K202119 · Weipeng (Suzhou) Medical Devices Co., Ltd. · IYO · Oct 7, 2020 · Radiology
Device Facts
Record ID
K202119
Device Name
AcuSee AS-P1000 System
Applicant
Weipeng (Suzhou) Medical Devices Co., Ltd.
Product Code
IYO · Radiology
Decision Date
Oct 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization. The AcuSee AS-P1000 system is intended to be used in a clinical setting.
Device Story
AcuSee AS-P1000 is an ultrasound-guided instrument tracking system. It uses stereo-vision optical detection to track markers on mounts attached to an ultrasound probe and a needle-like instrument. The system processes these inputs to calculate the relative position and trajectory of the needle in real-time. It outputs an overlay of the instrument's position and projected path onto the B-scan ultrasound image on a display. Used in clinical settings by physicians during ultrasound-guided procedures. The system is mounted on a movable cart and acts as an accessory to existing ultrasound machines. By providing real-time visual guidance, it assists the clinician in navigating the needle to the target, potentially improving procedural accuracy and patient outcomes.
Clinical Evidence
No clinical data. Bench testing only, including software verification/validation, system verification/validation, and compliance testing for electrical and electromagnetic safety (IEC 60601-1:2005+A1:2012, IEC 60601-1-2:2014).
Technological Characteristics
Optical tracking system consisting of a camera, PC, and tracking mounts. Uses stereo-vision optical detection to track markers on ultrasound probes and needles. Real-time processing of spatial geometry. Non-sterile, cart-mounted system. Complies with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for augmenting ultrasonic images of interventional needles or rigid needle-like devices (e.g., biopsy, aspiration, ablation needles) and predicting their future path on a display alongside B-scan ultrasound images. Intended for use in clinical settings where ultrasound visualization is required.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 7, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Weipeng (Suzhou) Medical Devices Co., Ltd. % Geetha Rao, Ph.D. Regulatory and Quality Representative Springborne Life Sciences 750 Menlo Avenue, Suite 200 MENLO PARK CA 94025
## Re: K202119
Trade/Device Name: AcuSee AS-P1000 System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO Dated: July 24, 2020 Received: July 30, 2020
Dear Dr. Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202119
Device Name AcuSee AS-P1000 System
#### Indications for Use (Describe)
The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization.
The AcuSee AS-P1000 system is intended to be used in a clinical setting.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"><b>X</b></div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"></div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
#### Submitter
| Name | Weipeng (Suzhou) Medical Devices Co., Ltd.<br>Unit 207, Building B2 |
|----------------|----------------------------------------------------------------------|
| Address | 218 Xinghu St., Suzhou Industrial Park<br>Suzhou, 215123, P.R. China |
| Phone Number | +86 0512-67311267 |
| Fax Number | NA |
| Contact Person | Bin Yang |
| Email | binyang@vibronixinc.com |
| Phone Number | +86 0512-67311267 |
| Fax Number | NA |
| Date Prepared | July 15, 2020 |
### Device Information
| Trade Name | AcuSee AS-P1000 System |
|--------------|--------------------------------------------------------|
| Common Name | Tracking or Guidance System |
| Product Code | IYO, Ultrasonic pulsed echo imaging system |
| Regulation | 21 CFR 892.1560, Ultrasonic pulsed echo imaging system |
| Device Class | II |
#### Predicate Information
| Device Name | Clear Guide ONE |
|---------------|-----------------|
| 510(k) Number | K141806 |
#### Device Overview
The AcuSee AS-P1000 system is a medical device that tracks instrument positioning during ultrasoundguided procedures to provide instrument guidance to the end user. With accessories attached to the transducer probe and needle or needle-like rigid device, the AcuSee AS-P1000 system identifies and tracks the probe and the device within the field of view through optical detection technology. By coupling positioning information with the ultrasound image, the projected instrument pathway is displayed to the user for guiding instrumentation. The AcuSee AS-P1000 system is an accessory to any compatible ultrasound machine and transducer probe.
#### Indications for Use
The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle, and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization.
{4}------------------------------------------------
# VibroniX
The AcuSee AS-P1000 system is intended to be used in a clinical setting.
### Technological Characteristics
The AcuSee AS-P1000 System applies optical tracking technology to augment the ultrasonic image of an interventional needle or needle-like rigid device and for predicting its future path on a display and also show the image of a B-scan (or similar display) of a medical ultrasound imaging system. The relative position and trajectory of the needle is calculated using stereo-vision optical detection of markers on tracking mounts attached to the ultrasound probe and needle device.
| Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| 510(k) Number | NA | K141806 | NA |
| Product Code | IYO | IYO | Same |
| Regulation No. | 892.1560 | 892.1560 | Same |
| Risk Class | II | II | Same |
| Regulatory<br>Pathway | 510(k) | 510(k) | Same |
| Indications for<br>Use | The AcuSee AS-P1000 System<br>is indicated for augmenting the<br>ultrasonic image of an<br>interventional needle or<br>needle-like rigid device, such as<br>a biopsy needle, an aspiration<br>needle, or ablation needle, and<br>for predicting its future path<br>on a display, which also shows<br>the image of a B-scan (or<br>similar display) of a medical<br>ultrasound imaging system.<br>The device is intended to be<br>used in procedures where<br>ultrasound is currently used for<br>visualization. | The Clear Guide ONE is<br>indicated for augmenting the<br>ultrasonic image of an<br>interventional needle or<br>needle-like rigid device, such as<br>a biopsy needle, an aspiration<br>needle, or ablation needle, and<br>for predicting its future path<br>on a display, which also shows<br>the image of a B-scan (or<br>similar display) of a medical<br>ultrasound imaging system.<br>The device is intended to be<br>used in procedures where<br>ultrasound is currently used for<br>visualization. | Identical |
| Intended Users | Physician | Physician | Same |
| Use<br>Environment | Clinical Setting | Clinical Setting | Same |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Movability | Main system mounted on<br>movable cart | Main system mounted on<br>movable cart | Same |
| Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment |
| Fundamental<br>Technology | Optical Detection | Optical Detection | Same |
| Device Form<br>Factor | Main system contains camera,<br>and PC with software on<br>movable cart, with power cord<br>for AC power,<br>Separate tracking accessories<br>consist of tracking mounts | Main system (Clear Guide<br>Core) contains hardware and<br>software with power adapter<br>and cord,<br>Separate tracking accessories<br>include camera (Clear Guide<br>SuperProbe) and custom<br>instruments | Similar primary<br>concept with different<br>physical configuration |
| Detection<br>Mechanism | Camera located at a distance<br>tracks custom tracking mounts<br>on ultrasound probe and<br>needle | Camera is directly mounted on<br>ultrasound probe and tracks<br>custom needle | Similar concept, with<br>a variant spatial<br>geometry |
| Tracking<br>Algorithm | Automatic | Automatic | Similar |
| Timing for<br>tracking | Real-time | Real-Time | Same |
| Information<br>Displayed | Overlay of the detected<br>instrument positioning and<br>projected path onto the<br>ultrasound image | Overlay of the detected<br>instrument positioning and<br>projected path onto the<br>ultrasound image | Same |
| Reusability | Reusable main system with<br>consumable tracking mounts | Reusable main system and<br>probe with consumable<br>tracking needle | Similar |
| Instruments<br>tracked | Any compatible needle-like<br>instrument | Requires custom needle (not<br>known what kind of needle) | Similar |
| Tracking of<br>instrument tip | Any compatible instrument | Only with custom instrument<br>with CLEAR GUIDE PercepTIP | Similar |
| Compatibility<br>with probe | Tracking mount fits any<br>compatible probe | Optical head with camera fits<br>any compatible probe | Same functionality |
| Tracking<br>instruments in<br>and out-of-plane | Within +/- 90 deg | Within +/- 45 deg | Subject device has<br>wider performance<br>range |
| Feature | Subject Device<br>AcuSee AS-P1000 System | Predicate Device<br>Clear Guide One | SE Comparison/<br>Comment |
| Electrical Safety<br>and<br>Electromagnetic<br>Compatibility | IEC 60601-1:2005+A1:2012 -<br>Medical electrical equipment –<br>Part 1: General requirements<br>for basic safety and essential<br>performance<br>IEC 60601-1-2:2014 - Medical<br>electrical equipment - Part 1-2:<br>General requirements for basic<br>safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests. | IEC 60601-1 3rd Edition:<br>Medical Electrical Equipment;<br>Part 1 General Requirements<br>for Safety<br>IEC 60601-1-2 :2007 – Medical<br>electrical equipment - Part 1-2:<br>General requirements for<br>safety – Collateral standard:<br>Electromagnetic compatibility | Similar - Subject<br>device meets updated<br>standard |
| Performance<br>Testing | Bench testing | Bench testing | Similar |
| Animal Testing | None | None | Same |
| Clinical Testing | None | None | Same |
### Device Comparison Table
{5}------------------------------------------------
# VibroniX
{6}------------------------------------------------
### Substantial Equivalence Overview
The AcuSee AS-P1000 System has the same intended use and indications for use, as well as similar technological characteristics and principles of operation compared to the previously legally cleared predicate device: Clear Guide ONE (K141806, cleared in 2014).
The AcuSee AS-P1000 System and the predicate device both apply the same optical detection technology in a similar device form factor. The AcuSee AS-P1000 system overlays the detected instrument positioning data onto an existing ultrasound image through proprietary software algorithms, In the same manner as the predicate device. Both devices track targeted instrumentation and provide real-time visual guidance during a clinical procedure. Both devices have reusable main components and some consumable accessories. The AcuSee AS-P1000 System was tested to demonstrate compliance to the same applicable safety standards as the predicate device, but for updated versions of the standards.
The AcuSee AS-P1000 System uses the same primary concept for the detection mechanism but uses a different physical configuration for the camera and tracked targets, resulting in a different spatial geometry. This variation does not alter the fundamental mechanism and raises no new questions of safety. Performance data was collected to demonstrate that the AcuSee AS-P1000 system achieves its intended function in a manner that is as effective as the predicate device.
The AcuSee AS-P1000 system uses tracking mounts that are designed to fit ultrasound probes and instruments that are available on the market. Although specific models of probes (and the associated ultrasound machines with which they are used) and instruments are individually validated to be compatible, the mounts themselves are standardized and may be used with any compatible probe or instrument. Hence unlike the predicate device, the system does not require special patterns to be marked or printed to the shaft of the needle or other tracked instrument.
{7}------------------------------------------------
# VibroniX
Although there are minor differences in technological characteristics of the AcuSee AS-P1000 System, they do not raise new issues of safety or effectiveness.
## Summary of Performance Testing
The performance testing for the AcuSee AS-P1000 System includes software verification/validation testing, system verification/validation testing and testing to compliance standards for electrical and electromagnetic safety. Traceability has been documented between the system specification to verification/validation protocols. The results of performance testing show that the AcuSee AS-P1000 system is as safe and as effective as the predicate device.
The AcuSee AS-P1000 system complies with the following recognized consensus standards:
- 1. IEC 60601-1:2005+A1:2012 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
- 2. IEC 60601-1-2:2014 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests).
# Conclusion
The performance of the AcuSee AS-P1000 System is substantially equivalent to that of the Clear Guide ONE system and raises no safety or effectiveness issues and performs as well or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.