← Product Code [LNH](/productcode/LNH) · K202014

# Magnetom Aera, Magnetom Skyra, Magnetom Prisma/Prisma fit (K202014)

_Siemens Medical Solutions USA, Inc. · LNH · Sep 8, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K202014

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Sep 8, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/ or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, vield information that may assist in diagnosis. Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

## Device Story

MAGNETOM Aera, Skyra, and Prisma/Prismafit are MR diagnostic systems (1.5T/3T) utilizing syngo MR XA30A software. Systems acquire MR signals to produce cross-sectional images/spectra of head, body, or extremities. New features include SVS EDIT (spectral editing), BEAT_FQ_nav (respiratory-gated 4D flow), injector coupling for synchronized contrast delivery, and Prostate Dot Engine for automated workflow/positioning. Modifications include optimized inversion recovery pulses for myocardial imaging and AbsoluteShim mode. Used in clinical settings by trained physicians/technicians. Output (images/spectra) interpreted by physicians to assist in diagnosis. Benefits include improved diagnostic quality, workflow reproducibility, and motion correction.

## Clinical Evidence

No clinical trials were conducted. Substantial equivalence is supported by bench testing, software verification/validation, and sample clinical images. Clinical publications were cited to support the utility of specific features like SVS_EDIT and the Prostate Dot Engine.

## Technological Characteristics

MR diagnostic system (1.5T/3T). Conforms to IEC 60601-1, IEC 60601-2-33, IEC 62304, ISO 14971, and NEMA MS standards. Features include phased array coils, gradient echo/inversion recovery pulse sequences, and automated workflow engines. Connectivity via DICOM. Software lifecycle follows IEC 62304.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- MAGNETOM Vida ([K192924](/device/K192924.md))

## Reference Devices

- MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/Prismafit with software syngo MR VE11C ([K153343](/device/K153343.md))
- syngo.via VB40 ([K192462](/device/K192462.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 8, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Siemens Medical Solutions USA, Inc. % Mr. Andrew Turner Regulatory Affairs Specialist 40 Liberty Boulevard, Mailcode 65-1A MALVERN PA 19355

# Re: K202014

Trade/Device Name: MAGNETOM Aera, MAGNETOM Skyra, Magnetom Prisma/Prismafit with syngo MR XA30A Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: July 17, 2020 Received: July 21, 2020

Dear Mr. Turner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K202014

Device Name

MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma Fit with syngo MR XA30A

#### Indications for Use (Describe)

Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal, and oblique cross sectional images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/ or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, vield information that may assist in diagnosis.

Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern. The background is a light gray color.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92.

- 1. General Information

| Establishment: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Registration Number: 2240869                                                                                                                                                                   |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | July 17, 2020                                                                                                                                                                                                                                                                                              |
| Manufacturer:  | Siemens Healthcare GmbH<br>Henkestrasse 127<br>91052 Erlangen<br>Germany<br>Registration Number: 3002808157<br>Siemens Shenzhen Magnetic Resonance LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number: 3004754211 |

#### 2. Contact Information

Andrew Turner, Regulatory Affairs Technical Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mail Code 65-1A Malvern, PA 19355, USA Phone: (610) 850-5627 Fax: (610) 448-1787 E-mail:

#### Traditional Premarket Notification 510(k)

July 17, 2020

Siemens MR Systems: MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

#### 3. Device Name and Classification

MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Device/Trade name: Prisma/ Prismafit with syngo SW XA30A

| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD) |
|-----------------------|---------------------------------------------|
| Classification Panel: | Radiology                                   |
| CFR Code:             | 21 CFR § 892.1000                           |
| Classification:       | II                                          |
| Product Code:         | Primary: LNH<br>Secondary: LNI. MOS         |

# 4. Legally Marketed Predicate Device

| Trade name:    | MAGNETOM Vida |
|----------------|---------------|
| 510(k) Number: | K192924       |

| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD) |
|-----------------------|---------------------------------------------|
| Classification Panel: | Radiology                                   |
| CFR Code:             | 21 CFR § 892.1000                           |
| Classification:       | II                                          |
| Product Code:         | Primary: LNH<br>Secondary: LNL MOS          |

#### 5. Intended Use

The indications for use for the subject devices are the same as the predicate device:

Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

#### Traditional Premarket Notification 510(k)

Siemens MR Systems: MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

# 6. Device Description

MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ with software syngo MR XA30A, include new and modified hardware and software compared to the predicate device, MAGNETOM Vida with software syngo MR XA20A. A high-level summary of the new and modified hardware and software is provided below:

#### Hardware

- New Computer

#### Software

New Features and Applications

- SVS EDIT is a special variant of the SVS SE pulse sequence type, which acquires two different spectra (one with editing pulses on resonance, one with editing pulses off resonance) within a single sequence.
- BEAT_FQ_nav is a pulse sequence that allows the user to make use of navigator echo based respiratory gating for flow imaging to acquire 4D flow data. Both navigator echo based respiratory gating and flow imaging are cleared features available on the predicate device. However, the combination of the two is new.
- Injector coupling is a software application that allows the connection of । certain contrast agent injectors to the MR system for simplified, synchronized contrast iniection and examination start.
- -The Prostate Dot Engine provides an assisted and guided workflow for prostate imaging. This automated workflow leads to higher reproducibility of slice angulation and coverage based on the segmentation algorithm described and cleared with syngo.via VB40; this may support exams not having to be repeated.

Modified Features and Applications

- i An optimized high bandwidth inversion recovery pulse is combined with gradient echo readout to improve diagnostic image quality when imaging myocardial tissue.
- The AbsoluteShim mode is a shimming procedure based on a 3-echo gradient echo protocol.

# Other Modifications and / or Minor Changes

- Elastography-AddIn synchronizes settings between the Elastography i sequence and the active driver.
- HASTE MoCo is an image-based motion correction function in the averagedimension for the HASTE pulse sequence type.
- Coil independent pulse sequences remove the coil information from the pulse sequences and generate this information during run-time from automatic coil detection and localization.

Siemens MR Systems: MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots in a circular pattern.

# 7. Substantial Equivalence

MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ with software syngo MR XA30A are substantially equivalent to the following predicate device:

| Predicate Device                     | FDA Clearance Number<br>and Date   | Product<br>Code | Manufacturer               |
|--------------------------------------|------------------------------------|-----------------|----------------------------|
| MAGNETOM Vida with<br>syngo MR XA20A | K192924,<br>cleared March 11, 2020 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH |

MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ with software syngo MR XA30A include hardware and software already cleared on the following reference devices:

| Reference Devices                                                                                     | FDA Clearance Number<br>and Date     | Product<br>Code | Manufacturer                            |
|-------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------|-----------------------------------------|
| MAGNETOM Aera,<br>MAGNETOM Skyra and<br>MAGNETOM Prisma/<br>Prismafit with software<br>syngo MR VE11C | K153343,<br>cleared April 15, 2016   | LNH<br>LNI, MOS | Siemens AG / Siemens<br>Healthcare GmbH |
| syngo.via VB40                                                                                        | K192462,<br>Cleared January 31, 2020 | LLZ, LNH        | Siemens Healthcare<br>GmbH              |

#### Technological Characteristics 8.

The subject devices, MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prismall with software syngo MR XA30A, are substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance.

The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards.

As seen above there are some differences in technological characteristics between the subject devices and predicate device, including different hardware and modified software. These differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.

Siemens MR Systems: MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/it (1.5 / 3T) with new Software syngo MR XA30A

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

# 9. Nonclinical Tests

The following performance testing was conducted on the subject devices.

| Performance Test                                                                                                                       | Tested Hardware or Software                                                                                                                        | Source/Rationale for test                                                                    |
|----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Sample clinical images                                                                                                                 | Coils, new and modified software features                                                                                                          |                                                                                              |
| Image quality assessments using sample clinical images. In some cases, a comparison of the image quality / quantitative data was made. | - new / modified pulse sequence types and algorithms.<br>- comparison images between the new / modified features and the predicate device features | Guidance for Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices |
| Performance bench test                                                                                                                 | new and modified hardware                                                                                                                          |                                                                                              |
| Software verification and validation                                                                                                   | new and modified software features                                                                                                                 | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices  |

The results from each set of tests demonstrate that the devices perform as intended and are therefore, substantially equivalent to the predicate device to which it has been compared.

# 10. Clinical Tests / Publications

No clinical tests were conducted to support substantial equivalence for the subject devices; however, as stated above, sample clinical images were provided. Clinical publications were referenced to provide information on the use of the following features and functions.

| Feature / Function  | Clinical Publication                                                                                                                                                                                                                                                                                      |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SVS_EDIT            | [1] Mescher et al, Simultaneous in vivo spectral editing and water<br>suppression, NMR Biomed 11, 266-272 (1998)                                                                                                                                                                                          |
|                     | [2] Mikkelsen et al, Big GABA: Edited MR spectroscopy at 24 research<br>sites, Neurolmage 159, 32-45 (2017)                                                                                                                                                                                               |
|                     | [3] Saleh et al, Multi-vendor standardized sequence for edited magnetic<br>resonance spectroscopy, Neurolmage 189, 425-431 (2019)                                                                                                                                                                         |
| Prostate Dot Engine | [7] Essner M. Zinsser D, Kündel M, et. al. Performance of an Automated<br>Workflow for Magnetic Resonance Imaging of the Prostate: Comparison<br>With a Manual Workflow. Invest Radiol. 2020 May;55(5):277-284. Doi:<br>10.1097.                                                                          |
|                     | [8] Horger W, Thoermer G, Weiland E, et. al. Prostate Dot Engine – a<br>system guided and assisted workflow to improve consistency in prostate<br>MR exams.                                                                                                                                               |
|                     | [9] Yadong C, Siyuan H, Chunmei Li, et. al. Performance and<br>Reproducibility of a Day Optimizing Throughput (Dot) Workflow Engine in<br>Automated Prostate MRI Positioning. Abstract accepted for the 28th<br>annual meeting of the International Society of Magnetic Resonance in<br>Medicine (ISMRM). |

#### Traditional Premarket Notification 510(k)

July 17, 2020

Siemens MR Systems: MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

## 11. Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.

Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.

MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ with software syngo MR XA30A conform to the following FDA recognized and international IEC, ISO and NEMA standards:

| Recognition<br>Number | Product<br>Area | Title of Standard                                                                                                                                                                                       | Reference<br>Number and date               | Standards<br>Development<br>Organization |
|-----------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4                  | General         | Medical electrical equipment -<br>part 1: general requirements for<br>basic safety and essential<br>performance                                                                                         | ES60601-<br>1:2005/(R) 2012<br>and A1:2012 | AAMI / ANSI                              |
| 19-8                  | General         | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests | 60601-1-2 Edition<br>4.0 2014-02           | IEC                                      |
| 12-295                | Radiology       | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis                | 60601-2-33 Ed. 3.2<br>B:2015               | IEC                                      |
| 5-40                  | General         | Medical devices - Application of<br>risk management to medical<br>devices                                                                                                                               | 14971: 2007                                | ISO                                      |
| 5-114                 | General         | Medical devices - Application of<br>usability engineering to medical<br>devices                                                                                                                         | 62366-1:2015                               | AAMI<br>ANSI<br>IEC                      |

#### Traditional Premarket Notification 510(k)

July 17, 2020

Siemens MR Systems: MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern. The background is a light gray color.

| 13-79  | Software             | Medical device software -<br>Software life cycle processes                                                                                | 62304 Edition 1.1<br>2015-06 | AAMI<br>ANSI<br>IEC |
|--------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|---------------------|
| 12-232 | Radiology            | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices                                           | MS 4:2010                    | NEMA                |
| 12-288 | Radiology            | Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images<br>(MRI)                                            | MS 9:2008                    | NEMA                |
| 12-300 | Radiology            | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology                                                  | PS 3.1 - 3.20<br>(2016)      | NEMA                |
| 2-220  | Biocompati<br>bility | biological evaluation of medical<br>devices - part 1: evaluation and<br>testing within a risk<br>management process<br>(Biocompatibility) | 10993-1: 2009                | AAMI<br>ANSI<br>ISO |

# 12. Conclusion as to Substantial Equivalence

MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/11 with software syngo MR XA30A have the same intended use and same basic technological characteristics as the predicate device system, MAGNETOM Vida with syngo MR XA20A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate device, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device and reference devices.

Siemens believes that MAGNETOM Aera, MAGNETOM Skyra and MAGNETOM Prisma/ Prismall with software syngo MR XA30A are substantially equivalent to the currently marketed device MAGNETOM Vida with software syngo MR XA20A (K192924, cleared on March 11, 2020).

#### Traditional Premarket Notification 510(k)

Siemens MR Systems: MAGNETOM Skyra and MAGNETOM Prisma/ Prisma/ Prismafit (1.5 / 3T) with new Software syngo MR XA30A

---

**Source:** [https://fda-staging.innolitics.com/device/K202014](https://fda-staging.innolitics.com/device/K202014)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
