Medline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs)

K202179 · Medline Industries, Inc. · LYZ · Oct 26, 2020 · General Hospital

Device Facts

Record IDK202179
Device NameMedline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs)
ApplicantMedline Industries, Inc.
Product CodeLYZ · General Hospital
Decision DateOct 26, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Medline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs) are a disposable device intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between the patient and examiner. The gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable vinyl examination gloves; worn by healthcare personnel to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Provides barrier protection in clinical settings. Warning: contraindicated for use with Carmustine and Thiotepa due to breakthrough times.

Clinical Evidence

Bench testing only. Permeation testing performed per ASTM D6978-05 for ten chemotherapy drugs. Breakthrough detection times reported for Carmustine (11.4 min) and Thiotepa (17.4 min); all other tested drugs exceeded 240 minutes.

Technological Characteristics

Vinyl material; non-powdered; disposable; form-fitting glove design. Tested for chemotherapy drug permeation per ASTM D6978-05 (Reapproved 2013).

Indications for Use

Indicated for use as a disposable medical glove worn on examiner's hands or fingers to prevent patient-examiner cross-contamination. Suitable for use with specific chemotherapy drugs, excluding Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 26, 2020 Medline Industries, Inc. Adam Ostrower Regulatory Affairs Sr. Specialist Three Lakes Drive Northfield, Illinois 60093 Re: K202179 Trade/Device Name: Medline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, OPJ Dated: August 3. 2020 Received: August 4, 2020 Dear Adam Ostrower: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202179 Device Name Medline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs) ## Indications for Use (Describe) The Medline Glide-On 3G Vinyl Examination Gloves (Tested for use with Chemotherapy Drugs) are a disposable device intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between the patient and examiner. The gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Chemotherapy | Drug Concentration | Minimum Breakthrough Detection | |----------------------------|--------------------|--------------------------------| | Carmustine | 3.3 mg/ml | 11.4 minutes | | Cisplatin | 1.0 mg/ml | >240 minutes | | Cyclophosphamide | 20.0 mg/ml | >240 minutes | | Dacarbazine | 10.0 mg/ml | >240 minutes | | Doxorubicin Hydrocholoride | 2.0 mg/ml | >240 minutes | | Etoposide (Toposar) | 20.0 mg/ml | >240 minutes | | Fluorouracil | 50.0 mg/ml | >240 minutes | | Methotrexate | 25.0 mg/ml | >240 minutes | | Paclitaxel (Taxol) | 6.0 mg/ml | >240 minutes | | Thiotepa | 10.0 mg/ml | 17.4 minutes | WARNING: DO NOT USE WITH CARMUSTINE DO NOT USE WITH THIOTEPA Type of Use (Select one or both, as applicable) | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...