The Cefaly® Dual is indicated for - The acute treatment of migraine with or without aura in patients 18 years of age or older. - The prophylactic treatment of episodic migraine in patients 18 years of age or older.
Device Story
Cefaly Dual is a supraorbital transcutaneous electrical nerve stimulator; applied to forehead via self-adhesive electrode with 2 conductive zones. Device powered by rechargeable LiPo battery; operated by single button to select/start automatic stimulation programs. Electrical impulses excite supratrochlearis and supraorbitalis nerves (upper branch of trigeminal nerve/V1) to generate analgesic effect for migraine treatment. Used by patients in home/non-clinical settings (OTC). Provider/patient uses device to manage migraine symptoms; output is electrical stimulation; intended to reduce migraine frequency/severity.
Clinical Evidence
No clinical performance testing performed. Substantial equivalence supported by identical technological characteristics and a human factors validation study for OTC use.
Technological Characteristics
Plastic casing; supraorbital electrode; rechargeable 3.7V LiPo battery. Biphasic, rectangular, fully compensated, symmetrical waveforms. Pulse width 250 µs. Program 1: 100 Hz, 60 min. Program 2: 60 Hz, 20 min. Max output 16 mA. Fixed program software.
Indications for Use
Indicated for acute treatment of migraine with or without aura and prophylactic treatment of episodic migraine in patients 18 years of age or older.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.
In combination with the general controls of the Food Drug & Cosmetic Act, the Transcutaneous Electrical Nerve Stimulator to Treat Headache is subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
9/29/2020
CEFALY Technology % Parul Chansoria CEO & Founder, Regulatory Consultant Elexes Medical Consulting 6494 Tralee Village Dr Dublin, California 94568
Re: K201895
Trade/Device Name: Cefaly Dual Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous Electrical Nerve Stimulator to Treat Headache Regulatory Class: Class II Product Code: PCC Dated: July 2, 2020 Received: July 8, 2020
Dear Parul Chansoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Xiaorui Tang, Ph.D. Interim Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201895
Device Name Cefaly® Dual
Indications for Use (Describe)
The Cefaly® Dual is indicated for
- The acute treatment of migraine with or without aura in patients 18 years of age or older.
- The prophylactic treatment of episodic migraine in patients 18 years of age or older.
| Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Registration Use (Part 21 CFR 201.321 Subpart D) <input checked="" type="checkbox"/> On The Container (21 CFR 201.321 Subpart C) |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CEFALY Technology. The logo consists of a red curved line above the words "CEFALY" and "TECHNOLOGY", which are stacked on top of each other. The text is in a sans-serif font and is gray in color.
Rue Louis Plescia, 34
4102 Seraing - BELGIUM
Tél.: +32 4 367 67 22
Fax: +32 4 367 67 02
www.cefaly-technology.com
## 5.1 SUBMITTER
CEFALY Technology, LIEGE Science Park, Rue Louis Plescia, 34, 4102 Seraing, BELGIUM
### Contact Person:
Parul Chansoria, MS, RAC, CQA CEO & Founder, Elexes Medical Consulting Telephone: + 408-475-8091 E-mail: parul@elexes.com
Summary Prepared: June 25, 2020
### 5.2 DEVICE
Common/Usual name/Regulation Description: Transcutaneous electrical nerve stimulator to treat headache Trade Name: Cefaly® Dual Classification Name: Stimulator, Nerve, Electrical, Transcutaneous, For Migraine Regulatory Class: Class II Classification Panel: Neurology Product Code: PCC Regulation Number: 21 CFR 882.5891
### 5.3 PREDICATE DEVICE
Cefaly® Dual is substantially equivalent to the following cleared device:
| Company | Product | 510(k) Number |
|-------------------|--------------|---------------|
| CEFALY Technology | Cefaly® Dual | K173006 |
### 5.4 DEVICE DESCRIPTION
The Cefaly® Dual device indicated for an OTC (also referred to as the Subject Device) is a supraorbital transcutaneous electrical nerve stimulator device to be applied on the forehead. A
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CEFALY Technology. The logo consists of the word "CEFALY" in a serif font, with the word "TECHNOLOGY" in a smaller sans-serif font underneath. Above the word "CEFALY" is a red graphic that resembles two curved lines meeting in the middle.
Rue Louis Plescia, 34
4102 Seraing - BELGIUM
Tél.: +32 4 367 67 22
Fax: +32 4 367 67 02
www.cefaly-technology.com
self-adhesive electrode with 2 conductive zones is placed on the forehead. This double electrode is directly connected to the device.
The Subject Device is operated by a rechargeable battery. Pressure on the single button allows selecting and starting a stimulation program, which runs automatically.
The electrical impulses generated by the Subject Device are transmitted transcutaneously via the supraorbital electrode to excite (trigger action potentials on) the supratrochlearis and supraorbitalis nerves. Supratrochlearis and supraorbitalis (or supratrochlear and supraorbital) nerves belong to the upper branch of the trigeminal nerve (V1). Therefore, the supraorbital neurostimulation is also known as external trigeminal nerve stimulation. The supraorbital neurostimulation generates an analgesic effect and is intended to treat migraine headaches.
### 5.5 INDICATIONS FOR USE
The indications for use of the Cefaly® Dual for an over-the-counter use are:
- The acute treatment of migraine with or without aura in patients 18 years of age or older; ●
- The prophylactic treatment of episodic migraine in patients 18 years of age or older. ●
### 5.6 TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Subject Device is identical to the legally marketed Predicate Cefaly® Dual (K173006) in terms of technological characteristics (design, material, and energy source).
The Subject Device is made of a plastic casing identical to that of the Predicate. It works with the same electrode as the Predicate Device. It is powered by the same battery. The electronics inside the device are also the same. The Subject Device delivers biphasic impulses of the same pulse shape and width (250 us) as the Predicate Device. The repetition frequency of the impulses of the Subject Device depends on the selected stimulation program. The frequencies delivered by the Subject Device for the treatment of Migraine is the same as that of the Predicate.
| Parameter | Subject Device<br>Cefaly® Dual<br>(indicated for OTC) | Predicate<br>Cefaly® Dual<br>(indicated for Rx) | Equivalence |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | CEFALY<br>Technology | CEFALY<br>Technology | - |
| Device Name | Cefaly® Dual | Cefaly® Dual | - |
| 510(k) number | - | K173006 | - |
| Product Code | PCC | PCC | Equivalent |
| Regulation<br>Number | 882.5891 | 882.5891 | Equivalent |
| Regulatory<br>Class | II | II | Equivalent |
| Indications<br>for use | The indications for use<br>of the Cefaly® Dual for<br>an over-the-counter<br>use are:<br>-The acute treatment of<br>migraine with or<br>without aura in patients<br>18 years of age or<br>older.<br>-The prophylactic<br>treatment of episodic<br>migraine in patients 18<br>years of age or older. | The indications for<br>use of the Cefaly®<br>Dual for an Rx use<br>are:<br>- The acute treatment<br>of migraine with or<br>without aura in<br>patients 18 years of<br>age or older.<br>- The prophylactic<br>treatment of episodic<br>migraine in patients<br>18 years of age or<br>older. | Different<br>While the clinical<br>application, the<br>intended users and<br>use remains<br>identical, the<br>difference is only in<br>terms of labelling<br>change where the<br>Subject Device is to<br>be indicated for an<br>OTC use instead of<br>Rx. |
| Power Source | 1 rechargeable LiPo<br>3.7 V<br>battery | 1 rechargeable LiPo<br>3.7 V<br>battery | Equivalent |
| Channels | 1 | 1 | Equivalent |
| Software provided | 2 fixed programs:<br>- 1 fixed program for<br>the acute treatment of<br>migraine attacks<br>(Program 1)<br>- 1 fixed program for | 2 fixed programs:<br>- 1 fixed program for<br>the acute treatment of<br>migraine attacks<br>(Program 1)<br>- 1 fixed program for | Equivalent |
| | prophylactic treatment<br>of migraine<br>(Program 2) | prophylactic<br>treatment of migraine<br>(Program 2) | |
| Program 1:<br>Max. output<br>current<br>Pulse width<br>Pulse frequency<br>Session duration | 16 mA<br>250 µs,<br>fixed 100 Hz,<br>fixed 60 minutes | 16 mA<br>250 µs,<br>fixed 100 Hz,<br>fixed 60 minutes | Equivalent |
| Program 2:<br>Max. output<br>current Pulse<br>width<br>Pulse frequency<br>Session duration | 16 mA<br>250 µs,<br>fixed 60 Hz,<br>fixed 20 minutes | 16 mA<br>250 µs,<br>fixed 60 Hz,<br>fixed 20 minutes | Equivalent |
| Waveform | Biphasic | Biphasic | Equivalent |
| Shape | Rectangular<br>Full compensated<br>Symmetrical | Rectangular<br>Full compensated<br>Symmetrical | Equivalent |
| Net charge (µC)<br>per pulse | 0 | 0 | Equivalent |
| Maximum output<br>current (mA):<br>At 500 ohms<br>At 2,000 ohms<br>At 10,000 ohms | 16<br>16<br>6 | 16<br>16<br>6 | Equivalent |
| Maximum current<br>density (mA/cm2,<br>r.m.s) at 500 ohms | 2.37 | 2.37 | Equivalent |
| Maximum<br>average power<br>density (W/cm2, | 0.000047 | 0.000047 | Equivalent |
Table 1: An overview of the Subiect Device w.r.t. thePredicate Device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for "CEFALY Technology". The logo consists of the text "CEFALY" in a larger, bolder font above the word "TECHNOLOGY" in a smaller font. Above the text is a stylized red graphic element that resembles a curved line or roof-like shape.
Rue Louis Plescia, 34
4102 Seraing - BELGIUM
Tél.: +32 4 367 67 22
Fax: +32 4 367 67 02
www.cefaly-technology.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for CEFALY Technology. The logo consists of the word "CEFALY" in large, bold, sans-serif font, with the word "TECHNOLOGY" in a smaller font size underneath. Above the word "CEFALY" is a red, stylized graphic that resembles two curved lines meeting in the middle, forming a shape similar to a stylized roof or eyebrow. The overall design is clean and modern, with a focus on the company name.
4102 S
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for "CEFALY TECHNOLOGY". The word "CEFALY" is in a larger font size than the word "TECHNOLOGY", and both words are in a sans-serif font. Above the words is a red graphic that looks like two curved lines meeting in the middle. The background is white.
Rue Louis Plescia, 34
4102 Seraing - BELGIUM
Tél.: +32 4 367 67 22
Fax: +32 4 367 67 02
www.cefaly-technology.com
| r.m.s) at 500 ohms | |
|--------------------|--|
|--------------------|--|
### Differences
While the clinical application, the intended users and use remains identical, the difference is only in terms of labelling change where the Subject Device is to be indicated for an OTC use instead of Rx.
### 5.7 NON-CLINICAL STUDY
The Subject Device is compliant to the same international standards as the legally marketed Predicate (K173006). To support the change in the indications for use from an Rx to an OTC use, a Human Factors validation study was conducted with the Cefaly® Dual (Subject Device).
### 5.8 SOFTWARE
The Subject Device software is identical to the software used in the Predicate Device.
### 5.9 CLINICAL STUDY
Not Applicable. Clinical performance testing was not performed with the Subject Device to support equivalence, as there are no differences between the Subject Device and the Predicate Device.
#### 5.10 CONCLUSION
The Subject Device is substantially equivalent to the Predicate Device in terms of technological characteristics, system operating ranges, and intended use. The Subject Device is as safe and effective as the Predicate Device (K173006).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.