K201868 · Wellpower Sporting Goods Co., Ltd. · FXX · Feb 22, 2021 · General, Plastic Surgery
Device Facts
Record ID
K201868
Device Name
Phyziofit Medical Surgical Mask
Applicant
Wellpower Sporting Goods Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Feb 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
Phyziofit Medical Surgical Masks are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
Device Story
Phyziofit Medical Surgical Mask is a flat-pleated, three-layer disposable face mask. Outer layer: polypropylene non-woven; middle layer: melt-blown non-woven; inner layer: absorbent non-woven. Features include PP plastic nose clips for fit and nylon/spandex elastic ear loops for securement. Used in clinical settings or general infection control to provide a physical barrier against blood, body fluids, and particulates. Worn by healthcare personnel and patients. Device is non-sterile and single-use. Provides protection by filtering particulates and resisting fluid penetration, reducing cross-contamination risk.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated through bench testing including differential pressure, fluid resistance (ASTM F1862), particulate filtration efficiency (ASTM F2299, 98%), bacterial filtration efficiency (ASTM F2101, 99.9%), flammability (16 CFR 1610, Class I), and biocompatibility (ISO 10993-5/10).
Technological Characteristics
Three-layer non-woven construction (polypropylene/melt-blown/absorbent non-woven). PP plastic nose clip. Nylon/spandex ear loops. Non-sterile. Flat-pleated design. Tested per ASTM F2299, ASTM F1862, ASTM F2101, and ISO 10993 standards.
Indications for Use
Indicated for healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material during infection control practices. Single-use, disposable, non-sterile.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Surgical Face Masks (Ear loops and Tie-on) (K160269)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
February 22, 2021
Wellpower Sporting Goods Co., Ltd % Mar Liu Regulatory Affairs Manager Shenzhen Reanny Medical Devices Management Consulting Co Ltd Room 2012#, Gebu commercial building, Hongxing community Songgang street Shenzhen, Guangdong 518105 China
Re: K201868
Trade/Device Name: Phyziofit Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: January 24, 2021 Received: February 1, 2021
Dear Mar Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K201868
Device Name Phyziofit Medical Surgical Mask
#### Indications for Use (Describe)
Phyziofit Medical Surgical Masks are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Wellpower Sporting Goods Co., Ltd. Phyziofit Medical Surgical Mask K201868
#### 1.0 Submitter:
Submitter's name : Submitter's address :
Phone number : Fax number : Name of contact person: Summary Preparation Date: Wellpower Sporting Goods Co., Ltd. No.12. Xinglong New Street, Quantang Industrial Zone, Liaobu Town, Dongguan City, Guangdong Province, 523425 China +86 13829158242 NA Carol Tseng Feb. 20th 2021
| 2.0 | Consultant Firm | Shenzhen Reanny Medical Devices Management Consulting Co., Ltd |
|-----|------------------|-----------------------------------------------------------------------------------------------------------|
| | Contact name | Reanny Wang |
| | Company address | Room 1813, Honghaigebu Business Building, Hongxing Community<br>Songgang Street, Baoan District, Shenzhen |
| | Telephone number | +86-755-27391220/86-15389012257 |
| | Contact email | counselor7@reanny.com |
counselor / @reanny.com
#### 3.0 Name of the Device
| Proprietary/Trade name: | Phyziofit Medical Surgical Mask |
|-------------------------|---------------------------------|
| Common Name: | Mask, Surgical |
| 510(k) Number | K201868 |
| Classification Name: | Mask, Surgical |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 878.4040 |
| Product Code: | FXX |
## 4.0 Predicate device
| Device Name: | Surgical Face Masks (Ear loops and Tie-on) |
|----------------|--------------------------------------------|
| Company name: | San-M Package Co., Ltd. |
| 510(K) Number: | K160269 |
## 5.0 Device Description:
The proposed device(s) are Blue color (colorant: Pigment blue 15:3/CAS No. 147-14-8), and Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose clips design for fitting the facemask around the nose. The proposed device(s) are manufactured with three layers, the outer layer is made of polypropylene non-woven fabric, the middle layer is made of melt blown non-woven fabric and the inner layer is made of absorbent nonwoven fabric.
The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose clips contained in the proposed device(s) is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of PP plastic material.
The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.
{4}------------------------------------------------
# 6.0 Indications for Use:
Phyziofit Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel form transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
# 7.0 Technological Characteristics Comparison:
#### Predicate Device Proposed Device Comparison (Same. Device Similar, Level 1 Level 1 Level 2 Level 3 Characteristic Different) Product name Surgical Face Masks (Ear Phyziofit Medical Surgical Mask N/A loops and Tie-on) K160269 K201868 510(K) No. N/A Product Owner San-M Package Co., Ltd. Wellpower Sporting Goods Co., N/A Ltd. Product Code FXX FXX Same Regulation 21 CFR 878.4040 21 CFR 878.4040 Same Class 11 11 Same The surgical face masks are Phyziofit Medical Surgical Indications for Similar Masks are intended to be worn intended to be worn to protect Use to protect both the patient and both the patient and healthcare personnel from transfer of healthcare personnel from microorganisms, body fluids, transfer of microorganisms. and particulate material. These body fluids and particulate face masks are intended for use material. These face masks are in infection control practices to intended for use in infection control practices to reduce the reduce the potential exposure to blood and body fluids. This potential exposure to blood and
body fluids. This is a single use, is a single-use, disposable device, provided non-sterile. disposable device(s), provided non-sterile. Level I Face Mask Models:# EL 10000, EL 10010, TO 10000, TÓ 10010
Level 2 Face Mask Models:# EL 20000, EL 20010, TO
20000, TO 20010, TO 20010
Level 3 Face Mask Models:# EL 30000, EL 300 I 0, TO 30000. TÓ 300 10 Color White or Blue Blue Similar Single Use, Disposable Single Use Single Use, Disposable Same Sterile Non-Sterile Non-Sterile Same Similar Materials Outer Material: Polypropylene Outer Layer: polypropylene non-Inner Material: Polypropylene woven fabric Filter Media: Polypropylene Middle layer: melt blown non-Spunbond and Polypropylene woven fabric Meltblown Inner layer: absorbent nonwoven
## Comparative Table
{5}------------------------------------------------
| | Nose Clamp: Polyethylene<br>coated<br>steel wire<br>Ear loops: Polyester, polyurethane | fabric<br>Nose Clips: PP plastic material<br>Ear loops: Nylon + spandex | |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|---------|
| Design features | Colors: white or blue<br>Flat-pleated | Colors: blue<br>Flat-pleated | Similar |
| Dimensions-<br>Length | $90\pm 3mm$ | $9.5cm \pm 1cm$ | Similar |
| Dimensions -<br>Palm Width | $175 \pm 5mm$ | $17.5cm \pm 1cm$ | Similar |
| Particulate<br>Filtration<br>Efficiency<br>( ASTM F2299-<br>03 (2017)) | Pass at 99.6% | Pass at 98% | Similar |
| Fluid resistance<br>( ASTM<br>F1862/F1862M-<br>17) | Pass at 80 mmHg | Pass at 80 mmHg | Same |
| Bacterial<br>Filtration<br>Efficiency<br>( ASTM F2101-<br>19) | Pass at >98% | Pass at >98% | Same |
| Flammability<br>(16 CFR1610) | Class 1 | Class 1 | Same |
| Differential<br>Pressure (EN<br>14683:2019+AC<br>: 2019 (E)) | Pass at $2.0mmH_2O/cm^2$ | Pass at $3.4 mmH_2O/cm^2$ | Similar |
| Biocompatibility<br>Cytotoxicity<br>(ISO10993-5) | Under the conditions of the<br>study, the subject device<br>extract was determined to be<br>non-cytotoxic. | Under the conditions of the<br>study, the subject device extract<br>was determined to be non-<br>cytotoxic. | Similar |
| Skin Irritation<br>(ISO10993-10) | Under the conditions of the<br>study, the subject device non-<br>polar and polar extracts were<br>determined to be non-irritating. | Under the conditions of the<br>study, the subject device non-<br>polar and polar extracts were<br>determined to be non-irritating. | |
| Skin<br>sensitization<br>(ISO10993-10) | Under the conditions of the<br>study, the subject device non-<br>polar and polar extracts were<br>determined to be non-<br>sensitizing. | Under the conditions of the<br>study, the subject device non-<br>polar and polar extracts were<br>determined to be non-<br>sensitizing. | |
{6}------------------------------------------------
# 8.0 Summary of Non-Clinical Performance Data
The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device, Surgical Face Masks (Ear loops and Tie-on) manufactured by San-M Package Co., Ltd.The subject device Phyziofit Medical Surgical Mask met the acceptance criteria as established by the Standard testing methods utilized as applicable. No new safety or efficacy concerns were raised during testing:
| Bench Testing | | | | |
|-------------------------------------------|-----------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Test Performed | Test Method | Acceptance<br>Criteria | Average | Results |
| Differential<br>Pressure | EN<br>14683:2019+AC<br>: 2019 (Ε),<br>Annex C | < 5.0mm<br>H2O/cm² | 3.4 mm H2O/cm²<br>(32/32) | Pass |
| Fluid resistance | ASTM<br>F1862/F1862M-<br>17 | No Penetration at<br>80 mm Hg | No Penetration at<br>80 mm Hg<br>(32/32) | Pass |
| Particulate<br>Filtration<br>Efficiency | ASTM F2299-03<br>(2017) | ≥95% | 98% (32/32) | Pass |
| Flammability | 16 CFR Part<br>1610 | Class I | Class I (32/32) | Pass |
| Bacterial<br>Filtration<br>Efficiency | ASTM F2101-19 | ≥95% | 99.9% (32/32) | Pass |
| Biocompatibility<br>cytotoxicity<br>Test | ISO 10993-<br>5:2009 | Non-Cytotoxic | Under the<br>conditions of the<br>study, the subject<br>device extract<br>was determined<br>to be non<br>cytotoxic | Pass |
| Biocompatibility<br>Skin irritation | ISO 10993-10:<br>2010 | Non-Irritating | Under the<br>conditions of the<br>study, the subject<br>device non-polar<br>and polar<br>extracts were<br>determined to<br>be non-irritating. | Pass |
| Biocompatibility<br>Skin<br>Sensitization | ISO 10993-10:<br>2010 | Non-Sensitizing. | Under the<br>conditions of the<br>study, the subject<br>device non-polar<br>and polar<br>extracts were<br>determined to<br>be non-<br>sensitizing. | Pass |
{7}------------------------------------------------
# 9.0 Summary of Clinical Performance Data:
Clinical data was not needed to demonstrate that the subject mask is substantially equivalent to the predicate. Therefore, determination of substantial equivalence is not based on an assessment of clinical performance data.
## 10.0 Conclusion:
The subject product has the same intended use and principle of operation, similar performance specifications as predicate device. The difference between Phyziofit Medical Surgical Mask and predicate device have been analyzed and demonstrates that the difference does not raise any new safety or effectiveness concerns.
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.