The ASTOPAD® Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The ASTOPAD Patient Warming System is indicated for use in all areas of healthcare facilities for preventing or treating hypothermia or maintaining normothermia. The warming blankets can be used over or under the patient (pediatric and adult) by appropriately trained healthcare professionals.
Device Story
ASTOPAD Patient Warming System provides conductive heat to patients via reusable warming blankets. System comprises control unit (DUO310-NA) and applied parts (blankets). Control unit connects to mains or optional battery; features two outputs for independent temperature control of blankets. Operator selects set temperature (32.0°C–39.0°C) via control panel. Integrated sensors monitor blanket temperature; control unit provides optical/acoustic alarms for errors or deviations. Safety features include redundant sensor monitoring, heater monitoring, and automatic power shut-off during failure conditions (overheating, sensor/heater defects). Used in clinical environments by trained healthcare professionals to maintain normothermia or treat hypothermia. Benefits include precise thermal regulation and patient warming during procedures.
Clinical Evidence
No clinical data or animal studies were required. Evidence consists of bench testing, including biocompatibility (cytotoxicity, sensitization, irritation), electrical safety, EMC, mechanical testing, thermal testing, human factors validation, and reprocessing validation (cleaning/disinfection).
Technological Characteristics
Conductive warming system; control unit with two outputs; reusable coated plastic blankets. Uses 8 temperature sensors per blanket. Powered by low-voltage electrical energy. Features: potential equalization connector, universal mounting clamp, optional battery (approx. 2-hour operation). Standards: IEC 60601-1 (defibrillation-proof), IEC 60601-1-2 (EMC), IEC 60601-1-6, IEC 60601-1-8, IEC 80601-2-35. Hardware-based temperature control and monitoring.
Indications for Use
Indicated for pediatric and adult patients in all healthcare facility areas to prevent or treat hypothermia or maintain normothermia.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
September 4, 2020
Stihler Electronic GmbH % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K202197
Trade/Device Name: Astopad Patient Warming System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II Product Code: DWJ Dated: August 4, 2020 Received: August 5, 2020
Dear Dave Yungvirt:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202197
Device Name ASTOPAD Patient Warming System
### Indications for Use (Describe)
The ASTOPAD® Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The ASTOPAD Patient Warming System is indicated for use in all areas of healthcare facilities for preventing or treating hypothermia or maintaining normothermia. The warming blankets can be used over or under the patient and adult) by appropriately trained healthcare professionals.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|-----------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | <input checked="true" type="checkbox"/> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# STIHLER ELECTRONIC
# 510(k) Summary K202197
| I. SUBMITTER | Stihler Electronic GmbH<br>Julius-Hoelder-Strasse 36<br>70597 Stuttgart<br>Germany |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Phone: +49-711-720670<br>Fax: +49-711-7206757 |
| Contact person: | Christina Miracle<br>Gentherm Medical, LLC<br>Director of Quality Assurance & Regulatory Affairs<br>phone: 513.326.5295<br>mail: <a href="mailto:christina.miracle@gentherm.com">christina.miracle@gentherm.com</a> |
| Date of Preparation: | 4 Sep 2020 |
| Manufacturer: | Stihler Electronic GmbH<br>Julius-Hoelder-Strasse 36<br>70597 Stuttgart<br>Germany |
| Establishment Registration Number: | 9617473 |
| II. DEVICE | |
| Trade name: | ASTOPAD® Patient Warming System |
| Model numbers: | DUO310-NA, COV070-NA, COV105-NA, COV150-NA, COV155-NA, COV180-NA |
| Common name: | Thermal Regulating System |
| Classification name: | System, Thermal Regulation (21 CFR Part 870.5900, Product Code DWJ) |
| Class: | Class II |
| III. PREDICATE DEVICE | HotDog Patient Warming System, K112488<br>This predicate has not been subject to a design-related recall. <sup>1</sup> |
| IV. DEVICE DESCRIPTION | The ASTOPAD Patient Warming System is a thermal regulating system which includes a connection cable that attaches to reusable conductive warming blankets for patient warming in clinical environments The |
ve warming blankets for patient warming in clinical environments. The system consists of a control unit (DUO310-NA) and one or optionally two applied parts (warming blankets/COVXXX-NA) with different sizes to fit different applications:
<sup>1</sup> Source: FDA Medical Device Recalls database, last update 02/14/2020
{4}------------------------------------------------
COV070-NA2 warming blanket, 680x480 mm COV105-NA warming blanket, 1050x500 mm COV150-NA warming blanket, 1500x500 mm COV155-NA Arm-chest-warming blanket, 1500x500 mm (with cut-outs) COV180-NA warming blanket 1800x800 mm
The ASTOPAD control unit is equipped with a universal mounting clamp for attaching to standard medical equipment rails or an infusion stand, two outputs for connecting one or two applied parts, connection to mains by detachable power supply cord, connector for potential equalization, the electronic boards and the control panel for the user/operator.
The control unit has two outputs (connecting sockets) A and B for connecting ASTOPAD applied parts. The desired set temperature can be selected in the range of 32.0 ℃ - 39.0 ℃ in 0.5 ℃ increments for each connected applied part, independent of each other, on the control panel of the control unit. The control unit can also be used with only one of the outputs, A or B. The selected set temperature and the current temperature are displayed individually for each applied part in the control panel.
The control unit supplies the applied part(s) with electrical current (low voltage) and monitors and controls the temperature of the applied part(s). When error conditions occur, the control unit draws the operator's attention to the error condition by means of alarm signals (optical & acoustical). In the event of failure conditions (overheating, heater defect, sensor defect), the control unit reacts immediately by shutting off the power supply to the applied part.
Optionally the control unit can be fitted with a battery. When the battery is installed, it is possible to operate the device independently from the mains for approximately two hours.
The ASTOPAD control unit does not control or indicate the actual temperature of the patient, but rather only the actual temperature of the applied part.
Temperature regulation of the individual applied parts is performed with several integrated sensors.
Safety of ASTOPAD is provided by the following measures per output:
- Several temperature sensors for each applied part ●
- Double independent sensor monitoring ●
- Heater monitoring .
- Time shut-off .
- Visual and acoustic alarm signals ●
- Overheating and low-temperature alarm if the contact surface ● temperature deviates from the temperature controller setting
<sup>2 &</sup>quot;NA" Refers to North America, with English labeling.
{5}------------------------------------------------
## V. INDICATIONS FOR USE
The ASTOPAD® Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The ASTOPAD Patient Warming System is indicated for use in all areas of healthcare facilities for preventing or treating hypothermia or maintaining normothermia. The warming blankets can be used over or under the patient (pediatric and adult) by appropriately trained healthcare professionals.
Both the subject device and the predicate device have the same intended use: to prevent or treat hypothermia and to provide warmth to patients.
The Indications for Use of the subject device are similar as that of the predicate device; however, the differences in wording do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device. Both the subject and predicate devices have the same intended use for the prevention or treatment of hypothermia and maintenance of normothermia for pediatric and adult patients in all areas of healthcare facilities.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Subject (New) Device
Control Unit
Image /page/5/Picture/9 description: The image shows a medical device, specifically an ASTOPAD system, which appears to be a temperature control unit. The device has a digital display showing a temperature of 38.5 degrees Celsius on both sides, labeled A and B. It features multiple buttons for adjustments and two connector ports at the front, suggesting it can be connected to other medical equipment. The device is white and has warning labels at the bottom.
Predicate Devices
Image /page/5/Picture/11 description: The image shows a blue HotDog warming system. The top portion of the device has a digital display showing the number 43. Below the display are several ports for connecting the warming blankets. The bottom of the device has the HotDog logo.
{6}------------------------------------------------
# STIHLER ELECTRONIC
### Stihler Electronic GmbH Traditional 510(k) Premarket Notification for ASTOPAD Patient Warming System
Image /page/6/Picture/2 description: The image shows medical warming blankets. On the left side of the image, there are three orange warming blankets. On the right side of the image, there is a blue and a black warming blanket.
- Heat transfer to the patient, using conductive heat, is the technological Comparison principle for both the subject and predicate devices. It is based on the use of applied parts which have thermal contact with the patient to provide heat.
At a high level, the subject and predicate devices are based on the following same technological elements:
- · Applied parts in the form of blankets are lightweight and flexible and both are made from coated plastics.
- · Applied parts are powered by electrical energy (low voltage).
- · Electrical energy is transformed to warmth by a heating element which opposes the current by an electrical resistance.
- · Temperature of the applied parts is controlled by temperature sensors.
- · The systems include a control unit and applied parts in different sizes.
- · Control units and applied parts are connected together by a detachable cable.
- · Control units include attachment device for IV-pole or medical rail, control and monitoring electronics, control panel for the user, outputs to connect applied parts simultaneously, and input for power supply cord.
- · User can select set temperature of the applied parts within a limited range.
- · Control panels show set temperature of the applied part.
- . The control units inform the user about alarm conditions by means of optical and acoustic signals.
- · Control units react with an alarm in the case of sensor defects
- · Control units include a connector for potential equalization.
- . Control units have a grip tray on the upper backside, with which the device can be transported with one hand.
The following technological differences exist between the subject and predicate devices:
- · The subject device uses 8 temperature sensors for temperature control and monitoring, whereas the predicate device uses 2 temperature sensors for temperature control and monitoring.
- · Each blanket of the subject device can be placed over or under the patient and can be used on either side, whereas the applied parts of the
ASTOPAD® Patient Warming System – 510(k) Summary Page 4 of 6
{7}------------------------------------------------
predicate device are either over blankets or underbody mattresses that cannot be used interchangeably and can only be used on one side.
- · The subject device is classified as "defibrillation-proof" according to the requirements of IEC 60601-1, whereas the predicate device is not.
- . The subject device shows the user selected set temperature and the current temperature of the applied part simultaneously, whereas in operation mode the predicate device shows only the set temperature of the applied part.
- Based on the requirements of IEC 80601-2-35 the subject device has a . set temperature range from 32℃ to 39℃ in 0.5℃ increments. The predicate device has a selectable set temperature range from 37°C to 43°C in 1°C increments for blankets and 35°C to 39°C for mattresses.
- · The subject device uses hardware for controlling and monitoring the temperature of the applied part, whereas the temperature control of the predicate device is managed by Microprocessor utilization.
- · The subject device can optionally be fitted with a chargeable battery which makes it possible to operate the device independently from the mains for approximately two hours.
#### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing The biocompatibility evaluation for the ASTOPAD Patient Warming System was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Use of International ISO 10993-1, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process".
The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation ●
Electrical Safety and Electrical safety and EMC testing were conducted on the ASTOPAD Patient Warming System, consisting of the control unit and different electromagnetic compatibility applied parts. Testing shows the system complies with the applicable (EMC) requirements of IEC 60601-1, IEC 60601-1-6, IEC 60601-1-8 and IEC 80601-2-35 standards for safety and IEC 60601-1-2 standard for EMC.
> Documentation for claiming electromagnetic compatibility was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices".
#### Software verification and validation testing were conducted, and Software Verification and Validation Testing documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
{8}------------------------------------------------
The software for this device was considered as a "minor" level of concern, since a failure or a latent flaw in the software could not result in injury to the patient or operator.
- Mechanical testing Mechanical testing was conducted on the ASTOPAD Patient Warming System within the testing of the standards IEC 60601-1 and IEC 80601-2-35.
Testing shows the system complies with the applicable mechanical requirements of the standards.
- Thermal testing was conducted on the ASTOPAD Patient Warming Thermal testing System within the testing of the standards IEC 60601-1 and IEC 80601-2-35.
Testing shows the system complies with the applicable thermal requirements of the standards.
Human Factors Validation Human factors validation testing for the ASTOPAD Patient Warming Testing System was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Applying Human Factors and Usability Engineering to Medical Devices".
> The ASTOPAD system has been found to be safe and effective for the intended users, use and use environments.
- Reprocessing Testing The ASTOPAD Patient Warming System is classified as a reusable medical device which requires cleaning and disinfecting after each use. Cleaning and disinfecting validation were conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling".
Required cleaning and disinfection validation testing were conducted on the control unit and on the applied part.
ASTOPAD Patient Warming System meets the applicable cleaning & disinfecting performance criteria, when cleaned and disinfected as labeled.
Animal Study Not required
- Clinical Study Not required
- CONCLUSION Electronic GmbH considers the ASTOPAD Stihler Patient VIII. Warming System substantially equivalent to the predicate device because they have the same intended use and the same operating principles (warming by convection), any differences between the subject and predicate device do not raise new questions of safety and effectiveness. Testing has been conducted and the subject device meets all requirements of the applicable standards demonstrating that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.