← Product Code [FXX](/productcode/FXX) · K202214

# IFM Surgical Mask Model SM3 (K202214)

_Fssc, LLC · FXX · Mar 2, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202214

## Device Facts

- **Applicant:** Fssc, LLC
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Mar 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The IFM Surgical Mask Model SM3 is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

## Device Story

IFM Surgical Mask Model SM3; four-layer, flat-pleated, single-use, non-sterile disposable face mask. Composed of polypropylene spunbond and melt-blown layers; includes spandex ear loops and malleable polyethylene nosepiece. Used in clinical settings or general infection control to provide physical barrier against blood, body fluids, and particulates. Protects both wearer and patient. No natural rubber latex. Device provides passive protection; no active components or software.

## Clinical Evidence

No clinical data. Bench testing only. Device demonstrated compliance with ASTM F2100, ASTM F1862 (fluid resistance), ASTM F2299 (PFE), ASTM F2101 (BFE), 16 CFR 1610 (flammability), and ISO 10993 (biocompatibility).

## Technological Characteristics

Four-layer construction: polypropylene spunbond outer/inner layers, polypropylene melt-blown insertion/filter layers. Flat-pleated form factor. Spandex ear loops. Polyethylene nosepiece. Non-sterile. ASTM F2100 Level 3 performance. Biocompatibility per ISO 10993-1, -5, -10.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- 3M™ High Fluid-Resistant Procedure Mask ([K191355](/device/K191355.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

March 2, 2021

Fssc, LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K202214

Trade/Device Name: IFM Surgical Mask Model SM3 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: February 3, 2021 Received: February 17, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202214

Device Name IFM Surgical Mask

#### Indications for Use (Describe)

The IFM Surgical Mask Model SM3 is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. The face mask is single use, disposable device, provided non-sterile.

| Type of Use (Select one or both, as applicable)            |
|------------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

_ | Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# K202214

IFM Surgical Mask Model SM3

| Sponsor Information:   | FSSC, LLC<br>3300 West Clark St.<br>Rensselaer, IN 47978                                                                                                                                                                                                                                                                                                                                             |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                        | Sponsor Contact:<br>Clayton Geyer<br>Operations Engineer<br>Phone Number: 219-863-8316<br>Email: clayton@ifmasks.com                                                                                                                                                                                                                                                                                 |
| Contact Person:        | Amy Fowler<br>Regulatory Counsel<br>Pathmaker FDA Law, PLLC<br>528 Hennepin Ave., Suite 503<br>Minneapolis, MN 55403                                                                                                                                                                                                                                                                                 |
| Date of Summary:       | March 1, 2021                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name:           | Surgical Mask                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name:   | Surgical Apparel                                                                                                                                                                                                                                                                                                                                                                                     |
| Proprietary Names:     | IFM                                                                                                                                                                                                                                                                                                                                                                                                  |
| Review Panel:          | General Hospital                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code:          | FXX                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Classification: | Class II per 21 CFR §878.4040                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device:      | 3M™ High Fluid-Resistant Procedure Mask (K191355) – Model<br>Number 1840                                                                                                                                                                                                                                                                                                                             |
| Intended Use:          | The IFM Surgical Mask Model SM3 is intended to be worn to<br>protect both the patient and healthcare personnel from transfer<br>of microorganisms, body fluids, and particulate material. These<br>face masks are intended for use in infection control practices to<br>reduce potential exposure to blood and body fluids. The face<br>mask is single use, disposable device, provided non-sterile. |
| Device Description:    | The IFM Surgical Mask Model SM3 is composed of four-layers<br>and is flat-pleated. The mask materials consist of an outer cover<br>web (polypropylene spunbond, white), insertion layer<br>(polypropylene, melt-blown, white), filter web (polypropylene<br>melt-blown, white) and inner cover web (polypropylene melt-<br>spunbond, white). Each mask contains ear loops to secure the              |

{4}------------------------------------------------

mask over the user's mouth and nose and includes a malleable nosepiece to provide a firm fit over the nose. The mask is a single use, disposable device, provided non-sterile.

This device is not made with natural rubber latex.

| Available Model Numbers | SM3 | IFM Surgical Mask |
|-------------------------|-----|-------------------|
|-------------------------|-----|-------------------|

## Comparison of Subject and Predicate Devices:

| Item(s)                                    | Subject Device<br>(K202214)<br>IFM Surgical Mask Model<br>SM3<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>(K191355)<br>1840 3M™ High Fluid-<br>Resistant Procedure Mask<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                              |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Intended Use /<br>Indications for Use      | The IFM Surgical Mask<br>Model SM3 is intended to<br>be worn to protect both<br>the patient and<br>healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce<br>potential exposure to<br>blood and body fluids.<br>The face mask is single<br>use, disposable device,<br>provided non-sterile. | The 3M™ High Fluid-<br>Resistant Surgical Mask<br>and 3M™ High Fluid-<br>Resistant Procedure Mask<br>is intended to be worn to<br>protect both the patient<br>and healthcare personnel<br>from transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face<br>masks are intended for<br>use in infection control<br>practices to reduce<br>potential exposure to<br>blood and body fluids.<br>The face mask is single<br>use, disposable device,<br>provided non-sterile.<br>1840 - 3M™ High Fluid-<br>Resistant Procedure<br>Mask1 | Same                                                    |
| Type of Use                                | Over-The-Counter Use<br>(21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                        | Over-The-Counter Use<br>(21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                    |
| Outer layer                                | Polypropylene Spunbond,<br>white                                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene Spunbond,<br>green                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar                                                 |
| Item(s)                                    | Subject Device<br>(K202214)<br>IFM Surgical Mask Model<br>SM3<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>(K191355)<br>1840 3M™ High Fluid-<br>Resistant Procedure Mask<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                              |
| Insertion                                  | Polypropylene Spunbond,<br>white                                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene Spunbond,<br>white                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                    |
| Middle layer                               | Polypropylene<br>Meltblown, white                                                                                                                                                                                                                                                                                                                                                                                                     | Polypropylene<br>Meltblown, white                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                    |
| Inner layer                                | Polypropylene<br>Meltblown, white                                                                                                                                                                                                                                                                                                                                                                                                     | Polypropylene Thermal-<br>bonded, white                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar                                                 |
| Nose Piece                                 | Polyethylene Strip                                                                                                                                                                                                                                                                                                                                                                                                                    | Polyethylene Coated<br>Steel Wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar                                                 |
| Edge wrap                                  | Polypropylene Spunbond,<br>white                                                                                                                                                                                                                                                                                                                                                                                                      | Polypropylene Spunbond,<br>white or Polyethylene<br>Terephthalate, white                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar                                                 |
| Ear Loops                                  | Spandex Elastic Cord                                                                                                                                                                                                                                                                                                                                                                                                                  | Spandex elastic cord<br>(polyurethane core with<br>polyethylene<br>terephthalate /nylon<br>cover)                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Similar                                                 |
| Colors                                     | White                                                                                                                                                                                                                                                                                                                                                                                                                                 | Green (Outer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different. Subject<br>device does not have<br>colorant. |
| Style                                      | Flat - Pleated                                                                                                                                                                                                                                                                                                                                                                                                                        | Flat - Pleated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                    |
| Multiple Layers                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                    |
| Single Use                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                    |
| Sterile                                    | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                           | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                    |
| Item(s)                                    | Subject Device<br>(K202214)<br>IFM Surgical Mask Model<br>SM3<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>(K191355)<br>1840 3M™ High Fluid-<br>Resistant Procedure Mask<br>ASTM F2100 Level 3                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison                                              |
| Length                                     | 6.63" ± 0.2"                                                                                                                                                                                                                                                                                                                                                                                                                          | 6.9" ± 0.2"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar                                                 |
| Width                                      | 3.5" ± 0.3"                                                                                                                                                                                                                                                                                                                                                                                                                           | 3.5" ± 0.3"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                    |
| Particulate Filtration<br>Efficiency (PFE) | Three lots passed at<br>≥98% @<br>0.1 micron<br>ASTM F2299                                                                                                                                                                                                                                                                                                                                                                            | 32/32 Passed at ≥98% @<br>0.1 micron<br>ASTM F2299                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Similar                                                 |
| Fluid Resistance                           | Three lots passed at<br>160mm Hg<br>ASTM F1862                                                                                                                                                                                                                                                                                                                                                                                        | 32/32 Passed at 160mm<br>Hg<br>ASTM F1862                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                    |
| Bacterial Filtration<br>Efficiency (BFE)   | Three lots passed at<br>≥98% ASTM F2101                                                                                                                                                                                                                                                                                                                                                                                               | 31/32 Passed at ≥98%<br>ASTM F2101                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                    |
| Differential Pressure                      | Three lots passed at <5<br>Delta-P (mmH₂O/cm²)<br>EN 14683:2019, Annex C<br>and ASTM F2100-19                                                                                                                                                                                                                                                                                                                                         | 32/32 Passed at <5<br>mmH2O/cm² MIL-<br>M36954C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar                                                 |
| Flammability                               | Three lots Passed. - Class<br>1 16 CFR<br>1610                                                                                                                                                                                                                                                                                                                                                                                        | 5/5 Passed ≥3 Seconds<br>burn time - Class 1 16 CFR<br>1610                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Similar                                                 |
| Biocompatibility                           | Under the conditions of<br>the study the device was<br>found non-cytotoxic;                                                                                                                                                                                                                                                                                                                                                           | Under the conditions of<br>the study the device was<br>found non-cytotoxic;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                    |
|                                            | Under the conditions of<br>the study the device was<br>found not an irritant;                                                                                                                                                                                                                                                                                                                                                         | Under the conditions of<br>the study the device was<br>found not an irritant;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                         |
|                                            | Under the conditions of<br>the study the device was<br>found not a sensitizer;                                                                                                                                                                                                                                                                                                                                                        | Under the conditions of<br>the study the device was<br>found not a sensitizer;                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                         |

<sup>1</sup> The predicate device has four cleared model numbers. The subject device is only being compared to model 1840 3M™ High Fluid-Resistant Procedure Mask

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

# Performance Testing (non-clinical)

The following non-clinical performance testing standards have been met for the IFM Surgical Mask Model SM3:

| ASTM F2100       | Standard Specification for Performance of Materials Used in Medical Face<br>Masks                                                                                  |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM F1862       | Standard Test Method for Resistance of Medical Face Masks to Penetration by<br>Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)         |
| ASTM F2299       | Standard Test Method for Determining the Initial Efficiency of Materials Used<br>in Medical Face Masks to Penetration by Particulates Using Latex Spheres          |
| ASTM F2101       | Standard Method for Evaluating the Bacterial Filtration Efficiency (BFE) of<br>Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus<br>aureus |
| 16 CFR Part 1610 | Standard for the Flammability of Clothing                                                                                                                          |
| ISO 10993-1      | Biological evaluation of medical devices - Part 1: Evaluation and testing within<br>a risk management process                                                      |
| ISO 10993-5      | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity<br>of medical devices                                                           |
| ISO 10993-10     | Biological evaluation of medical devices — Part 10: Tests for irritation and skin<br>sensitization                                                                 |

# Performance Testing (clinical)

No Clinical performance testing conducted.

## Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K202214](https://fda-staging.innolitics.com/device/K202214)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).