K201945 · Karl Storz Endoscopy America, Inc. · KCT · Feb 25, 2021 · General Hospital
Device Facts
Record ID
K201945
Device Name
KARL STORZ UDEL Sterilization Trays
Applicant
Karl Storz Endoscopy America, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The KARL STORZ UDEL Sterilization Trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in STERRAD and steam sterilization systems as indicated in the attached table. The system may be used with flexible endoscopes with lumen diameters ≥ 1.2mm and a maximum length of 845mm. KARL STORZ UDEL Sterilization Trays must be used in conjunction with an FDA cleared sterilization wrap or container.
Device Story
UDEL Sterilization Trays are rigid, perforated containers designed to organize and protect specific KARL STORZ reusable medical instruments during sterilization. Trays consist of a Udel polysulfone base and lid, often featuring silicone mats or holders to secure instruments. The device is used in clinical settings by healthcare staff. Instruments are placed inside the tray, which is then double-wrapped in an FDA-cleared sterilization wrap to maintain a microbial barrier. The tray design allows for sterilant penetration (steam or STERRAD) through perforations. The device does not maintain sterility independently; it serves as a containment system to facilitate sterilization and protect instruments from damage during handling and processing. Benefits include organized instrument storage and protection of delicate endoscopes and camera heads during the sterilization cycle.
Clinical Evidence
Bench testing only. Efficacy validated for STERRAD (100S, NX, 100NX) and steam (pre-vacuum, gravity) cycles achieving 10^-6 SAL. Cleaning efficacy validated for protein (< 6.4 µg/cm²) and hemoglobin (< 2.2 µg/cm²) removal. Biocompatibility confirmed per ISO 10993. Material integrity verified after 200 STERRAD cycles and 2100 steam cycles.
Technological Characteristics
Materials: Udel polysulfone polymer (base/lid), silicone (holders/mats). Construction: Thermoformed or injection-molded perforated trays with latches. Sterilization: Steam (pre-vacuum/gravity) and STERRAD (100S, NX, 100NX). Standards: AAMI ST77, ISO 14937, ISO 17665-1, AAMI TIR 30, ISO 10993. Connectivity: None (mechanical device).
Indications for Use
Indicated for encasing and protecting specific KARL STORZ reusable medical devices (flexible endoscopes, rigid telescopes, camera heads, light cables) during sterilization in STERRAD and steam systems. Compatible with flexible endoscopes with lumen diameters ≥ 1.2mm and max length 845mm. Must be used with FDA-cleared sterilization wrap.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
KARL STORZ Endoscope Sterilization Trays (K090818)
Symmetry Medical Polyvac Surgical Instrument Delivery System (K012105)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2021
Karl Storz EndoscopyAmerica Inc % David Furr Consultant Toscano Consulting 8708 Capehart Cove Austin, Texas 78733
Re: K201945
Trade/Device Name: KARL STORZ UDEL Sterilization Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: January 26, 2021 Received: January 28, 2021
Dear David Furr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Clarence W. Murray, III, Ph.D Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201945
Device Name KARL STORZ UDEL Sterilization Trays
#### Indications for Use (Describe)
The KARL STORZ UDEL Sterilization Trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in STERRAD and steam sterilization systems as indicated table. The system may be used with flexible endoscopes with lumen diameters ≥ 1.2mm and a maximum length of 845mm. KARL STORZ UDEL Sterilization Trays must be used in conjunction with an FDA cleared sterilization wrap or container.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Tray Name | Intended Content<br>KARL STORZ Instruments Only | STERRAD® 100S | STERRAD® NX<br>Standard Cycle | STERRAD® NX<br>Advanced Cycle | STERRAD® 100NX<br>Standard Cycle | STERRAD® 100NX<br>Flex Cycle | Steam<br>Pre-vacuum 132°C 4 minutes<br>See Dry Time in Table | Steam<br>Gravity Cycle 121°C 30 minutes<br>See Dry Time in Table | Product Load and Maximum Weight |
|----------------------------------------------------------------------------------------------------|-------------------------------------------------|----------------|-------------------------------|---------------------------------|----------------------------------|------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Flexible Endoscope Tray,<br>P/N 39401AS | Flexible Endoscope | Non-Lumen Only | | Note: Tray does not fit chamber | | | Note: Flexible endoscopes cannot be autoclaved | Note: Flexible endoscopes cannot be autoclaved | 1 flexible fiberscope<br>5.76 lbs. |
| Flexible Endoscope Tray,<br>P/N 39402AS | Flexible Endoscope | Non-Lumen Only | | ✓ | | ✓ | Note: Flexible endoscopes cannot be autoclaved | Note: Flexible endoscopes cannot be autoclaved | 1 flexible fiberscope<br>1 fiber optic light cable<br>8.18 lbs. |
| Flexible Video Endoscope Tray,<br>P/N 39403AS | Flexible CCD Video Endoscope | Non-Lumen Only | | ✓ | | ✓ | Note: Flexible video endoscopes cannot be autoclaved | Note: Flexible video endoscopes cannot be autoclaved | 1 flexible CCD video endoscope<br>5.28 lbs. |
| Flexible Endoscope Tray,<br>P/N 39406AS | Flexible CMOS Video Endoscope | Non-Lumen Only | | ✓ | | ✓ | Note: Flexible video endoscopes cannot be autoclaved | Note: Flexible video endoscopes cannot be autoclaved | 1 flexible CMOS video endoscope<br>5.23 lbs. |
| Camera Trays<br>P/N 39301HCTS<br>P/N 39301PHTS<br>P/N 39301BCTS | Non-autoclave-able camera heads | | ✓ | ✓ | | ✓ | Note: Steam is not indicated when used with non-autoclavable camera heads | Note: Steam is not indicated when used with non-autoclavable camera heads | 1 camera head<br>2.82 lbs. (HCTS)<br>2.64 lbs. (PHTS)<br>4.32 lbs. (BCTS) |
| Camera Trays<br>P/N 39301HCTS<br>P/N 39301ACTS<br>P/N 39301PHTS | Autoclave-able camera heads | | ✓ | ✓ | | ✓ | Minimum Dry Time 50 minutes | Minimum Dry Time 80 minutes | 1 camera head<br>2.82 lbs. (HCTS)<br>3.08 lbs. (ACTS)<br>2.64 lbs. (PHTS) |
| Rigid Telescope Trays<br>P/N 39301AS<br>P/N 39301BS<br>P/N 39301CS<br>P/N 39301DS | Rigid Telescopes<br>(Non-Lumen only) | | ✓ | ✓ | | ✓ | Minimum Dry Time 30 minutes | Minimum Dry Time 80 minutes | 2 Telescopes<br>2.71 lbs. (AS)<br>3.21 lbs. (CS)<br>3.71 lbs. (DS)<br>4 Telescopes<br>6.86 lbs. (BS) |
| Rigid Telescope Tray<br>P/N 39301C1S | Rigid Telescopes<br>(Non-Lumen only) | | ✓ | ✓ | | ✓ | Pre-Vacuum Only Minimum Dry Time 30 minutes | | 1 rigid telescope<br>1 fiber optic cable<br>3.17 lbs. |
| Rigid Telescope Trays<br>P/N 39311AS<br>P/N 39314FS | Rigid Telescopes<br>(Non-Lumen only) | | ✓ | ✓ | | ✓ | Pre-Vacuum Only Minimum Dry Time 30 minutes | | 4 rigid telescopes<br>1 fiber optic cable<br>6.86 lbs. (AS)<br>8.12 lbs. (FS) |
| Rigid Telescope Tray,<br>P/N 39311BS | Rigid Telescopes<br>(Non-Lumen Only) | | ✓ | ✓ | | ✓ | Pre-Vacuum Only Minimum Dry Time 30 minutes | | 2 rigid telescopes<br>1 fiber optic cable<br>4.9 lbs. |
| Basket-Style Rigid Telescope Trays<br>P/N 39305C1S<br>P/N 39305C2S<br>P/N 39305L1S<br>P/N 39305L2S | Rigid Telescopes<br>(Non-Lumen only) | | ✓ | ✓ | | ✓ | Pre-Vacuum Only Minimum Dry Time 30 minutes | | 1 rigid telescope<br>1.58 lbs (C1S)<br>1.58 lbs (L1S)<br>2 rigid telescopes<br>2.56 lbs. (C2S)<br>2.56 lbs. (L2S) |
{4}------------------------------------------------
# 510(k) Summary - K201945
## Date: February 24, 2021
| 1. | Submitted By: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Ave.<br>El Segundo, California 90245<br>424-218-8376 |
|----|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Contact: | David C. Furr<br>Toscano Consulting Group, Inc.<br>8708 Capehart Cove<br>Austin, Texas 78733<br>512-906-9654 |
| 3. | Product: | KARL STORZ UDEL Sterilization Trays<br>Product code: KCT - Class II (21 CFR 880.6850) |
| 4. | Common/Classification<br>Name:<br>Predicate devices: | Sterilization wrap/container<br>KARL STORZ- Endoscope Sterilization Trays K090818<br>Symmetry Medical Polyvac Surgical Instrument Delivery System K012105 |
## Description:
The KARL STORZ UDEL Sterilization Trays are intended only for use to encase specific KARL STORZ reusable medical devices for sterilization in steam and specified STERRAD Sterilization System cycles. The sterilization trays are not intended to maintain sterility by themselves. Prior to sterilization, the trays must be double-wrapped with an appropriate FDA-cleared sterilization wrap to provide a microbial barrier which allows sterilant to permeate throughout the interior of the loaded tray.
The tray configurations, available in various sizes, are designed to encase specific KARL STORZ medical devices, such as fiber optic light cable, flexible endoscopes, and camera heads. Some of the trays are specifically molded to fit and accommodate KARL STORZ light cables, flexible endoscopes, camera heads and rigid telescopes, allowing the instruments to be arranged in an organized manner when placed in the tray. All systems consist of a Udel plastic base and Udel plastic lid. Lids are attached to the trays with assembled hardware. Baskets are injection molded and trays are thermoformed from Udel polysulfone polymer (sourced from Solvay Specialty Polymers), which is compatible with all indicated sterilization modes.
The sterilization trays are constructed with a perforated lid and base to allow for permeation of sterilant during sterilization. Some of the tray configurations are simply molded into the form of a mesh. The trays have latches designed to fasten the lid onto the base. Other tray components include silicone holders and mats (depending on the tray type) to secure instruments and provide protection of the medical devices in the sterilization tray. The holders and mats are manufactured from silicone material, which is also compatible with all sterilization modes.
{5}------------------------------------------------
# Intended Use:
The KARL STORZ UDEL Sterilization Trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in STERRAD and steam sterilization systems as indicated in the attached table. The system may be used with flexible endoscopes with lumen diameters ≥ 1.2mm and a maximum length of 845mm.
KARL STORZ UDEL Sterilization Trays must be used in conjunction with an FDA cleared sterilization wrap or container.
| Tray Name | Intended<br>Content | STERRAD®<br>100S | STERRAD® NX | | STERRAD®<br>100NX | | Steam<br>Pre-vacuum<br>132°C 4 minutes<br>See Dry Time in<br>Table | Steam<br>Gravity Cycle<br>121°C 30<br>minutes<br>See Dry Time in<br>Table | Product Load and<br>Maximum Weight |
|-------------------------------------------------------------------------------------------------------------|--------------------------------------------|----------------------------------------------------------------|-------------------------------------------|-------------------------------------------|-------------------------------------------|-------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | KARL STORZ<br>Instruments<br>Only | | Standard<br>Cycle | Advanced<br>Cycle | Standard<br>Cycle | Flex<br>Cycle | | | |
| Flexible Endoscope<br>Tray,<br>P/N 39401AS | Flexible<br>Endoscope | <span style="text-align:center;">✓</span><br>Non-Lumen<br>Only | | Note: Tray<br>does not fit<br>chamber | | <span style="text-align:center;">✓</span> | Note:<br>Flexible endoscopes<br>cannot be autoclaved | Note:<br>Flexible endoscopes<br>cannot be autoclaved | 1 flexible fiberscope<br>5.76 lbs. |
| Flexible Endoscope<br>Tray, P/N 39402AS | Flexible<br>Endoscope | <span style="text-align:center;">✓</span><br>Non-Lumen<br>Only | | <span style="text-align:center;">✓</span> | | <span style="text-align:center;">✓</span> | Note:<br>Flexible endoscopes<br>cannot be autoclaved | Note:<br>Flexible endoscopes<br>cannot be autoclaved | 1 flexible fiberscope<br>1 fiber optic light<br>cable<br>8.18 lbs. |
| Flexible Video<br>Endoscope Tray,<br>P/N 39403AS | Flexible CCD<br>Video<br>Endoscope | <span style="text-align:center;">✓</span><br>Non-Lumen<br>Only | | <span style="text-align:center;">✓</span> | | <span style="text-align:center;">✓</span> | Note:<br>Flexible video<br>endoscopes cannot<br>be autoclaved | Note:<br>Flexible video<br>endoscopes cannot<br>be autoclaved | 1 flexible CCD video<br>endoscope<br>5.28 lbs. |
| Flexible Endoscope<br>Tray,<br>P/N 39406AS | Flexible<br>CMOS Video<br>Endoscope | <span style="text-align:center;">✓</span><br>Non-Lumen<br>Only | | <span style="text-align:center;">✓</span> | | <span style="text-align:center;">✓</span> | Note:<br>Flexible video<br>endoscopes cannot<br>be autoclaved | Note:<br>Flexible video<br>endoscopes cannot<br>be autoclaved | 1 flexible CMOS<br>video endoscope<br>5.23 lbs. |
| Camera Trays<br>P/N 39301HCTS<br>P/N 39301PHTS<br>P/N 39301BCTS | Non-<br>autoclave-able<br>camera heads | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Note: Steam is not<br>indicated when used<br>with non-<br>autoclavable camera<br>heads | Note: Steam is not<br>indicated when used<br>with non-<br>autoclavable camera<br>heads | 1 camera head<br>2.82 lbs (HCTS)<br>2.64 lbs. (PHTS)<br>4.32 lbs. (BCTS) |
| Camera Trays<br>P/N 39301HCTS<br>P/N 39301ACTS<br>P/N 39301PHTS | Autoclave-able<br>camera heads | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Minimum Dry Time<br>50 minutes | Minimum Dry Time 80<br>minutes | 1 camera head<br>2.82 lbs (HCTS)<br>3.08 lbs. (ACTS)<br>2.64 lbs. (PHTS) |
| Rigid Telescope<br>Trays<br>P/N<br>39301AS<br>P/N<br>39301BS<br>P/N<br>39301CS<br>P/N 39301DS | Rigid<br>Telescopes<br>(Non-Lumen only) | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Minimum Dry Time<br>30 minutes | Minimum Dry Time 80<br>minutes | 2 Telescopes<br>2.71 lbs. (AS)<br>3.21 lbs. (CS)<br>3.71 lbs. (DS) 4<br>Telescopes<br>6.86 lbs. (BS) |
| Rigid Telescope<br>Tray<br>P/N<br>39301C1S | Rigid<br>Telescopes<br>(Non-Lumen only) | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Pre-Vacuum Only<br>Minimum Dry Time<br>30 minutes | | 1 rigid telescope 1<br>fiber optic cable<br>3.17 lbs. |
| Rigid Telescope<br>Trays<br>P/N 39311AS<br>P/N 39314FS | Rigid<br>Telescopes<br>(Non-Lumen only) | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Pre-Vacuum Only<br>Minimum Dry Time<br>30 minutes | | 4 rigid telescopes 1<br>fiber optic cable<br>6.86 lbs. (AS)<br>8.12 lbs. (FS) |
| Rigid Telescope<br>Tray,<br>P/N 39311BS | Rigid<br>Telescopes<br>(Non-Lumen<br>Only) | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Pre-Vacuum Only<br>Minimum Dry Time<br>30 minutes | | 2 rigid telescopes<br>1 fiber optic cable<br>4.9 lbs. |
| Basket-Style Rigid<br>Telescope Trays<br>P/N 39305C1S<br>P/N<br>39305C2S<br>P/N<br>39305L1S<br>P/N 39305L2S | Rigid<br>Telescopes<br>(Non-Lumen only) | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | <span style="text-align:center;">✓</span> | Pre-Vacuum Only<br>Minimum Dry Time<br>30 minutes | | 1 rigid telescope<br>1.58 lbs (C1S)<br>1.58 lbs (L1S)<br>2 rigid telescopes<br>2.56 lbs. (C2S)<br>2.56 lbs. (L2S) |
{6}------------------------------------------------
## Comparison of Technological Characteristics:
The KARL STORZ UDEL Sterilization Trays are intended to protect medical device instrumentation and facilitate the sterilization process by sterilant penetration and air removal.
The KARL STORZ UDEL Sterilization Trays are predicated on the original KARL STORZ Endoskope Sterilization Trays (K090818). Three of the devices, 39301HCTC, 39402AS and 39301BS, are identical to this predicate with the exception of additional indications. In addition the Symmetry Medical Polyvac Surgical Instrument Delivery System (K012105) is included as a predicate for steam applications.
Tray configurations and loads are the same as the devices which were included in KARL STORZ previous 510(k) K090818. Additional tray configurations have been added and validated based on comparison of worst-case vent to volume factors and load.
Sterilant enters the KARL STORZ UDEL sterilization trays through perforations in the tray base and lid. After sterilization, sterility is maintained by an FDA cleared sterilization wrap. All of these characteristics are the same as the predicate devices.
All trays including all the predicate devices are manufactured by the same supplier using the same manufacturing processes.
{7}------------------------------------------------
| Element of<br>Comparison | 510(k) Device:<br>KARL STORZ UDEL<br>Sterilization Trays | Predicate Device:<br>KARL STORZ<br>Endoskope Sterilization<br>Trays (K090818) | Predicate Device:<br>Symmetry Medical Polyvac<br>Surgical Instrument<br>Delivery System<br>(K012105) | Explanation of Differences |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation<br>and Product<br>Classification<br>Code | 21 CFR 880.6850<br>KCT | 21 CFR 880.6850<br>KCT | 21 CFR 880.6850<br>KCT | None |
| Indications<br>for Use | The KARL STORZ<br>UDEL Sterilization Trays<br>are intended only for use<br>to encase and protect<br>specific KARL STORZ<br>reusable medical devices<br>for sterilization in<br>STERRAD and steam<br>sterilization systems as<br>indicated in the attached<br>table. The system may be<br>used with flexible<br>endoscopes with lumen<br>diameters ≥ 1.2mm and a<br>maximum length of<br>845mm.<br>KARL STORZ UDEL<br>Sterilization Trays must<br>be used in conjunction<br>with an FDA cleared<br>sterilization wrap or<br>container. | The KARL STORZ<br>Endoskope Sterilization<br>Trays are intended only<br>for use to encase and<br>protect specific KARL<br>STORZ reusable medical<br>devices for sterilization in<br>STERRAD 100S and NX<br>sterilization systems as<br>indicated. STERRAD<br>Sterilization Systems are<br>pre-set and cycle<br>parameters cannot be<br>adjusted. Table 1<br>provides sterilization<br>compatibility for<br>STERRAD Sterilization<br>Cycles and max load with<br>KARL STORZ-<br>Endoskope Sterilization<br>Trays & Instruments.<br>(Table provided in IFU) | Polyvac's Delivery Systems<br>consist of perforated trays<br>with lids, which are<br>intended to enclose and<br>protect medical device<br>instrumentation, and to<br>facilitate the sterilization<br>processing by allowing<br>steam penetration and air<br>removal, when used in<br>conjunction with an<br>approved sterilization wrap.<br>Sterility of the enclosed<br>medical devices maintained<br>until used. Polyvac's<br>Delivery Systems are to be<br>sterilized in one of the<br>following cycles: pre<br>vacuum steam, 132°C 4<br>minutes minimum, gravity<br>steam 132ºC 30 minutes<br>minimum and gravity steam<br>121°C 55 minutes<br>minimum. | The new system can be used for<br>any of the three sterilization<br>methods cited in the predicates.<br>The new indications for use<br>includes all three methods. |
| Principal<br>Material of<br>Construction | Thermoformed or<br>injection molded Udel<br>Polysulfone | Thermoformed or<br>injection molded Udel<br>Polysulfone | Thermoformed Radel<br>Polyphenylsulfone or<br>aluminum | Udel trays are made from the<br>identical material as the KARL<br>STORZ predicate; Most<br>Symmetry Medical predicates are<br>made from Radel, a similarly<br>performing polymer in steam<br>cycles |
| Element of<br>Comparison | 510(k) Device:<br>KARL STORZ<br>UDEL<br>Sterilization Trays | Predicate Device:<br>KARL STORZ<br>Endoskope Sterilization<br>Trays (K090818) | Predicate Device:<br>Symmetry Medical Polyvac<br>Surgical Instrument<br>Delivery System<br>(K012105) | Explanation of Differences |
| KARL<br>STORZ<br>Model<br>Numbers | 20 models<br>• 39301PHTS<br>• 39301HCTS<br>• 39301ACTS<br>• 39301BCTS<br>• 39401AS<br>• 39402AS<br>• 39403AS<br>• 39406AS<br>• 39311BS<br>• 39301DS<br>• 39314FS<br>• 39311AS<br>• 39301C1S<br>• 39301CS<br>• 39301AS<br>• 39301BS<br>• 39305C1S<br>• 39305C2S<br>• 39305L1S<br>• 39305L2S | 3 models<br>• 39301HCTS<br>• 39402AS<br>• 39301BS | N/A | Additional models have been<br>added for this submission.<br>Symmetry Medical model<br>numbers from previous<br>submissions are not related. |
| Dimensional<br>Configuration<br>Range | Approximate sizes<br>available in inches (see<br>product description chart<br><b>Section 4.2</b> for exact<br>dimensions by catalog<br>number):<br>• 15x10x2.5<br>• 27x7x4<br>• 21x10x4<br>• 21x10x3<br>• 26x4x3<br>• 20x9x4<br>• 13x9x2<br>• 21x3.5x2<br>• 12x3.5x2<br>• 17x3.5x2<br>• 19x4x9 | Approximate sizes<br>available in inches:<br>• 15x10x2.5<br>• 22x10x4<br>• 17x3.5x2 | Approximate sizes<br>available in inches:<br>• 7x2x1<br>• 7x3x1<br>• 8x4x1<br>• 11x7x1<br>• 21x10x4<br>• 17x10x4<br>• 17x8x2<br>• 15x10x1.5<br>• 20.5x9.7x5<br>• 20x10x3<br>• 26x9x6<br>• 17x3.5x1.5<br>• And others | 3 of the KARL STORZ trays are<br>identical to the KARL STORZ<br>predicate and the others fall into<br>very similar size ranges. |
| Device<br>Manufacturer | Contract manufactured<br>for KARL STORZ by<br>Tecomet | Contract manufactured<br>for KARL STORZ by<br>Tecomet | Manufactured by Tecomet<br>(formerly Symmetry<br>Medical) | None. All trays are manufactured<br>in the same location by Tecomet |
| Sterilization<br>Cycles | Prevacuum Steam<br>4 minute cycle 132°C,<br>Gravity Steam 30 mins<br>121 °C<br>STERRAD 100S, NX,<br>and 100NX cycles | STERRAD 100S, NX,<br>and 100NX cycles | Prevacuum Steam and<br>Gravity Steam Cycles | Subject device can be used in the<br>predicate device cycles as<br>specified in the Indications. |
| Load | Trays are to be loaded<br>with specific KARL<br>STORZ instruments | Trays are to be loaded<br>with specific KARL<br>STORZ instruments | Various loads up to 25 lbs. | KARL STORZ trays are used with<br>specific instruments and<br>Symmetry predicates are general |
{8}------------------------------------------------
# 510(k) Premarket Notification
{9}------------------------------------------------
Summary of Non-Clinical Testing:
The following testing was conducted or is referenced to establish efficacy. Worst case tray configurations were used to represent all trays.
| Type of Testing | Primary Standard(s) Used<br>(as applicable) | Test Criteria<br>and Result |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| STERRAD 100S sterilization efficacy | AAMI ST77 Containment Devices for Reusable Medical Device Sterilization<br><br>ISO 14937 Sterilization of Health Care Products – General Requirements for the Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices | 10-6 SAL<br><br>PASSED |
| STERRAD NX standard and advanced cycle sterilization efficacy | AAMI ST77 Containment Devices for Reusable Medical Device Sterilization<br><br>ISO 14937 Sterilization of Health Care Products – General Requirements for the Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices | 10-6 SAL<br><br>PASSED |
| STERRAD 100NX flex cycle sterilization efficacy | AAMI ST77 Containment Devices for Reusable Medical Device Sterilization<br><br>ISO 14937 Sterilization of Health Care Products – General Requirements for the Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices | 10-6 SAL<br><br>PASSED |
| Pre-vacuum sterilization efficacy | AAMI ST77 Containment Devices for Reusable Medical Device Sterilization<br><br>ISO 17665-1 Sterilization of Health Care Products – Moist Heat - Part 1 Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices | 10-6 SAL<br><br>PASSED |
| 510(k) Premarket Notification<br>KARL STORZ UDEL Sterilization Trays | | |
| Type of Testing | Primary Standard(s) Used<br>(as applicable) | Test Result |
| Pre-vacuum dry time | AAMI ST77 Containment Devices for<br>Reusable Medical Device Sterilization<br>ISO 17665-1 Sterilization of Health<br>Care Products – Moist Heat - Part 1<br>Requirements for the Development,<br>Validation, and Routine Control of a<br>Sterilization Process for Medical<br>Devices | Establish<br>Minimum Dry<br>Time<br>ESTABLISHED |
| Gravity sterilization efficacy | AAMI ST77 Containment Devices for<br>Reusable Medical Device Sterilization<br>ISO 17665-1 Sterilization of Health<br>Care Products – Moist Heat - Part 1<br>Requirements for the Development,<br>Validation, and Routine Control of a<br>Sterilization Process for Medical<br>Devices | 10-6 SAL<br>PASSED |
| Gravity dry time | AAMI ST77 Containment Devices for<br>Reusable Medical Device Sterilization<br>ISO 17665-1 Sterilization of Health<br>Care Products – Moist Heat - Part 1<br>Requirements for the Development,<br>Validation, and Routine Control of a<br>Sterilization Process for Medical<br>Devices | Establish<br>Minimum Dry<br>Time<br>ESTABLISHED |
| Manual Cleaning – Protein,<br>Hemoglobin | AAMI TIR 30 A Compendium of<br>Processes, Materials, Test Methods, and<br>Acceptance Criteria for Cleaning<br>Reusable Medical Devices | < 6.4 µg/cm2<br>protein and<br>< 2.2 µg/cm2<br>hemoglobin on<br>device after<br>cleaning<br>PASS |
| Mechanical Cleaning - Protein,<br>Hemoglobin | AAMI TIR 30 A Compendium of<br>Processes, Materials, Test Methods, and<br>Acceptance Criteria for Cleaning<br>Reusable Medical Devices | < 6.4 µg/cm2<br>protein and<br>< 2.2 µg/cm2<br>hemoglobin on<br>device after<br>cleaning<br>PASS |
| Material biocompatibility | ISO 10993 Biological Evaluation of<br>Medical Devices | Cytotoxicity –<br>No evidence of<br>lysis<br>PASS |
| Material stability 200 STERRAD<br>Cycles and 2100 Steam Cycles | Internal Protocols and AAMI TIR 12<br>Designing, Testing, and Labeling<br>Reusable Medical Devices for<br>Reprocessing in Health Care Facilities | Material<br>Maintains<br>Integrity<br>PASS |
{10}------------------------------------------------
{11}------------------------------------------------
## Conclusion:
The conclusions drawn from the non-clinical tests demonstrate substantial equivalence in that the proposed device is as safe, as effective, and performed as well as or better than the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.