← Product Code [EZL](/productcode/EZL) · K202134

# Well Lead All Silicone Foley Catheter with Temperature Sensor (K202134)

_Well Lead Medical Co., Ltd. · EZL · Apr 8, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202134

## Device Facts

- **Applicant:** Well Lead Medical Co., Ltd.
- **Product Code:** [EZL](/productcode/EZL.md)
- **Decision Date:** Apr 8, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

All Silicone Foley Catheter with Temperature Sensor is intended for use in the drainage/collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgical or post-surgical intervals.

## Device Story

Device is a medical-grade silicone Foley catheter; includes shaft, drainage funnel, inflation funnel, balloon, valve, and X-ray opaque line; integrates temperature sensor for core body temperature monitoring. Used in clinical settings (surgical/post-surgical) by healthcare professionals. Input is physiological thermal data; output is temperature reading via external monitor. Device facilitates bladder drainage while providing continuous temperature monitoring to assist clinical decision-making during procedures. Provided sterile (ethylene oxide), single-use. Subject device is MR Conditional, differing from predicate's 'Safety in MRI Not Evaluated' status.

## Clinical Evidence

Bench testing only. Testing confirmed conformance to ASTM F2052-15 (displacement force), ASTM F2119-07 (image artifacts), ASTM F2182-19 (RF-induced heating), ASTM F2213-17 (torque), and ASTM F2503-13 (marking).

## Technological Characteristics

Medical grade silicone construction; includes integrated temperature sensor; X-ray opaque line; color-coded by size; ethylene oxide sterilization. MR Conditional per ASTM F2052-15, F2119-07, F2182-19, F2213-17, and F2503-13.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Well Lead All Silicone Foley Catheter with Temperature Sensor ([K091516](/device/K091516.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services logo. The FDA logo is in blue and features the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. The Department of Health & Human Services logo is on the left and features a stylized human figure.

April 8, 2021

Well Lead Medical CO., LTD. Jenny Zhu RA Specialist C-4# Jinhu Industrial Estate, Hualong, Panyu Guangzhou. Guangdong 511434 China

Re: K202134

> Trade/Device Name: All Silicone Foley Catheter with Temperature Sensor Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: March 5. 2021 Received: March 11, 2021

Dear Jenny Zhu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Sharon M. Andrews -S

For Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202134

### Device Name

All Silicone Foley Catheter with Temperature Sensor

Indications for Use (Describe)

All Silicone Foley Catheter With Temperature Sensor is intended for use in the drainage/collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgical intervals.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☒</span> Prescription Use (21 CFR 201.2) | ☐ Over-The-Counter Use (21 CFR 201.2) |
|-----------------------------------------------------------------------------------|---------------------------------------|
|-----------------------------------------------------------------------------------|---------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary – K202134

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

#### 1. Submitter Information

| Date Prepared:     | April 6, 2021                                                                                                                                        |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:         | WELL LEAD MEDICAL CO., LTD.<br>Address: C-4 # Jinhu Industrial Estate, Hualong, Panyu, Guangzhou<br>511434, P.R. China<br>Jenny Zhu<br>RA Specialist |
| Contact Person:    | WELL LEAD MEDICAL CO., LTD.<br>Email: jenny_zhu@welllead.com.cn<br>Tel: +86-20-84758878                                                              |
| Device Name:       | All Silicone Foley Catheter with Temperature Sensor                                                                                                  |
| Common Name:       | Foley Catheter                                                                                                                                       |
| Regulation Number: | 21 CFR 876.5130                                                                                                                                      |
| Regulation Name:   | Urological catheter and accessories                                                                                                                  |

{4}------------------------------------------------

| Product Code:        | EZL (catheter, retention type, balloon)                               |
|----------------------|-----------------------------------------------------------------------|
| Regulatory Class:    | Class II                                                              |
| Predicate Device(s): | K091516-Well Lead All Silicone Foley Catheter with Temperature Sensor |

# 2. Intended Use

All Silicone Foley Catheter with Temperature Sensor is intended for use in the drainage/collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgical or post-surgical intervals.

# 3. Device Description

The All Silicone Foley Catheter with Temperature Sensor is made from medical grade silicone, consists of temperature sensor for monitoring core body temperature and Foley catheter including a shaft, drainage funnel, inflation funnel, balloon, valve, and X-ray opaque line. The device is provided sterile by ethylene oxide and is for single use only. It is provided in a variety of sizes and color-coded by size.

# 4. Substantial Equivalence-Comparison to Predicate Device

The All Silicone Foley Catheter with Temperature Sensor has the same intended use as the predicate device. It is intended for use in the drainage/collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgical intervals. The subject device uses the same design, materials, and manufacturing technique as the predicate device; and therefore, the subject and predicate device have technological characteristics.

The All Silicone Foley Catheter with Temperature Sensor differs from the predicate device in MR compatibility. The predicate device is labeled "Safety in MRI Not Evaluated" and the subject device is labeled "MR Conditional."

#### 5. Summary of Non-Clinical Testing

Bench-top testing was conducted to assure conformance to the following standards:

{5}------------------------------------------------

 ASTM F2052-15, Standard Test Method for Measurement of Magnetically Induced . Displacement Force on Medical Devices in the Magnetic Resonance Environment

♦ ASTM F2119-07 (Reapproved 2013), Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants

♦ ASTM F2182-19, Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging

• ASTM F2213-17, Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment

ASTM F2503-13, Standard Practice for Marking Medical Devices and Other Items for Safety � in the Magnetic Resonance Environment

The performance data support that the subject device can be labeled "MR Conditional."

# 6. Conclusion

The All Silicone Foley Catheter with Temperature Sensor is substantially equivalent to predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K202134](https://fda-staging.innolitics.com/device/K202134)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
