, Suite 102<br>Lawrenceville, GA 30043 |
| CONTACT PERSON: | Christy Coleman, OD, MPH<br>Vice President, Regulatory Affairs, Irrimax®<br>Corporation<br>Tel: (770) 807-3355 |
| DEVICE TRADE NAME: | Irrisept® Wound Debridement and<br>Cleansing System |
| CLASSIFICATION NAME: | Jet Lavage |
| DEVICE CLASSIFICATION: | Class II (21 CFR 880.5475)<br>Unclassified (Pre-amendment) |
| PRODUCT CODE: | FQH; FRO |
| PREDICATE DEVICE: | Irrisept® Wound Debridement and Cleansing System<br>(K080779)<br>• Product Code FQH (Jet Lavage);<br>Class II (21 CFR 880.5475)<br>• Product Code FRO (Dressing, Wound,<br>Drug); Unclassified (pre-amendment)<br>Applicant: Irrimax Corporation |
#### DEVICE DESCRIPTION:
Irrisept® Wound Debridement and Cleansing System is a single-use, manual, self-contained irrigation device comprised of a 450mL bottle of 0.05% Chlorhexidine Gluconate (CHG) in 99.95% Sterile Water for Irrigation, United States Pharmacopeia (USP) and an applicator (labeled Irriprobe®). The solution is aseptically-filled in a Blow-Fill-Seal (BFS) bottle. The CHG acts as a preservative to inhibit microbial growth in the solution.
### INTENDED USE/INDICATIONS FOR USE:
The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.
{4}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The Irrisept that is the subject of this 510(k) is substantially equivalent to its predicate, predecessor K080779 Irrisept. This is a catch-up 510(k) submission intended to update the Irrisept record at FDA with the changes that have been implemented in the subject product. As such, the intended use/indications remain identical to that for the cleared product. Further, the implemented changes were primarily to be responsive to user feedback, capture the now available longer-term shelf-life data, or reflect a need to change a supplier or manufacturing contractor. These are captured in the comparison table below.
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS | | |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Feature | Cleared Irrisept Product<br>(K080779) | Subject Irrisept Product<br>(K202222) |
| Intended Use/<br>Indications for Use | The Irrisept Wound<br>Debridement and Cleansing<br>System is a wound cleansing<br>delivery system. The<br>mechanical action effectively<br>loosens and removes wound<br>debris. | Identical |
| Rx only | Prescription Product | Identical |
| Single Use | Single Use | Identical |
| Mechanism of Action | The mechanical action of fluid<br>across the wound removes<br>wound debris. | Identical. |
| Components and<br>Materials | Polypropylene bottleSolutionPolycarbonate applicator | New supplier for<br>Chlorhexidine GluconateNew supplier for<br>polycarbonate |
| Solution Composition | 0.05% Chlorhexidine Gluconate<br>in 99.95% Sterile Water for<br>Irrigation, USP | Identical |
| Container Design | 450 mL BFS bottle delivers fluid<br>by manual compression | Identical |
| Applicator Design | Multi-port applicator that<br>threads onto bottle. | Multi-port applicator that<br>threads onto bottle and with<br>a longer tip.Change to a higher capacity<br>manufacturing mold |
| Biocompatibility | Conforms to ISO 10993-1 for a<br>surface device with breached or<br>compromised surface contact<br>and a limited contact duration<br>(≤ 24 hours) | Identical |
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS | | |
| Comparison Feature | Cleared Irrisept Product<br>(K080779) | Subject Irrisept Product<br>(K202222) |
| Manufactured using<br>Aseptic Techniques | Solution is manufactured using Blow-Fill-Seal (BFS) aseptic processing. Solution labeled as Sterile. | Identical BFS process New contract manufacturer Changed to a higher capacity BFS manufacturing mold |
| Non-Pyrogenic | Solution | Solution and applicator<br>Endotoxin Limit ≤ 20 EU/device<br>(<0.045 EU/mL for 450 mL) |
| Packaging | 1. Packaged applicator in Tyvek® film pouch<br>2. Protective PETG tube added over bottle cap to protect bottle cap during shipment<br>3. Sterilized applicator and bottle-filled solution are then packaged in a shelf box. | Changed manufacturing final step for protective PETG tube Bottle-filled solution and applicator are packaged together in CSR wraps and Tyvek® header pouch for sterilization New contract manufacturer |
| Sterilization | Packaged applicator conforms to ISO 11135 and ISO 10993-7 for EO sterilization, whereas the solution is aseptically processed | The bottle and applicator are EO-sterilized and conform to ISO 11135 and ISO 10993-7, whereas the solution is aseptically processed |
| Shelf-Life | Six-month shelf-life. | Two-year shelf-life |
| Labeling | Labeling consists of: Bottle Label Applicator label Package Label - Shelf Box Label Shipping Box Label | Labeling consists of: Bottle Label Package Label - Front Package Label - Back Shipping Box Label |
{5}------------------------------------------------
The differences between the two versions of Irrisept do not raise different questions of safety and effectiveness and have been confirmed through testing.
{6}------------------------------------------------
## PERFORMANCE TESTING:
Performance testing demonstrates that the subject Irrisept is as safe and as effective as the predicate, predecessor Irrisept product. These tests include:
- Preservative Antimicrobial Effectiveness ●
- o USP <51> Antimicrobial Effectiveness Testing
- Aseptic Processing
- o ISO 13408 Aseptic Processing of Health Care Products
- Sterilization
- ISO 11135 Sterilization of Health-Care Products Ethylene Oxide O
- ANSI AAMI ISO 10993-7 Ethylene oxide sterilization residuals O
- USP <71> Sterility Tests O
- AAMI TIR 28 Product adoption and process equivalence for ethylene oxide O sterilization
- Endotoxins and Pyrogens ●
- USP <85> Bacterial Endotoxins Test O
- USP <151> Pyrogen Test (USP Rabbit Test) O
- USP <161> Medical Devices Bacterial Endotoxin and Pyrogen Tests O
#### ● Packaging and Shelf-Life
- ISO 11607 Packaging for Terminally Sterilized Medical Devices O
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier O Systems for Medical Devices
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging о by Internal Pressurization (Bubble Test)
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous o Medical Packaging by Dye Penetration
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping o Containers and Systems
- Biocompatibility ●
- o ISO 10993- Biological Evaluation of Medical Devices
#### CONCLUSIONS:
In conclusion, Irrimax has demonstrated through performance testing that the subject Irrisept is substantially equivalent to its predicate Irrisept product in intended use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.