AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are dental prosthetics placed onto a dental implant or screw retained abutment, to provide support for dental prosthetic restorations. The Base Abutments and Adhesive Sleeves include: Titanium bases/sleeves with a pre-machined implant connection or screw retained abutment connection on which the CAD/CAM designed upper structure may be fitted to complete a two-piece dental abutment. The Base Abutments and Adhesive Sleeves include an abutment/sleeve screw for fixation to the underlying implant/screw retained abutment. These dental prosthetics may be used for single-tooth or multiple-unit (bridges and bars) restorations and are compatible for use with AB Dental Implants with internal connection and screw retained connections for standard and narrow platforms. All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.
Device Story
Device consists of titanium alloy (Ti-6Al-4V ELI) base abutments and adhesive sleeves used as part of a two-piece dental abutment system. Inputs include patient intra-oral scan data used to design a custom CAD/CAM zirconia superstructure. The titanium base/sleeve is fixed to the implant or screw-retained abutment via a screw; the CAD/CAM superstructure is then cemented to the titanium base using dental cement. Used in dental clinics by dentists/prosthodontists. Design and manufacturing occur at a validated milling center. Output is a final prosthetic restoration providing support for single or multiple-unit teeth. Benefits include patient-specific fit and restoration of dental function.
Clinical Evidence
No clinical data were included in this submission. Evidence consists of bench testing, including fatigue testing per ISO 14801, cytotoxicity testing per ISO 10993-5, and MRI safety testing per ASTM F2052, F2213, F2182, and F2119.
Technological Characteristics
Materials: Ti-6Al-4V ELI titanium alloy. Components: Anti-rotational and rotational base abutments, adhesive sleeves. Connectivity: CAD/CAM digital workflow requiring validated milling center. Sterilization: End-user sterilized. MR Conditional per ASTM F2503-20.
Indications for Use
Indicated for patients requiring single-tooth or multiple-unit (bridges and bars) dental prosthetic restorations supported by dental implants or screw-retained abutments.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
A.B. Dental Device Ltd. % Nissim Shaked RA Specialist Nissim Shaked RA Specialist 17th Tel-Hay St. Raanana, 4340525 ISRAEL
Re: K202144
Trade/Device Name: A.B. Dental Devices® Dental Implants System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: August 1, 2022 Received: August 23, 2022
Dear Nissim Shaked:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202144
## Device Name
A.B. Dental Devices® Dental Implants System
## Indications for Use (Describe)
AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are dental prosthetics placed onto a dental implant or screw retained abutment, to provide support for dental prosthetic restorations. The Base Abutments and Adhesive Sleeves include: Titanium bases/sleeves with a pre-machined implant connection or screw retained abutment connection on which the CAD/CAM designed upper structure may be fitted to complete a two-piece dental abutment. The Base Abutments and Adhesive Sleeves include an abutment/sleeve screw for fixation to the underlying implant/screw retained abutment. These dental prosthetics may be used for single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with AB Dental Implants with internal connection and screw retained connections for standard and narrow platforms.
All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY A.B. Dental Devices® CADCAM Products
Sponsor:
A.B. Dental Devices Ltd. 19 Hayalomim St., Ashdod 7761117 Israel
Contact Person: Gabi Krauss
RA Manager Phone: 972-8-8531388 Facsimile: 972-8-8522562
| Date Prepared: | September 22, 2022 |
|-----------------------|------------------------------------------------------|
| Name of Device: | A.B. Dental Devices® Dental Implants System |
| Common or Usual Name: | Base Abutment/ Adhesive sleeve |
| Classification Name: | Endosseous dental implant abutment; 21 CFR §872.3630 |
| Product Code: | NHA |
Class II
Predicate Devices:
Regulatory Class:
K181580 Cortex CAD/CAM Abutments
### Device Description
A.B. Dental Devices® CADCAM Products consists of; Base Abutments and Adhesive Sleeves which are used with Standard Platform (Internal Hex connection) implants, Narrow Platform implants and Screw retained (P64 and P14/P16 MultiUnit) abutments.
Base Abutments (standard or narrow platform) and Adhesive Sleeves (screw retained abutment platform) are prefabricated abutments/sleeves made to be used as final abutments/ sleeves that will be used as adhesive connection (cement retained) of the prosthetic restoration on implant platform or on Multiunit platforms.
The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral scanning using the manufacturing. All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.
This submission adds the following Base Abutments and Adhesive sleeves:
| Device | Description |
|------------------------------------------------------------------|-----------------|
| Ti Base Anti-Rotational Abutments (Standard or Narrow platforms) | Base Abutment |
| Ti Base Rotational Abutments (Standard or Narrow platforms) | Base Abutment |
| P64 Ti Conical Adhesive Sleeve | Adhesive Sleeve |
| Ti Straight Adhesive Sleeves | Adhesive Sleeve |
### Base Abutments;
The titanium bases are used as part of a two-piece abutment, where the base is premanufactured from titanium alloy (Ti-6Al-4V ELI) and the top half is a CAD/CAM zirconia superstructure, milled at a validated milling center. These pieces are cemented together to form the final abutment.
- > Raw material Zirconium blanks
Dental Direkt Zirconia Blanks by Dental Direkt GmbH, cleared under K183569
- > Dental Cement
PANAVIA SA Cement Universal by Kuraray Noritake Dental Inc., cleared under K183537
{4}------------------------------------------------
# 510(K) Summary Page 2 of 5
#### Zirconia structure Details
| Spec. | Zirconium Blank | |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Trade name | Dental Direkt Zirconia Blanks | |
| Common name | Dental zirconia blanks/blocks | |
| Manufacturer | Dental Direkt GmbH | |
| 510K number | K183569 | |
| Product code | EIH | |
| Regulatory class | Class II | |
| Intended use | DD cube ONE ML dental zirconia blanks are indicated for crowns, multi-unit bridges and inlay bridges. Applications include both, anterior and posterior bridges. | |
#### Adhesive Sleeves
The adhesive sleeves are all Rotational (non-engage) for bridges restoration.
The adhesive sleeves are made of Titanium alloy (Ti-6Al-4V ELI) and will be installed on straight or angular screw-retained abutments.
The Multiunit (screw retained abutments already cleared devices) are P14/P16 platform: The below listed devices are previously cleared items that the sleeves are intended to be used with:
| Ref No. |
|-----------------|
| P14-3,17-1 |
| P14-3,17-3 |
| P14-3,30-1 |
| P14-3,30-3 |
| P14-3.75,17-1 |
| P14-3.75,17-3 |
| P14-3.75,30-1 |
| P14-3.75,30-3 |
| P16-3.75,1 |
| P16-3.75,2 |
| P16-3.75,3 |
| P16-3.75,4 |
| P16-3.75,5 |
| P64-3,1 |
| P64-3,17-0.5 |
| P64-3,17-2 |
| P64-3,2 |
| P64-3,3 |
| P64-3,30-0.5 |
| P64-3,30-2 |
| P64-3.75,1 |
| P64-3.75,2 |
| P64-3.75,3 |
| P64-3.75,4 |
| P64-3.75,5 |
| P64-3.75,17-0.5 |
| P64-3.75,17-2 |
| P64-3.75,17-3 |
| P64-3.75,30-0.5 |
| P64-3.75,30-2 |
| P64-3.75.30-3 |
#### Indications for Use
AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are dental prosthetics placed onto a dental implant or screw retained abutment, to provide support for dental prosthetic restorations. The Base Abutments and Adhesive Sleeves include: Titanium bases/sleeves with a pre-machined implant connection or screw retained abutment connection on which the CAD/CAM designed upper structure may be fitted to complete a two-piece dental abutment. The Base Abutments and Adhesive Sleeves include an abutment/sleeve screw for fixation to the underlying implant/screw retained abutment.
{5}------------------------------------------------
# 510(K) Summary Page 3 of 5
These dental prosthetics may be used for single-unit (bingle-unit (bridges and bars) restorations and are compatible for use with AB Dental Implants with internal connection and screw retained connections for standard and narrow platforms.
All digitally designed abutments and/or copings for use with AB Dental CAD/CAM Base Abutments and Adhesive Sleeves are intended to be sent to AB Dental validated milling center for manufacture.
#### Technological Characteristics Comparison
| | Subject Device | Predicate |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Base Abutment/ Adhesive Sleeves | Ti-Base Abutment 2-CONnect Abutment |
| 510(k) number | K202144 | K181580 |
| Manufacture | A.B. Dental Devices Ltd. | Cortex Dental Implants Industries Ltd. |
| Classification | 872.3630 | 872.3630 |
| Indications for<br>Use | AB Dental CAD/CAM Base Abutments and<br>Adhesive Sleeves are dental prosthetics placed<br>onto a dental implant or screw retained<br>abutment, to provide support for dental<br>prosthetic restorations. The Base Abutments<br>and Adhesive Sleeves include: Titanium<br>bases/sleeves with a pre-machined implant<br>connection or screw retained abutment<br>connection on which the CAD/CAM designed<br>upper structure may be fitted to complete a<br>two-piece dental abutment. The Base<br>Abutments and Adhesive Sleeves include an<br>abutment/sleeve screw for fixation to the<br>underlying implant/screw retained abutment.<br>These dental prosthetics may be used for single-<br>unit (single-tooth) or multiple-unit (bridges and<br>bars) restorations and are compatible for use<br>with AB Dental Implants with internal<br>connection and screw retained connections for<br>standard and narrow platforms.<br>All digitally designed abutments and/or copings<br>for use with AB Dental CAD/CAM Base<br>Abutments and Adhesive Sleeves are intended<br>to be sent to AB Dental validated milling center<br>for manufacture. | Cortex CAD/CAM Abutments are dental<br>abutments placed onto a dental implant to<br>provide support for dental prosthetic<br>restorations. The abutments include: 1) Titanium<br>abutment blanks with a pre-machined implant<br>connection where the upper portion may be<br>custom- milled in accordance with a patient-<br>specific design using CAD/CAM techniques; and<br>2) Titanium bases with a pre-machined implant<br>connection upon which a CAD/CAM designed<br>superstructure may be fitted to complete a two-<br>piece dental abutment. The abutments include<br>an abutment screw for fixation to the underlying<br>implant. The abutments may be used for single-<br>unit (single-tooth) or multiple-unit (bridges and<br>bars) restorations and are compatible for use<br>with Cortex Dental Implants with internal hex-<br>connection mating platform and conical<br>connection platform diameters.<br>All digitally designed abutments and/or copings<br>for use with Cortex CAD/CAM Abutments are<br>intended to be sent to a Cortex-validated milling<br>center for manufacture. Cortex abutments<br>designed using CAD/CAM techniques must fulfill<br>the Cortex allowable range of design parameters. |
The indications for the subject device and the indications cleared in K181580 are provided above.
The difference between the subject device and the predicate is that the highlighted sections were removed as this submission doesn't include the Titanium abutment blanks.
In addition, in the predicate devices the description of the Ti-Base refers also to screw retained sleeves (they refer to them as abutment or 2-component abutment).
The language was also adjusted to match AB Dental products throughout the text.
The sleeves are intended to be used with devices that are angled at 0°, 17°, and 30°. No additional angulation provided by these sleeves or the CAD/CAM ceramic restorations.
The purpose of this traditional 510(k) is to expand the current product line to include the new abutments listed below. No substantive changes are being made to the indications. Comparison tables for each modification are provided below.
#### 1. Anti-Rotational Base Abutments (P3,TI)
| Spec. | Subject P3-X,TI,XX ; P3H-X,TIT; (X=3.75<br>or 3) (XX=Different Lengths) | Predicate<br>CORTEX |
|-------------|-------------------------------------------------------------------------|---------------------|
| 510k number | K202144 | K181580 |
{6}------------------------------------------------
| Spec. | Subject P3-X,TI,XX ; P3H-X,TIT; (X=3.75<br>or 3) (XX=Different Lengths) | Predicate<br>CORTEX |
|----------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Diagram | Image: dental implant | Image: dental implant |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform<br>(Internal implant<br>connection) | Standard Platform,<br>Narrow platform | Internal Hex,<br>Conical Connection |
| Diameter, mm | 4.3, 5.2<br>4.3, 5.2 | 5.2 |
| Overall Length, mm | 6.45 ,7.85 ,8.85, 7.05<br>6.8, 8.0, 9.0, 7.3 | Internal Hex<br>Collar Height:1, 3 Body Height: 4<br>Conical Connection<br>Collar Height:1, 2, 3 Body Height: 4 |
| Angulation | Minimum and Maximum abutment<br>angle (°): Straight structure with 0°<br>angle | Minimum and Maximum abutment angle (°):<br>Straight structure with 0° angle |
| Design | Anti-rotation | Anti-rotation |
| Zirconia | FDA Cleared Zirconia<br>To be manufactured by validated milling<br>center | FDA Cleared Zirconia<br>To be manufactured by validated milling center |
| Sterility | End user sterilized | End user sterilized |
#### Rotational Base Abutment (P3, TIT) 2.
| Spec. | Subject P3N-X,TIT,XX ; P3-X,TIT;<br>(X=3.75 or 3) (XX=Different Lengths) | Predicate<br>CORTEX |
|----------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Picture | Image: Subject | Image: Cortex |
| 510k number | Subject | K181580 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform<br>(Internal implant<br>connection) | Standard Platform,<br>Narrow platform | Internal Hex,<br>Conical Connection |
| Diameter, mm | 4.3, 5.2<br>4.1, 5.2 | 4.6 |
| Overall Length, mm | 6.85 ,7.29 ,8.29, 6.85 | Internal Hex<br>Collar Height:1, 3 Body Height: 4<br>Conical Connection<br>Collar Height:1, 2, 3 Body Height: 4 |
| Angulation | Minimum and Maximum abutment<br>angle (°): Straight structure with 0°<br>angle | Minimum and Maximum abutment angle (°):<br>Straight structure with 0° angle |
| Design | Rotational (Non engaging) | Rotational (Non engaging) |
| Zirconia | FDA Cleared Zirconia<br>To be manufactured by validated<br>milling center | FDA Cleared Zirconia<br>To be manufactured by validated milling center |
| Sterility | End user sterilized | End user sterilized |
{7}------------------------------------------------
#### 3. Adhesive Sleeves - Cement Retained, Non-Angled Sleeves (P64, P14)
| Spec. | Subject: P64 and P14 Adhesive Sleeves | Predicate CORTEX (refer Multi Unit Ti-Base) |
|--------------------------------------------|----------------------------------------|-----------------------------------------------------------------------|
| Diagram | Image: (P14-bTs)<br>Image: (P64-bTs-C) | Image: Diagram of body and collar height<br>Image: Two Cortex sleeves |
| 510k number | Subject | K181580 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Platform<br>(screw retained<br>connection) | P64<br>P14 | Multiunit<br>2-CONnect |
| Diameter, mm | 4.9<br>4.4 | 5 |
| Length, mm | 5.5 | Collar Height:0.3, 0.7<br>Body Height: 4 |
| Design | Sleeve | Sleeve |
| Sterility | End user sterilized | End user sterilized |
#### Performance Data
Non-clinical data submitted to demonstrate substantial equivalence included:
- Engineering analysis to determine if additional components constitute a new mechanical worst-case. .
- Cytotoxicity testing per ISO 10993-5 (Nelson Lab. Study Number 1502319-S01). ●
- . Fatigue testing at 40° per ISO 14801 on the two worst case constructs of the subject submission.
- . Sterilization validation according to ANSI/AAMI/ISO 17665 (Sterilization of health care products- Moist heat- Part 1 and 2, ISO 11138- Sterilization of health care products- Biological indicators – Part 1, 3 and 7)
- . The conducted MRI testing conducted per standards listed hereunder demonstrates that the claimed products are MR Conditional:
- ASTM F2052-21 (Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment)
- ASTM F2213-17 (Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment)
- ASTM F2182-19 (Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging)
- ASTM F2119-2013 (Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants) A patient with this device can be scanned safely in an MR system under the conditions as stated in the labeling (Recommended MRI labeling based on the document; American Society of Testing and Materials, International, ASTM F2503-20. Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR)
Environment, Guidance for Industry and Food and Drug Administration Staff, Document issued on May 20, 2021.) No clinical data were included in this submission.
### Conclusions
Based on the data included in this submission, A.B. Dental Implants System is substantially equivalent in indication and design principles to the predicate device indicated above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.