K201927 · Maquet GmbH · FLE · Mar 1, 2021 · General Hospital
Device Facts
Record ID
K201927
Device Name
Getinge GSS610N Series Steam Sterilizer
Applicant
Maquet GmbH
Product Code
FLE · General Hospital
Decision Date
Mar 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The steam sterilizer is used by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
Device Story
Steam sterilizer for healthcare facilities; processes wrapped/unwrapped porous and nonporous heat/moisture-stable items (surgical instruments, linens). Uses gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Control system sequences cycle phases, monitors parameters, and provides audible/visual notifications of deviations. Operated by healthcare staff. Benefits include effective sterilization of medical equipment to support infection control. Output consists of sterilized items; device status/alerts provided to operator.
Clinical Evidence
Bench testing only. Validation performed per AAMI ST8:2013 (R2018) standards. Testing included biological performance (SAL 10^-6) for fabric PCDs, liquid loads, wrapped instrument PCDs, and IUSS cycles. Physical performance testing confirmed chamber temperature profiles, mechanical air removal, air leak rates, and moisture retention met acceptance criteria.
Technological Characteristics
Steam sterilizer; stainless steel chamber (SA240-316L) and jacket (SA240-304). Pressure vessel built to ASME Section VIII, Div. 1. Features vacuum pump for air removal. Horizontal door design with electric motor. Control system provides cycle sequencing and monitoring. Connectivity/interlocks via electro-mechanical logic. Sterilization via pressurized steam.
Indications for Use
Indicated for use by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items, including surgical instruments and linens, via pressurized steam.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 1, 2021
Maquet GmbH % Mark Smith Regulatory Affairs Manager Getinge 45 Barbour Pond Road Wayne, New York 07470
Re: K201927
Trade/Device Name: Getinge GSS610N Series Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE Dated: February 1, 2021 Received: February 2, 2021
Dear Mark Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201927
Device Name Getinge GSS610N Series Steam Sterilizer
Indications for Use (Describe)
{3}------------------------------------------------
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary - K201927 Getinge GSS610N Series Steam Sterilizer
| Submitted by | Maquet GmbH_<br>Kehler Strasse 31<br>Rastatt DE-BW<br>Germany 76437 |
|--------------|---------------------------------------------------------------------|
|--------------|---------------------------------------------------------------------|
- Mark N. Smith Contact person Manager, Regulatory Affairs Phone: 585-272-5274 Email: mark.n.smith@getinge.com
- K201927 510(k) number
- Date Prepared February 26, 2021
# Proprietary Device Name
- Trade Name: Getinge GSS610N Series Steam Sterilizer
610N10 40.9 inch (1000 mm) long chamber Models: 610N14 51.1 inch (1400 mm) long chamber 610N15 60.6 inch (1540 mm) long chamber
Common Name: Steam Sterilizer
- Classification: Steam Sterilizer (21CFR880.6880, Product Code 80 FLE)
# Predicate Devices
[K172159] Getinge GSS67N Series Steam Sterilizer Getinge. Date 01/30/2018
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Getinge, a company that provides equipment and systems for healthcare and life sciences. The logo consists of the word "GETINGE" in a bold, sans-serif font, followed by a stylized symbol that resembles a four-pointed star or a stylized flower. The logo is simple and modern, and the colors are muted, giving it a professional and trustworthy appearance.
# Description of Device
The steam sterilizer is used by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.
The Getinge GSS610N Series Steam Sterilizer employs both with gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters
# Indications for Use
The Getinge GSS610N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS610N Series Steam Sterilizer is available in 3 models differentiated by chamber length: GSS610N Model 610N10 (40.9 inch chamber), GSS610N Model 610N14 (51.1 inch chamber), and GSS610N Model 610N15 (60.6 inch chamber).
# Cycles and Load Chart
| Cycle Type | Factory Settings | | Load Configuration<br>(Note 1) | Maximum Items per<br>Chamber Length | | | |
|------------------|----------------------|------------------|--------------------------------|----------------------------------------------------------------------------|------------------------------|------------------------------|------------------------------|
| | Exp.<br>Temp. | Exp.<br>Time | Drying<br>Time | | 610N10<br>1040 mm<br>40.9 in | 610N14<br>1400 mm<br>51.1 in | 610N15<br>1540 mm<br>60.6 in |
| P1<br>PREVAC 1 | 135.0°C<br>(275.0°F) | 3 min | 16 min | Full instrument packs | 18 | 24 | 24 |
| P2<br>PREVAC 2 | 135.0°C<br>(275.0°F) | 3 min | 3 min | Full fabric packs | 40 | 40 | 60 |
| P3<br>PREVAC 4 | 132.2°C<br>(270.0°F) | 4 min | 30 min | Full instrument packs | 18 | 24 | 24 |
| P4<br>PREVAC 5 | 132.2°C<br>(270.0°F) | 4 min | 5 min | Full fabric packs | 40 | 40 | 60 |
| P5<br>B & D TEST | 134°C<br>(273.0°F) | 3 min,<br>30 sec | 0 min | 1 B&D Test Pack in an<br>EMPTY chamber (other<br>than loading accessories) | 1 Test<br>Pack | 1 Test<br>Pack | 1 Test<br>Pack |
| P6<br>GRAVITY 1 | 121.1°C<br>(250.0°F) | 30 min | 45 min | Full instrument packs<br>Full fabric packs | 18<br>40 | 24<br>40 | 24<br>60 |
| P7<br>GRAVITY 2 | 135.0°C<br>(275.0°F) | 10 min | 45 min | Full instrument packs<br>Full fabric packs | 18<br>40 | 24<br>40 | 24<br>60 |
| P8<br>GRAVITY 3 | 132.2°C<br>(270.0°F) | 15 min | 45 min | Full instrument packs<br>Full fabric packs | 18<br>40 | 24<br>40 | 24<br>60 |
| P9<br>IUSS 1 | 135.0°C<br>(275.0°F) | 3 min | 1 min | Full instrument packs | 1 | 1 | 1 |
# Getinge GSS610N Series Steam Sterilizer Cvcles and Load Chart
510k Summary GSS610N K201927
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Getinge. The word "GETINGE" is written in a bold, sans-serif font in dark blue. To the right of the word is a stylized symbol, also in blue, that resembles four curved arrows pointing inward towards a central point. The logo is simple and modern.
| Vac | | | | | Full fabric packs | 1 | 1 | 1 |
|--------------------------|----------------------|--------|-------------------------------------------|--|-------------------------------------------------------------|---|---|---|
| P10<br>IUSS 2<br>Grav | 135.0°C<br>(270.0°F) | 10 min | 30 sec | | Full instrument packs | 1 | 1 | 1 |
| | | | | | Full fabric packs | 1 | 1 | 1 |
| P11<br>IUSS 3<br>Grav | 132.2°C<br>(270.0°F) | 4 min | 1 min | | Full instrument packs | 1 | 1 | 1 |
| | | | | | Full fabric packs | 1 | 1 | 1 |
| P12<br>Vented<br>Bottles | 121.1°C<br>(250.0°F) | 45 min | 3kPa/min<br>(0.44<br>psi/min)<br>(Note 2) | | Each container 1000 mL<br>(34 fl oz) or smaller<br>(Note 3) | 3 | 3 | 3 |
#### NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.
## TABLE NOTES:
1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAMI Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAM ST8 and were made to be consistent with the packs described in ANSI/AAM ST8). For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to sterility assurance in health care facilities.
2. Cooldown rate
3. Use vented or open containers only.
# Technological Characteristics Table:
Shown below is a comparison of the subject device Getinge GSS610N Series Steam Sterilizer versus the predicate device (Getinge GSS67N Series Steam Sterilizer):
| Item | Getinge GSS610N<br>[Subject device] | Getinge GSS67N<br>[Predicate Device]<br>K172159 | Comparison |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Pressure Vessel | | | |
| Chamber Sizes | Model 610N10: 26.4" x<br>42.5" x 40.9" (672 x1080<br>x1040 mm) | Model 6710: 26.4" x 27.6" x<br>39.4" (672 x700 x1000 mm) | Different Chamber<br>Sizes |
| | Model 610N14: 26.4" x<br>42.5" x 55.1" (672 x1080<br>x1400 mm) | Model 6713: 26.4" x 27.6" x<br>51.2" (672 x700 x1300 mm) | Different Chamber<br>Sizes |
| | Model 610N15: 26.4" x<br>42.5" x 60.6" (672<br>x1080 x1540 mm) | Model 6717: 26.4" x 27.6" x<br>66.9" (672 x700 x1700 mm) | Different Chamber<br>Sizes |
| Chamber Volumes | Model 610N10: Single Door<br>26.52 Cu Ft (751L), Double<br>Door 26.59 Cu Ft (753L) | Model 6710: Single Door 17<br>Cu Ft (481L), Double Door<br>16.5 Cu Ft (468L) | The chamber volumes<br>correspond to the<br>chamber sizes |
| | Model 610N14: Single Door<br>35.73 Cu Ft (1012L) | | |
| | Double Door 35.77 Cu Ft<br>(1013L)<br><br>Model 610N15: Single Door<br>39.3 Cu Ft (1113L), Double<br>Door 39.34 Cu Ft (1114L) | Model 6713: Single Door 22<br>Cu Ft (621L), Double Door<br>21.5 Cu Ft (609L)<br><br>Model 6717: Single Door 28.6<br>Cu Ft (809L), Double Door<br>28.11 Cu Ft (796L) | |
| ASME Pressure Vessel | All pressure vessels are<br>built to ASME Sect. VIII,<br>Div. 1 | All pressure vessels are built<br>to ASME Sect. VIII, Div. 1 | Same |
| "U" Stamped Unfired<br>Pressure Vessel) | Chamber: 45 psi<br>Jacket: 45 psi | Chamber: 43.5 psi<br>Jacket: 51 psi | The pressure rating<br>differs slightly |
| Material of Jacket | Stainless Steel (SA240- 304 ) | Stainless Steel (SA240- 304 ) | Same |
| Material of Chamber | Stainless Steel (SA240-<br>316L ) | Stainless Steel (SA240-316L ) | Same |
| Vacuum method | Vacuum pump standard. | Vacuum pump standard. | Same |
| Chamber Closure | | | |
| Door Operation | Horizontal opening/closing<br>door design operated by<br>an electrical motor. | Vertical opening/closing door<br>design with pneumatic<br>cylinder actuator | Different door<br>design. |
| Safety and Interlocks | | | |
| Door switch<br>system/steam to<br>chamber interlock | Electro-mechanical logic | Electro-mechanical logic | Same |
| Cycles | | | |
| Types of cycles offered | Prevac 132.2°C, 4<br>min; 135°C, 3 min,<br>Gravity 121°C, 30 min;<br>135°C 10 min; 132.2°<br>15 min;<br>IUSS 135°C, 3 min,;<br>135C, 10 min; 132.2°C<br>4 min<br>Vented Bottles 121°C<br>45min.<br>BD Test | Prevac 132.2°C, 4 min;<br>135°C, 3 min,<br>Gravity 121°C, 30 min;<br>135°C 10 min; 132.2° 15<br>min;<br>IUSS 135°C, 3 min,;<br>135C, 10 min; 132.2°C 4<br>min<br>Vented Bottles 121°C<br>45min.<br>BD Test | Same |
| Maximum Load<br>Capacities | Reference chart for<br>maximum loads within<br>GSS610N - up to 25lbs<br>per tray | Reference chart for<br>maximum loads within<br>GSS67N - up to 25lbs per<br>tray | Same |
| Utility Requirements | | | |
| Primary<br>Electrical<br>Connection | GSS610N are available for<br>connection to:<br>208V 3ph 60Hz (Not<br>GSS61015N with integrated<br>electrical heated steam<br>boiler)<br>460V 3ph 60Hz<br>480V 3ph 60Hz | GSS67N are available for<br>connection to:<br>208V 3ph 60Hz<br>460V 3ph 60Hz<br>480V 3ph 60Hz<br>600V 3ph 60Hz (Not with<br>integrated electrical heated<br>steam boiler) | Similar electrical<br>connections |
{7}------------------------------------------------
# GETINGE
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "GETINGE" in a sans-serif font. To the right of the word is a stylized four-pointed star or asterisk-like symbol. The text and symbol are in a dark blue color.
# Summary of Non-Clinical Testing
Shown below is a summary of the non-clinical testing that was performed with this device:.
| Test<br>Performed | Device<br>Description | Test Method | Acceptance<br>Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|--------------------------------------|------------------------------------------------------------------------------------------------|---------|
| <b>Biological Performance Tests</b> | | | | |
| Sterilization<br>Efficacy Validation<br>Biological<br>Performance with<br>a fabric PCD | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.5.2 | Sterility Assurance<br>Level (SAL) 10-6 | Pass |
| Biological<br>Performance with<br>liquid loads | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.5.3 | Sterility Assurance<br>Level (SAL) 10-6 | Pass |
| Biological<br>Performance with<br>a wrapped<br>instrument PCD | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.5.4 | Sterility Assurance<br>Level (SAL) 10-6 | Pass |
| Biological<br>performance of<br>immediate-use<br>steam sterilization<br>for single-wrapped<br>or unwrapped<br>nonporous items | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.5.5 | Sterility Assurance<br>Level (SAL) 10-6 | Pass |
| <b>Physical Performance Test</b> | | | | |
| Chamber<br>Temperature<br>Profile | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.4.2.5 | +3°C (or +6°F) and -<br>0°C (or -0°F) of the<br>selected sterilization<br>exposure temperature | Pass |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "GETINGE" in a bold, sans-serif font. The text is a dark blue color. To the right of the word is a stylized logo, also in blue, that resembles four curved arrows pointing inward towards a central point.
| Mechanical Air<br>Removal Test | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.6.1 | Load reaching<br>exposuire temperature<br>witin 10 secs<br><br>Color change on BD<br>chemical indicator<br>sheet | Pass |
|--------------------------------|-------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|------|
| Air Leak Rate Test | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.6.2 | Average leak rate of 1<br>mmHg (0.13 kPa) (0.019<br>psia) per minute or less<br>over the measured time<br>interval. | Pass |
| Moisture<br>Retention Rest | GSS610N10<br>GSS610N14<br>GSS610N15 | AAMI ST8:2013<br>(R2018)<br>§5.7 | < 3% increase in<br>presterilization test<br>pack weight for fabric<br>pack & <20% increase<br>for wrapped instrument<br>pack | Pass |
# Conclusion:
The conclusion drawn from the nonclinical test demonstrates that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device, Class II (21 CFR 880.6880, Product code FLE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.