← Product Code [FXX](/productcode/FXX) · K202096

# 3-ply EcoGuard B with Earloop, 3-ply EcoGuard B with Tie-On (K202096)

_Ecoguard, Inc. · FXX · Feb 2, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202096

## Device Facts

- **Applicant:** Ecoguard, Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Feb 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Following EcoGuard Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and bodily fluid. These surgical masks are single use, disposable devices provided non-sterile.

## Device Story

3-ply surgical mask; pleated design; available in earloop (ECO01) or tie-on (ECO02) configurations. Composed of inner/outer spunbond polypropylene layers and middle melt-blown polypropylene filter layer. Includes malleable polyethylene wire nose piece for fit. Single-use; disposable; non-sterile. Worn by healthcare personnel and patients to reduce exposure to blood, bodily fluids, and particulates. Provides physical barrier protection in clinical/infection control environments.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified per ASTM F2100-11 Level 3: Synthetic blood penetration (ASTM F1862, 160 mmHg), Bacterial Filtration Efficiency (ASTM F2101, 99.9%), Particulate Filtration Efficiency (ASTM F2299, 99.8%), Differential Pressure (Mil-M-36954C, 4.4 mm H2O/cm2), and Flammability (16 CFR 1610, Class 1). Biocompatibility testing (cytotoxicity, irritation, sensitization) passed.

## Technological Characteristics

3-ply construction; spunbond polypropylene (inner/outer), melt-blown polypropylene (filter). Malleable polyethylene wire nose piece. Ear loops (polyester/nylon/spandex) or polypropylene ties. ASTM F2100-11 Level 3 compliant. Non-sterile. No software or energy source.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Surgical Face Mask ([K153496](/device/K153496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

February 2, 2021

EcoGuard Inc. % Nickita Alexiades Consultant mdi Consultants. Inc. 55 Northern Blvd. Suite 200 Great Neck, New York 11021

Re: K202096

Trade/Device Name: 3-ply EcoGuard B with Earloop, 3-ply EcoGuard B with Tie-On Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 5, 2021 Received: January 13, 2021

Dear Nickita Alexiades:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202096

Device Name

3-ply EcoGuard B with Earloop; 3-ply EcoGuard B with Tie-on

Indications for Use (Describe)

The Following EcoGuard Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and bodily fluid. These surgical masks are single use, disposable devices provided non-sterile.

3-ply EcoGuard B with Earloop, Model Number: ECO01 3-ply EcoGuard B with Tie-on, Model Number: ECO02

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

The assigned 510(k) number is: K202096.

#### 1. Submitter's Identification:

Manufacturer: EcoGuard Inc. Address: 700 S BATTLEGROUND AVE, SUITES 103 GROVER, NC 28073

Contact Person: Ms. Fang Wang Ecoguard Inc.

Date Summary Prepared: January 29, 2021

Official Correspondent: Mr. Nickita Alexiades mdi Consultants, Inc.

#### 2. Name of the Device:

| Device Name(s):  | 3-ply EcoGuard B with Earloop, | Model Number: ECO01 |
|------------------|--------------------------------|---------------------|
| Model Number(s): | 3-ply EcoGuard B with Tie-on,  | Model Number: ECO02 |

Common Name: Surgical Mask Regulation Number: 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX

#### 3. Information for the 510(k) Cleared Device (Predicate Device):

Predicate Device: Disposable Surgical Face Mask K153496 Xiantao Rayxin Medical Products Co., Ltd.

#### 4. Device Description:

The proposed surgical masks are available in two models with three-ply pleated earloop or tie-on (EcoGuard B) masks with ear loops and nose piece. Detail configurations of them are presented in Table I: Surgical Masks Description.

| Product Model Number: | Product Model                 | ASTM Level |
|-----------------------|-------------------------------|------------|
| ECO01                 | 3-Ply EcoGuard B with Earloop | Level 3    |
| ECO02                 | 3-Ply EcoGuard B with Tie-On  | Level 3    |

## Table 1: Surqical Masks Description

The inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops or ties are held in place over the users' mouth and nose by two elastic ear loops or four ties welded to the surgical mask. Neither the elastic ear loops, nor ties are made with natural rubber latex. The nose piece in the layers of surgical mask is to allow the user to fit the surgical 

{4}------------------------------------------------

mask around their nose, which is made of malleable polyethylene wire. The surgical masks are sold nonsterile and are intended to be single use, disposable devices.

#### 5. Indications for Use:

The following EcoGuard Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from the transfer of microorqanisms, body fluids, and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and bodily fluid. These surgical masks are single use, disposable devices provided non-sterile.

3-ply EcoGuard B with Earloop, Model Number: ECO01 3-ply EcoGuard B with Tie-on, Model Number: ECO02

#### 6. Comparison to the 510(k) Cleared Device (Predicate Device):

| Item                        | Subject Device                                                                                                                                                                                                                                                                                                                                                                | Predicate Device                                                                                                                                                                                                                                                                                                                                                                        | Similar or Different |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacturer                | EcoGuard Inc.                                                                                                                                                                                                                                                                                                                                                                 | Xiantao Rayxin Medical Products Co., Ltd.                                                                                                                                                                                                                                                                                                                                               | -                    |
| 510K number                 | K202096                                                                                                                                                                                                                                                                                                                                                                       | K153496                                                                                                                                                                                                                                                                                                                                                                                 | -                    |
| Model Name                  | 3-ply EcoGuard B with Earloop<br>Model Number:<br>ECO01<br><br>3-ply EcoGuard B with Tie-on<br>Model Number:<br>ECO02                                                                                                                                                                                                                                                         | Disposable Surgical Face Mask                                                                                                                                                                                                                                                                                                                                                           | Similar              |
| Classification              | Class II Device<br>Product Code: FXX<br>(21 CFR878.4040)                                                                                                                                                                                                                                                                                                                      | Class II Device<br>Product Code: FXX<br>(21 CFR878.4040)                                                                                                                                                                                                                                                                                                                                | Similar              |
| Indications for Use         | The Following EcoGuard Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These surgical masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. These surgical masks are single use, disposable | The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided nonsterile. | Similar              |
|                             | devices provided<br>non-sterile.                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                         |                      |
| Model(s) &<br>Style         | EcoGuard B<br>Ear Loops<br>Flat Pleated, 3 layers<br>ECO01<br>Tie-On<br>Flat Pleated, 3 layers<br>ECO02                                                                                                                                                                                                                                                                       | Ear Loops<br>Tie-On<br>Flat Pleated<br>3 layers                                                                                                                                                                                                                                                                                                                                         | Similar              |
| Outer Facing<br>Layer       | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                    | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                              | Similar              |
| Middle Layer                | Melt blown<br>polypropylene filter                                                                                                                                                                                                                                                                                                                                            | Melt blown<br>polypropylene filter                                                                                                                                                                                                                                                                                                                                                      | Similar              |
| Inner Facing<br>Layer       | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                    | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                                                                                                              | Similar              |
| Nose Piece                  | Malleable<br>Polyethylene wire                                                                                                                                                                                                                                                                                                                                                | Malleable Aluminum<br>wire                                                                                                                                                                                                                                                                                                                                                              | Different            |
| Ear Loops or<br>Tie-On      | Earloops:<br>Polyester<br>Nylon/Spandex<br>Tie-On:<br>Polypropylene                                                                                                                                                                                                                                                                                                           | Earloops:<br>Polyester                                                                                                                                                                                                                                                                                                                                                                  | Similar              |
| Color                       | Blue                                                                                                                                                                                                                                                                                                                                                                          | Blue                                                                                                                                                                                                                                                                                                                                                                                    | Similar              |
| Dimension<br>(Width)        | $17.5cm±2cm$<br>$6.89" ± 0.40"$                                                                                                                                                                                                                                                                                                                                               | $17.5cm±1cm$                                                                                                                                                                                                                                                                                                                                                                            | Similar              |
| Dimension<br>(Length)       | $9.5cm±2cm$<br>$3.74" ± 0.40"$                                                                                                                                                                                                                                                                                                                                                | $9.5cm±1cm$                                                                                                                                                                                                                                                                                                                                                                             | Similar              |
| OTC use                     | Yes                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                     | Similar              |
| Sterility                   | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                   | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                             | Similar              |
| Use                         | Single Use,<br>Disposable                                                                                                                                                                                                                                                                                                                                                     | Single Use,<br>Disposable                                                                                                                                                                                                                                                                                                                                                               | Similar              |
| ASTM F2100<br>Level         | Level 3                                                                                                                                                                                                                                                                                                                                                                       | Level 2                                                                                                                                                                                                                                                                                                                                                                                 | Similar              |
| Non-Clinical<br>Testing     | ASTM F1862<br>ASTM F2299<br>ASTM F2101<br>ASTM F2100<br>MIL-M369454C<br>16 CFR 1610                                                                                                                                                                                                                                                                                           | ASTM F1862<br>ASTM F2299<br>ASTM F2101<br>ASTM F2100<br>MIL-M369454C<br>16 CFR 1610                                                                                                                                                                                                                                                                                                     | Similar              |
| Biocompatibility<br>Testing | Cytotoxicity<br>Irritation<br>Sensitization                                                                                                                                                                                                                                                                                                                                   | Cytotoxicity<br>Irritation<br>Sensitization                                                                                                                                                                                                                                                                                                                                             | Similar              |

Table 2: Comparison to Predicate Device

{5}------------------------------------------------

The difference in the materials does not raise additional questions for safety and effectiveness. Performance testing including biocompatibility evaluation has been performed on the final finished device

### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

{6}------------------------------------------------

The EcoGuard Surgical Masks have been tested accordance with ASTM 2100-11 Standard Specification for Performance of Materials Used in Medical Face Masks. See Table 2 for a summary of other non-clinical testing.

| Test                               | Purpose                                                                | Acceptance<br>Criteria per ASTM<br>F2100-11 Level 3<br>(AQL = 4.0%) | Subject Device Test Results-<br>EcoGuard B                                        |                   |
|------------------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------|
|                                    |                                                                        |                                                                     | ASTM F2100-11<br>Level 3                                                          | Average           |
| ASTM F1862<br>Synthetic<br>Blood   | Determine<br>synthetic<br>blood<br>penetration<br>resistance           | Level III                                                           | Pass at 160<br>mmHg (95/96)<br>32 Samples<br>each from 3 non-<br>consecutive lots | N/A               |
| ASTM F2101<br>BFE                  | Determine<br>the<br>bacterial<br>filtration<br>efficiency              | ≥ 98%                                                               | Pass (96/96)<br>32 Samples<br>each from 3 non-<br>consecutive lots                | 99.9%             |
| ASTM F2299<br>PFE at 0.1<br>micron | Determine<br>submicron<br>particulate<br>filtration<br>efficiency      | ≥ 99%                                                               | Pass (96/96)<br>32 Samples<br>each from 3 non-<br>consecutive lots                | 99.8%             |
| Mil-M-<br>36954C<br>Delta P        | Determine<br>breathing<br>resistance<br>or<br>differential<br>pressure | < 6.0 mm<br>H2O/cm2                                                 | Pass (96/96)<br>32 Samples<br>each from 3 non-<br>consecutive lots                | 4.4 mm<br>H2O/cm2 |
| 16 CFR 1610<br>Flammability        | Determine<br>flammability<br>or flame<br>spread                        | Class 1                                                             | Class 1 Pass<br>(96/96)<br>32 Samples<br>each from 3 non-<br>consecutive lots     | N/A               |

|  | Table 2: Summary of Non-Clinical Testing |  |
|--|------------------------------------------|--|
|  |                                          |  |

All testing results have met ASTM F2100-11 2019 Level 3 acceptance criteria. The testing information demonstrates that the safety and effectiveness of the EcoGuard B Surgical Masks in their intended environment of use is supported by testing that was conducted in accordance with the FDA 2004 guidance "Surgical Masks - Premarket Notification [510(k)] Submissions", which outlines performance requirements for surgical masks.

| Test                        | Testing Findings                                                      | Result           |
|-----------------------------|-----------------------------------------------------------------------|------------------|
| Cytotoxicity<br>(Standard)  | Under the conditions of the<br>study, the device is<br>noncytotoxic.  | EcoGuard B: Pass |
| Irritation<br>(Standard)    | Under the conditions of the<br>study, the device is<br>nonirritating. | EcoGuard B: Pass |
| Sensitization<br>(Standard) | Under the conditions of the<br>study, the device is<br>nonsensitizing | EcoGuard B: Pass |

Table 3: Summary of Biocompatibility testing

{7}------------------------------------------------

None of the biocompatibility testing results demonstrated any safety hazards or any design characteristics that violated the requirements set forth in the Guidance for Industry regarding Premarket Notification Submissions of Surgical Masks. It was our conclusion that both 3-ply EcoGuard B with Earloop and 3-ply EcoGuard B with Tie-on surgical masks met all relevant testing requirements.

#### 8. Discussion of Clinical Tests Performed:

No Clinical testing was performed.

#### 9. Conclusions:

Based on the comparison and analysis above and the nonclinical tests performed, the subject device, EcoGuard B Surgical Masks, are as safe, as effective, and performs as well as the legally marketed predicate device, Xianto Rayxin Medical Products Disposable Surgical Mask cleared under K153496.

---

**Source:** [https://fda-staging.innolitics.com/device/K202096](https://fda-staging.innolitics.com/device/K202096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).