K201935 · Xiamen Weiyou Intelligent Technology Co., Ltd. · IRP · Jun 22, 2021 · Physical Medicine
Device Facts
Record ID
K201935
Device Name
Air Pressure Therapy System
Applicant
Xiamen Weiyou Intelligent Technology Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Jun 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
Device Story
Air Pressure Therapy System VU-IPC04B is an intermittent pneumatic compression device consisting of a pump unit connected to a multi-chambered cuff (6 chambers) via hoses. The pump inflates and deflates chambers in specific sequences to provide compression from the body's distal end toward the center. Operated by the user in home or rehabilitation settings, the device allows adjustment of cycle time and pressure to ensure comfort. The system provides four distinct compression modes (A, B, C, D) to facilitate muscle relaxation and circulation. By applying sequential pressure, the device aims to relieve muscle soreness and improve blood flow, aiding in recovery and rehabilitation.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified through electrical safety, electromagnetic compatibility, and biocompatibility testing (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Molded ABS enclosure; 6-chamber inflatable cuff; compressor and valve system; pressure range 30-110mmHg; power 110-240V, 50/60Hz; dimensions 22x14x9cm; weight 1.6kg. Standards: IEC 60601-1-11, IEC 60601-1-2, ANSI AAMI ES60601-1, ISO 10993-5, ISO 10993-10, IEC 62133-2.
Indications for Use
Indicated for individuals seeking temporary relief of minor muscle aches and/or pains and temporary increase in circulation to treated areas. Suitable for home use for circulation improvement and exercise rehabilitation settings for pre-exercise muscle relaxation and post-exercise muscle soreness relief.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2021
Xiamen Weiyou Intelligent Technology Co., Ltd % Sam Lin Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609 Room, No.133 Shengang Avenue, Pudong New District Shanghai. 201306 China
Re: K201935
Trade/Device Name: Air Pressure Therapy System: VU-IPC04B Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: February 3, 2021 Received: June 2, 2021
Dear Sam Lin:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director. Acute Injurv Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201935
Device Name
Air Pressure Therapy System: VU-IPC04B
Indications for Use (Describe)
Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Type of submission | Traditional |
|--------------------------------------------|--------------------------------------------------------------------------------|
| Date prepared | February 3, 2021 |
| Submission sponsor | |
| Manufacturer name | Xiamen Weiyou Intelligent Technology Co., Ltd. |
| Address | Unit 101-401, No.6 Xianghong Road, Xiang'an District, Xiamen,<br>Fujian, China |
| Tel | 86-0592-6251545 |
| Email | 17916220@qq.com |
| Contact person | Yong Sun |
| Device identification | |
| Classification name | Massager, Powered Inflatable Tube |
| Trade name | Air Pressure Therapy System: VU-IPC04B |
| Device classification | Class II |
| Regulation number | 21 CFR 890.5650 |
| Panel | Physical Medicine |
| Product code | IRP |
| Previous submissions | None |
| Application correspondent | |
| Company name | Shanghai Spica Management Consulting Co., Ltd. |
| Address | 609 Room, No.133 Shengang Avenue, Pudong New District,<br>Shanghai, China |
| Tel | 86-15626132181 |
| Email | sam@spicagloble.com |
| Contact person | Sam Lin |
| Predicate and reference device information | |
| | |
| Sponsor | NORMATEC INDUSTRIES, LP X |
|-------------------|---------------------------------------|
| Trade/Device name | NormaTec Pulse and NormaTec Pulse Pro |
| 510K number | K160608 |
| Regulation number | 21 CFR 890.5650 |
{4}------------------------------------------------
### Indications for use of the device
Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
### Device description
Air Pressure Therapy System has a pump. The pump is connected to the dedicated cuff via the hose series, one cuff has 6 compression chambers. The pump compresses and inflates air into the chambers continuously to produce compression from the body tail end to body center and release the air after compression as one cycle process. Cycle time and pressure can be adjusted by purpose to avoid any discomfort.
Air Pressure Therapy System intermittent pneumatic compression could be used in family for improving blood circulation in the treated area, in exercise rehabilitation field for muscle relaxation before exercise and muscle rehabilitation after exercise to relieve muscle soreness.
### Performance testing - clinical
Not Applicable.
### Performance testing - animal
Not Applicable.
{5}------------------------------------------------
| | Subject device | Predicate device | Differences discussion | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device name | Air Pressure Therapy System: VU-IPC04B | NormaTec Pulse and NormaTec Pulse<br>Pro | | |
| 510(k) number | K201935 | K160608 | N/A | |
| Manufacturer | Xiamen Weiyou Intelligent Technology Co., Ltd. | NORMATEC INDUSTRIES, LP X | N/A | |
| Product regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | Same | |
| Classification name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Same | |
| Regulation class | 2 | 2 | Same | |
| Product code | IRP | IRP | Same | |
| Indications for use | Air Pressure Therapy System VU-IPC04B is<br>intended to temporarily relieve minor muscle aches<br>and/or pains, and to temporarily increase circulation<br>to the treated areas. | The NormaTec Pulse and Pulse Pro is<br>an air pressure massager intended to<br>temporarily relieve minor muscle<br>aches and/or pains, and to temporarily<br>increase circulation to the treated<br>areas. | Same | |
| Rx or OTC | OTC | OTC | Same | |
| Pressure range | 30-110mmHg | 30-110mmHg | Same | |
| Treatment time | 5-99mins | Stays on until the user turns it off or<br>can be set up to turn off in a range of<br>10 mins to continuous / User<br>controlled 10 minutes to 175 minutes<br>or continuous -total time over 4 | Similar<br>The treatment time of subject<br>device is smaller than<br>predicate device (K160608)<br>so the difference of treatment | |
| | | | | |
| | | segments. | time would not raise adversely<br>impact on safety and<br>effectiveness. | |
| Standard | IEC 60601-1-11,<br>ES60601-1:2005/(R)2012<br>C1:2009/(R)2012 and<br>60601-1-2, ISO 10993-10,<br>62133-2 | ANSI AAMI A1:2012,<br>and A2:2010/(R)2012, IEC<br>ISO 10993-5, IEC | ES 60601-1, IEC 60601-1-2, IEC<br>60601-1-11 | Similar |
| Mode of compression | Sequential | Sequential | Same | |
| Power source | 110-240V; 50Hz/60Hz | 100- 240 VAC input | Similar | |
| Power consumption | 35W | 14W | Similar | |
| Dimensions(W*H*D) | 22*14*9CM | 4" x 5" x 9" | Similar | |
| Weight | 1.6KG | 3.6 pounds | Similar | |
| Photo | Image: a black device with buttons | N/A | Similar | |
| Size and appearance of<br>sleeves (leg part) | Image: two black sleeves and a black round object<br>M: 91x65cm<br>XL: 110x70cm(overlapping)<br>XXL: 125x76cm(overlapping) | Short: 14" x 43"<br>Standard: 14" x 48"<br>Tall: 14" x 60" | Similar | |
| Housing materials | Molded ABS enclosure | Molded ABS enclosure | Same | |
| Number of chambers | 6 chambers | 5 chambers | Similar | |
| | Mode A: | Sequential mode: | Although the subject device | |
| | In this mode, only a single chamber is inflated at a<br>time. Starting from the chamber ① and working up | Starting with the distal chamber and | provides 4 kinds of work | |
| | to the chamber ⑥. Then the cycle repeats. | progressing up the proximal chamber, | mode, the Mode A and Mode | |
| | Mode B: | each section compresses and the | C are the same with predicate | |
| | In this mode, the chamber ① stays inflated. It | pressure gradually rises to the | device (K160608), while the | |
| | gradually adds a chamber until all six chambers are | pre-determined air pressure level, | other work modes of subject | |
| | filled with air. Then the cycle repeats. | then decompresses and the air<br>pressure drops. | device just have difference | |
| | Mode C: | Once the top section decompresses,<br>the cycle begins again. again. | about inflatable order of the | |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.